HiPace 700 P with TC 400 — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Operating instructions for the Pfeiffer Vacuum HiPace 700 P process turbopump with TC 400 electronic drive unit, 48 V DC, covering three HV flange variants: DN 160 ISO-K, DN 160 CF-F, DN 160 ISO-F. Related documents: TC 400 variants (standard PT 0203 BN, Profibus PT 0244 BN, SEMI E74 PT 0302 BN, DeviceNet PT 0352 BN, EtherCAT PT 0452 BN).
Section-by-Section Technical Breakdown
Risk assessment per Machinery Directive 2006/42/EC and EN ISO 12100 covering transport, installation, operation, maintenance and malfunction hazards, including rotor-jam torque hazards per ISO 27892. Defines proper use (vacuum generation with suitable backing pump, dry/inert gases; corrosive gases only with sealing gas) and lists foreseeable misuse. Ambient and operating limits are tabulated.
- Mains disconnection device per SEMI-S2 required (pump/TC 400 have no mains switch); circuit breaker with interruption rating ≥ 10,000 A
- Power supply pack must meet double isolation requirements per IEC 61010-1 / IEC 60950-1 / IEC 62368-1
- Never exceed 1500 hPa (absolute) inlet pressure on suction side or venting/sealing gas connection
- Ambient: 530–1060 hPa, altitude max. 5000 m, ambient temp 5–35 °C air-cooled / 5–40 °C water-cooled, protection class III, overvoltage category II, IP54/Type 12
- Max. surrounding magnetic field 6 mT; max. irradiated thermal output 4.2 W; max rotor temperature 90 °C; max bakeout temp at HV flange 120 °C
- Rel. humidity max. 80% at T < 31 °C, up to 50% at T < 40 °C; pollution degree 2
- Do not stack product during transport; carry with both hands
The HiPace 700 P is a process-version turbopump ('P' = insensitive to dust and corrosive gases) forming a compact unit with the TC 400 electronic drive unit (48 V DC). It is a hybrid-bearing pump with wear-free permanent magnetic bearing on the HV side and ceramic-ball bearing with operating fluid reservoir on the fore-vacuum side.
- Cooling: water cooling standard, air cooling optional
- Bearing: permanent magnetic (HV side), ball bearing with ceramic balls (fore-vacuum side), lubricated by operating fluid reservoir
- Pump housing: stainless steel (rotor area) and aluminum; venting screw, sealing gas connection, DN 25 ISO-KF fore-vacuum connection
- Shipment includes pump with TC 400, protective covers, mating plugs for remote/E74 (type-dependent), operating instructions
- Product designation: family + size + attribute; P = process version
Covers fastening to the mounting plate, HV flange connections (ISO-K/ISO-F/CF-F), fore-vacuum connection, cooling water hookup, accessories and electrical supply/grounding. Detailed counter-flange requirements derive from ISO 27892 burst tests. Mechanical installation must always precede electrical connection.
- Mounting: M8 screws, 6 pcs, strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, torque 25 Nm ± 10%, mounting plate ≥ 3 mm thick, tensile strength > 270 MPa
- Counter flange: flatness ± 0.05 mm; min tensile strength 270 N/mm² at 1.5 × d or 170 N/mm² at 2.5 × d engagement depth
- Max torque in case of burst: 4200 Nm; max permissible axial load on HV flange: 1000 N (100 kg), no one-sided load
- Splinter shield DN 160 reduces pumping speed by 20% (N₂)/23% (Ar); protective screen DN 160 by 6% (N₂)/7% (Ar)
- ISO-K flange bracket/hex screws: tighten crosswise in 3 steps 5, 15, 25 ± 2 Nm
- With oil-sealed backing pumps: fore-vacuum connection downward vertical, deviation ± 25° max; all orientations permitted with dry backing pumps
- Vibration compensator: contact Pfeiffer Vacuum first; secure pump additionally to HV flange
- Create flexible connection to HV flange when fastened via bottom part to avoid reactive forces
Commissioning, operating modes (standalone, remote D-Sub, E74, Pfeiffer control unit, fieldbus), switching on, monitoring via LED and redundant temperature monitoring, switching off and venting including delayed venting factory settings.
- Electronic drive unit reduces then switches off drive output if max permissible operating temperature is exceeded
- LED operating mode display on TC 400; behavior/meaning table provided
- Factory settings for delayed venting documented in table
- Do not fully unscrew venting screw during venting; keep distance from automatic venting valves
Maintenance intervals checklist, replacement of the operating fluid reservoir (fore-vacuum side bearing lubrication) and replacement of the TC 400 electronic drive unit with confirmation of nominal speed specification afterwards.
- Maintenance levels: level 1 customer (trained specialist), level 2 customer with technical education, level 3 customer with Pfeiffer service training
- Decontaminate parts exposed to toxic process media before maintenance; observe operating fluid safety data sheet
- Characteristic nominal rotation speeds table used to confirm speed after drive unit replacement
Shutdown procedures for longer periods, recommissioning guidance and disposal of turbopumps per general disposal regulations.
Troubleshooting table (Tbl. 16) covering malfunctions of the turbopump.
Spare parts list for HiPace 700 P (Tbl. 17) with exploded view, and accessories table (Tbl. 18) with ordering information.
Technical data for HiPace 700 P | 48 V (Tbl. 21), materials in contact with process media (Tbl. 22), pressure/gas throughput unit conversion tables, and dimension drawings for DN 160 ISO-K, CF-F and ISO-F variants with TC 400.
- Rating plate example: 49,200 1/min (820 Hz) nominal rotor speed
- Dimension drawings given for TC 400 with DN 160 ISO-K, CF-F and ISO-F HV flanges
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Mounting to bottom part: 6 × M8 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, tightening torque 25 Nm ± 10%, mounting plate min. 3 mm thick with tensile strength > 270 MPa p.22 ↗ | Mounting plate must be dimensioned accordingly; use calibrated torque wrench (factor ≤ 1.6) | HiPace 700 P (all variants) |
| Maximum torque occurring in case of rotor burst: 4200 Nm (calculated per ISO 27892) p.22 ↗ | Customer-side mounting and flange connections must absorb this torque; use flexible HV connections | HiPace 700 |
| Max permissible axial load on the HV flange: 1000 N (100 kg), one-sided load not permitted; counter flange flatness ± 0.05 mm p.23 ↗ | Limit superstructural parts on the HV flange; design counter flange to shape tolerances to avoid housing stresses/leaks | HiPace 700 |
| Flange material tensile strength requirement: 170 N/mm² at 2.5 × d or 270 N/mm² at 1.5 × d screw engagement depth p.23 ↗ | Counter flange material and tapped-hole depth must satisfy these minima in all operating states | HiPace 700 counter flange design |
| Splinter shield DN 160 reduces pumping speed by 6% (H₂), 9% (He), 20% (N₂), 23% (Ar); protective screen DN 160 by 1–7% p.23 ↗ | Factor this loss into effective pumping speed sizing; splinter shield recommended for ISO flanges | HiPace 700 P DN 160 variants |
| ISO-K flange connections: tighten bracket/hex screws crosswise in 3 steps at 5, 15, 25 ± 2 Nm p.25 ↗ | Prevents flange distortion; twisting during rotor blockage may occur but does not jeopardize leak-tightness | HiPace 700 P ISO-K / ISO-F installations |
| Max permissible inlet pressure: 1500 hPa absolute on suction side, venting and sealing gas connections p.12 ↗ | Overpressure protection required upstream; exceeding destroys the pump | HiPace 700 P |
| Ambient temperature limits: 5–35 °C with air cooling, 5–40 °C with water cooling; operation at higher temperatures possible with reduced backing pressure/gas throughput p.15 ↗ | Choose cooling type and cabin cooling design accordingly | HiPace 700 P |
| Max permissible rotor temperature 90 °C; max bakeout temperature at HV flange 120 °C; max surrounding magnetic field 6 mT; max irradiated thermal output 4.2 W p.15 ↗ | Bakeout above 120 °C at the flange and insufficient magnetic/thermal shielding are not permitted | HiPace 700 P |
| Electrical: 48 V DC supply via TC 400, protection class III, overvoltage category II, IP54 / Type 12 UL 50E p.8 ↗ | Power supply pack must meet double isolation per IEC 61010-1/60950-1/62368-1; external mains disconnection device per SEMI-S2 with ≥ 10 kA breaking capacity required | HiPace 700 P with TC 400 |
| Rotor speed: 49,200 1/min (820 Hz) per rating plate example p.8 ↗ | Nominal speed value used for confirmation after electronic drive unit replacement (Tbl. 15) | HiPace 700 P |
| With oil-sealed backing pumps, fore-vacuum connection must point vertically downward, deviation ± 25° max; all orientations allowed with dry compressing backing pumps p.24 ↗ | Prevents oil backflow from fore-vacuum range; support tubing so no piping forces act on the pump | HiPace 700 P |
| Vibration compensator use requires prior consultation with Pfeiffer Vacuum; rotor jam per ISO 27892 can shear the pump off its flange p.12 ↗ | Additional on-site torque compensation and securing of the pump to the HV flange are mandatory | HiPace 700 P with vibration compensator |
| Hybrid bearing: permanent magnetic bearing (HV side, wear-free) and ceramic-ball bearing (fore-vacuum side) lubricated by an replaceable operating fluid reservoir p.18 ↗ | Operating fluid reservoir replacement is a defined maintenance task; observe its safety data sheet and disposal rules | HiPace 700 P |
| Process version 'P' is insensitive to dust and corrosive gases; corrosive gases allowed only with sealing gas; pumping explosive media, fluids, or dust without sealing gas is improper use p.15 ↗ | Verify process media compatibility and enable sealing gas for corrosive/dusty applications | HiPace 700 P |
| Before any maintenance or work on the HV connection: switch off, wait for rotor standstill (f = 0), unplug mains, secure against restart p.13 ↗ | Prevents electric shock and cut injuries from sharp-edged rotor; seal open flanges with original covers | HiPace 700 P with TC 400 |
| Mechanical installation must be completed before electrical connection; mating plugs may enable automatic pump run-up on power-on p.11 ↗ | Only connect mating plugs after mechanical installation to avoid cut injuries at the open HV flange | HiPace 700 P with TC 400 |
| Damaging or removing the factory warranty seal voids the warranty p.8 ↗ | Do not open sealed areas; use authorized maintenance levels (2/3 with Pfeiffer training) | HiPace 700 P |