HiPace 300 M with TM 700 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Operating instructions (EN, 62 pages) for Pfeiffer Vacuum HiPace 300 M turbopump with 5-axis active magnetic bearings, covering three variants: DN 100 ISO-K, DN 100 ISO-F and DN 100 CF-F high vacuum flange, each with integrated electronic drive unit TM 700 (48 V DC) and magnetic bearing controller. Applicable documents include TM 700 standard (PT 0329 BN), Profibus (PT 0354 BN), SEMI E74 (PT 0355 BN), DeviceNet (PT 0356 BN) and EtherCAT (PT 0503 BN) drive manuals.

Section-by-Section Technical Breakdown

pp. 15
Limits of Use and Ambient Conditions

Permissible ambient conditions table: installation altitude, humidity, temperatures by cooling type, magnetic field, bakeout and rotor temperature limits.

  • Ambient temperature: 5–30 °C convection (no gas throughput), 5–35 °C air cooling, 5–40 °C water cooling
  • Max. installation altitude 5000 m (with sufficient cooling), air pressure 530–1060 hPa
  • Rel. humidity max 80% at T < 31 °C, up to 50% at T < 40 °C
  • Max. surrounding magnetic field 5.0 mT; max irradiated thermal output 2.4 W
  • Max rotor temperature 110 °C; max bakeout temperature at HV flange 120 °C; protection degree IP54
pp. 18-19
Product Description

HiPace M series uses active 5-axis magnetic bearings making it maintenance-free and hydrocarbon-free. Compact unit with TM 700 electronic drive unit (48 V DC); cooling options are convection, water, or optional air cooling.

  • Active 5-axis magnetic bearing: maintenance-free, no lubricants
  • Pre-assembled venting valve for automatic venting after switch-off or power failure
  • Scope of delivery: pump with drive unit, venting valve, protective caps (HV and fore-vacuum), operating instructions
  • Designation key: M = active magnetic bearing; flange materials aluminium (ISO-K, ISO-F), stainless steel (CF-F)
pp. 21-34
Installation

Covers counter-flange dimensioning requirements, earthquake protection, splinter shields, vibration compensators, mounting orientations, ISO-K/ISO-F/CF flange connections with tightening torques, fore-vacuum and cooling water connections, venting valve, accessories, grounding and electrical connection.

  • Counter flange: max burst torque 3000 Nm, max axial load 500 N (50 kg, no one-sided load), flatness ±0.05 mm, flange material tensile strength ≥170 N/mm² at 2.5×d or 270 N/mm² at 1.5×d screw engagement
  • Splinter shield DN 100 reduces pumping speed: H2 5%, He 7%, N2/Ar 24%; protective screen: H2/He 2%, N2 10%, Ar 8%
  • ISO-K bracket screws torque: 5, 15, 25 ±2 Nm in 3 cross-wise steps; ISO-K/ISO-F screw connections: 5, 10, 16 ±1 Nm
  • With oil-sealed backing pumps: mount horizontally, fore-vacuum connection vertical downwards, deviation max ±25°
  • Cooling water composition and connection requirements per Tbl. 7/8; accessories pre-set on TM 700 per Tbl. 9
  • Mechanical installation must always precede electrical connection; grounding required
pp. 35-40
Operation

Commissioning, operating modes (standalone, remote D-Sub, E74, Pfeiffer control unit, fieldbus), switching on, operation monitoring via LEDs, temperature and safety-bearing monitoring, switching off with electric brake, emergency power supply and venting.

  • Operating voltage 48 V DC via TM 700; factory settings per Tbl. 10
  • LED operating mode display on electronic drive unit (Tbl. 11)
  • Redundant temperature monitoring reduces then shuts off drive power on overheating
  • Delayed venting factory settings per Tbl. 12
  • Venting via pre-assembled venting valve after switch-off or power failure
pp. 41-44
Maintenance

Magnetic bearings make the pump maintenance-free for life; maintenance intervals/responsibilities by level, and replacement procedure for the electronic drive unit TM 700 including speed-specification confirmation.

  • Characteristic nominal rotation speeds per Tbl. 13
  • Electronic drive unit removal/installation procedure (Figs. 19-20)
  • Maintenance levels: 1 trained specialist, 2 technically educated customer / PF service, 3 PF-trained customer / PF service
pp. 45-46
Decommissioning, Recycling and Disposal

Shutting down for longer periods and recommissioning procedure; general disposal information and turbopump disposal (decontamination of toxic-media-exposed parts required).

  • Decontaminate parts exposed to toxic process media before maintenance/disposal
pp. 47-54
Malfunctions, Service and Spare Parts

Troubleshooting table (Tbl. 14), Pfeiffer Vacuum service solutions, spare parts overview for HiPace 300 M (Tbl. 15), and accessory ordering information (Tbl. 16).

  • Only use accessories/spare parts listed in the instructions
  • Malfunction diagnostics via troubleshooting table on p. 48
pp. 55-58
Technical Data and Dimensions

Pressure and gas throughput unit conversion tables, technical data for HiPace 300 M with TM 700 (Tbl. 19), materials in contact with process media (Tbl. 20) and dimensional drawings for DN 100 ISO-K, ISO-F and CF-F versions.

  • Dimensions shown for all three HV flange variants (Figs. 22-24)
  • Process media contact materials listed in Tbl. 20 (incl. FKM seals)
pp. 59-60
EC and UK Declarations of Conformity

Manufacturer's EC Declaration of Conformity and UK Declaration of Conformity enclosed with the manual.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Turbopump uses active 5-axis magnetic bearings — maintenance-free over the entire service life and no hydrocarbon contamination risk on malfunction. p.18 ↗ No lubrication/oil management required; suitable for oil-free, contamination-sensitive applications; eliminates oil_capacity concerns. whole series (HiPace M)
Operating voltage of the electronic drive unit TM 700 is 48 V DC; supply must be via a power supply pack with double isolation per IEC 61010-1/60950-1/62368-1. p.11 ↗ Specify a certified 48 V DC PSU with double isolation; non-compliant supplies risk fatal electric shock and void safety compliance. all variants with TM 700
Pump and drive unit have no mains disconnection device; a mains disconnection device per SEMI-S2 and a circuit breaker with interruption rating ≥ 10,000 A must be installed on-site. p.11 ↗ Panel design must include an external disconnect and adequately rated breaker before commissioning. whole series
Maximum permissible inlet pressure on suction side and venting/sealing gas connection is 1500 hPa absolute. p.12 ↗ Prevent process overpressure from entering the pump; exceedance destroys the turbopump with ejected-object hazard. whole series
Maximum torque occurring on rotor burst is 3000 Nm; maximum permissible axial load on the HV flange is 500 N (50 kg), no one-sided load; counter-flange flatness ±0.05 mm. p.22 ↗ Counter flange and chamber connections must be dimensioned to absorb 3000 Nm burst torque per ISO 27892; superstructure weight limited to 50 kg. HiPace 300 M all flange variants
Counter flange material tensile strength requirement: ≥170 N/mm² at 2.5×d screw engagement or ≥270 N/mm² at 1.5×d. p.22 ↗ Verify flange material and thread engagement depth before machining tapped holes. customer-supplied high vacuum counter flange
ISO-K flange connection with bracket screws: tighten cross-wise in 3 steps at 5, 15, 25 ±2 Nm (torque wrench factor ≤1.6, WAF 15). p.24 ↗ Use calibrated torque wrench and staged cross-wise tightening to avoid housing stress and leaks. DN 100 ISO-K variant
ISO-K to ISO-F screw connections: tighten in 3 steps at 5, 10, 16 ±1 Nm. p.25 ↗ Staged cross-wise tightening with calibrated wrench required for all hex/stud screw variants. DN 100 ISO-K / ISO-F connections
Splinter shield DN 100 reduces pumping speed by up to 24% (N2, Ar); protective screen DN 100 by up to 10% (N2). p.23 ↗ Account for pumping-speed loss in system sizing when foreign-particle protection is used. DN 100 flange variants
With oil-sealed backing pumps the pump must be mounted horizontally with the fore-vacuum connection aligned vertically downwards, permissible deviation ±25°; dry compressing backing pumps permit all orientations. p.24 ↗ Choose mounting orientation based on backing pump type to prevent oil backflow into the turbo. whole series
Ambient temperature limits: 5–30 °C with convection cooling without gas throughput, 5–35 °C with air cooling, 5–40 °C with water cooling; cooling options are convection, water, optional air. p.15 ↗ Select cooling type based on installation environment; higher ambient temperatures require reduced backing pressure or gas throughput. whole series
Maximum permissible rotor temperature 110 °C; maximum bakeout temperature at the high vacuum flange 120 °C. p.15 ↗ Limit bakeout jackets to 120 °C at the flange; drive unit throttles power then shuts off if rotor overtemperature occurs. whole series
Maximum permissible surrounding magnetic field 5.0 mT; maximum irradiated thermal output 2.4 W. p.15 ↗ Install magnetic and thermal shielding if process environment exceeds these values. whole series
Rated rotation speed 60,000 1/min (1000 Hz) per rating plate; mass 13 kg; protection degree IP54 (Type 12 per UL 50E). p.8 ↗ Rating data for mounting design, transport handling and enclosure selection. HiPace 300 M
Before installing a vibration compensator, Pfeiffer Vacuum must be contacted; the pump must be additionally secured to the HV flange and the compensator installed only with vertical passage. p.12 ↗ Vibration compensator use carries pump shearing risk on rotor jamming (ISO 27892 torques); on-site torque compensation measures are mandatory. whole series
Proper use limited to evacuation of dry and inert gases indoors, with a suitable backing pump; improper use includes pumping explosive, corrosive or condensing media, dust, fluids, and operation in potentially explosive areas or ionizing radiation. p.15 ↗ Do not specify this pump for corrosive/hazardous processes; corrosive-gas versions are separate (HiPace C designation). whole series
Mains disconnection before maintenance: switch off, wait until rotation speed f = 0, remove mains plug, secure against restart. p.13 ↗ Rotor must be fully stopped before any work on the HV connection; magnetic bearing rotor can restart unexpectedly. whole series
Removing or damaging the factory warranty seal voids the warranty; modifications by the operator invalidate the EC Declaration of Conformity. p.8 ↗ Do not open the drive unit or modify the pump; system-level conformity must be re-evaluated by the operator after integration. whole series
Operation monitoring includes LED status display, redundant temperature monitoring, safety-bearing wear and balance monitoring; delayed venting has defined factory settings (Tbl. 12). p.36 ↗ Integrate LED/error signals and venting behavior into system PLC logic; venting valve vents automatically on switch-off or power failure. whole series with TM 700
Flange materials: aluminium for DN 100 ISO-K and ISO-F variants, stainless steel for DN 100 CF-F. p.19 ↗ Material choice affects temperature behavior and process compatibility when selecting the variant. variant-specific