HiPace 300 Plus — instruction manual with tc 400

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiPace 300 Plus turbopump operating instructions (126 pages), covering six variants: DN 100 high-vacuum flange in ISO-K, ISO-F, or CF-F, each with electronic drive unit TC 400 in 24 V DC or 48 V DC versions. 'Plus' designates the low-vibration, low-magnetic-stray-field version.

Section-by-Section Technical Breakdown

pp. 72-79
Safety and Limits of Use

Risk assessment per Machinery Directive 2006/42/EC and EN ISO 12100. Covers transport, installation, operation, maintenance and fault hazards (electric shock, rotor burst torques per ISO 27892, toxic media, cuts from sharp rotor edges). Defines permissible ambient conditions and proper/improper use.

  • Mains disconnection device required per SEMI-S2; circuit breaker with interruption rating ≥ 10,000 A
  • Power supply pack must meet double isolation per IEC 61010-1 / 60950-1 / 62368-1
  • Max. absolute inlet pressure on suction side and venting/sealing gas connection: 1500 hPa
  • Ambient: 530–1060 hPa, altitude ≤ 5000 m, protection class III, overvoltage category II, min. IP54 / Type 12
  • Ambient temperature: 5–30 °C convection cooling (no gas throughput), 5–35 °C air cooling, 5–40 °C water cooling
  • Limits: surrounding magnetic field 5.5 mT, irradiated thermal output 2.4 W, rotor temp 90 °C, bakeout temp at HV flange 120 °C
  • Proper use: vacuum generation only, dry/inert gases, indoor, with suitable backing pump; improper use includes explosive, corrosive, condensing, dusty media and explosive atmospheres
  • Maintenance levels 1–3 defined; Pfeiffer offers training for levels 2 and 3
pp. 80-81
Product Description

Describes the HiPace 300 Plus design, cooling options, hybrid rotor bearing, and TC 400 drive in 24 V or 48 V. Scope of delivery and product identification (rating plate, naming system, features) are given.

  • 'Plus' = particularly low-vibration operation and low magnetized stray field
  • Cooling: water cooling standard, air cooling optional; drive power auto-reduced on excess temperature
  • Hybrid bearing: wear-free permanent magnetic bearing on HV side, ball bearing with ceramic balls on fore-vacuum side, lubricated by operating fluid reservoir
  • HV flange material: aluminium (ISO-K, ISO-F), stainless steel (CF-F)
  • Scope of delivery: pump with TC 400, protective covers for HV and fore-vacuum connections, operating instructions
pp. 83-96
Installation

Mechanical installation must precede electrical connection. Covers fastening to the mounting plate, high vacuum flange connections (ISO-K/ISO-F/CF-F with torque specs), fore-vacuum connection, cooling water connection, accessories and earthing. Safety against rotor-burst torques per ISO 27892 governs all fastening design.

  • Fastening: M8 screws, 5 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, torque 25 Nm ± 10%, mounting plate ≥ 2 mm thick, tensile strength > 270 MPa
  • Counter flange flatness ± 0.05 mm; min. tensile strength 170 N/mm² at 2.5 × d, 270 N/mm² at 1.5 × d engagement depth
  • Max. burst torque (ISO 27892): 2000 Nm; max. axial load on HV flange 500 N (50 kg), one-sided load not permitted
  • Splinter shield DN 100 reduces pumping speed: H2 5%, He 7%, N2/Ar 24%; protective screen DN 100: H2/He 2%, N2 10%, Ar 8%
  • Mounting orientation only with rotor axis vertical, HV flange up; upside-down operation forbidden
  • ISO-K/ISO-K bracket screws torque in 3 steps: 5, 15, 25 ± 2 Nm; ISO-K/ISO-F hex screws: 5, 10, 16 ± 1 Nm; calibrated torque wrench with tightening factor ≤ 1.6
  • Vibration compensator: contact Pfeiffer Vacuum before installing; vertical passage only; additional HV flange securing required
  • Cooling water composition and connection requirements specified (Tbl. 8/9); pre-set accessory connections listed for TC 400 (Tbl. 10)
  • Fore-vacuum connection DN 16 ISO-KF
pp. 97-101
Operation

Commissioning, operating modes (standalone without operating unit, remote D-Sub, E74, Pfeiffer control unit, field bus), switching on, monitoring via LEDs and temperature sensors, switching off and venting. Factory settings of the electronic drive unit and delayed venting defaults are tabulated.

  • Operation via: multi-function 'remote' 26-pole D-Sub socket, E74 connection, Pfeiffer control unit, or field bus
  • LED operating mode display on TC 400 (behavior table, Tbl. 12)
  • Temperature monitoring: drive output first reduced, then shut off on exceeding max. permissible temperature
  • Factory settings for delayed venting tabulated (Tbl. 13); factory drive settings in Tbl. 11
pp. 102-108
Maintenance

Maintenance intervals and checklist, replacement of the operating fluid reservoir (fore-vacuum side bearing lubrication) and replacement of the TC 400 electronic drive unit, including confirmation of the nominal speed specification after replacement.

  • Maintenance intervals per Tbl. 14
  • Characteristic nominal rotation speeds per Tbl. 15 (rating plate: 60,000 1/min, 1000 Hz)
  • Drive unit replacement: remove/install TC 400, then confirm speed specification
pp. 109-110
Decommissioning, Recycling and Disposal

Shutting down for longer periods and recommissioning procedures, plus general disposal information and turbopump-specific disposal (contamination by pumped media must be considered).

  • Decontaminate parts exposed to toxic media before disposal
pp. 111-113
Malfunctions

Troubleshooting table (Tbl. 16) for turbopump faults, plus Pfeiffer Vacuum service solutions overview.

  • Fault causes and remedies compiled in troubleshooting table Tbl. 16
pp. 115-117
Spare Parts and Accessories

Spare parts overview for HiPace 300 Plus (Tbl. 17) and accessory ordering information, including accessories for HiPace 300 | 48 V DC (Tbl. 18/19).

  • Only accessories/spare parts listed in the manual may be used
pp. 118-122
Technical Data and Dimensions

Technical data tables for HiPace 300 Plus 24 V (Tbl. 22) and 48 V (Tbl. 23), materials in contact with process media (Tbl. 24), unit conversion tables, and dimension drawings for DN 100 ISO-K, ISO-F and CF-F configurations.

  • Materials in contact with media tabulated (Tbl. 24)
  • Dimension drawings for TC 400 with DN 100 ISO-K / ISO-F / CF-F (Figs. 24–26)
  • Rating plate data example: 60,000 1/min, 1000 Hz, 7.0 kg

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Fastening to mounting plate: 5 × M8 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, tightening torque 25 Nm ± 10%; mounting plate min. 2 mm thick with tensile strength > 270 MPa. p.84 ↗ Undersized plates or wrong torque risk pump shear-off under rotor burst loads; use calibrated torque wrench (factor ≤ 1.6). HiPace 300 Plus (all variants)
Maximum burst torque per ISO 27892: 2000 Nm; maximum permissible axial load on HV flange: 500 N (50 kg), one-sided load not permitted. p.84 ↗ Customer-side structure must absorb burst torques; no cantilevered loads on the HV flange. HiPace 300 Plus (all variants)
Counter flange flatness tolerance ± 0.05 mm; flange material min. tensile strength 170 N/mm² at 2.5 × d or 270 N/mm² at 1.5 × d screw engagement. p.85 ↗ Non-conforming counter flanges cause housing stress, leakage and degraded running behavior. HiPace 300 Plus (all variants)
Splinter shield DN 100 reduces pumping speed by 5% (H2), 7% (He), 24% (N2/Ar); protective screen DN 100 by 2% (H2/He), 10% (N2), 8% (Ar). p.85 ↗ Include these reductions when sizing effective pumping speed for the application. HiPace 300 Plus DN 100
Maximum absolute inlet pressure on suction side and on venting/sealing gas connection: 1500 hPa. p.74 ↗ Higher process overpressures destroy the pump and eject debris; must be prevented by process design. HiPace 300 Plus (all variants)
Ambient temperature limits: 5–30 °C with convection cooling (no gas throughput), 5–35 °C with air cooling, 5–40 °C with water cooling. p.77 ↗ Water cooling extends ambient envelope to 40 °C; reduce backing pressure/gas throughput for higher ambient temperatures. HiPace 300 Plus (all variants)
Limits: surrounding magnetic field 5.5 mT, irradiated thermal output 2.4 W, max. rotor temperature 90 °C, max. bakeout temperature at HV flange 120 °C. p.77 ↗ Install magnetic/thermal shielding if process exceeds these; drive auto-reduces output then shuts off on over-temperature. HiPace 300 Plus (all variants)
Installation altitude up to 5000 m, ambient air pressure 530–1060 hPa, relative humidity max. 80% at T < 31 °C (50% at T < 40 °C), min. protection IP54 / Type 12. p.77 ↗ Defines permissible installation locations and enclosure requirements for cabinets. HiPace 300 Plus (all variants)
Power supply pack must provide double isolation between mains input and output per IEC 61010-1 / IEC 60950-1 / IEC 62368-1; mains disconnection device per SEMI-S2 with circuit breaker ≥ 10,000 A interruption rating required. p.73 ↗ Non-approved supplies risk fatal electric shock; installation lacks its own mains switch. HiPace 300 Plus with TC 400 (24 V / 48 V)
Rotor bearing: permanent magnetic bearing (wear-free) on HV side; ceramic-ball bearing on fore-vacuum side lubricated by an replaceable operating fluid reservoir. p.80 ↗ Fore-vacuum bearing requires periodic operating fluid reservoir replacement (section 7.3) as preventive maintenance. HiPace 300 Plus (all variants)
Rated rotor speed 60,000 1/min (1000 Hz) per rating plate; drive available as TC 400 in 24 V DC or 48 V DC. p.70 ↗ Voltage variant must be chosen to match available DC supply; speed confirmation required after drive unit replacement. HiPace 300 Plus, TC 400 24 V / 48 V
ISO-K to ISO-K bracket screws tightened cross-wise in 3 steps: 5, 15, 25 ± 2 Nm; ISO-K to ISO-F hexagon screws: 5, 10, 16 ± 1 Nm. p.87 ↗ Cross-wise staged tightening prevents flange distortion and leaks; use calibrated torque wrench. HiPace 300 Plus DN 100 ISO-K / ISO-F
Pump must be mounted with rotor axis vertical and HV flange pointing up; upside-down operation is forbidden. p.86 ↗ Any non-vertical system layout is invalid for this pump; process design must accommodate this. HiPace 300 Plus (all variants)
Fore-vacuum connection is DN 16 ISO-KF; permissible media are dry and inert gases only. p.80 ↗ Corrosive, explosive, condensing, fluid or dusty media invalidate warranty and are foreseeable improper use. HiPace 300 Plus (all variants)
Vibration compensator use likely causes pump shear-off under rotor jamming (ISO 27892 torques); contact with Pfeiffer Vacuum required before installation, plus additional HV flange securing and vertical passage. p.86 ↗ Extra on-site torque compensation measures are mandatory for vibration-sensitive installations. HiPace 300 Plus (all variants)
Damaging or removing the ex-factory warranty seal results in loss of warranty. p.70 ↗ Do not open the pump housing; only authorized service may intervene internally. HiPace 300 Plus (all variants)
Pump weight is 7.0 kg per rating plate; keep masses connected to the pump as low as possible and use flexible lines. p.70 ↗ Rigid heavy piping on connections can be damaged by fault-induced twisting, leaking toxic media. HiPace 300 Plus (all variants)
Maintenance levels: level 1 = trained customer specialist, level 2 = customer with technical education / Pfeiffer service, level 3 = customer with Pfeiffer service training. p.79 ↗ Assign maintenance tasks per personnel qualification level; Pfeiffer offers training courses for levels 2 and 3. HiPace 300 Plus (all variants)
Operation is possible without an operating unit, via the 26-pole 'remote' D-Sub socket, E74 connection, Pfeiffer control unit or field bus (Profibus/DeviceNet/EtherCAT variants of TC 400). p.98 ↗ Control interface choice determines which TC 400 manual (PT 0203/0244/0302/0352/0452 BN) applies. HiPace 300 Plus with TC 400
Keep lines and cables away from hot surfaces (> 70 °C); never operate with open HV connection; mechanical installation always before electrical connection. p.76 ↗ Wiring routing and installation sequence requirements for safe commissioning. HiPace 300 Plus (all variants)