Induztr
Engineering Digest Document Viewer Related Models

IDP-10 Dry Scroll Vacuum Pump User Manual

Contents

  1. Overview pp. 3-4
    Table of contents listing manual sections and languages
  2. Warning pp. 5-6
    Safety guidelines for dry scroll vacuum pumps
  3. Procedure pp. 7-8
    Cleaning instructions for the pump
  4. Overview pp. 9-10
    Chinese-language operating instructions
  5. Warning pp. 11-12
    Japanese safety guidelines for dry scroll pumps
  6. Procedure pp. 13-14
    English cleaning instructions
  7. Overview pp. 15-38
    Italian-language operating instructions and product description
  8. Overview pp. 39-62
    German-language operating instructions and product description
  9. Overview pp. 63-86
    French-language operating instructions and product description
  10. Overview pp. 87-110
    Spanish-language operating instructions and product description
  11. Overview pp. 111-134
    Chinese-language operating instructions and product description
  12. Overview pp. 135-158
    Japanese-language operating instructions and product description
  13. Overview pp. 159-182
    English-language operating instructions and product description
  14. Spec pp. 183-184
    Technical information and specifications
  15. Spec pp. 185-186
    Detailed technical specifications table
  16. Spec pp. 187-188
    IDP-10 outline drawing and dimensions
  17. pp. 189-190
    Electrical connection details and pin functions
  18. Spec pp. 191-192
    Inrush current data table
  19. Procedure pp. 193-194
    Pump operation instructions
  20. Procedure pp. 195-196
    Troubleshooting guide table
  21. Procedure pp. 197-204
    Tip seal replacement procedure and notes
  22. p. 205
    Customer care contact information
  23. p. 185 table, 18 rows
  24. p. 189 table, 9 rows
  25. p. 196 table, 4 rows
  26. p. 176 table, 9 rows
  27. p. 183 table, 19 rows
  28. p. 186 table, 11 rows

Table 15 Technical Specifications

ModelIDP-10 Dry Scroll Single Hermetic Vacuum Pump
Interface dimensionsH=297 mm x W=260 mm x L= 420mm
Peak pumping speed170L/min (10.2 m3/h)
MediaNo corrosive, explosive or particulate forming gases
Ultimate pressure*1.5 x 10-2 Torr (2.0 x 10-2 mbar)
Maximum recommended continuous inlet pressure300 Torr (400 mbar)
Maximum inlet pressure at nominal voltage1.0 atmosphere (0 psig)
Maximum outlet pressure115V range: 5 psig, 115 VAC 50/60HZ or greater, derate linearly to 0 psig at 100VAC below 115VAC 230V range: 5 pisg, 200 VAC 50/60HZ or greater
Inlet connectionNW25
Exhaust connectionFemale 1/4" National Pipe Thread (NW16)
Gas ballastFemale 1/8" National Pipe Thread, (20 micron sintered plug provided)
Ambient operating temperature5 °C to 40 °C (41 °F to 104 °F)
Storage temperature–20 °C to 60 °C (–4 °F to 140 °F)
Motor rating1/2 HP (370 W)
Operating voltages**Operates with inverter - functions with several power inputs: 1 phase/100-120VAC 50/60Hz, 9 A typical; 200-240 50/60Hz, 5 A typical
Motor thermal protectionAutomatic thermal protection
Operating speedFull speed is 1900 RPM; RPM can be adjusted between 1350 to 1900 RPM. Factory setting for X3810-64000 and X3810-64010 is 1900 rpm, there is optional default setting at 1800 rpm available.
Cooling systemAir-cooled
WeightPump only: 24.7 kg (55 lbs) Shipping weight: 28.1 kg (62 lbs)

*Ultimate pressure; **Operating voltages

Table 16 IDP-10 Power Cord Selection

CountryPower Cord SpecificationOrder
Europe10 A / 220-230 VAC, 2.5 m656494220
Denmark10 A / 220-230 VAC, 2.5 m656494225
Switzerland10 A / 230 VAC, 2.5 m656494235
UK/Ireland13 A / 230 VAC, 2.5 m656494250
India10 A / 220-250 VAC, 2.5 m656494245
Israel10 A / 230 VAC, 2.5 m656494230
Japan12 A / 100 VAC, 2.3 m656494240
North America15 A / 125 VAC, 2.0 m656458203
10 A / 230 VAC, 2.5 m656494255

Table 20 Service Options

Factory Service PlansModel Number
IDP-10 Exchange PumpX3810-69000
IDP-10 Exchange with Inlet ValveX3810-69010
IDP-10 Full Pump RebuildX3810-69000-RP
IDP-10 Full Pump (with Inlet Valve) RebuildX3810-69010-RP

Table 14 DSUB9 Pin Connections

Pin #Function
1Analog speed control input voltage 0-10 VDC
2External pump On/Off input, active LO
3NC
4NC
5Ground
6NC
7+10VDC output
8Reserved for Factory use
9Ground

Table

РазделСтраница
Description of the IDP-10 Dry Scroll Vacuum Pump184
Technical Specifications185
IDP-10 Dry Scroll Vacuum Pump Outline Drawing187
Electrical Connections189
Power Cord189
Grounding Instructions189
Extension Cords190
Inrush Current191
Mechanical Connections191
Isolation Valve191
Pump Inlet191
Pump Exhaust191
Gas Ballast192
Pump Operations193
Mounting of the IDP-10 Scroll Vacuum Pump193
Protection Features194
Troubleshooting195
Kits and Service Options196
Tip Seal Replacement197

Technical Information

CharacteristicValue
Noise Level (per ISO 11201)53 +/- 2 dB(A)
Vibration level at inlet (per ISO 10816-1)Class 1B, 7 mm/s
Fuse typeM5 x 20 mm, 10 A, Slo-Blo, Littelfuse Model 215010 or equivalent, not user replaceable
Hour meterIntegral hour meter provided, displays running time to closest 1 hour
Conformance standardsConforms to EN61010 3rd edition, CSA 22.2 for US and Canada, CE marked.
InstallationIndoor use, Installation Category II, Pollution Degree 2
Altitude2000 m (6562’)
Emissions and ImmunityEN 61236-1:2006 as a Class A device
Water Vapor Handling120 g/hr
Ambient Relative Humidity5-95%; non-condensing
SafetyEN /IEC 61010 3rd Edition CSA certified for US and Canada

*If the pump and/or vacuum system is new, or has been exposed to atmosphere for an extended period of time, it is possible that water can be present which can affect the pump’s ability to achieve ultimate pressure. To remove water from the system it is recommended to open GB1 for 15 minutes and/or flush with dry nitrogen. Ultimate pressure will varies for different configuration and RPM setting. **Mains supply voltage fluctuations shall not exceed 10% of the nominal voltage.

Engineering Facts

11 facts · every claim linked to a page
Installation

The pump must be kept within -20 to +60 °C and 0–95% non-condensing RH during transport and storage.

Sets the environmental limits for shipping and warehousing before installation.

Source
“Temperatur: -20 °C bis +60 °C (-4 °F bis 140 °F)”
Installation

The IDP-10 alone weighs 24.7 kg, so lifting aids and proper handling techniques are required to avoid injury.

Determines the lifting equipment and manpower needed for safe installation.

Source
“La pompe IDP-10 pèse à elle seule 24,7 kg (55 lb).”
Installation

A clear space of at least 60 cm must be maintained around the equipment for ventilation, service access, and main breaker access.

Defines the minimum bench/floor clearance required when laying out the installation.

Source
“Il doit y avoir un espace bien dégagé d’au moins 60 cm (24") sur l’intégralité du pourtour de l’équipement.”
Compliance

The unit is Class A Group 1 ISM equipment, so EMC compatibility may be difficult to guarantee outside industrial environments.

Restricts where the pump may be installed and warns about interference in residential or sensitive environments.

Source
“Ce dispositif est conforme aux exigences de CISPR11, Groupe 1, Classe A en tant qu’équipement professionnel de radiation.”
Operation

Pin 2 is a TTL-compatible logic input where a closed switch to signal ground or a logic LO runs the pump, and the slide switch must be in position 0 when using it.

Specifies the remote start/stop wiring and the mandatory switch setting to avoid misoperation.

Source
“Notez que l'interrupteur coulissant DOIT être dans la position 0 lorsque vous utilisez cette sortie.”
Operation

Pin 7 provides a 10 VDC reference for a speed-control potentiometer, and a 5 kOhm potentiometer is recommended.

Gives the correct potentiometer value for manual speed control wiring.

Source
“Un potentiomètre de 5 Kohm est recommandé.”
Operation

Pin 8 must be left unconnected, and the shell (Pin 9) is chassis ground and is not connected to the signal ground pins.

Prevents wiring errors and clarifies the grounding scheme for the control connector.

Source
“Broche n°8 : n'effectuez pas de raccordement à cette broche.”
Safety

Vacuum pumps and components may be contaminated by toxic, reactive, or radioactive substances, so adequate PPE must be worn during maintenance and reinstallation.

Mandates PPE selection based on process contamination risk during service work.

Source
“Wear adequate protective equipment during maintenance and repairs or in case of reinstallation.”
Safety

Any equipment heavier than 20 kg must be moved with a suitable lifting device, and steel-toe safety shoes per EN 347 are required.

Sets the manual-handling threshold and PPE requirement for moving the pump.

Source
“Any equipment heavier than 20 kg should be transported using a suitable lifting device.”
Maintenance

The pump must be fully stopped before servicing or before mounting/dismounting from the system, and hands must not be put inside the high-vacuum flange.

Defines the lockout/standstill prerequisite for safe maintenance and removal.

Source
“Before servicing the pump or before any mounting/ dismounting action of the pump from the system, wait for the complete standstill of the pump.”
Maintenance

Higher-level maintenance must be performed only by specially trained personnel, not by the ordinary user.

Clarifies the boundary between user-level and trained-service maintenance tasks.

Source
“must not be carried out by the user.”