Pfeiffer Vacuum HiPace 300 with TC 110
Pfeiffer Vacuum ·Made in Germany · In Active Production

Pfeiffer Vacuum HiPace 300 with TC 110

Pfeiffer Vacuum HiPace 300 with TC 110 — Turbomolecular Vacuum Pumps with Pumping speed 936 m³/h.

The Pfeiffer Vacuum HiPace 300 is a compact turbomolecular pump driven by the TC 110 converter, delivering 260 l/s (N₂), 255 l/s (Ar, He) and 220 l/s (H₂) with a DN 100 ISO-K/ISO-F/CF-F inlet. Positioned mid-series between the 13 l/s HiPace 10 Neo and the 1250–1450 l/s HiPace 1200–1800 family, it targets leak detection, lab vacuum and semiconductor process tools.

Pumping speed
936 m³/h

Direct drop-in replacements

Agilent Technologies
Turbo V 1001
63%
match
⚖️ Compare
Agilent Technologies
Turbo V 551
63%
match
⚖️ Compare
Agilent Technologies
Turbo V 701
63%
match
⚖️ Compare
Agilent Technologies
TwisTorr 84 FS
47%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Pumping speed 260.0 l/s (N₂)
Ar: 255.0 l/s · He: 255.0 l/s · H₂: 220.0 l/s
Compression ratio 1*10^(11) (N₂)
Ar: 1*10^(11) · He: 1*10^(8) · H₂: 9*10^(5)
Fore-vacuum max 15.0 hPa (N₂)
Gas throughput 2.0 hPa l/s (Ar @ final rotational speed)
He @ final rotational speed: 8.0 hPa l/s · H₂ @ final rotational speed: 14.0 hPa l/s · N₂ @ final rotational speed: 5.0 hPa l/s
Run-up time 3.5 min
Communication interfaces optional mfr Profibus, Devicenet,Semi E74
Dimensions, Weight & Mechanical Design
Weight 6.2 kg
Mounting orientation mfr arbitrary
Interfaces, Connections & Control
Inlet flange DN 100 ISO-K/ISO-F/CF-F
Outlet flange DN 16 ISO-KF /G 1/4"
Operating Conditions, Environment & Protection
Noise level 50.0 dBA
Standards, Safety & General Data
Communication interfaces standard mfr RS-485, remote

Description

Architecture and positioning within the Hipace series

Behavior under load and across gases

Installation, connection and control

Integration & Operating Notes

Process Limits
Keep foreline pressure below 15 hPa under full gas load

The pump loses its compression function if backing pressure at the DN 16 outlet exceeds 15 hPa (N₂) at maximum throughput. Size the backing pump for the worst-case process gas load, especially for hydrogen processes where the low H₂ compression ratio (9·10⁵) already makes ultimate pressure forevacuum-dependent.

Factory Specification Sheet, HiPace 300 with TC 110
Process Limits
Respect per-gas continuous throughput ceilings

At final rotational speed the pump tolerates 14 hPa·l/s (H₂), 8 hPa·l/s (He), 5 hPa·l/s (N₂) and 2 hPa·l/s (Ar). Process recipes with pulsed or continuous gas inlet must be checked against these limits, as exceeding them risks rotor instability and loss of pumping action in the molecular-flow range.

Factory Specification Sheet, HiPace 300 with TC 110
Mounting
Any mounting orientation, but plan flange and venting hardware

The pump can be installed in any orientation, simplifying compact chamber layouts. Chamber interface is DN 100 ISO-K/ISO-F/CF-F; use the optional mounting kit for standardized installation, and specify the safety valve where inlet sealing against oil backstreaming on shutdown is required. Weight is only 6.2 kg, so flange support is normally sufficient.

HiPace 300 product leaflet, Interfaces and Mounting, p.3
Control & I/O
Default control via RS-485/remote; fieldbus only as option

The TC 110 converter offers RS-485 and hardwired remote as standard. Profibus, DeviceNet and SEMI E74 interfaces are optional variants — confirm the correct converter configuration at ordering. Sequence the process start so gas load is applied only after the 3.5 min run-up to final speed.

HiPace 300 product leaflet, Interfaces and Mounting, p.3
Maintenance
Plan venting protection for unattended shutdowns

Hybrid bearings with dedicated lubrication give a long bearing service life, but sudden power loss must be handled by venting to protect pump and chamber. Specify the power-failure venting unit for systems operating unattended; standard venting accessories are available for controlled venting procedures.

HiPace 300 product leaflet, Accessories, p.1
Electrical
Cables and power supply kit are not in base delivery

The base pump with TC 110 does not include mains/connecting cables, the power supply kit with display unit, or fieldbus hardware — these are ordered separately as accessories. Verify the complete electrical accessory set before commissioning to avoid delays at installation.

HiPace 300 product leaflet, Accessories, p.1

Selection & Integration Q&A

When should I choose the HiPace 300 with TC 110 instead of a HiPace 1200/1500/1800 or the HiPace 10 Neo?

Choose the HiPace 300 for ~260 l/s (N₂) needs in leak detection, load-lock or lab/process applications with compact 6.2 kg footprint. Step up to HiPace 1200–1800 (1250–1450 l/s, TC 1200) for high-throughput chambers; pick HiPace 10 Neo (13 l/s) only for small analyzers or ion pumps.

What backing pump and foreline hardware must I pair with this pump for hydrogen-containing processes?

The foreline must keep pressure below 15 hPa (N₂) at full gas load; because the H₂ compression ratio is only 9·10⁵, ultimate H₂ pressure depends heavily on backing quality. Use a DN 16 ISO-KF or G 1/4" foreline sized for up to 14 hPa·l/s H₂ throughput, with a backing pump holding low foreline pressure.

Are there mounting, vibration or duty restrictions I should design around?

Mounting orientation is arbitrary. Laser-balanced rotor keeps noise at 50 dBA and vibration low, so no enclosure is normally needed; add the optional vibration damper for vibration-sensitive instruments (electron optics, surface analysis). Allow 3.5 min run-up before applying process gas load.

What must be ordered besides the pump itself, and how is it controlled?

Not included in base delivery: mains/connecting cables, power supply kit with display unit, mounting kit, safety valve, venting accessories, power-failure venting unit, vibration damper and fieldbus hardware. Control is via RS-485/remote standard; Profibus, DeviceNet or SEMI E74 must be specified as options.

Markets & Applications

Automotive and Transport
  • 3D printing
  • Leak Detection and EoL Testing
Battery
  • Leak Detection and EoL Testing (Master)
Chemical
  • Laboratory Vacuum
Medical
  • Lab Vacuum
  • Mass spectromer
Power Generation
  • Green Energy Generation
R&D
  • Cryogenics
  • High Energy Physics
  • mass spectrometry
  • Ultra-High Vacuum
Scientific Instruments
  • Surface Analysis
Semiconductor
  • ALD
  • Inspection & Metrology
  • Ion Implantation
  • Light deposition

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
HiPace 300 with TC 110 — product leaflet
Datasheet ·512,5 КБ ·3 pp.
View PDF