Contents

  1. Overview p. 1
    Introduces the Unistat P920w unit and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 2 table, 26 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Circulation pumpMK pump
max. delivery191 l/min
max. delivery pressure5,5 bar
Delivery at 1,0 bar167 l/min
Delivery at 2,0 bar140 l/min
Delivery at 3,0 bar107 l/min
Delivery at 4,0 bar67 l/min
Delivery at 5,0 bar22 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C730 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity12 l
Filling capacity expansion tank29 l
Overall dimensions WxDxH **950x1205x1650 mm
Net weight916 kg
Power supply requirement400V 3~ 50Hz
max. current38 A
Fuse3x40A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S460015

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicUnistat P920w
Order-No.1061.0011.01
Operating temperature range-90...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C11 kW
Cooling power with Thermooil at 100°C11 kW
Cooling power with Thermooil at 20°C11 kW
Cooling power with Ethanol at 0°C11 kW
Cooling power with Ethanol at -20°C11 kW
Cooling power with Ethanol at -40°C10 kW
Cooling power with Ethanol at -60°C8 kW
Cooling power with Ethanol at -80°C2 kW
Cooling power with Ethanol at -90°C1 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity4,55 kg
CO2 equivalent9,7 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage1,9 kg
CO2 equivalent 2nd stage28,1 t

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is a water-cooled refrigerated heating circulator with a magnetic-coupled stainless steel pump, suitable for both externally closed and externally open applications.

Confirms cooling medium and application compatibility before specifying the unit.

Source
“Refrigerated Heating Circulator with water-cooled cooling machine and optical level indicator. Magnetic coupled circulation pump made of stainless steel.”
Selection

The 'P' model pump delivers high discharge pressure for systems with high pressure drops such as flow-through chemistry or semiconductor processes.

Determines whether this pump variant suits a high-pressure-drop loop.

Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation

Cooling water must be supplied between 5°C and 20°C; below the condensation point the cooling water pipes must be insulated.

Sets the facility cooling water requirement and insulation work at installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Rated output data assume 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure; higher cooling water temperature reduces cooling power.

Prevents overestimating cooling capacity when site conditions differ from the rating basis.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility checks for the cooling water supply.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Defines acceptable supply quality for safe commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

Leakage current exceeds 3.5 mA, which affects protective earthing and RCD selection.

Requires appropriate earth connection and residual current device planning.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be provided on site.

Affects electrical installation scope and ordering of the mains connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Pump connections use bore shape Y (60°) per DIN 3863 with ball socket per DIN 3863 and sleeve nut per DIN 3870.

Ensures correct mating fittings and hoses are procured for the piping.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

Confirms built-in safety function required for unattended operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The refrigerant is R-452A with GWP 2141 and 4.55 kg charge (9.7 t CO2e), plus R-23 second stage with GWP 14800 and 1.9 kg (28.1 t CO2e).

Relevant for refrigerant handling, leak checks and environmental reporting.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280)”