HiPace 10 Neo — instruction manual with tc 80

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiPace 10 Neo turbopump operating instructions (54 pages, EN). Single variant covered: HiPace 10 Neo, DN 40 ISO-KF high vacuum flange, with integrated electronic drive unit TC 80. Companion document: TC 80 standard operating instructions PT 0659 BN. Delivery includes pump with TC 80, blank flange DN 40 ISO-KF, 2x C-clamp DN 40, sealing plug for fore-vacuum connection, and operating instructions.

Section-by-Section Technical Breakdown

pp. 10-16
Safety, limits of use and improper use

Risk assessment per Machinery Directive 2006/42/EC Annex I and EN ISO 12100. Key hazards: electric shock (power pack must meet double isolation per IEC 61010-1/60950-1/62368-1), cuts from sharp rotor parts at the open HV flange, pump break-away torques per ISO 27892, overpressure destruction, burns, and toxic media contamination. Proper use restricted to vacuum generation of dry, inert gases indoors with a suitable backing pump.

  • No built-in mains disconnection device: install one per SEMI-S2 with a circuit breaker rated ≥10,000 A interruption
  • Never exceed 1500 hPa (absolute) at suction side, venting or sealing gas connection
  • Mechanical installation always before electrical connection; never start with open vacuum connections
  • Improper use includes pumping explosive/corrosive media, condensing vapors, fluids, dust, use in explosive atmospheres or ionizing radiation
  • Mains plug removal and rotor standstill (f = 0) required before any service work
pp. 14
Permissible ambient conditions

Table 3 defines environmental limits: indoor weatherproof installation, 5-35 °C ambient, altitude to 5000 m, IP20, overvoltage category II, pollution degree 2. Exceeding rotor temperature triggers automatic drive power reduction then shutdown (redundant temperature monitoring).

  • Ambient temperature 5 °C to 35 °C; reduced backing pressure/gas throughput allows higher ambient temperatures
  • Max rotor temperature 110 °C
  • Max surrounding magnetic field 3 mT
  • Max irradiated thermal output 0.6 W
  • Rel. humidity max 80% at T<31 °C, up to 50% at T<40 °C
pp. 20-26
Installation: high vacuum side, fore-vacuum side, accessories, electrical supply

Covers counter-flange dimensioning per ISO 27892 burst tests, earthquake securing, mesh screens, mounting orientations, DN 40 ISO-KF connection, fore-vacuum hose connection, AccessLink/X3 accessory hookup, grounding (M4 thread) and power supply connection. Rotor burst safety measures and flange loading limits are specified numerically.

  • Counter flange: max burst torque 80 Nm, max axial HV flange load 200 N (20 kg, no one-sided load), flatness ±0.05 mm, flange material ≥170 N/mm² with screw engagement 2.5x d (≥270 N/mm² allows 1.5x d)
  • C-clamp hexagon socket screws tightened evenly to 2 Nm with calibrated torque wrench (factor ≤1.6), Allen key WAF 5
  • Mesh screen DN 40 pumping speed loss: H2 6%, He 9%, N2 17%, Ar 18%
  • With oil-sealed backing pumps: fore-vacuum connection aligned downwards, deviation ±25°, all orientations allowed with dry compressing pumps
  • Fore-vacuum line cross section ≥ nominal diameter of fore-vacuum connection; prevent operating fluid/condensate backflow
  • Functional earth via M4 internal thread; power pack requires double isolation per IEC 61010-1/60950-1/62368-1
  • Connection cable with blocking diode prevents power feed-back after pump switch-off; automatic ramp-up via bridges on X3 pins 2, 5, 7
pp. 27-31
Operation: commissioning, operating modes, monitoring, venting

Parameters stored in TC 80, configurable via RS-485 with Pfeiffer protocol ([P:xxx]). Gas mode must be set via [P:027]; gases with molecular mass >80 require consultation with Pfeiffer. Multiple operating modes (standalone with bridge cable, via X3, RS-485 control unit/PC, E74, field bus). LEDs indicate operating mode; temperature monitoring throttles and then shuts down the drive.

  • Factory settings: GasMode [P:027]=0 (heavy gases), AccA1 [P:035]=0 (fan continuous), AccB1 [P:036]=1 (venting valve), run-up time [P:700]=8 min, speed switch point 1 [P:701]=80%, setpoint speed mode [P:707]=65%, power set value [P:708]=100%
  • Delayed venting factory: VentSpd [P:720]=50%, VentTime [P:721]=3600 s
  • Automatic start when X3 pins 2/5/7 bridged and supply voltage applied
  • Simultaneous high drive power, heat radiation, and magnetic fields can overheat the rotor — consult Pfeiffer before combining loads
pp. 32-39
Maintenance: operating fluid reservoirs, electronic drive unit, speed verification

Maintenance levels assigned to customer or Pfeiffer service. Covers replacement of the two operating fluid reservoirs (bearing side 1 and side 2), replacement of the TC 80 electronic drive unit, and confirmation of nominal rotation speed (Table 10 lists characteristic nominal speeds; rating plate 90000 1/min / 1500 Hz).

  • Two operating fluid reservoirs, one per bearing side, replaceable per illustrated procedures
  • Electronic drive unit TC 80 replaceable as a unit
  • Speed specification confirmation procedure with characteristic nominal rotation speeds in Tbl. 10
pp. 40-41
Decommissioning, recycling and disposal

Shutting down for longer periods and recommissioning procedures; general disposal rules and turbopump-specific disposal, with decontamination requirements for toxic media contamination.

  • Decontaminate parts exposed to toxic process media before disposal
  • Observe operating fluid safety data sheet when disposing of fluid reservoirs
pp. 42-44
Malfunctions and service solutions

Troubleshooting table (Tbl. 11) for turbopump malfunctions and description of Pfeiffer Vacuum service offerings.

  • Troubleshooting table Tbl. 11 on p. 43
  • Pfeiffer Vacuum service solutions described in section 11
pp. 46-48
Spare parts and accessories

Spare parts list with figure (Tbl. 12/Fig. 15) and accessory ordering information (Tbl. 13): connection cables, power supply packs (TPS), control units, mesh screen centering rings, venting valve etc. Accessories connect via AccessLink sockets or X3 with suitable cables/adapters; configuration via RS-485.

  • Connection cable variants: with RS-485 + bridges (auto ramp-up pins 2,5,7); with RS-485 + M8 accessory connector + blocking diode; bridge-only variants
  • Accessory output configured via RS-485 using control unit or PC
pp. 49-51
Technical data, materials and dimensions

Section 14 provides general data (Tbl. 16 HiPace 10 Neo), unit conversion tables for pressure and gas throughput, materials in contact with process media (Tbl. 17), and dimension drawing for HiPace 10 Neo | TC 80 | DN 40 ISO-K (Fig. 16).

  • Rating/nominal speed: 90000 1/min (1500 Hz), mass 1.3 kg, IP20, protection class III
  • Materials in contact with media listed in Tbl. 17
  • Dimension drawing for DN 40 ISO-K variant on p. 51
pp. 52-53
EC and UK Declarations of Conformity

EC Declaration of Conformity (p. 52) and UK Declaration of Conformity (p. 53) are included in the document.

  • Conformity declared per relevant European Directives; invalid if product modified or non-original equipment added

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Rated rotor speed 90000 1/min (1500 Hz); pump mass 1.3 kg; protection degree IP20 p.8 ↗ Compact pump for light-duty mounting; verify counter flange load capacity and IP rating for the installation environment HiPace 10 Neo
Max permissible axial load on high vacuum flange is 200 N (20 kg equivalent), one-sided load not permitted; max burst torque 80 Nm per ISO 27892 p.21 ↗ Superstructures on the HV flange must stay ≤20 kg total; account for burst torque in flange design HiPace 10 Neo
Counter flange flatness tolerance ±0.05 mm; flange material ≥170 N/mm² needs screw engagement 2.5x d, ≥270 N/mm² allows 1.5x d p.21 ↗ Dimension customer counter flanges to these values to avoid housing stress, leaks, and running degradation customer-specific HV connection
C-clamp (DN 40 ISO-KF) hexagon socket screws tightened evenly to 2 Nm with torque wrench tightening factor ≤1.6, Allen key WAF 5 p.22 ↗ Prevent over- or under-torquing of the flange connection HiPace 10 Neo HV connection
Mesh screen DN 40 reduces pumping speed: H2 6%, He 9%, N2 17%, Ar 18% p.22 ↗ Include conductance loss in effective pumping speed calculations when foreign-particle protection is used DN 40 ISO-KF with mesh screen centering ring
Permissible inlet pressure max 1500 hPa absolute at suction side and venting/sealing gas connection p.12 ↗ Overpressure above this destroys the pump; ensure process cannot impose higher pressures whole series
Ambient temperature 5 °C to 35 °C; max rotor temperature 110 °C; excess temperature first reduces then cuts drive output p.14 ↗ Cooling and heat shielding must maintain these limits; lower loads permit higher ambient temperatures HiPace 10 Neo (convection/optional air cooling)
Max surrounding magnetic field 3 mT and max irradiated thermal output 0.6 W p.20 ↗ Install magnetic and thermal shielding if process environment exceeds these values whole series
Installation altitude max 5000 m; air pressure 770-1060 hPa; humidity max 80% at <31 °C, 50% at <40 °C; overvoltage category II, pollution degree 2, protection class III p.14 ↗ Site selection and electrical supply design must respect these environmental classes whole series
No integrated mains disconnection device — a mains disconnection device per SEMI-S2 and a circuit breaker with ≥10,000 A interruption rating must be installed p.11 ↗ Mandatory external disconnect/breaker in the system design pump + TC 80 installation
Power supply pack must have double isolation between mains input and output per IEC 61010-1/IEC 60950-1/IEC 62368-1 p.25 ↗ Only approved/original (e.g., TPS) power packs or equivalent certified units may be used TC 80 supply
Automatic pump start occurs when X3 pins 2, 5 and 7 are bridged and supply voltage is applied p.26 ↗ Do not connect mating plugs before mechanical installation is complete — risk of unexpected rotor start operation via connection cables with bridges
Functional earth connection via M4 internal thread on the pump; grounding cable recommended for discharge of applicative interferences p.25 ↗ Wire FE connection per local provisions during electrical installation HiPace 10 Neo
Gas mode set via parameter [P:027]; factory setting 0 = heavy gases; gases with molecular mass >80 require prior consultation with Pfeiffer Vacuum p.27 ↗ Wrong gas mode can destroy the pump — verify [P:027] matches the process gas TC 80 configuration
Factory parameter set: run-up time [P:700]=8 min, speed switch point 1 [P:701]=80%, setpoint speed [P:707]=65%, power set value [P:708]=100%, delayed venting [P:720]=50%, [P:721]=3600 s, accessory A1=fan, B1=venting valve p.27 ↗ Baseline configuration for commissioning; adjust via RS-485/Pfeiffer protocol as required TC 80 as delivered
Mounting orientation unrestricted with dry compressing backing pumps; with oil-sealed pumps, fore-vacuum connection must point downwards (deviation ±25°) to prevent oil backflow p.22 ↗ Select orientation and backing pump type jointly; support piping so no forces act on the pump HiPace Neo series
Fore-vacuum line cross section must be at least equal to the nominal diameter of the fore-vacuum connection; measures against fluid/condensate backflow required p.23 ↗ Size backing lines accordingly and use hose on the quick-fitting coupling fore-vacuum installation
Pump is ball-bearing mounted with one ball bearing per shaft end, lubricated by two replaceable operating fluid reservoirs (bearing side 1 and 2) p.17 ↗ Plan maintenance access for reservoir replacement per intervals; observe fluid safety data sheet HiPace 10 Neo
Earthquake/external vibration protection: all flange and customer-side safety connections must absorb resulting forces; secure vacuum chamber against displacement and tipping p.21 ↗ Add safety connections in seismic or vibrating installations whole series
Proper use limited to evacuation of dry, inert gases indoors; improper uses (corrosive, explosive, condensing media, dust, fluids, ionizing radiation) void warranty/liability p.15 ↗ Qualify process compatibility before deployment; decontaminate before service/disposal if toxic media were pumped whole series