Unistat T402 Temperature Control System Datasheet

Contents

  1. Overview p. 1
    Product introduction and main features of Unistat T402
  2. Spec pp. 1-2
    Technical data, performance and operating parameters per DIN 12876
  3. p. 2 table, 5 rows
  4. p. 1 table, 23 rows

Technical data according to DIN 12876

CharacteristicValue
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S622126

Unistat T402 Technical data according to DIN 12876

CharacteristicValue
Operating temperature range80... 425 °C
Temperature stability at 70°C0,05 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Circulation pump: max. delivery45 l/min
Circulation pump: max. delivery pressure0,9 bar
Pump connectionM24x1,5 male
Cooling water connectionG1/2 male
min. filling capacity1,45 l
Filling capacity expansion tank2,8 l
Overall dimensions WxDxH **506x400x765 mm
Net weight62 kg
Power supply (3 Phase)400V 3~N 50/60Hz
max. current (3 Phase)17 A
Fuse (3 phase)3x20A

** The dimensions specified below are approximate and may be subject to change.

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit requires a three-phase 400 V supply with up to 17 A per phase, so the electrical infrastructure must be sized accordingly.

Determines whether the site has a suitable three-phase supply and correct breaker/fuse rating before purchase.

Source
“Power supply (3 Phase) 400V 3~N 50/60Hz”
Installation

Three-phase devices drawing less than 63 A are delivered with cable but without a plug, so the customer must provide and install the plug.

Affects installation planning and the need to procure a suitable plug and have a qualified electrician connect it.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The pump connection is an M24x1.5 male thread, so mating hoses or adapters must match this thread.

Prevents ordering incompatible hoses or fittings and avoids leaks or delays during commissioning.

Source
“Pump connection M24x1,5 male”
Installation

The cooling water connection is G1/2 male, so the site cooling water supply must be fitted with a matching G1/2 female coupling.

Ensures the cooling water circuit can be connected without adapters and that the correct thread is available on site.

Source
“Cooling water connection G1/2 male”
Installation

The expansion tank lid can be fixed and has connections for a nitrogen blanket at a maximum pressure of 0.2 bar.

Relevant when inert gas blanketing is required; exceeding 0.2 bar could damage the tank or create a safety hazard.

Source
“The lid has connections for a nitrogen blanket (maximum pressure 0,2 bar)”
Operation

The ambient temperature must stay between 5 °C and 40 °C for safe and reliable operation.

Sites outside this range may experience degraded performance, condensation, or overheating and should be assessed before installation.

Source
“min. ambient temperature 5 °C”
Operation

The maximum ambient temperature is 40 °C, so the installation location must not exceed this limit.

Exceeding this limit can reduce cooling capacity and cause overheating or protective shutdowns.

Source
“max. ambient temperature 40 °C”
Safety

The unit has an adjustable overtemperature protection according to DIN 12876, which must be set correctly for the application.

Correct setting of the overtemperature protection is essential to prevent overheating of the heat transfer fluid and a fire hazard.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety

The unit is classified as safety class III / FL according to DIN 12876, which defines the permitted fluids and operating conditions.

The safety class determines which heat transfer fluids may be used and what protective measures are required.

Source
“Safety classification III / FL”
Performance

The circulation pump delivers up to 45 l/min at a maximum pressure of 0.9 bar, which must be sufficient for the connected external circuit.

Ensures the pump can overcome the pressure drop of the external loop and maintain the required flow for heat transfer.

Source
“max. delivery 45 l/min”
Performance

The heating power is 6 kW, which determines heat-up time and the maximum heat load that can be handled.

Needed to verify that the unit can reach the required temperature within the desired time and support the process heat load.

Source
“Heating power 6 kW”
Maintenance

The minimum filling capacity is 1.45 l and the expansion tank capacity is 2.8 l, which must be respected when filling the system.

Underfilling can cause pump cavitation and poor temperature control; overfilling can cause overflow when the fluid expands.

Source
“min. filling capacity 1,45 l”
Compliance

The unit is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011, which may require additional filtering in residential areas.

Class A equipment is not intended for residential use without additional measures; this affects site selection and EMC compliance.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The overall dimensions are 506 x 400 x 765 mm and the net weight is 62 kg, so the installation location must provide adequate space and floor loading.

Needed to plan the installation footprint, access for service, and ensure the floor can support the weight.

Source
“Overall dimensions WxDxH ** 506x400x765 mm”