HiPace 700 P with TC 400 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Operating instructions for the Pfeiffer Vacuum HiPace 700 P process turbopump with TC 400 electronic drive unit, 48 V DC, covering three HV flange variants: DN 160 ISO-K, DN 160 CF-F, DN 160 ISO-F. Related documents: TC 400 variants (standard PT 0203 BN, Profibus PT 0244 BN, SEMI E74 PT 0302 BN, DeviceNet PT 0352 BN, EtherCAT PT 0452 BN).

Section-by-Section Technical Breakdown

pp. 10-17
Safety, Personnel Qualification and Limits of Use

Risk assessment per Machinery Directive 2006/42/EC and EN ISO 12100 covering transport, installation, operation, maintenance and malfunction hazards, including rotor-jam torque hazards per ISO 27892. Defines proper use (vacuum generation with suitable backing pump, dry/inert gases; corrosive gases only with sealing gas) and lists foreseeable misuse. Ambient and operating limits are tabulated.

  • Mains disconnection device per SEMI-S2 required (pump/TC 400 have no mains switch); circuit breaker with interruption rating ≥ 10,000 A
  • Power supply pack must meet double isolation requirements per IEC 61010-1 / IEC 60950-1 / IEC 62368-1
  • Never exceed 1500 hPa (absolute) inlet pressure on suction side or venting/sealing gas connection
  • Ambient: 530–1060 hPa, altitude max. 5000 m, ambient temp 5–35 °C air-cooled / 5–40 °C water-cooled, protection class III, overvoltage category II, IP54/Type 12
  • Max. surrounding magnetic field 6 mT; max. irradiated thermal output 4.2 W; max rotor temperature 90 °C; max bakeout temp at HV flange 120 °C
  • Rel. humidity max. 80% at T < 31 °C, up to 50% at T < 40 °C; pollution degree 2
  • Do not stack product during transport; carry with both hands
pp. 18-19
Product Description and Identification

The HiPace 700 P is a process-version turbopump ('P' = insensitive to dust and corrosive gases) forming a compact unit with the TC 400 electronic drive unit (48 V DC). It is a hybrid-bearing pump with wear-free permanent magnetic bearing on the HV side and ceramic-ball bearing with operating fluid reservoir on the fore-vacuum side.

  • Cooling: water cooling standard, air cooling optional
  • Bearing: permanent magnetic (HV side), ball bearing with ceramic balls (fore-vacuum side), lubricated by operating fluid reservoir
  • Pump housing: stainless steel (rotor area) and aluminum; venting screw, sealing gas connection, DN 25 ISO-KF fore-vacuum connection
  • Shipment includes pump with TC 400, protective covers, mating plugs for remote/E74 (type-dependent), operating instructions
  • Product designation: family + size + attribute; P = process version
pp. 21-34
Installation (Mounting, Flanges, Cooling, Electrical)

Covers fastening to the mounting plate, HV flange connections (ISO-K/ISO-F/CF-F), fore-vacuum connection, cooling water hookup, accessories and electrical supply/grounding. Detailed counter-flange requirements derive from ISO 27892 burst tests. Mechanical installation must always precede electrical connection.

  • Mounting: M8 screws, 6 pcs, strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, torque 25 Nm ± 10%, mounting plate ≥ 3 mm thick, tensile strength > 270 MPa
  • Counter flange: flatness ± 0.05 mm; min tensile strength 270 N/mm² at 1.5 × d or 170 N/mm² at 2.5 × d engagement depth
  • Max torque in case of burst: 4200 Nm; max permissible axial load on HV flange: 1000 N (100 kg), no one-sided load
  • Splinter shield DN 160 reduces pumping speed by 20% (N₂)/23% (Ar); protective screen DN 160 by 6% (N₂)/7% (Ar)
  • ISO-K flange bracket/hex screws: tighten crosswise in 3 steps 5, 15, 25 ± 2 Nm
  • With oil-sealed backing pumps: fore-vacuum connection downward vertical, deviation ± 25° max; all orientations permitted with dry backing pumps
  • Vibration compensator: contact Pfeiffer Vacuum first; secure pump additionally to HV flange
  • Create flexible connection to HV flange when fastened via bottom part to avoid reactive forces
pp. 35-39
Operation (Commissioning, Operating Modes, Monitoring, Venting)

Commissioning, operating modes (standalone, remote D-Sub, E74, Pfeiffer control unit, fieldbus), switching on, monitoring via LED and redundant temperature monitoring, switching off and venting including delayed venting factory settings.

  • Electronic drive unit reduces then switches off drive output if max permissible operating temperature is exceeded
  • LED operating mode display on TC 400; behavior/meaning table provided
  • Factory settings for delayed venting documented in table
  • Do not fully unscrew venting screw during venting; keep distance from automatic venting valves
pp. 40-46
Maintenance (Operating Fluid Reservoir, Electronic Drive Unit Replacement)

Maintenance intervals checklist, replacement of the operating fluid reservoir (fore-vacuum side bearing lubrication) and replacement of the TC 400 electronic drive unit with confirmation of nominal speed specification afterwards.

  • Maintenance levels: level 1 customer (trained specialist), level 2 customer with technical education, level 3 customer with Pfeiffer service training
  • Decontaminate parts exposed to toxic process media before maintenance; observe operating fluid safety data sheet
  • Characteristic nominal rotation speeds table used to confirm speed after drive unit replacement
pp. 47-48
Decommissioning, Recycling and Disposal

Shutdown procedures for longer periods, recommissioning guidance and disposal of turbopumps per general disposal regulations.

pp. 49-50
Malfunctions / Troubleshooting

Troubleshooting table (Tbl. 16) covering malfunctions of the turbopump.

pp. 53-56
Spare Parts and Accessories

Spare parts list for HiPace 700 P (Tbl. 17) with exploded view, and accessories table (Tbl. 18) with ordering information.

pp. 57-60
Technical Data and Dimensions

Technical data for HiPace 700 P | 48 V (Tbl. 21), materials in contact with process media (Tbl. 22), pressure/gas throughput unit conversion tables, and dimension drawings for DN 160 ISO-K, CF-F and ISO-F variants with TC 400.

  • Rating plate example: 49,200 1/min (820 Hz) nominal rotor speed
  • Dimension drawings given for TC 400 with DN 160 ISO-K, CF-F and ISO-F HV flanges

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Mounting to bottom part: 6 × M8 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, tightening torque 25 Nm ± 10%, mounting plate min. 3 mm thick with tensile strength > 270 MPa p.22 ↗ Mounting plate must be dimensioned accordingly; use calibrated torque wrench (factor ≤ 1.6) HiPace 700 P (all variants)
Maximum torque occurring in case of rotor burst: 4200 Nm (calculated per ISO 27892) p.22 ↗ Customer-side mounting and flange connections must absorb this torque; use flexible HV connections HiPace 700
Max permissible axial load on the HV flange: 1000 N (100 kg), one-sided load not permitted; counter flange flatness ± 0.05 mm p.23 ↗ Limit superstructural parts on the HV flange; design counter flange to shape tolerances to avoid housing stresses/leaks HiPace 700
Flange material tensile strength requirement: 170 N/mm² at 2.5 × d or 270 N/mm² at 1.5 × d screw engagement depth p.23 ↗ Counter flange material and tapped-hole depth must satisfy these minima in all operating states HiPace 700 counter flange design
Splinter shield DN 160 reduces pumping speed by 6% (H₂), 9% (He), 20% (N₂), 23% (Ar); protective screen DN 160 by 1–7% p.23 ↗ Factor this loss into effective pumping speed sizing; splinter shield recommended for ISO flanges HiPace 700 P DN 160 variants
ISO-K flange connections: tighten bracket/hex screws crosswise in 3 steps at 5, 15, 25 ± 2 Nm p.25 ↗ Prevents flange distortion; twisting during rotor blockage may occur but does not jeopardize leak-tightness HiPace 700 P ISO-K / ISO-F installations
Max permissible inlet pressure: 1500 hPa absolute on suction side, venting and sealing gas connections p.12 ↗ Overpressure protection required upstream; exceeding destroys the pump HiPace 700 P
Ambient temperature limits: 5–35 °C with air cooling, 5–40 °C with water cooling; operation at higher temperatures possible with reduced backing pressure/gas throughput p.15 ↗ Choose cooling type and cabin cooling design accordingly HiPace 700 P
Max permissible rotor temperature 90 °C; max bakeout temperature at HV flange 120 °C; max surrounding magnetic field 6 mT; max irradiated thermal output 4.2 W p.15 ↗ Bakeout above 120 °C at the flange and insufficient magnetic/thermal shielding are not permitted HiPace 700 P
Electrical: 48 V DC supply via TC 400, protection class III, overvoltage category II, IP54 / Type 12 UL 50E p.8 ↗ Power supply pack must meet double isolation per IEC 61010-1/60950-1/62368-1; external mains disconnection device per SEMI-S2 with ≥ 10 kA breaking capacity required HiPace 700 P with TC 400
Rotor speed: 49,200 1/min (820 Hz) per rating plate example p.8 ↗ Nominal speed value used for confirmation after electronic drive unit replacement (Tbl. 15) HiPace 700 P
With oil-sealed backing pumps, fore-vacuum connection must point vertically downward, deviation ± 25° max; all orientations allowed with dry compressing backing pumps p.24 ↗ Prevents oil backflow from fore-vacuum range; support tubing so no piping forces act on the pump HiPace 700 P
Vibration compensator use requires prior consultation with Pfeiffer Vacuum; rotor jam per ISO 27892 can shear the pump off its flange p.12 ↗ Additional on-site torque compensation and securing of the pump to the HV flange are mandatory HiPace 700 P with vibration compensator
Hybrid bearing: permanent magnetic bearing (HV side, wear-free) and ceramic-ball bearing (fore-vacuum side) lubricated by an replaceable operating fluid reservoir p.18 ↗ Operating fluid reservoir replacement is a defined maintenance task; observe its safety data sheet and disposal rules HiPace 700 P
Process version 'P' is insensitive to dust and corrosive gases; corrosive gases allowed only with sealing gas; pumping explosive media, fluids, or dust without sealing gas is improper use p.15 ↗ Verify process media compatibility and enable sealing gas for corrosive/dusty applications HiPace 700 P
Before any maintenance or work on the HV connection: switch off, wait for rotor standstill (f = 0), unplug mains, secure against restart p.13 ↗ Prevents electric shock and cut injuries from sharp-edged rotor; seal open flanges with original covers HiPace 700 P with TC 400
Mechanical installation must be completed before electrical connection; mating plugs may enable automatic pump run-up on power-on p.11 ↗ Only connect mating plugs after mechanical installation to avoid cut injuries at the open HV flange HiPace 700 P with TC 400
Damaging or removing the factory warranty seal voids the warranty p.8 ↗ Do not open sealed areas; use authorized maintenance levels (2/3 with Pfeiffer training) HiPace 700 P