Peter Huber Kältemaschinenbau Unistat P920w Dynamisches
Peter Huber Kältemaschinenbau Unistat P920w Dynamisches — Temperature control units.
The Unistat P920w Dynamisches is a water-cooled refrigerated heating circulator with a magnetically coupled stainless steel pump, covering -90 to 200 °C with 12 kW heating. Its high-discharge-pressure 'P' pump targets high-pressure-drop loops in flow-through chemistry and semiconductor processes. It serves as a dynamic temperature control source for externally closed or open applications.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P920w Dynamisches Current | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 29 l |
| Filling volume | 12 l |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Weight | 916 kg |
| Current consumption | 38 A (max.) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 730 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigerated heating circulator with optical level indicator; magnetic-coupled stainless steel circulation pump.
- The 'P' designation denotes a pump variant with high discharge pressure for applications with high pressure drops (flow-through chemistry, semiconductor industry).
- Sibling models in the Unistat w Dynamisches series (705w, 815w, 1005w, 1015w) share the water-cooled architecture; the P920w is distinguished by its high-pressure pump and 12 kW heating capacity.
- Pump connections are M38x1.5 AG male; cooling water connection is G3/4 male.
- Safety class III/FL with adjustable overtemperature protection per DIN 12876.
Behavior and media
- Temperature range -90 to 200 °C; display resolution 0.01 K.
- Heating capacity 12 kW. Cooling water consumption 730 l/h at 15 °C inlet with 0 °C flow; minimum differential pressure 1 bar, maximum cooling water pressure 6 bar.
- Rated output data assume 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure; higher cooling water temperature reduces cooling power.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible cooling water may be used.
- Refrigerant R-452A (GWP 2141, 4.55 kg, 9.7 t CO2e) plus R-23 second stage (GWP 14800, 1.9 kg, 28.1 t CO2e).
Installation, control and operation
- Controller Pilot ONE with POKO ONE control output; interfaces Ethernet, USB (Host and Device), RS232; external Pt100 sensor connection.
- Power supply 400 V 3~ 50 Hz; max current consumption 38 A; protection rating IP20.
- Three-phase devices below 63 A are delivered with cable but without plug; a suitable plug must be provided on site.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%; certain combinations such as -5% voltage with +2% frequency are not allowed.
- Leakage current exceeds 3.5 mA, affecting protective earthing and RCD selection.
- Class A, Group 1 electromagnetic disturbance to EN55011.
- Cooling water intake temperature must be between 5 °C and 20 °C; below the condensation point the cooling water pipes must be insulated.
- Pump connections use bore shape Y (60°) per DIN 3863 with ball socket per DIN 3863 and sleeve nut per DIN 3870.
- Ambient temperature limit 5 to 40 °C.
- Filling volume 12 l; expansion volume 29 l; weight 916 kg.
Integration & Operating Notes
The unit must not be installed or started up within an ATEX zone; without an Ex px cabinet the unit is not protected against explosion. Site selection must exclude ATEX zones, and Ex px cabinet operation requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable.
Only transport the temperature control unit in an upright position; horizontal transport damages the compressor. Only an emptied unit may be moved; filled transport risks thermal fluid overflow. Casters (if present) are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be unscrewed/activated and parking brakes engaged before commissioning.
This temperature control unit is designed to operate with cooling water intake temperature between 5 °C and 20 °C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit. Rated output data assume 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure; higher cooling water temperature reduces cooling power.
Materials used in the cooling water circuit include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Only compatible cooling water may be used; water quality must be checked against these materials before connection.
In Process safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles of the temperature control. This mode requires the compressor to be set to Permanently ON to keep cooling active at the OT trigger.
If temperature control is active during a power outage, the process will automatically continue after the power outage when Power Failure Auto is enabled. Choose the power-interruption behavior according to whether unattended restart is acceptable for the process.
Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g. overnight) to prevent water damage and unnecessary load on the cooling water circuit.
Check periodically for leaks and the quality of the hose and hose connections and take suitable measures (replace) as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors.
Refrigerant R-452A with GWP 2141 and 4.55 kg charge (9.7 t CO2e), plus R-23 second stage with GWP 14800 and 1.9 kg (28.1 t CO2e). Relevant for refrigerant handling, leak checks and environmental reporting.
Selection & Integration Q&A
Choose the P920w when the loop has high pressure drop (flow-through chemistry, semiconductor processes) requiring the high-discharge-pressure 'P' pump. Sibling w models share water cooling but lack the P pump designation; select them for lower pressure-drop duty.
Pump connections are M38x1.5 AG male using bore shape Y (60°) per DIN 3863 with ball socket per DIN 3863 and sleeve nut per DIN 3870. Cooling water connection is G3/4 male; supply 5–20 °C at min. 1 bar differential, max. 6 bar, 730 l/h at 15 °C.
Supply 400 V 3~ 50 Hz, max. 38 A, with ±5% voltage and ±2% frequency tolerance (not all combinations allowed). Leakage current >3.5 mA; Class A EN55011. Ambient 5–40 °C; IP20; must not be installed in an ATEX zone.
Base delivery includes cable but no plug for three-phase <63 A; provide a suitable plug. Periodically check hoses/connections and replace as needed; shut off cooling water overnight. Refrigerant work requires approved contractors. A drip tray must be installed below the unit.