HiPace 1800 with TC 1200 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Operating instructions PT 0242 BEN/F (1603) for the Pfeiffer Vacuum HiPace 1800 turbopump with integrated electronic drive unit TC 1200. Variants covered: HiPace 1800 (standard), HiPace 1800 C (corrosive gas version), HiPace 1800 U (upside-down installation orientation), HiPace 1800 U C (upside-down + corrosive gases). TC 1200 connection panel options: standard, PB (Profibus), E74 (SEMI E74), DN (DeviceNet).

Section-by-Section Technical Breakdown

pp. 9-12
Product Description and Identification

The HiPace 1800 is a hybrid-bearing turbopump (permanent magnetic bearing on high-vacuum side, ceramic ball bearing on fore-vacuum side) driven by a TC 1200 electronic drive unit with integrated power supply. Nameplate codes identify model, properties (U = upside-down, C = corrosive, P/M/T/E/H/I optional properties), pumping speed class, oil type, S(N2) and nominal speed. Scope of delivery includes drive unit, flange covers, mating plugs, sealing gas valve, 50 ml operating fluid with syringe, cooling-water nozzles, and eye bolts.

  • HV flange: DN 200 ISO-K or DN 200 ISO-F, aluminium
  • Certified to CAN/CSA-C22.2 No. 61010-1, 2nd edition; TUVdotCOM-ID 0000021320
  • Drive: electronic drive unit TC 1200; cooling: water cooling
  • Nameplate example (HiPace 300 shown): S(N2), nominal speed and frequency, weight, oil type listed
pp. 12
Range of Application (Ambient Conditions)

Specifies installation location and permissible ambient conditions: indoors, IP 54, +5 °C to +40 °C, humidity limits, magnetic field ≤ 7 mT, altitude ≤ 2000 m. Connection voltage 100-120/200-240 V AC ±10 %. Above the maximum operating temperature the TC 1200 first reduces drive power, then switches off.

  • Ambient temperature +5 °C to +40 °C; relative humidity max. 80 % at ≤31 °C, max. 50 % at ≤40 °C
  • Atmospheric pressure 750-1060 hPa; installation altitude ≤ 2000 m; degree of pollution 2; overvoltage category II
  • Permissible surrounding magnetic field ≤ 7 mT (shielding required above)
  • Connection voltage 100-120/200-240 V AC (± 10 %)
pp. 13-17
Installation, Mounting and Flange Connection

Covers set-up conditions, mounting orientations, high-vacuum flange attachment (ISO-K with 24 claw clamps at 5/15/25 ±2 Nm; ISO-F with 12 screws/studs at 10/20/38 ±3 Nm in three crosswise steps), and precautions against rotor-blockage torques up to 16,000 Nm. Horizontal mounting requires the fore-vacuum flange pointing downward within ±25°. Counter-flange evenness must be ≤ 0.05 mm; axial HV flange load ≤ 2000 N.

  • Sudden rotor blocking can generate torques up to 16,000 Nm — only Pfeiffer Vacuum mounting kits allowed
  • ISO-K DN 200: 24 claw clamps, crosswise tightening 5, 15, 25 ±2 Nm
  • ISO-F DN 200: 12 hex/stud screws, crosswise tightening 10, 20, 38 ±3 Nm; flange tensile strength ≥ 270 N/mm² (hex-in-thread) or ≥ 170 N/mm² general
  • Counter-flange flatness: max. 0.05 mm over whole surface
  • Max axial load on HV flange: 2000 N (200 kg), no one-sided loads; attachment at pump bottom prohibited
  • Splinter shield DN 200 reduces N2 pumping speed 26 %; protection screen DN 200 reduces N2 8 %
  • Horizontal mounting: fore-vacuum flange vertically downward ±25°; vibration damper max temperature 100 °C
pp. 17-18
Filling Up the Operating Fluid

The pump is delivered without operating fluid; fill only after on-site installation, using the supplied 50 ml with filling syringe. Fill via the higher-positioned filler screw (8.23a), never via the drain screw (8.23b); observe O-ring 8.35. Capacity depends on orientation: 40 ml vertical, 50 ml horizontal.

  • Operating fluid capacity: 40 ml (vertical mounting), 50 ml (horizontal mounting)
  • Never heat synthetic fluids (e.g. F3) above 300 °C — toxic vapours; keep away from tobacco products
  • Type and amount of operating fluid: see rating plate
pp. 18-20
Electrical, Remote and Accessory Connections

Mains connection via HAN3A socket (Pin 1 = Phase L, Pin 2 = Neutral, PE = protective earth); mains cable ordered separately. The supplied mating plug bridges remote pins 1/2/3/4/14 so the pump starts automatically on power-up. Up to 4 accessories connect via preconfigured M12 sockets: A1 sealing gas valve, A2 backing pump, B1 venting valve, B2 casing heating unit.

  • Remote plug bridges Pin 2 (enable venting on), Pin 3 (motor pump on), Pin 4 (pumping station on), Pin 14 (remote access request) to supply voltage Pin 1
  • CAUTION: automatic start occurs on applying supply voltage with the bridged plug
  • Factory presets: sealing gas valve → accessory A1, backing pump → A2, venting valve → B1, casing heating → B2; Y-connector allows two units per connection
  • Back-up pump connected electrically via relay box; default accessory A2
pp. 21-22
Venting Valve and Sealing Gas Connection

Pfeiffer venting valves enable automatic venting at shut-down or power failure; permissible connection pressure max. 1500 hPa abs. Sealing gas is mandatory above 50 % of maximum gas throughput (rotor cooling), with flow rate 17.5-20 sccm; supplied via sealing gas valve (accessory-controlled) or uncontrolled throttle, inlet G 1/8".

  • Venting valve and sealing gas valve connection pressure: max. 1500 hPa absolute
  • Sealing gas flow rate HiPace 1800: 17.5-20 sccm
  • Sealing gas required when operating > 50 % of max gas throughput
pp. 23
Water Cooling

Water cooling is standard with TC 1200; connection via G1/4" internal thread or supplied screw-in nozzles for 7-8 mm hoses (tightening torque max. 15 Nm). Defines cooling water quality (pH 7-9, chloride ≤ 100 mg/kg, hardness ≤ 10 °dH) and fore-line overpressure ≤ 6000 hPa.

  • Hose nozzle G 1/4", hose internal diameter 7-8 mm with hose clamp, nozzle torque max. 15 Nm
  • Water quality: pH 7-9, O2 ≤ 4 mg/kg, chloride ≤ 100 mg/kg, hardness ≤ 10 °dH / 178 ppm CaCO3, KMnO4 consumption ≤ 10 mg/kg, CO2/NH3 undetectable
  • Fore-line overpressure max. 6000 hPa; dirt trap recommended in fore-line
pp. 24-26
Commissioning, Operation Modes and Monitoring

Factory-default TC 1200 parameters are listed (gas mode 0 = heavy gases, run-up time monitoring 8 min, rotation speed switchpoint 80 %, etc.). Operation modes: standalone, remote (26-pin D-sub), RS-485 with DCU/HPU, or fieldbus. LEDs indicate pumping-station status, warnings, malfunctions; drive power depends on mains voltage (700-930 W at 100-120 V, 1200 W at 200-240 V).

  • Factory parameters: [P:027] gas mode 0 (heavy gases); [P:700] max run-up time 8 min; [P:701] speed switchpoint 80 %; [P:707] set speed 65 %; [P:708] power consumption 100 %; [P:720] venting speed 50 %; [P:721] venting time 3600 s
  • Contact Pfeiffer Vacuum before using gases with molecular mass > 80; wrong gas mode can destroy pump
  • Green LED: flashing = station OFF speed ≤ 60 min⁻¹; invers flashing = ON, set speed not attained; blinking = OFF with speed > 60 min⁻¹
  • Delayed venting: valve opens at 50 % nominal speed for 3600 s (also on mains power failure)
pp. 26
Switching Off and Venting

After switch-off, vent the pump to prevent backstreaming from the fore-vacuum side: close fore-vacuum, switch off pump, then vent manually (venting screw ~1 turn) or via venting valve. Keep cooling water on until the rotor stops (< 1 Hz). Rapid venting: max 15 hPa/s pressure rise rate for 20 seconds, valve cross-section adapted to chamber volume.

  • Manual venting: open venting screw about one turn
  • Keep cooling water flowing until rotor speed < 1 Hz
  • Rapid venting: pressure rise rate max. 15 hPa/s, 20 s duration, two-step procedure
pp. 28-30
Maintenance and Replacement

User-serviceable tasks: external cleaning (lint-free cloth + industrial alcohol), operating fluid change, and electronic drive unit replacement. Operating fluid and bearings must be changed at least every 4 years; bearing change via Pfeiffer Service. Fluid is drained via the lower drain screw (8.23b); decontaminate contaminated parts before maintenance and dispose of fluid per regulations.

  • Change operating fluid at least every 4 years; change turbopump bearing at least every 4 years (Pfeiffer Service)
  • Replace electronic drive unit yourself; all other repair work via Pfeiffer Service
  • Never fill via drain screws; only use the higher-positioned filler screw 8.23a and lower drain screw 8.23b
pp. 34-36
Service, Spare Parts and Accessories

The manual references spare parts for HiPace 1800 and accessories (mains cable, venting valve, sealing gas throttle/valve, splinter shields, protection screens, vibration dampers, Y-connectors, relay boxes). Only original Pfeiffer Vacuum accessory and spare parts are permitted.

  • Mains cable must be ordered separately (accessories)
  • Splinter shields and protection screens optional in HV flange mounting kits
  • Related document: Operating instructions 'Electronic drive unit TC 1200', PT 0239 BN
pp. 37-40
Technical Data and Dimensions

Section 13 provides general data plus tables for HiPace 1800 / HiPace 1800 U and HiPace 1800 C / HiPace 1800 U C, and dimension drawings; Declaration of Conformity follows. Detailed numeric data (pumping speed, speeds, cooling water consumption, temperatures) are in the referenced tables.

  • Separate data tables for standard/U versions and corrosive-gas C/U C versions
  • Declaration of conformity included in the document (p. 40)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Sudden rotor blocking can produce torques up to 16,000 Nm able to rip the pump off its mounting. p.13 ↗ Only use Pfeiffer Vacuum original mounting kits on the HV flange; never attach via the pump bottom; design chamber attachment to absorb blockage torque. whole series
ISO-K DN 200 flange connection: 24 claw clamps tightened crosswise in three steps at 5, 15, 25 ±2 Nm. p.16 ↗ Follow the staged crosswise tightening sequence to avoid flange warping and leaks. whole series (DN 200 ISO-K)
ISO-F DN 200 connection: 12 hex/stud screws tightened crosswise at 10, 20, 38 ±3 Nm; flange tensile strength ≥ 270 N/mm² (hex-in-threaded-hole) or ≥ 170 N/mm² minimum in all cases. p.15 ↗ Verify counter-flange material strength before selecting the ISO-F fastening variant. whole series (DN 200 ISO-F)
Counter-flange flatness must be ≤ 0.05 mm over the whole surface. p.16 ↗ Machine/inspect customer counter-flanges; exceeding this can warp the pump casing, causing leaks and poor running. whole series
Operating fluid capacity: 40 ml for vertical mounting, 50 ml for horizontal mounting; pump ships unfilled. p.17 ↗ Fill only after final mounting, using the higher-positioned filler screw; over/underfilling per orientation risks pump destruction. whole series
Operating fluid and bearings must be changed at least every 4 years. p.28 ↗ Include 4-year fluid/bearing service in preventive maintenance planning; contact Pfeiffer Service for extreme-duty interval shortening. whole series
Mains supply 100-120/200-240 V AC ±10 %; drive power 700-930 W at 100-120 V and 1200 W at 200-240 V. p.12 ↗ Size mains supply and breaker per the lower-power figure if wired for 100-120 V; full 1200 W drive power requires 200-240 V. whole series with TC 1200
Ambient conditions: +5 °C to +40 °C, IP 54, humidity max. 80 % (≤31 °C) / 50 % (≤40 °C), altitude ≤ 2000 m, magnetic field ≤ 7 mT, overvoltage category II, pollution degree 2. p.12 ↗ Provide magnetic shielding above 7 mT and climate control outside these limits; indoor, weather-protected installation only. whole series
Max permissible rotor temperature 120 °C; radiated heat input from process must not exceed 24 W. p.13 ↗ Install screening sheets for high-temperature processes exceeding 24 W radiant load (design info from Pfeiffer). whole series
Max axial load on high vacuum flange: 2000 N (200 kg), no one-sided loading; horizontal mounting requires fore-vacuum flange pointing downward within ±25°. p.15 ↗ Limit pipe weight/loads on the inlet flange; verify pump orientation before horizontal installation. whole series
Splinter shield DN 200 reduces pumping speed (N2: -26 %, H2: -10 %); protection screen DN 200 (N2: -8 %, H2: -3 %). p.14 ↗ Factor speed losses into sizing when particle protection is used in the HV flange. whole series, DN 200 accessories
Sealing gas flow rate 17.5-20 sccm; sealing gas mandatory when operating above 50 % of maximum gas throughput; connection pressure max. 1500 hPa abs. p.21 ↗ Install sealing gas supply (valve or throttle) for high-throughput or dirty processes to ensure rotor cooling and protection. whole series (especially C versions)
Cooling water spec: pH 7-9, chloride ≤ 100 mg/kg, hardness ≤ 10 °dH (178 ppm CaCO3), O2 ≤ 4 mg/kg, fore-line overpressure ≤ 6000 hPa; hoses 7-8 mm ID, nozzle torque max. 15 Nm. p.23 ↗ Qualify cooling water quality and use dirt trap in the fore-line; observe torque limit to avoid damaging G1/4" connections. whole series (water cooling standard)
Factory TC 1200 parameters: gas mode 0 (heavy gases) [P:027], max run-up time monitoring 8 min [P:700], speed switchpoint 80 % [P:701], venting time 3600 s [P:721]. p.24 ↗ Verify/adjust gas mode before pumping gases with molecular mass > 80 (contact Pfeiffer first); delayed venting opens valve for 1 h at 50 % nominal speed. whole series with TC 1200
Remote mating plug bridges pins 1/2/3/4/14 on the 26-pin 'remote' D-sub, causing automatic pump start-up when mains power is applied. p.19 ↗ Only connect mains immediately before operation to avoid unintended start; integrate in emergency-off circuit (breaker ≥ 10,000 A interruption rating per SEMI-S2). whole series with TC 1200
Factory-preset accessory assignments: A1 = sealing gas valve, A2 = backing pump, B1 = venting valve, B2 = casing heating unit (max 4 accessories via M12 sockets). p.19 ↗ Connect units to the matching preconfigured port for plug-and-play operation; other accessories require TC 1200 reconfiguration. TC 1200 all panel versions
Rapid venting: two-step, max pressure rise rate 15 hPa/s for 20 s, valve cross-section matched to chamber volume; venting valve connection pressure max. 1500 hPa abs. p.21 ↗ Prevents particle backstreaming and rotor/particle damage; adapt venting orifice to chamber size. whole series
Keep cooling water on after switch-off until rotor speed < 1 Hz; do not operate with open HV flange; annual run-up check of the overspeed safety system required. p.26 ↗ Include these in operating procedures to prevent bearing overheat, injury, and safety-function degradation. whole series
Venting valve opens automatically for 3600 s also on mains power failure; restoring mains aborts the venting procedure. p.26 ↗ Account for this behavior in UPS/power-recovery logic to avoid aborted or repeated venting cycles. whole series with venting valve on accessory B1
Storage -25 °C to +55 °C indoors; transport only with vertical rotor axis and max 60° lifting angle toward pump longitudinal axis via both eye bolts. p.8 ↗ Follow transport/storage rules to avoid operating-fluid spillage and bearing damage; never tilt a fluid-filled pump. whole series