Unistat T345 HT Technical Data

Contents

  1. Overview p. 1
    Product name and general description of the Unistat T345 HT unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 25 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range65... 300 °C
Temperature stability at 70°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power72 kW
Cooling power at 300°C30 kW
Circulation pump:MK pump
max. delivery200 l/min
max. delivery pressure5,5 bar
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Filling capacity expansion tank56 l
Overall dimensions WxDxH **800x1007x2200 mm
Power supply (3 Phase)400V 3~ 50Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/25

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a high-temperature circulator covering 65 to 300 °C, so it suits processes requiring heating up to 300 °C but not sub-ambient cooling.

Defines the process window and excludes applications needing temperatures below 65 °C.

Source
“Operating temperature range 65... 300 °C”
Performance

With 72 kW heating and 30 kW cooling at 300 °C, the unit can handle large thermal loads but requires adequate facility power and heat rejection.

Sizing the electrical supply and cooling water/heat rejection must match these capacities.

Source
“Heating power 72 kW”
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable 3-phase feed.

Confirms the electrical infrastructure needed before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices under 63 A are delivered with cable but without a plug, so the customer must provide and fit the site plug.

Affects electrical installation planning and who supplies the plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±5% and ±2% but not simultaneously in opposite directions, so supply quality must be checked.

Prevents commissioning on a supply that would violate the allowed tolerance combination.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for high leakage.

Impacts electrical safety design and choice of residual current protection.

Source
“Attention: leakage current > 3,5mA”
Installation

Pump connections use DIN 3863 ball-socket fittings with DIN 3870 sleeve nuts, so hoses and pipework must match this standard.

Ensures correct mating fittings and avoids leaks at the M38x1.5 pump connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The unit is rated for ambient temperatures from 5 to 40 °C, so the installation room must stay within this range.

Room conditions outside this range can affect performance and reliability.

Source
“min. ambient temperature 5 °C”
Installation

The enclosure is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Prevents placing the unit in washdown or outdoor environments.

Source
“Degree of Protection IP20”
Selection

The pump delivers up to 200 l/min at 5.5 bar, so external circuit pressure drop must be checked against this curve.

Ensures the circulation pump can overcome the external loop resistance.

Source
“max. delivery pressure 5,5 bar”
Selection

The maximum permissible kinematic viscosity is 50 mm²/s, so the heat transfer fluid must be selected within this limit at operating temperature.

Prevents pump overload and poor circulation with overly viscous fluids.

Source
“max. permissible kin. viscosity 50 mm²/s”
Compliance

The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Affects installation location and EMC compatibility with sensitive equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has adjustable overtemperature protection per DIN 12876 and safety class III/FL, so it is suitable for flammable liquids with the appropriate setup.

Confirms built-in safety features required for unattended operation and flammable media.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The expansion tank holds 56 l and is not thermoregulated, so fluid volume and expansion must be planned accordingly.

Determines fluid fill volume and expansion allowance for the system.

Source
“Filling capacity expansion tank 56 l”