Unistat P925w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat P925w unit
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 26 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Circulation pumpMK pump
max. delivery191 l/min
max. delivery pressure5,5 bar
Delivery at 1,0 bar167 l/min
Delivery at 2,0 bar140 l/min
Delivery at 3,0 bar107 l/min
Delivery at 4,0 bar67 l/min
Delivery at 5,0 bar22 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1300 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity12 l
Filling capacity expansion tank29 l
Overall dimensions WxDxH **950x1205x1650 mm
Net weight994 kg
Power supply requirement400V 3~ 50Hz
max. current52 A
Fuse3x63A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S517345

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicUnistat P925w
Order-No.1081.0003.01
Operating temperature range-90...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C16 kW
Cooling power with Thermooil at 100°C16 kW
Cooling power with Thermooil at 20°C16 kW
Cooling power with Ethanol at 0°C16 kW
Cooling power with Ethanol at -20°C16 kW
Cooling power with Ethanol at -40°C15 kW
Cooling power with Ethanol at -60°C13,5 kW
Cooling power with Ethanol at -80°C3,5 kW
Cooling power with Ethanol at -90°C1,8 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity5 kg
CO2 equivalent10,7 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage2,2 kg
CO2 equivalent 2nd stage32,6 t

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V 3~ 50 Hz supply with a maximum current of 52 A and 3x63 A fusing, so the electrical installation must be sized accordingly.

Determines the site power distribution, cable cross-section and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 52 A Fuse 3x63A”
Installation

Cooling water must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and the specified flow is 1300 l/h at 15 °C.

Sizing the cooling water circuit and verifying available pressure/flow is mandatory for achieving rated cooling capacity.

Source
“Consumption at water 15°C, flow 0°C 1300 l/h min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Prevents corrosion and fouling that would degrade heat transfer and damage the water-cooled refrigeration machine.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.

Avoids condensation, dripping and corrosion on the unit when using cold cooling water.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The pump connection is M38x1.5 male and the cooling water connection is G3/4 male, so matching hoses and adapters must be provided.

Ensures correct mechanical interface for process piping and cooling water hook-up.

Source
“Pump connection M38x1,5 male”
Installation

Three-phase devices with current consumption less than 63 A are delivered with cable but without plug, so a suitable plug must be sourced on site.

Affects electrical commissioning planning and the bill of materials for the installation.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The leakage current exceeds 3.5 mA, which requires attention to protective earthing and residual current device selection.

Impacts electrical safety design and choice of RCD type for the supply circuit.

Source
“Attention: leakage current > 3,5mA”
Safety

The refrigerant R-452A has a GWP of 2141 and 5 kg charge (10.7 t CO2e), while the second stage R-23 has a GWP of 14800 and 2.2 kg charge (32.6 t CO2e).

Relevant for F-gas compliance, leak monitoring obligations and environmental reporting.

Source
“Global Warming Potential (GWP) 2141 Refrigerant quantity 5 kg CO2 equivalent 10,7 t”
Safety

The unit has safety classification III / FL and adjustable overtemperature protection according to DIN 12876.

Confirms suitability for flammable or high-risk temperature control applications and defines the safety concept.

Source
“Safety classification III / FL”
Operation

The operating temperature range is -90 to 200 °C with a display resolution of 0.01 K and a Pt100 internal sensor plus external Pt100 connection.

Defines the process window and the need for an external sensor for process control.

Source
“Operating temperature range -90...200 °C”
Performance

Cooling power drops sharply at low temperatures, from 16 kW at -20 °C to 3.5 kW at -80 °C and 1.8 kW at -90 °C.

Critical for sizing the unit to the actual heat load at the lowest required process temperature.

Source
“at -80°C 3,5 kW at -90°C 1,8 kW”
Performance

The MK circulation pump delivers up to 191 l/min and 5.5 bar, with 22 l/min still available at 5.0 bar.

Determines whether the pump can overcome the pressure drop of the external process loop.

Source
“max. delivery 191 l/min max. delivery pressure 5,5 bar”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Limits the choice of heat transfer fluid, especially at low temperatures where viscosity rises.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit has IP20 protection and must be operated between 5 °C and 40 °C ambient temperature.

Restricts installation location to dry indoor areas with adequate ventilation and climate control.

Source
“Degree of Protection IP20”