Agilent Technologies IDP 10 Dry Scroll Pump

Agilent Technologies IDP 10 Dry Scroll Pump — Liquid chromatography with Motor power 0.746 kW, Ultimate pressure 2 mbar.

The IDP-10 is a single-phase dry scroll vacuum pump with a peak pumping speed of 170 L/min and an ultimate pressure of 1.5 × 10⁻² Torr. Its 1 HP motor runs at a factory-set 1900 RPM (adjustable 1350–1900 RPM), and it handles up to 120 g/hr of water vapor with both gas ballast ports open. It is intended for oil-free evacuation of clean, non-corrosive gases in laboratory and analytical systems.

Motor power
0.746 kW
Ultimate pressure
2 mbar

Full specifications

Core Performance & Functional Specifications
Motor power 1 HP
Ultimate pressure 1.5 Torr
Coverslip type No corrosive, explosive or particulate forming gases
Operating Speed 1350...1900 RPM
Water Vapor Handling Per Pneurop 6602 120 g/hr
Peak pumping speed mfr 170 L/min
Maximum recommended continuous inlet pressure mfr 300 Torr
Maximum outlet pressure mfr 115 psig
Exhaust connection mfr 1 Thread
Gas ballast mfr 1
Motor thermal protection mfr Automatic thermal protection protection
Factory setting operating speed mfr 1900 RPM
Hour meter mfr 1 hour
Conformance standards mfr 61010 marked.
Installation mfr 2
Emissions and Immunity mfr 61236 device
Safety mfr 61010 Canada
Europe mfr 10 m
Denmark mfr 10 m
Switzerland mfr 10 m
UK/Ireland mfr 13 m
India mfr 10 m
Israel mfr 10 m
Japan mfr 12 m
North America mfr 15 m
IDP-10 Exchange Pump mfr 3810
IDP-10 Exchange with Inlet Valve mfr 3810
IDP-10 Full Pump Rebuild mfr 3810 RP
IDP-10 Full Pump (with Inlet Valve) Rebuild mfr 3810 RP
Description of the IDP-10 Dry Scroll Vacuum Pump mfr 184
Technical Specifications mfr 185
IDP-10 Dry Scroll Vacuum Pump Outline Drawing mfr 187
Electrical Connections mfr 189
Power Cord mfr 189
Grounding Instructions mfr 189
Extension Cords mfr 190
Mechanical Connections mfr 191
Isolation Valve mfr 191
Pump Inlet mfr 191
Pump Exhaust mfr 191
Gas Ballast mfr 192
Pump Operations mfr 193
Protection Features mfr 194
Troubleshooting mfr 195
Kits and Service Options mfr 196
Tip Seal Replacement mfr 197
Minimum clearance around equipment mfr 60 cm
Maximum speed at 10 VDC mfr 1900 RPM
Default speed (no input / below 1 V) mfr 1800 tr/min
Pin 1 input impedance mfr 10 Kohm
Recommended potentiometer mfr 5 Kohm
Manual lifting threshold mfr 20 kg
Dimensions, Weight & Mechanical Design
Weight 24.7 kg
Mounting of the IDP-10 Scroll Vacuum Pump mfr 193
Packaged weight mfr 28.1 kg
Electrical, Power & Drive Specifications
Inrush current 191 ampere
Maximum inlet pressure at nominal voltage mfr 1.0 atmosphere
Operating voltages mfr 1 VAC
Fuse type mfr 5 replaceable
Pin 7 reference voltage mfr 10 Vcc
Interfaces, Connections & Control
Inlet flange 25
Interface dimensions mfr 297 mm
Operating Conditions, Environment & Protection
Cooling System Air-cooled
Storage temperature 20 °C
Altitude mfr 2000 m
Ambient Relative Humidity mfr 5 condensing
Standards, Safety & General Data
Noise Level (per ISO 11201) mfr 53 dB
Vibration level at inlet (per ISO 10816-1) mfr 1 mm/s

Description

Construction and Series Positioning

The IDP-10 is a dry scroll pump available in two configurations: the standard model X3810-64000 and the X3810-64010 with an integral inlet valve. The inlet valve opens 30 seconds after start and closes immediately on power loss, providing system isolation. The pump weighs 24.7 kg and uses air cooling. It is not intended for corrosive, explosive, or particulate-forming gases.

Performance and Operating Behavior

Peak pumping speed is 170 L/min (10.2 m³/h), and ultimate pressure is 1.5 × 10⁻² Torr (2.0 × 10⁻² mbar). The maximum recommended continuous inlet pressure is 300 Torr (400 mbar), and the maximum inlet pressure at nominal voltage is 1.0 atm. The maximum outlet pressure is 115 psig. The operating speed is adjustable from 1350 to 1900 RPM; the factory setting is 1900 RPM, with an optional 1800 RPM setting. Under heavy load, overload control reduces motor speed to 1740 RPM after a time delay. Overtemperature protection shuts off the motor at approximately 85 °C while the fan continues running; the pump restarts automatically when cooled. Water vapor handling capacity is 120 g/hr with both gas ballast ports open. Noise level is 53 ± 2 dB(A) per ISO 11201.

Electrical, Control, and Installation

The pump operates on 1-phase 100–120 VAC 50/60 Hz (9 A typical) or 200–240 VAC 50/60 Hz (5 A typical). Inrush current can reach 50 A at 100–127 V, 60 Hz, or 49 A at 200 V, 50 Hz. A DSUB9 connector provides remote on/off (switch closure to ground or logic LO on pin 2) and remote speed control (0–10 VDC input; 1900 RPM at 10 V, 1350 RPM at 6.9 V; input impedance 10 kΩ). The slide switch must be in the 0 position for remote operation. Gas ballast port 1 is for moderate water vapor loads; port 2 is for high loads. For dry nitrogen gas ballast, use a 1/4 NPT connection at approximately 5 L/min. The pump requires at least 60 cm (24 in) of clear space on all sides for ventilation and access. Mounting holes accept four 1/4" or M6 through holes; rubber feet are secured by M3 hardware. The fuse (M5 × 20 mm, 10 A, Slo-Blo) is not user replaceable.

Integration & Operating Notes

Electrical
Circuit must withstand 50 A inrush current

The pump draws up to 50 A inrush at 100–127 V, 60 Hz (49 A at 200 V, 50 Hz). Select circuit breakers, wiring, and connectors rated for this starting current to avoid nuisance trips or voltage drop during startup.

User Manual, Inrush Current, p.191
Control & I/O
Remote control requires slide switch in 0 position

For remote on/off via the DSUB9 connector, the front-panel slide switch must be set to 0. Use a shielded cable and ferrite bead to maintain EMC compliance. The remote speed input expects 0–10 VDC into 10 kΩ.

User Manual, Description of Pin Functions, p.177 and Pump Operations, p.193
Process Limits
Continuous inlet pressure must stay below 300 Torr

Operating continuously above 300 Torr (400 mbar) inlet pressure can overload the pump. The maximum inlet pressure at nominal voltage is 1.0 atm, but continuous operation should remain within the recommended limit to avoid overheating and shortened tip seal life.

User Manual, Technical Specifications, p.185
Cooling & Media
Maintain 60 cm clearance on all sides for air cooling

The pump is air-cooled and requires at least 60 cm (24 in) of free space on all sides for ventilation and access. Insufficient clearance can cause overtemperature shutdown at approximately 85 °C and reduced service life.

User Manual, Environmental Conditions, p.181
Maintenance
Fuse is not user replaceable; contact service

The 10 A Slo-Blo fuse (M5 × 20 mm, Littelfuse 215010 or equivalent) is internal and must not be replaced by the user. If the pump fails to power on, contact Agilent service rather than attempting fuse replacement.

User Manual, Technical Specifications, p.186
Process Limits
Water vapor load limited to 120 g/hr with both ballast ports open

Exceeding 120 g/hr of water vapor can shorten tip seal life. Use gas ballast port 1 for moderate loads and port 2 for high loads; when both ports are open, install the sintered filter plug in port 2 to prevent debris ingress.

User Manual, Gas Ballast, p.192
Mounting
Four through holes accept 1/4" or M6 hardware

The base has four through holes for 1/4" or M6 bolts. Rubber feet are secured by M3 hardware and must be removed before rigid mounting. Ensure the mounting surface is level and can support the 24.7 kg weight.

User Manual, Outline Drawing, p.188

Selection & Integration Q&A

When should I choose the IDP-10 with integral inlet valve (X3810-64010) instead of the standard model (X3810-64000)?

Choose the X3810-64010 when the system requires automatic isolation on power loss: its inlet valve closes immediately on power loss and opens 30 seconds after start. The standard X3810-64000 lacks this valve and is suitable when external isolation is provided or not needed.

What power supply and circuit protection are required for the IDP-10?

The IDP-10 operates on 1-phase 100–120 VAC 50/60 Hz (9 A typical) or 200–240 VAC 50/60 Hz (5 A typical). Circuit breakers and wiring must handle inrush currents up to 50 A at 100–127 V, 60 Hz. Use a grounded outlet and a power cord rated for the current; extension cords must be at least 16-gage and no longer than 25 ft (7.6 m).

What are the ambient and installation requirements for the IDP-10?

Install the pump indoors at altitudes up to 2000 m, with ambient temperatures between 5 °C and a maximum not exceeding the pump's rating (refer to manual), and relative humidity 5–95% non-condensing. Leave at least 60 cm (24 in) clearance on all sides for air cooling and access. The pump is air-cooled and must not be enclosed without ventilation.

What maintenance items and accessories are not included with the base pump?

The base delivery does not include power cord, inlet and exhaust fittings, gas ballast filter plug, or remote-control cable. Tip seal replacement kits and service options are available separately. The fuse is not user replaceable; contact Agilent service for internal fuse replacement. Regular maintenance includes exterior cleaning and tip seal replacement per the manual.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
IDP-10 Dry Scroll Vacuum Pump User Manual
Manual ·206 pp.