HiPace 300 H with TC 110 — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Operating instructions (DE/EN, 122 pages) for Pfeiffer Vacuum HiPace 300 H turbopump with TC 110 electronic drive unit, 24 V version. Three variants covered: HiPace 300 H DN 100 ISO-K, DN 100 ISO-F, and DN 100 CF-F (stainless steel flange on CF-F). Related applicable documents: TC 110 standard (PT 0204 BN), TC 110 PB (PT 0245 BN), TC 110 E74 (PT 0301 BN), TC 110 RS (PT 0351 BN).
Section-by-Section Technical Breakdown
The HiPace 300 H is a compact turbopump with integrated TC 110 drive unit. 'H' denotes high compression ratios especially for light gases; no sealing gas connection is fitted. Hybrid bearing: permanent magnetic bearing on the HV side, ball bearing with ceramic balls on the fore-vacuum side lubricated by an operating fluid reservoir.
- Cooling options: convection (standard), air cooling (optional), water cooling (optional); drive power is automatically reduced at excessive temperatures
- Drive: TC 110 electronic drive unit
- HV flange materials: aluminium (ISO-K, ISO-F), stainless steel (CF-F)
- Fore-vacuum connection DN 16 ISO-KF; venting screw; multifunction connection X3; LED operating mode display
- Scope of delivery: pump with drive unit, protective covers for HV and fore-vacuum connections, operating instructions
Mechanical installation must always precede electrical connection. Fastening to the mounting plate uses 5 M8 screws at 25 Nm ±10% into a plate ≥2 mm thick with >270 MPa tensile strength. High vacuum side is connected via ISO-K, ISO-F or CF-F flanges with defined cross-wise 3-step tightening sequences; customer counter flange must meet flatness ±0.05 mm and minimum tensile strength requirements.
- Fastening HiPace 300: M8, 5 screws, screw engagement ≥ 1.3 × d, torque 25 Nm ±10%, plate ≥2 mm / >270 MPa, screws strength class ≥8.8 galvanized, washers DIN EN ISO 7090/7092, torque wrench tightening factor ≤1.6
- Max burst torque (rotor shaft breakage, ISO 27892): 2000 Nm; max axial load on HV flange 500 N (50 kg), no one-sided loading
- Flange material min tensile strength: 170 N/mm² at 2.5 × d engagement, 270 N/mm² at 1.5 × d
- Pumping speed reduction with accessories (DN 100): splinter shield H2 5%, He 7%, N2 24%, Ar 24%; protective screen H2 2%, He 2%, N2 10%, Ar 8%
- With oil-sealed backing pumps: mount with fore-vacuum connection pointing vertically downward (±25° max deviation)
- ISO-K to ISO-K bracket screws: cross-wise in 3 steps, 5 / 15 / 25 ±2 Nm (WAF 15)
- ISO-K to ISO-F hexagon head screw with tapped hole: cross-wise in 3 steps, 5 / 10 / 16 ±1 Nm (13 WAF)
- Mating plugs enable automatic pump start-up on power-on: only connect after mechanical installation
- Vibration compensator: vertical passage only, additional securing of pump to HV flange required, contact Pfeiffer Vacuum first
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Turbopump rated speed 60000 1/min (1000 Hz) and mass 6.2 kg per rating plate p.68 ↗ | Rotor kinetic energy is high; fastening and counter flange must absorb up to 2000 Nm burst torque per ISO 27892 | HiPace 300 H all variants |
| Maximum permissible inlet pressure 1500 hPa (absolute) on suction side and venting/sealing gas connection p.72 ↗ | Never expose pump to process overpressure above this limit; pump destruction with ejected debris otherwise possible | whole series |
| Fastening to bottom part: 5 × M8 screws, strength class ≥8.8, engagement ≥1.3 × d, tightening torque 25 Nm ±10%, mounting plate ≥2 mm thick with >270 MPa tensile strength p.82 ↗ | Undersized plates or insufficient screw depth compromise rotor-jam torque absorption | HiPace 300 |
| Maximum permissible axial load on the high vacuum flange is 500 N (50 kg), one-sided loading not permitted p.82 ↗ | Limit superstructural weight on the HV flange; support any attached fittings independently | HiPace 300 |
| Counter flange flatness tolerance ±0.05 mm; flange material minimum tensile strength 170 N/mm² at 2.5 × d and 270 N/mm² at 1.5 × d screw engagement p.83 ↗ | Uneven counter flanges induce housing stress, leakage and degraded running behavior | HiPace 300 custom HV connections |
| Ambient temperature limits: 5–30°C (convection cooling, no gas throughput), 5–35°C (air cooling), 5–40°C (water cooling) p.75 ↗ | Selection of cooling option determines permissible installation environment; higher temperatures require reduced backing pressure or gas throughput | whole series |
| Maximum bakeout temperature at the high vacuum flange is 120°C; maximum rotor temperature 90°C p.75 ↗ | Bakeout above 120°C damages the pump; temperature monitoring reduces then shuts off drive output when limits are exceeded | whole series |
| Maximum permissible surrounding magnetic field 5.5 mT and maximum irradiated thermal output 2.4 W p.75 ↗ | Install magnetic/thermal shielding if the process environment exceeds these values | whole series |
| No mains switch on pump or drive unit; a mains disconnection device per SEMI-S2 with a circuit breaker rated ≥10,000 A interruption capacity is required p.71 ↗ | Customer must integrate an emergency-off circuit and external disconnect for safety compliance | whole series |
| Power supply pack must satisfy double isolation between mains input and output per IEC 61010-1 / IEC 60950-1 / IEC 62368-1 p.71 ↗ | Non-compliant PSUs create electric shock hazard; use original Pfeiffer power supply packs | whole series |
| Pumping speed reduction with splinter shield DN 100: H2 5%, He 7%, N2 24%, Ar 24%; with protective screen DN 100: H2 2%, He 2%, N2 10%, Ar 8% p.83 ↗ | Account for gas-type-dependent speed loss when sizing systems with foreign-matter protection | DN 100 flange variants |
| ISO-K to ISO-K flange: bracket screws tightened cross-wise in 3 steps at 5, 15, 25 ±2 Nm p.85 ↗ | Use calibrated torque wrench (factor ≤1.6), WAF 15; sequence prevents flange distortion | HiPace 300 H DN 100 ISO-K |
| ISO-K to ISO-F flange: hexagon head screws cross-wise in 3 steps at 5, 10, 16 ±1 Nm (13 WAF) p.85 ↗ | Observe minimum flange material tensile strength and screw engagement depth per Tbl. 6 | HiPace 300 H DN 100 ISO-K with ISO-F counter flange |
| With oil-sealed backing pumps the fore-vacuum connection must point vertically downward (±25° max deviation); dry compressing backing pumps allow any orientation p.84 ↗ | Prevents oil backflow into the turbopump from the fore-vacuum side | whole series |
| Bearing system: wear-free permanent magnetic bearing on the HV side, ceramic-ball bearing lubricated by an operating fluid reservoir on the fore-vacuum side p.78 ↗ | Maintenance includes periodic replacement of the operating fluid reservoir (section 7.3) | whole series (hybrid bearing) |
| Mating plugs cause automatic pump start-up when power is applied p.72 ↗ | Connect mating plugs only after mechanical installation to avoid cut injuries from exposed sharp-edged rotor | whole series |
| The 'H' property denotes high compression ratios especially for light gases; this model has no sealing gas connection p.78 ↗ | Suitable for light-gas (H2/He) applications; no sealing gas fitting to plan in layouts | HiPace 300 H |
| Fore-vacuum connection is DN 16 ISO-KF; HV flange options DN 100 ISO-K / ISO-F (aluminium) or DN 100 CF-F (stainless steel) p.78 ↗ | Determines counter flange, centering ring and clamping hardware selection; CF-F variant only for metal-seal applications | HiPace 300 H variants |
| Maximum permissible installation altitude 5000 m; air pressure 530–1060 hPa; protection degree IP44 / UL 50E Type 12 p.75 ↗ | Verify site conditions against limits before installation; do not use outdoors (weatherproof indoor space only) | whole series |
| Pump is sealed ex-factory with a warranty seal; damaging or removing it voids the warranty p.68 ↗ | Do not open the pump; repairs must use authorized service levels (level 1–3) | whole series |