Peter Huber Kältemaschinenbau Unistat P925w Dynamisches

Peter Huber Kältemaschinenbau Unistat P925w Dynamisches — Temperature control units.

The Unistat P925w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 200 °C. Its 12 kW heating and staged cascade refrigeration (R-452A/R-23) deliver 16 kW at -20 °C falling to 1.8 kW at -90 °C. It serves closed-loop process temperature control where high cooling demand at low temperature and 191 l/min pump flow are required.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat P925w Dynamisches Current —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Expansion volume 29 l
Filling volume 12 l (min)
Heating capacity 12 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -90...200 °C
Dimensions, Weight & Mechanical Design
Weight 994 kg (net)
Electrical, Power & Drive Specifications
Current consumption 52 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 1300 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must be sized for 52 A with 3x63 A fusing

The 400 V 3~ 50 Hz supply must be dedicated and rated for the 52 A maximum current; the 3x63 A fuse rating defines upstream breaker and cable cross-section. The unit is delivered with cable but without plug, so the plug and any hard-wired connection must be provided by an electrician on site.

Unistat P925w Technical Data, page 2
Cooling & Media
Cooling water must deliver 1300 l/h at 1–6 bar

The cooling water circuit requires a minimum differential pressure of 1 bar and must not exceed 6 bar; the specified flow is 1300 l/h at 15 °C water and 0 °C flow. Only water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM may be used, and pipes must be insulated if water temperature falls below the condensation point.

Unistat P925w Technical Data, page 2
Process Limits
Heat transfer fluid viscosity must not exceed 50 mm²/s

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s. At low process temperatures, viscosity rises and may exceed this limit, restricting fluid choice and potentially affecting pump performance and heat transfer.

Unistat P925w Technical Data, page 2
Mounting
Transport and store upright; acclimate 2 hours before power-on

The unit must only be transported and stored upright; horizontal transport damages the compressor. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, page 22 and page 29
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; otherwise the safety function will not operate as intended.

Unistat Operation Manual, page 19
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the P925w over the smaller 705w or 815w Dynamisches models?

Choose the P925w when the process requires up to 12 kW heating and cooling capacity down to -90 °C with high pump flow (191 l/min, 5.5 bar). The 705w and 815w have lower capacities; select them only if the heat load and pressure drop are within their smaller envelopes.

What are the requirements for the cooling water supply and the process loop connections?

Cooling water: 5–20 °C, 1300 l/h at 15 °C, minimum 1 bar differential pressure, maximum 6 bar, G3/4 male connection. Process loop: M38x1.5 male pump connection. Ensure water compatibility with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.

What ambient and installation conditions does the manual require?

Install indoors only (IP20), ambient 5–40 °C, upright transport and storage, leveling feet activated, 2-hour acclimation after temperature changes. Do not install in an ATEX zone. A drip tray for condensation water/heat transfer fluid must be installed below the unit.

What maintenance and accessories are required beyond the base delivery?

The unit is delivered with cable but without plug; a suitable plug must be sourced. Periodic checks of hoses and connections for leaks and quality are required, with replacement as needed. Cooling water supply should be shut off during downtimes. Refrigerant-circuit work must be by approved contractors.