Peter Huber Kältemaschinenbau Unistat T345 HT Dynamisches
Peter Huber Kältemaschinenbau Unistat T345 HT Dynamisches — Temperature control units.
The Unistat T345 HT Dynamisches is a high-temperature dynamic temperature control system with a 65–300 °C working range, 72 kW heating and 300 kW cooling at 30 °C. It is the larger-capacity HT modification in the Unistat T3xx series, positioned above the T305 variants. It serves closed-loop heating and cooling of large process loads requiring precise, dynamic temperature control up to 300 °C.
Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection
| Model | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C | Expansion volume, m³ |
|---|---|---|---|---|---|---|
| Unistat T345 HT Dynamisches Current | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Chili Dynamisches | 0.01 | 200-240V 1~/2~ 50/60Hz | M24x1,5 AG male | Pressure pump, variable | 65–300 | 0.002 |
| Unistat T305 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305 HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305w HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T320 Dynamisches | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T320w HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330w HT Dynamisches | 0, 01 K | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T340 Dynamisches | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T340 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T350 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T402 Dynamisches | 0, 01 | 400V 3~N 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 80–425 | — |
| Unistat TR401 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 50–400 | 0.005 |
| Unistat TR402 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 80–425 | — |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 56 l |
| Heating capacity | 72 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 65...300 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 300 kW (at 30 °C) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The T345 HT is a dynamic (flow-through) temperature control unit with a magnetically coupled stainless-steel pump, 56 l expansion volume and M38x1.5 male pump connections.
- Within the Unistat T3xx series it is the higher-capacity HT variant; the T305 and T305w HT models are the smaller siblings, while the Chili Dynamisches is a different line.
- Heating capacity is 72 kW; cooling capacity is 300 kW at 30 °C. The unit is rated for 400 V 3~ 50 Hz and IP20 enclosure.
Working range and media behavior
- Operating temperature range is 65–300 °C, so the unit is intended for heating duties and cannot provide sub-ambient cooling.
- Maximum permissible kinematic viscosity is 50 mm²/s; the heat transfer fluid must be selected to stay within this limit at operating temperature.
- The pump delivers up to 200 l/min at 5.5 bar maximum delivery pressure; external circuit pressure drop must be checked against this curve.
- Safety class III/FL with adjustable overtemperature protection per DIN 12876 supports use with flammable liquids when correctly set up.
Control, installation and operation
- Control is via Pilot ONE with POKO ONE control output; display resolution 0.01 K; interfaces include Ethernet, USB (host and device) and RS232; an external Pt100 sensor can be connected.
- The unit must be installed upright, level and with parking brakes/leveling feet engaged; it must not be placed in an ATEX zone.
- For externally closed applications the expansion vessel must remain unlocked; for externally open applications it must be locked.
- Process Safety mode requires the compressor set to Permanently ON; Power Failure Auto resumes temperature control after a power outage if it was active.
Integration & Operating Notes
The unit requires 400 V 3~ 50 Hz with ±5% voltage and ±2% frequency tolerance, but these tolerances must not be applied simultaneously in opposite directions. Verify supply quality before commissioning; three-phase devices under 63 A are delivered with cable but without a plug, so the site plug must be provided and fitted by an electrician.
For externally closed applications the expansion vessel must remain unlocked; locking it will damage the application due to overpressure. For externally open applications the expansion vessel must be locked to prevent fluid loss. Confirm the application type before commissioning.
The maximum permissible kinematic viscosity is 50 mm²/s. Select the heat transfer fluid so that its viscosity at the lowest operating temperature remains within this limit; otherwise pump load rises and circulation may become inadequate.
The unit must be transported and stored upright only; horizontal transport damages the compressor. Only an emptied unit may be moved, since filled transport risks thermal fluid overflow. Casters are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck.
If Process Safety mode is used to protect against exothermic reactions, the compressor must be configured to Permanently ON so that cooling remains active when the overtemperature trigger occurs. Power Failure Auto resumes temperature control automatically after a power outage if it was active.
Shut off the cooling water supply to the unit even for shorter downtimes such as overnight, to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality and replace as required.
Selection & Integration Q&A
Choose the T345 HT when the process needs the larger 72 kW heating and 300 kW cooling capacity at 30 °C and a 65–300 °C range. The T305/T305w HT are smaller-capacity siblings; select them only if their lower capacity covers the load.
Pump connections are M38x1.5 male with bore shape Y (60°) per DIN 3863; hoses need ball sockets per DIN 3863 and sleeve nuts per DIN 3870. External pressure drop must be checked against the pump curve of up to 200 l/min at 5.5 bar.
Install indoors within 5–40 °C ambient, IP20 enclosure, never inside an ATEX zone. Transport and store upright only, level the unit and engage parking brakes before operation, and allow 2 h acclimation after a temperature change before power-on.
The unit ships with cable but without a plug for three-phase supplies under 63 A; the site must provide the plug. Plan periodic hose/connection inspection and replacement, shut off cooling water during downtime, and provide heat transfer fluid within the 50 mm²/s viscosity limit.