HiPace 700 Plus with TC 400 — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiPace 700 Plus turbopump operating instructions (124 pages), covering three variants: HiPace 700 Plus with DN 160 ISO-K, DN 160 CF-F, and DN 160 ISO-F high vacuum flanges, each with TC 400 electronic drive unit (24 V DC or 48 V DC versions). 'Plus' designates low-vibration, low magnetic stray field models. Drive unit variants have separate manuals (TC 400 standard PT 0203 BN, Profibus PT 0244 BN, E74 PT 0302 BN, DeviceNet PT 0352 BN, EtherCAT PT 0452 BN).
Section-by-Section Technical Breakdown
Defines the covered variants, target group and conventions, then details safety instructions per lifecycle phase based on Machinery Directive 2006/42/EC and EN ISO 12100 risk assessment. Key hazards: electric shock (no mains switch on pump — external disconnection device required per SEMI-S2), rotor burst torques per ISO 27892, cuts from sharp rotor parts in open HV flange, and poisoning from toxic media leaks. Also defines proper/improper use and personnel qualification levels.
- No mains disconnection device built in: install a mains switch per SEMI-S2 with circuit breaker rated at least 10,000 A interruption capacity
- Power supply pack must meet double insulation requirements per IEC 61010-1, IEC 60950-1 and IEC 62368-1
- Never exceed 1500 hPa (absolute) at suction side or venting/sealing gas connections — risk of pump destruction
- Improper use includes pumping explosive, corrosive, condensing, fluid or dusty media, and operation in explosive atmospheres or ionizing radiation
- Warranty seal: breaking/removing it voids the warranty
- Maintenance levels: level 1 (trained specialist), level 2 (technically educated customer or PV technician), level 3 (PV-service-trained customer or PV technician)
Table of environmental limits for installation and operation. Ambient temperature limits depend on cooling type; rotor and bakeout temperatures are separately limited. Redundant temperature monitoring makes the pump intrinsically safe — the drive unit reduces then shuts off drive power on overtemperature.
- Ambient temperature: 5–35 °C with air cooling, 5–40 °C with water cooling
- Air pressure 530–1060 hPa; installation altitude max. 5000 m
- Relative humidity max. 80% at T < 31 °C, up to max. 50% at T < 40 °C
- Permissible surrounding magnetic field: 6 mT; max. irradiated thermal output 4.2 W
- Max. permissible rotor temperature 90 °C; max. bakeout temperature at HV flange 120 °C
- Protection degree IP54 / Type 12 (UL 50E); protection class III; overvoltage category II; degree of pollution 2
The turbopump is a compact unit with TC 400 electronic drive unit; Pfeiffer power supply packs provide the DC supply. Hybrid bearing: wear-free permanent magnetic bearing on the high vacuum side, ceramic-ball bearing lubricated by an operating fluid reservoir on the fore-vacuum side. Water cooling is standard, air cooling optional.
- Rotor bearing: permanent magnetic (HV side, wear-free) + ball bearing with ceramic balls (fore-vacuum side), lubricated by operating fluid reservoir
- Drive unit TC 400: operating voltage 24 V DC or 48 V DC variants
- Rating plate example: nominal speed 49,200 1/min (820 Hz)
- HV flange DN 160 ISO-K or ISO-F with aluminium flange material; DN 160 CF-F with stainless steel
- Fore-vacuum connection: DN 25 ISO-KF; sealing gas connection and venting screw present
- Shipment includes pump with drive unit, protective covers, remote/E74 mating plugs (type-dependent), operating instructions
Fasten pump bottom part to a customer mounting plate (min. 3 mm thickness, > 270 MPa tensile strength) with 6× M8 screws, ≥ class 8.8, torque 25 Nm ± 10%, screw-in depth ≥ 1.3 × d. Counter flange must meet flatness ± 0.05 mm; max. axial load on HV flange 1000 N (100 kg), no one-sided loading. ISO-K flange joints are tightened cross-wise in 3 steps (5, 15, 25 ± 2 Nm). With oil-sealed backing pumps, fore-vacuum connection must point down vertically (± 25° deviation allowed).
- Maximum torque in the event of a rotor burst: 4200 Nm (calculated per ISO 27892; not reached in any experimental test) — mounting plate on operator side must absorb it
- Flange material minimum tensile strength: 170 N/mm² at 2.5 × d engagement depth, 270 N/mm² at 1.5 × d
- Splinter shield DN 160 reduces pumping speed: H2 −6%, He −9%, N2 −20%, Ar −23%; protective screen DN 160: H2 −1%, He −2%, N2 −6%, Ar −7%
- Vibration compensator use requires prior consultation with Pfeiffer Vacuum and on-site torque compensation measures
- Always perform mechanical installation before electrical connection; use flexible HV connection if pump is fastened at the bottom part
- Do not start the pump with open vacuum connections; wear protective gloves; remove protective covers only immediately before flanging
Water cooling standard (air cooling optional); cooling water quality and connection requirements are specified in tables (Tbl. 8/9). Accessory connections are pre-set at the TC 400 electronic drive unit (Tbl. 10).
- Cooling water composition and connection requirements defined in Tbl. 8 and Tbl. 9 (p. 93)
- Accessory connections pre-assigned to electronic drive unit TC 400 per Tbl. 10 (p. 94)
Covers grounding of the turbopump (ground terminal, protection class I sticker marking the ground position) and connecting the TC 400 electronic drive unit to the DC power supply pack. Only power supply packs meeting double-insulation safety standards (IEC 61010-1 / 60950-1 / 62368-1) or original Pfeiffer packs may be used.
- Ground the turbopump via the designated ground terminal (protection class 1)
- Connect TC 400 to power supply pack via DCin connector
- Install a mains disconnection device per SEMI-S2 with ≥ 10,000 A interruption rating
Commissioning, operating modes (standalone without operating unit, via 'remote' 26-pole D-Sub, via E74 connection, via Pfeiffer control unit, via field bus), switch-on procedure, operation monitoring via LEDs and temperature monitoring, and switching off/venting. Factory settings of the drive unit are documented at delivery (Tbl. 11), delayed venting factory settings in Tbl. 13.
- Operating mode indicated by LEDs on the electronic drive unit (Tbl. 12, p. 100)
- Remote connection: 26-pole D-Sub socket on TC 400
- Mating plugs enable automatic pump run-up on power-on — only connect after mechanical installation
- On overtemperature the drive unit first reduces drive output, then switches off
Maintenance intervals and responsibilities by level; procedures for removing/reassembling the operating fluid reservoir (fore-vacuum side bearing lubrication) and for replacing the TC 400 electronic drive unit. After exchanging the drive unit, the nominal speed specification must be confirmed — via parameter [P:777] set once with the enclosed Pfeiffer Vacuum SpeedConfigurator, or per Tbl. 14 characteristic nominal speeds (e.g., HiPace 700 Plus: 49,200 rpm / 820 Hz).
- Operating fluid reservoir lubricates fore-vacuum-side ball bearing; replace per section 7.3 procedures
- After replacing the electronic drive unit TC 400, confirm speed specification via [P:777] (SpeedConfigurator included with exchange units)
- Decontaminate parts exposed to toxic media before maintenance; observe operating fluid safety data sheet
- De-energize fully: switch off mains, wait for f = 0 (rotor standstill), unplug, secure against restart
Shutting down for longer periods and recommissioning procedures, plus disposal information for turbopumps in accordance with general disposal regulations.
- Seal open connections with original protective covers after removal
- Dispose of pump and operating fluid reservoir per applicable regulations
Troubleshooting table (Tbl. 16/15) for turbopump malfunctions, Pfeiffer Vacuum service solutions, and spare parts list for HiPace 700 Plus with accessories ordering information (Tbl. 17).
- Troubleshooting table Tbl. 15 (p. 111)
- Spare parts for HiPace 700 Plus listed in Tbl. 16 (p. 114)
- HiPace 700 accessories ordering table Tbl. 17 (p. 117)
Technical data table for HiPace 700 Plus (Tbl. 20), materials in contact with process media (Tbl. 21), pressure and gas throughput unit conversion tables, and dimension drawings for the three TC 400 variants (DN 160 ISO-K, CF-F, ISO-F). Includes EC and UK Declarations of Conformity.
- Dimension figures for HiPace 700 Plus | TC 400 with DN 160 ISO-K (p. 120), DN 160 CF-F and DN 160 ISO-F (p. 121)
- Materials in contact with process media listed in Tbl. 21
- Pressure unit and gas throughput conversion tables provided
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal rotor speed of HiPace 700 Plus is 49,200 rpm (820 Hz) per rating plate example p.70 ↗ | Use for speed confirmation after drive unit exchange via parameter [P:777] and for vibration/balance calculations | HiPace 700 Plus (all flange variants) |
| Pump must be fastened with 6× M8 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, tightening torque 25 Nm ± 10%, into mounting plate ≥ 3 mm thick with > 270 MPa tensile strength p.84 ↗ | Dimension the customer mounting plate and fixings to absorb rotor-burst torques; use calibrated torque wrench (factor ≤ 1.6) | HiPace 700 (plus HiPace 350/400/450/800) |
| Maximum torque occurring in event of rotor burst: 4200 Nm (ISO 27892); maximum axial load on HV flange: 1000 N (100 kg), one-sided load not permitted p.84 ↗ | Counter flange and superstructure must absorb burst torque on operator side; keep attached mass ≤ 100 kg | HiPace 700 |
| Counter flange flatness tolerance: ± 0.05 mm; flange material tensile strength ≥ 170 N/mm² at 2.5 × d or ≥ 270 N/mm² at 1.5 × d screw engagement p.85 ↗ | Prevents housing stress, leakage and degraded running characteristics from uneven counter flanges | HiPace 700 |
| ISO-K flange connections are tightened cross-wise in 3 steps: 5, 15, 25 ± 2 Nm, WAF 15 p.86 ↗ | Specified torque sequence ensures leak-tight joint without flange stress | DN 160 ISO-K / ISO-F variants |
| Splinter shield DN 160 reduces pumping speed by up to 23% (Ar) / 20% (N2); protective screen DN 160 only 6-7% for N2/Ar p.85 ↗ | For sizing effective pumping speed choose protective screen where particle protection permits, to minimize speed loss | DN 160 HV flange variants |
| Max. permissible inlet pressure 1500 hPa absolute at suction side and at venting/sealing gas connections p.74 ↗ | Exceeding it destroys the pump; interlock process overpressure upstream | whole series |
| Ambient temperature limits: 5–35 °C air-cooled, 5–40 °C water-cooled; max. rotor temperature 90 °C; max. bakeout temperature at HV flange 120 °C p.77 ↗ | Defines cooling choice and bakeout process envelope; drive reduces power then shuts off on overtemperature | whole series |
| Permissible surrounding magnetic field 6 mT and max. irradiated thermal output 4.2 W p.77 ↗ | Install magnetic/thermal shielding if process environment exceeds these values | whole series |
| TC 400 electronic drive unit operating voltage: 24 V DC or 48 V DC variants; no built-in mains switch — external mains disconnection device per SEMI-S2 with ≥ 10,000 A interruption rating required p.73 ↗ | Power supply design must include customer-side breaker and double-insulated supply pack per IEC 61010-1/60950-1/62368-1 | whole series with TC 400 |
| Fore-vacuum connection is DN 25 ISO-KF; with oil-sealed backing pumps the pump must be mounted with fore-vacuum connection pointing vertically down (deviation max. ± 25°) p.86 ↗ | Prevents oil backflow from fore-vacuum line into the turbopump | whole series with oil-sealed backing pumps |
| Hybrid bearing: wear-free permanent magnetic bearing on HV side, ceramic-ball bearing with operating fluid reservoir lubrication on fore-vacuum side p.80 ↗ | Schedule operating fluid reservoir replacement (section 7.3); HV-side bearing is maintenance-free | whole series |
| Vibration compensator installation requires prior consultation with Pfeiffer Vacuum; rotor jamming generates destructive torques per ISO 27892 that can shear off the pump p.74 ↗ | Design on-site torque compensation before using a vibration compensator on vibration-sensitive systems | whole series |
| Protection degree IP54 / Type 12 (UL 50E), protection class III, overvoltage category II, pollution degree 2; installation altitude max. 5000 m, air pressure 530–1060 hPa p.77 ↗ | Defines permissible installation environments and enclosure requirements | whole series |
| Venting must not exceed permissible venting rates; do not fully unscrew the venting screw during venting (vacuum contact injury risk) p.74 ↗ | Use controlled/automatic venting within factory delayed-venting settings (Tbl. 13) | whole series |
| After replacing the electronic drive unit, nominal speed must be confirmed via parameter [P:777]; a SpeedConfigurator for one-time setting is included with exchange units p.107 ↗ | Mandatory commissioning step after TC 400 exchange; otherwise pump runs at incorrect speed preset | whole series with TC 400 |
| Only dry and inert gases may be pumped; pumping explosive, corrosive, condensing, fluid or dusty media is improper use and voids warranty/liability p.77 ↗ | Corrosive processes require the separate C (corrosive gas) version, not this standard pump | whole series |
| Damaging or removing the factory warranty seal voids the warranty p.70 ↗ | Do not open the pump housing; send to service for internal repairs | whole series |
| Mating plugs on the TC 400 enable automatic pump run-up at power-on; they must only be connected after mechanical installation and the pump switched on immediately prior to operation p.74 ↗ | Prevents unexpected rotor start-up with open HV flange — cut/severance hazard | whole series with TC 400 |