HiPace 300 P with TC 400 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Operating instructions (EN/DE, 126 pages) for the Pfeiffer Vacuum HiPace 300 P process-version turbopump with electronic drive unit TC 400. Covers three variants: HiPace 300 P DN 100 ISO-K, DN 100 ISO-F and DN 100 CF-F, all with TC 400. Related TC 400 drive manuals (standard, Profibus, E74, DeviceNet, EtherCAT) are referenced but not included.

Section-by-Section Technical Breakdown

pp. 80-81
Product description

Describes the compact turbopump/TC 400 unit, cooling options, hybrid bearing and drive. The 'P' (process) feature makes pumps insensitive to dust and corrosive gases. Shipment contents and product identification via rating plate are defined.

  • Cooling: water cooling standard, air cooling optional; drive power auto-reduced at excessive temperature
  • Bearing: permanent magnetic bearing on HV side (wear-free), ceramic-ball bearing on fore-vacuum side lubricated by operating fluid reservoir
  • Drive: TC 400 electronic drive unit, 24 V DC or 48 V DC versions
  • Fore-vacuum connection DN 16 ISO-KF; sealing gas and venting connections on housing
  • Flange materials: ISO-K and ISO-F aluminium, CF-F stainless steel
pp. 83-96
Installation

Detailed mechanical installation: fastening to mounting plate, high vacuum flange connections (ISO-K, ISO-F, CF-F), fore-vacuum connection, cooling water, accessories and electrical supply. Includes torque values, counter-flange flatness requirements, splinter-shield pumping-speed losses and mounting orientations. Rotor-jamming torques must be absorbed by operator-side construction.

  • Fastening: M8 screws, 5 screws, ≥1.3×d engagement, 25 Nm ±10%, mounting plate ≥2 mm / >270 MPa
  • Max torque on burst (ISO 27892, theoretical): 2000 Nm; max axial load on HV flange: 500 N (50 kg), no one-sided load
  • Counter flange flatness ±0.05 mm
  • Splinter shield DN 100 reduces pumping speed: H2 5%, He 7%, N2 24%, Ar 24%; protective screen: H2/He 2%, N2 10%, Ar 8%
  • ISO-K bracket screws: crosswise in 3 steps, 5/15/25 ±2 Nm
  • ISO-K to ISO-F hexagon screw (tapped hole): 3 steps 5/10/16 ±1 Nm
  • Oil-sealed backing pumps: fore-vacuum connection aligned vertically downwards, ±25° max deviation
  • Vibration compensator: only after consulting Pfeiffer Vacuum; requires additional HV-flange securing
pp. 97-101
Operation

Commissioning and operating modes (without operating unit, via 'remote' multi-pin socket, E74, Pfeiffer control unit, fieldbus), switching on, operation monitoring via LED and temperature monitoring, switching off and venting. Factory settings of the TC 400 and delayed venting are tabulated.

  • Operating modes: standalone, 'remote' D-Sub, 'E74', Pfeiffer control unit, fieldbus
  • Temperature monitoring is redundant; drive output first reduced, then switched off on over-temperature
  • LED display shows operating mode; venting via venting screw or automatic venting valve
pp. 102-108
Maintenance

Maintenance intervals per checklist, replacement of the operating fluid reservoir and the electronic drive unit TC 400, plus confirmation of the nominal speed specification after drive replacement. Maintenance levels 1–3 with personnel qualification requirements.

  • Replace operating fluid reservoir (lubricates fore-vacuum-side bearing)
  • Replace electronic drive unit TC 400; confirm speed specification afterwards (nominal 60000 1/min / 1000 Hz)
  • Decontaminate toxic-contaminated parts before service; observe operating-fluid safety data sheet
pp. 109-110
Decommissioning, Recycling and Disposal

Shutting down for longer periods and recommissioning; general disposal information and disposal of turbopumps according to applicable regulations, considering possible toxic contamination.

  • Long-term shutdown and recommissioning procedures provided
  • Dispose of turbopumps and operating fluid reservoir per applicable regulations
pp. 111-117
Malfunctions, Service, Spare Parts and Accessories

Troubleshooting table for turbopump malfunctions, Pfeiffer Vacuum service solutions, spare parts list for HiPace 300 P and accessory ordering information including splinter shields/protective screens and TC 400 accessories.

  • Troubleshooting table (Tbl. 16) covers pump malfunctions
  • Spare parts overview for HiPace 300 P (Tbl. 17)
  • Accessories include splinter shields/protective screens DN 100 and other HiPace 300 48 V DC accessories
pp. 118-124
Technical data and dimensions

Technical data tables for the 24 V and 48 V variants, materials in contact with process media, dimension drawings for DN 100 ISO-K / ISO-F / CF-F with TC 400, plus unit conversion tables and Declarations of Conformity (EC and UK).

  • Separate technical data for HiPace 300 P | 24 V and HiPace 300 P | 48 V
  • Materials-in-contact table (Tbl. 24) for process media compatibility
  • Dimension drawings for TC 400 with DN 100 ISO-K, ISO-F, CF-F

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal rotor speed is 60000 1/min (1000 Hz); pump mass 6.4–8.4 kg per rating plate. p.70 ↗ Anchoring and handling design must account for high-speed rotor energy and ~7 kg mass. HiPace 300 P all variants
Max permissible pressure on suction side, venting and sealing gas connections: 1500 hPa absolute. p.74 ↗ Never allow process overpressure above 1500 hPa abs into the pump — risk of destruction. whole series
Fastening to mounting plate: 5× M8 screws, strength class ≥8.8, engagement depth ≥1.3×d, torque 25 Nm ±10%, plate min. 2 mm thick with tensile strength >270 MPa. p.84 ↗ Mounting plate must be dimensioned accordingly; torque must be verified with calibrated torque wrench (factor ≤1.6). HiPace 300
Maximum torque in event of burst (ISO 27892): 2000 Nm; maximum axial load on HV flange: 500 N (50 kg), one-sided load not permitted. p.84 ↗ Operator-side construction must absorb 2000 Nm jamming torque; superstructures limited to 50 kg evenly loaded. HiPace 300
Counter flange flatness requirement: ±0.05 mm. p.85 ↗ Machining/verification of customer counter flange needed to avoid housing stress and leaks. whole series
Splinter shield DN 100 reduces pumping speed: H2 5%, He 7%, N2/Ar 24%; protective screen DN 100: H2/He 2%, N2 10%, Ar 8%. p.85 ↗ Size the pump and backing train with these speed-loss figures for contamination protection. DN 100 variants
ISO-K bracket screw connection: tighten crosswise in 3 steps, 5 / 15 / 25 ±2 Nm. p.87 ↗ Use calibrated torque wrench and staged crosswise tightening for leak-tight, safe flange joint. DN 100 ISO-K
ISO-K to ISO-F hexagon screw in tapped hole: crosswise in 3 steps, 5 / 10 / 16 ±1 Nm; WAF 13. p.87 ↗ Observe flange material tensile strength and screw depth per Tbl. 6 (170 N/mm² at 2.5×d, 270 N/mm² at 1.5×d). DN 100 ISO-K/ISO-F
With oil-sealed backing pumps, fore-vacuum connection must point vertically downwards, deviation max ±25°, with supported piping and no one-sided HV flange load. p.86 ↗ Prevents oil backflow into the turbopump; plan piping layout accordingly. whole series with oil-sealed backing pumps
Ambient conditions: 530–1060 hPa, altitude ≤5000 m, ambient temp 5–30 °C (convection, no gas throughput), 5–35 °C air cooling, 5–40 °C water cooling, permissible IP54 / Type 12, protection class III, overvoltage category II. p.77 ↗ Installation site selection and enclosure design must match these limits; higher ambient allowed only with reduced backing pressure/gas throughput. whole series
Max surrounding magnetic field 5.5 mT; max irradiated thermal output 2.4 W; max rotor temperature 110 °C (Tbl. 2) / 90 °C for HV connection design (Tbl. 6); max bakeout at HV flange 120 °C. p.77 ↗ Provide magnetic and thermal shielding near processes with heaters or magnetic fields; bakeout limited to 120 °C. whole series
Mains disconnection device required per SEMI-S2 with circuit breaker interruption rating ≥10,000 A; pump/drive have no built-in mains switch. p.73 ↗ Integrator must supply compliant disconnection and Emergency-Off circuit integration. whole series with TC 400
Power supply pack must satisfy double isolation between mains input and output per IEC 61010-1 / IEC 60950-1 / IEC 62368-1. p.73 ↗ Only approved or equivalent power packs may be used; non-compliant packs risk fatal electric shock. whole series
Bearing concept: wear-free permanent magnetic bearing (HV side) + ceramic-ball bearing (fore-vacuum side) lubricated by replaceable operating fluid reservoir. p.80 ↗ Fore-vacuum side requires periodic operating fluid reservoir replacement (Section 7.3); HV side is maintenance-free. HiPace 300 P (hybrid bearing)
TC 400 drive available in 24 V DC and 48 V DC versions; separate technical data tables for each. p.81 ↗ Verify supply voltage variant before ordering spares and dimensioning the power pack. TC 400 electronic drive unit
Corrosive gases permitted only with sealing gas; dust- and corrosive-gas-insensitive by virtue of the 'P' process feature. p.77 ↗ For corrosive or dusty processes, connect and configure sealing gas; improper use without it voids warranty. HiPace 300 P
Vibration compensator installation requires prior consultation with Pfeiffer Vacuum and additional securing to the HV flange, due to ISO 27892 break-away risk. p.74 ↗ Do not fit vibration compensators ad hoc; engineer torque compensation on site. whole series
Keep lines and cables away from hot surfaces >70 °C; never operate with open HV connection or with cleaning agent residues. p.76 ↗ Cable routing and cleaning procedures in system design must respect these constraints. whole series
Before maintenance: switch off, wait for f = 0 (complete rotor standstill), disconnect mains plug, secure against restart; only then open HV connection. p.75 ↗ Lockout procedure must include rotor spin-down wait time; hot or coasting rotor is a cutting/severing hazard. whole series
Product certification: UL/CSA 61010-1, TÜV Rheinland Type 12, IPx4 (sticker); Declaration of Conformity invalid if operator modifies product or adds equipment. p.70 ↗ Post-installation, the system integrator must re-evaluate conformity of the overall system per EU directives. whole series