Induztr
Engineering Digest Document Viewer Related Models

HiPace 400 with TC 400 24 V — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Original operating instructions (DE/EN, 134 pp) for the Pfeiffer Vacuum HiPace 400 turbopump with electronic drive unit TC 400. Covers six variants: HiPace 400 with DN 100 ISO-K, DN 100 CF-F, or DN 100 ISO-F high vacuum flange, each available with TC 400 in 24 V DC or 48 V DC. Related TC 400 drive unit manuals (standard PT 0203 BN, Profibus PT 0244 BN, SEMI E74 PT 0302 BN, DeviceNet PT 0352 BN, EtherCAT PT 0452 BN) are listed as separate documents.

Section-by-Section Technical Breakdown

pp. 76-83
Safety, limits of use and personnel qualification

Risk assessment per Machinery Directive 2006/42/EC and EN ISO 12100, with warnings on electric shock, rotor burst torque (ISO 27892), hot surfaces, toxic media and falling objects. Defines permissible ambient conditions and lists foreseeable improper use. Personnel are classified into maintenance levels 1-3.

  • Mains disconnection device per SEMI-S2 required; circuit breaker with interruption rating ≥ 10,000 A
  • Power supply pack must meet double isolation requirements per IEC 61010-1/60950-1/62368-1
  • Max permissible inlet pressure on suction side / venting / sealing gas: 1500 hPa (absolute)
  • Ambient: 530-1060 hPa, altitude ≤ 5000 m, magnetic field ≤ 6 mT, irradiated heat ≤ 4.2 W
  • Ambient temperature 5-35°C air-cooled, 5-40°C water-cooled; permissible IP54 / Type 12
  • Max rotor temperature 90°C; max bakeout temperature at HV flange 120°C
  • Proper use only for dry, inert gases with a suitable backing pump, indoors
  • Vibration compensators require prior consultation with Pfeiffer Vacuum and additional securing of the pump
pp. 84-85
Product description

The HiPace 400 forms a compact unit with the TC 400 electronic drive unit. Describes cooling options, hybrid rotor bearing and drive electronics, plus scope of shipment and product designation logic.

  • Hybrid bearing: wear-free permanent magnetic bearing on HV side, ball bearing with ceramic balls on fore-vacuum side, lubricated by an operating fluid reservoir
  • Cooling: water cooling standard, air cooling optional; drive power auto-reduced on overtemperature
  • Fore-vacuum connection DN 25 ISO-K; housing in aluminum and stainless steel depending on flange variant
  • Shipment includes pump with TC 400, protective covers, remote/E74 mating plugs (type-dependent)
pp. 87-100
Installation: fastening, flange connections, cooling water and electrical supply

Mechanical installation must always precede electrical connection. Details mounting to the bottom part, counter flange design, splinter shield speed losses, flange-specific tightening torques, cooling water requirements and electrical/grounding connection. Safety relevant because burst torques per ISO 27892 act on the anchoring within fractions of a second.

  • Fastening: 6× M8 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3×d, torque 25 Nm ±10%, mounting plate ≥3 mm thick with >270 MPa tensile strength
  • Max torque in event of rotor burst: 4200 Nm; max axial load on HV flange 1000 N (100 kg), no one-sided load
  • Counter flange flatness ± 0.05 mm; material strength 170 N/mm² at 2.5×d or 270 N/mm² at 1.5×d screw engagement
  • Pumping speed reduction with DN 100 splinter shield: 5% (H2), 7% (He), 24% (N2, Ar); protective screen: 2% (H2, He), 10% (N2), 8% (Ar)
  • With oil-sealed backing pumps: fore-vacuum connection vertically down ±25°, support piping, no forces on pump
  • ISO-K/ISO-K bracket screws: tighten cross-wise in 3 steps 5/15/25 ±2 Nm; ISO-K to ISO-F hex screws: 5/10/16 ±1 Nm
  • Use calibrated torque wrench with tightening factor ≤ 1.6; only approved Pfeiffer mounting kits
pp. 101-105
Operation and commissioning

Covers commissioning, operating modes (stand-alone, remote multi-pin connection, E74, Pfeiffer control unit, fieldbus), switching on, LED-based operating mode display, temperature monitoring, and switching off/venting incl. delayed venting factory settings.

  • Factory settings of the TC 400 electronic drive unit documented in Tbl. 11 (p.101)
  • Operating mode shown via LED on TC 400 (Tbl. 12); overtemperature first reduces drive output, then shuts off
  • Venting screw must not be fully unscrewed during manual venting; delayed venting factory settings in Tbl. 13
pp. 106-112
Maintenance

Maintenance intervals and responsibilities by level, replacement of the operating fluid reservoir (fore-vacuum side bearing lubrication) and replacement of the electronic drive unit TC 400, including confirmation of the nominal speed specification afterwards.

  • Replace operating fluid reservoir and electronic drive unit TC 400 per described procedures (pp. 107-111)
  • After drive unit replacement, confirm nominal rotation speed (characteristic speeds in Tbl. 14, p.111)
  • Decontaminate parts exposed to toxic media before maintenance; observe operating fluid safety data sheet
pp. 113-114
Decommissioning, recycling and disposal

Instructions for shutting down for longer periods and recommissioning, plus general disposal information and turbopump-specific disposal.

  • Distinct procedures for extended shutdown (8.1) and recommissioning (8.2)
  • Disposal per general and product-specific rules in section 9
pp. 115-125
Malfunctions, service, spare parts and accessories

Troubleshooting table (Tbl. 15), Pfeiffer Vacuum service offerings, spare parts list (Tbl. 16) and accessory ordering tables for DN 100 ISO-K, DN 100 CF and DN 100 ISO-F.

  • Accessories listed per flange type: Tbl. 17 ISO-K (p.122), Tbl. 18 CF (p.123), Tbl. 19 ISO-F (p.125)
pp. 126-132
Technical data and dimensions

Technical data tables for HiPace 400 with TC 400 in 48 V and 24 V versions, pressure/gas throughput unit conversion tables, materials in contact with process media, and dimension drawings for ISO-K, ISO-F and CF-F versions. Includes EC and UK Declarations of Conformity.

  • Tbl. 22: HiPace 400 | 48 V technical data (p.128); Tbl. 23: 24 V version (p.129)
  • Materials in contact with process media in Tbl. 24 (p.129)
  • Dimension drawings for TC 400 with DN 100 ISO-K, ISO-F and CF-F (pp. 130)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Fastening to mounting plate requires 6× M8 screws of strength class ≥ 8.8 with screw-in depth ≥ 1.3×d, tightening torque 25 Nm ±10%; plate minimum 3 mm thick with >270 MPa tensile strength p.88 ↗ The customer mounting plate must be dimensioned and torqued to absorb rotor-burst torques; underspecified fastening risks pump shearing off in a rotor jam event HiPace 400 (and HiPace 350/450/700/800)
Maximum torque occurring in event of rotor burst: 4200 Nm; maximum permissible axial load on high vacuum flange: 1000 N (100 kg), no one-sided load permitted p.88 ↗ Counter flange and chamber structure must withstand burst torque; attachments above the HV flange must stay under 100 kg total weight and be symmetrically loaded HiPace 400
Counter flange flatness tolerance ± 0.05 mm; flange material tensile strength 170 N/mm² at 2.5×d or 270 N/mm² at 1.5×d screw engagement depth p.89 ↗ Uneven counter flanges induce housing stresses leading to leaks or degraded running behavior; verify flatness and material before installation HiPace 400 high vacuum connection
Never exceed 1500 hPa (absolute) inlet pressure on suction side or venting/sealing gas connection p.78 ↗ Protect pump from process overpressure events; exceeding this destroys the pump and ejects fragments whole series
Maximum permissible bakeout temperature at the high vacuum flange: 120°C; maximum permissible rotor temperature: 90°C p.81 ↗ Bakeout above 120°C at the flange is not allowed; redundant temperature monitoring reduces then cuts drive output on overtemperature whole series
Ambient temperature 5-35°C with air cooling and 5-40°C with water cooling; permissible surrounding magnetic field 6 mT; irradiated thermal output max 4.2 W p.81 ↗ Sites with hot processes or strong magnetic fields need shielding or reduced backing pressure/gas throughput to operate legally whole series
Installation altitude max 5000 m; air pressure 530-1060 hPa; protection class III, overvoltage category II, permissible IP54 / Type 12 (UL 50E) p.81 ↗ Outdoor or unprotected installation is improper use; enclosure must achieve at least IP54 whole series
Pumping speed reduction with splinter shield DN 100: H2 5%, He 7%, N2 24%, Ar 24%; with protective screen DN 100: H2 2%, He 2%, N2 10%, Ar 8% p.89 ↗ Size the effective pumping speed in system calculations for the actual gas; the shield choice significantly changes light-gas throughput HiPace 400 DN 100 flanges
ISO-K/ISO-K bracket screws tightened cross-wise in 3 steps: 5, 15, 25 ±2 Nm; ISO-K to ISO-F hex head screws: 5, 10, 16 ±1 Nm; torque wrench tightening factor ≤ 1.6 p.91 ↗ Use a calibrated torque wrench and staged cross-wise tightening to ensure leak-tight, stress-free flange joints DN 100 ISO flange variants
Hybrid bearing: wear-free permanent magnetic bearing on high vacuum side, ceramic-ball bearing lubricated by operating fluid reservoir on fore-vacuum side p.84 ↗ Fore-vacuum side bearing lifetime depends on the operating fluid reservoir — its replacement is a required maintenance task whole series
With oil-sealed backing pumps, the fore-vacuum connection must point vertically downward with max deviation ±25°, piping supported and no forces transmitted to pump p.90 ↗ Prevents oil backmigration and piping stress; all orientations are allowed only with dry compressing backing pumps whole series
Vibration compensator use requires prior consultation with Pfeiffer Vacuum and additional securing of the turbopump to the high vacuum flange; vertical passage only p.90 ↗ A vibration compensator likely causes pump shear-off during rotor jam (ISO 27892 torques); on-site torque compensation measures are mandatory whole series
Mains disconnection device per SEMI-S2 must be installed; circuit breaker with interruption rating of at least 10,000 A; power supply pack must meet double isolation per IEC 61010-1/60950-1/62368-1 p.77 ↗ Neither pump nor TC 400 has a mains switch — the customer's electrical design must provide compliant disconnection and a certified PSU whole series with TC 400
Turbopump is only completely de-energized when mains plug is disconnected and rotation speed f = 0; secure against restart before any work p.79 ↗ Maintenance lockout must wait for full rotor standstill; exposed sharp-edged rotor blades pose cut/amputation risk in open HV flange whole series
Proper use limited to dry and inert gases, indoors, with a suitable backing pump; pumping explosive, corrosive or condensing media, dust, fluids, or use in explosive atmospheres or ionizing radiation is improper use p.81 ↗ Corrosive applications require the C (corrosive gas) version and gas-mode-appropriate presets; standard pump voids warranty otherwise whole series
Rotor nominal speed on the rating plate example: 49,200 1/min (820 Hz); after drive unit replacement the speed specification must be confirmed (Tbl. 14) p.74 ↗ Verify programmed nominal speed against Tbl. 14 / rating plate whenever the TC 400 electronic drive unit is exchanged HiPace 400 with TC 400
Mass connected to the turbopump must be kept as low as possible and flexible lines used, because sudden rotor jam causes twisting that can damage rigid fittings and leak toxic media p.77 ↗ For toxic processes, use flexible fore-vacuum lines and minimize fitting mass to prevent lethal leaks during fault events whole series
Connected cables/lines must be kept away from hot surfaces > 70°C; device must never be filled or operated with cleaning agents or their residues p.80 ↗ Route cabling with thermal separation; improper cleaning invalidates safety and can damage the pump whole series