Peter Huber Kältemaschinenbau Unistat P825w Dynamisches

Peter Huber Kältemaschinenbau Unistat P825w Dynamisches — Temperature control units.

The Unistat P825w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -85 to 250 °C. Its 'P' designation targets applications with high pressure drop, delivering up to 3 bar and 67 l/min. It serves closed-loop process temperature control in industrial and laboratory systems.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat P825w Dynamisches Current —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 3 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -85...250 °C
Dimensions, Weight & Mechanical Design
Weight 237 kg
Electrical, Power & Drive Specifications
Current consumption 15 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Supply tolerance is narrow and asymmetric

Voltage may deviate ±5% only while frequency deviates ±2%, and the combination of -5% voltage with +2% frequency is explicitly not allowed. Site power quality must be verified before connection, since out-of-tolerance supply can damage the compressor and electronics.

Unistat P825w Technical Data, page 2
Electrical
Three-phase cable supplied without plug

Units below 63 A are delivered with a cable but no plug, so a suitable plug or hard-wired connection must be provided on site. Hard-wired connection must be carried out by an electrician, and the supply must match the rating plate.

Unistat P825w Technical Data, page 2; Unistat Operation Manual, page 37
Cooling & Media
Cooling water window is 5–20 °C at 1–6 bar

Inlet water must be between 5 °C and 20 °C with at least 1 bar differential pressure and no more than 6 bar. Higher water temperature reduces cooling power and increases flow demand; pipes below dew point must be insulated to prevent condensation.

Unistat P825w Technical Data, page 2
Cooling & Media
Cooling water circuit materials restrict water quality

The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used. Aggressive or untreated water can corrode or foul these materials and degrade heat transfer.

Unistat P825w Technical Data, page 2
Process Limits
Cooling capacity collapses at low temperature

Cooling power is 2.2 kW at 0 °C but only 0.3 kW at -80 °C. Process loads that are acceptable at mid-range temperatures may exceed capacity near the lower limit, so load sizing must be done at the actual target temperature.

Unistat P825w Technical Data, page 1
Process Limits
FL rating limits heat transfer fluid choice

For FL-rated (flammable) fluid the fire point must be at least 65 °C; ethanol is permitted only under supervised operation. Fluid selection and supervision must respect the DIN 12876 FL classification.

Unistat Operation Manual, Section 1.7, page 19
Mounting
Expansion vessel must stay unlocked for closed loops

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open (bath) applications it must be locked to prevent fluid loss. The setting must match the application type before commissioning.

Unistat Operation Manual, page 33
Mounting
Upright transport and emptied movement only

The unit must be transported and stored upright only; horizontal transport damages the compressor. Only an emptied unit may be moved, since a filled unit risks thermal fluid overflow. Casters are not for transport and carry only 25% of total mass symmetrically.

Unistat Operation Manual, page 22
Mounting
Acclimation and leveling before first start

Allow about 2 hours acclimation when moving between cold and warm environments before power-on to avoid condensation-induced electrical faults. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, pages 22 and 29
Control & I/O
Process Safety mode needs compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles. This mode requires the compressor to be set to Permanently ON, so the controller configuration must be checked before relying on it for exothermic protection.

Unistat Operation Manual, page 19
Control & I/O
Restart behavior after power loss is configurable

Pilot ONE can be set to Off/Standby (default, manual restart) or ON/Temperature control active (automatic restart after power interruption, with a brief INFO allowing cancellation). The choice must match whether unattended restart is acceptable for the process.

Unistat Operation Manual, Section 1.8.1, page 21
Maintenance
Shut off cooling water even for overnight stops

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Unistat Operation Manual, page 29
Maintenance
Refrigerant work requires certified contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit contains R-452A (GWP 2141, 0.95 kg) and R-23 (GWP 14800, 0.39 kg), about 7.8 t CO2 equivalent, which also triggers leak-monitoring and shipping obligations.

Unistat Operation Manual, Section 1.4.2, page 19; Unistat P825w Technical Data, page 1

Selection & Integration Q&A

When should I choose the P825w Dynamisches over the 815w or 1005w Dynamisches in the same series?

Pick the P825w when the application loop has high pressure drop and needs up to 3 bar / 67 l/min, or when -85 °C reach with 0.3 kW at -80 °C is sufficient. The 815w and 1005w differ in cooling capacity and pump characteristics; select them when their capacity curve better matches the load at the target temperature.

What must be provided on site for the electrical and cooling-water connections before commissioning?

A 400 V 3~ 50 Hz supply with 3x16 A fuse, within ±5% voltage and ±2% frequency (the -5% voltage/+2% frequency combination is not allowed). The unit arrives with cable but no plug. Cooling water: 5–20 °C inlet, ≥1 bar differential, ≤6 bar, G1/2 male connection, 180 l/h at 15 °C water.

What mounting and environment limits apply to the P825w Dynamisches?

IP20, ambient 5–40 °C, not in an ATEX zone. Transport and store upright only and move only when emptied; casters are not for transport. Allow about 2 h acclimation before power-on, activate leveling feet and parking brakes, and install a drip tray below the unit.

What maintenance and consumables are required, and what is not included in base delivery?

Not included: a suitable three-phase plug, cooling-water hoses and fittings (G1/2 male), pump-loop fittings (M30x1.5 male), heat transfer fluid, and a drip tray. Periodic tasks: check hoses and connections for leaks and replace as needed, clean liquefier fins and the hat-type strainer, inspect/replace heat transfer fluid, and shut off cooling water even overnight.