HiPace 700 H with TC 400 — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Original operating instructions (128 pp.) for the Pfeiffer Vacuum HiPace 700 H turbopump with electronic drive unit TC 400, covering three variants: DN 160 ISO-K, DN 160 CF-F, and DN 160 ISO-F high vacuum flanges (all with TC 400, 48 V DC drive). Related TC 400 drive manuals (standard/Profibus/E74/DeviceNet/EtherCAT) are referenced separately. 'H' = high compression version, especially for light gases.
Section-by-Section Technical Breakdown
Risk assessment per Machinery Directive 2006/42/EC and EN ISO 12100. Covers electrical, mechanical (rotor burst per ISO 27892), and toxic-media hazards across the product life cycle, plus permissible ambient conditions and improper-use prohibitions. Personnel graded into maintenance levels 1-3.
- Mains disconnection device required per SEMI-S2; circuit breaker with interruption rating ≥ 10,000 A
- Power supply pack must meet double isolation per IEC 61010-1 / IEC 60950-1 / IEC 62368-1
- Never exceed 1500 hPa (absolute) inlet pressure on suction side or venting/sealing gas connection
- Permissible ambient: 5-35°C air cooling, 5-40°C water cooling; altitude max. 5000 m; air pressure 530-1060 hPa
- Max. surrounding magnetic field 6 mT; max. irradiated thermal output 4.2 W; max. rotor temperature 90°C; max. bakeout temperature at HV flange 120°C
- Protection degree IP54 / Type 12 (UL 50E), protection class III, overvoltage category II
- Proper use only for dry and inert gases, indoors, with suitable backing pump
The turbopump forms a compact unit with TC 400 electronic drive unit (48 V DC). Hybrid bearing: wear-free permanent magnetic bearing on high vacuum side, ceramic-ball bearing lubricated by operating fluid reservoir on fore-vacuum side. Water cooling standard, air cooling optional.
- Fore-vacuum connection DN 25 ISO-K
- Rated speed 49,200 1/min (820 Hz) per rating plate
- HV flange materials: aluminium (ISO-K, ISO-F), stainless steel (CF-F)
- Shipment includes pump with TC 400, protective covers, remote/E74 mating plugs (type-dependent), manual
Detailed mechanical installation: fastening to mounting plate, counter flange design, ISO-K/ISO-F/CF flange connections with tightening torques, fore-vacuum connection, cooling water requirements, and electrical connection/grounding. Safety against rotor-burst torque (ISO 27892) drives most fastening requirements.
- Fastening: 6× M8 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3×d, torque 25 Nm ±10%, mounting plate min. 3 mm / >270 MPa tensile strength
- Max. burst torque 4200 Nm; max. axial load on HV flange 1000 N (no one-sided load); counter flange flatness ±0.05 mm; flange material ≥170 N/mm² at 2.5×d or ≥270 N/mm² at 1.5×d screw engagement
- ISO-K connections: tighten cross-wise in 3 steps 5/15/25 ±2 Nm, calibrated torque wrench (factor ≤1.6)
- Splinter shield DN 160 reduces pumping speed: H2 -6%, He -9%, N2 -20%, Ar -23%; protective screen: H2 -1%, He -2%, N2 -6%, Ar -7%
- With oil-sealed backing pumps: fore-vacuum connection downwards vertical, deviation max. ±25°
- Vibration compensator requires prior consultation with Pfeiffer Vacuum and additional securing to HV flange
- Keep lines/cables away from surfaces >70°C; mechanical installation before electrical connection
Commissioning and operating modes (stand-alone, remote multi-function connector, E74, Pfeiffer control unit, field bus), switching on/off, LED monitoring, temperature monitoring, and venting. Factory settings and delayed venting are tabulated.
- LED operating mode display and temperature monitoring with automatic drive-power reduction then shutdown on overtemperature
- Venting via venting screw; do not fully unscrew venting screw during venting
- Delayed venting factory settings tabulated (Tbl. 13)
Maintenance intervals and responsibilities per maintenance level, replacement of the operating fluid reservoir (fore-vacuum-side bearing lubrication), replacement of the electronic drive unit TC 400, and confirming the nominal speed specification on replacement units via parameter [P:777].
- A Pfeiffer SpeedConfigurator is supplied with replacement drives for one-time setting of [P:777] nominal speed confirmation
- Decontaminate parts exposed to toxic process media before maintenance; observe operating fluid safety data sheet
- Before any work: switch off, wait for f = 0, unplug mains, secure against restart
Shutting down for longer periods, recommissioning procedure, and disposal of turbopumps according to general disposal regulations.
Troubleshooting table (Tbl. 15), Pfeiffer Vacuum service solutions, spare parts list for HiPace 700 (Tbl. 16, Fig. 23), and accessories with ordering information (Tbl. 17).
Technical data table for HiPace 700 H (Tbl. 20), compression ratio vs. supply pressure characteristics (Fig. 24), materials in contact with process media (Tbl. 21), and dimensional drawings for DN 160 ISO-K / CF-F / ISO-F variants with TC 400. Includes pressure and gas throughput unit conversion tables.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Rated rotational speed is 49,200 rpm (820 Hz) as stated on the rating plate. p.72 ↗ | Use for run-up time checks and confirming [P:777] nominal speed parameter on replacement TC 400 drive units. | HiPace 700 H (all flange variants) |
| Maximum permissible inlet pressure is 1500 hPa (absolute) on the suction side and on venting/sealing gas connections. p.76 ↗ | Interlock process side to prevent overpressure from destroying the pump. | whole series |
| Permissible ambient temperature: 5-35°C with air cooling, 5-40°C with water cooling; max. altitude 5000 m. p.79 ↗ | Choose cooling type based on cabinet/room temperature; reduced backing pressure/gas throughput allows higher ambient temperatures. | whole series |
| Maximum permissible rotor temperature 90°C; maximum bakeout temperature at the HV flange 120°C; permissible surrounding magnetic field 6 mT; max. irradiated thermal output 4.2 W. p.79 ↗ | Bakeout limited to 120°C at the flange; shield the pump if process radiation or magnetic fields exceed these limits. | whole series |
| Fastening to mounting plate: 6× M8 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3×d, tightening torque 25 Nm ±10%, plate ≥ 3 mm thick with >270 MPa tensile strength. p.86 ↗ | Under-dimensioned mounting compromises rotor-burst safety (ISO 27892); torque with calibrated wrench (factor ≤1.6). | HiPace 350/400/450/700/800 group incl. HiPace 700 |
| Maximum torque occurring on rotor burst: 4200 Nm; maximum permissible axial load on HV flange: 1000 N (100 kg), one-sided load not permitted. p.86 ↗ | Design the customer-side mounting and chamber interface to absorb 4200 Nm reaction torque; keep superstructure weight ≤100 kg, evenly distributed. | HiPace 700 |
| Counter flange flatness tolerance ±0.05 mm; flange material tensile strength ≥170 N/mm² at 2.5×d or ≥270 N/mm² at 1.5×d screw engagement. p.87 ↗ | Specify counter flange machining and material accordingly to avoid housing stress and leakage. | HiPace 700 |
| ISO flange connections (bracket or hexagon screws) tightened cross-wise in 3 steps: 5, 15, 25 ±2 Nm. p.89 ↗ | Use this torque sequence for ISO-K/ISO-F installations with a calibrated torque wrench (tightening factor ≤1.6). | DN 160 ISO-K and ISO-F variants |
| Splinter shield DN 160 reduces pumping speed by 6% (H2) to 23% (Ar); protective screen by 1% (H2) to 7% (Ar). p.87 ↗ | Account for conductance loss when sizing the pump; with ISO flanges use centering rings with splinter shield/screen. | DN 160 flange variants |
| With oil-sealed backing pumps, fore-vacuum connection must point vertically downwards with max. ±25° deviation; with dry compressing backing pumps all mounting orientations are permissible. p.88 ↗ | Prevents oil backmigration from the fore-vacuum side; support tubing so no forces act on the pump. | whole series |
| Rotor bearing is a hybrid design: wear-free permanent magnetic bearing on the HV side, ceramic-ball bearing with operating fluid reservoir lubrication on the fore-vacuum side. p.82 ↗ | Operating fluid reservoir replacement is a scheduled maintenance item; oil type per oil sticker on pump. | whole series |
| Electronic drive unit TC 400 operates at 48 V DC supply voltage; power supply pack must meet double isolation per IEC 61010-1 / IEC 60950-1 / IEC 62368-1. p.75 ↗ | Only use approved/double-insulated power supply packs for the DCin connection. | all variants with TC 400 |
| The pump and drive have no mains disconnection device; a circuit breaker with interruption rating ≥ 10,000 A per SEMI-S2 must be installed. p.75 ↗ | Include an external mains disconnection device and integrate the pump into an Emergency Off circuit in the system design. | whole series |
| Fore-vacuum connection is DN 25 ISO-K. p.82 ↗ | Size backing pump and fore-line to deliver the required maximum fore-vacuum pressure via this connection. | HiPace 700 H |
| Replacement TC 400 drive units ship with a Pfeiffer SpeedConfigurator for one-time setting of nominal speed confirmation parameter [P:777]. p.46 ↗ | Run this configuration after drive replacement, otherwise the pump will not correctly confirm rated speed. | all variants with TC 400 |
| Vibration compensator installation requires prior consultation with Pfeiffer Vacuum and additional securing of the pump to the HV flange; vertical passage only. p.76 ↗ | Unsecured use risks pump shearing off under rotor-jam torque (ISO 27892). | whole series |
| Only dry, inert gases may be pumped; explosive, corrosive media, condensing vapors, fluids, and dust are prohibited uses. p.79 ↗ | For corrosive processes a different variant (e.g. 'C' corrosive gas version) would be required, not this model. | whole series |
| Permissible protection degree IP54 / Type 12 (UL 50E), protection class III, overvoltage category II, pollution degree 2; relative humidity max. 80% at T<31°C, max. 50% at T<40°C. p.79 ↗ | Enclose the installation accordingly in wet or dusty environments. | whole series |
| Damaging or removing the factory closure seal results in loss of warranty. p.72 ↗ | Do not open the pump housing; only perform user-level maintenance (fluid reservoir, drive unit). | whole series |
| HV flange materials: aluminium for DN 160 ISO-K and ISO-F variants, stainless steel for DN 160 CF-F variant. p.83 ↗ | CF-F variant is the choice where metal-sealed, bakeable/stainless interfaces are required. | variant-specific |