HiPace 700 with TC 400 — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum + Fab Solutions HiPace 700 turbomolecular vacuum pump operated with the TC 400 electronic drive unit (Document: Instruction Manual HiPace 700 TC 400_EN_en, 72 pages). Covers flange variants DN 160 ISO-K, DN 160 ISO-F and DN 160 CF-F; forevacuum connection DN 25 ISO-KF; cooling variants water cooling and air cooling.
Section-by-Section Technical Breakdown
Defines risk levels (DANGER to NOTE), general safety precautions, personnel qualification levels, intended use (vacuum generation of dry inert gases with a suitable backing pump, indoor only) and foreseeable misuse. Gives the permissible ambient conditions table for the pump. The pump has redundant temperature monitoring and reduces then shuts off drive output on over-temperature.
- Installation location weatherproof indoor; air pressure 530–1060 hPa; altitude max. 5000 m
- Rel. humidity max. 80% at T < 31 °C, up to 50% at T < 40 °C; protection class III, overvoltage category II
- Permissible protection degree IP54 / Type 12 (UL 50E); degree of pollution 2
- Ambient temperature 5–35 °C (air cooling), 5–40 °C (water cooling)
- Max. radial magnetic field 6 mT; max. irradiated thermal output 4.2 W
- Max. rotor temperature 90 °C; max. bakeout temperature at high vacuum flange 120 °C
- Maintenance levels 1–3 defined; level 2/3 require technical education or Pfeiffer training
Describes the HiPace 700 design with TC 400 electronic drive unit as compact unit, powered by Pfeiffer power supply packs. Cooling is water or air; rotor is hybrid-bearing with permanent magnetic bearing on the high vacuum side and ceramic-ball bearing on the forevacuum side, lubricated by an operating fluid reservoir. Lists scope of delivery including mating plugs for 'remote' (26-pole D-Sub) and 'E74' connections.
- High vacuum connection DN 160 CF-F; forevacuum connection DN 25 ISO-KF
- LED operating mode display; venting screw; sealing gas connection; cooling water coupling
- Hybrid bearing: wear-free permanent magnetic bearing (HV side), ball bearing with ceramic balls (FV side)
- Drive power automatically reduced on excessive temperatures
Rotor safety against crash and bursting is verified per ISO 27892; fastening must absorb rotor-block torques on the operator-side mounting plate. Defines required screws, washers, tools and the exact fastening parameters for HiPace 700. Mechanical installation must always precede electrical connection.
- Fastening: M8 thread, 6 screws, screw-in depth ≥ 1.3 × d, tightening torque 25 Nm ± 10%
- Mounting plate min. thickness 3 mm, tensile strength > 270 MPa; screws strength class ≥ 8.8 galvanized
- Washers DIN EN ISO 7090 or 7092; hex key WAF 6; calibrated torque wrench (tightening factor ≤ 1.6)
- Create flexible connection to high vacuum flange when bolted to bottom part
Specifies counter flange design requirements (flatness, material strength, axial load) and burst torque. Covers earthquake protection, splinter shields/protective screens with pumping speed reduction values, vibration compensator precautions, mounting orientation (±25° with oil-sealed backing pumps) and step-by-step ISO-K, ISO-F and CF-F flange connection procedures with torques.
- Burst torque (ISO 27892): 4200 Nm; max axial load on HV flange 1000 N (100 kg), one-sided load not permitted
- Counter flange flatness ± 0.05 mm; min tensile strength 170 N/mm² at 2.5 × d, 270 N/mm² at 1.5 × d
- Splinter shield DN 160 pumping speed reduction: H2 6%, He 9%, N2 20%, Ar 23%
- Protective screen DN 160 reduction: H2 1%, He 2%, N2 6%, Ar 7%
- With oil-sealed backing pumps: forevacuum connection downwards vertical, deviation max. ± 25°
- ISO-K and ISO-K→ISO-F connections: tighten crosswise in 3 steps, 5/15/25 ± 2 Nm, WAF 15
- ISO-F to ISO-F connections: crosswise in 3 steps, 5/15/22 ± 2 Nm
- Vibration compensator requires prior consultation with Pfeiffer + Fab Solutions and additional HV-flange securing
Section covers connecting the forevacuum side (DN 25 ISO-KF example given), the cooling water supply, and accessory connections. Sections 5.4–5.6 exist in the table of contents with corresponding graphics for forevacuum connection and cooling water supply.
- Forevacuum connection DN 25 ISO-KF with protective cover included in delivery
- Cooling water connection via connection coupling (Graphic 16, p.29)
Covers earthing of the pump via earthing connection and establishing the electric connection between the TC 400 electronic drive unit and the power supply pack (Graphic 18, p.32). PE connection must always be secure and plug connections must never be disconnected during operation.
- Earthing cable connection shown in Graphic 17 (p.31)
- DC voltage supply via connecting plug on electronic drive unit connecting panel
Commissioning and operating modes: operation without operating unit, via multi-function 'remote' connection (26-pole D-Sub), via 'E74' connection, via Pfeiffer control unit, or via field bus. Includes switching on, LED operating mode display, temperature monitoring, switching off and venting procedures.
- Operating modes: standalone, remote (26-pole D-Sub), E74, control unit, field bus
- LED operating mode display on electronic drive unit
- Before working on HV connection wait until rotor speed f = 0
- Never operate with high vacuum connection open
Contains a maintenance checklist, and procedures for replacing the operating fluid reservoir (which lubricates the forevacuum-side bearing) and the TC 400 electronic drive unit, including confirming the speed specification after drive unit replacement. Warranty seal damage/removal voids warranty.
- Inspection/maintenance checklist (Table p.43)
- Operating fluid reservoir removal/installation procedures (p.45-47)
- Electronic drive unit removal/installation (p.48-50)
- Speed specification must be confirmed after drive unit replacement (p.51)
Shutdown for longer periods and recommissioning procedures, plus general disposal information and specific disposal guidance for turbomolecular vacuum pumps.
- Shutting down for longer periods and recommissioning procedures (p.52)
- Disposal of turbomolecular pumps (p.53)
Malfunction tables (p.54), service solutions from Pfeiffer Vacuum + Fab Solutions (p.58), spare parts list for HiPace 700 (Graphic 23, p.59) and accessories with ordering information (p.60-62). Only accessories/spare parts listed in this manual may be used.
- Malfunction troubleshooting table (p.54-57)
- Spare parts diagram HiPace 700 (p.59)
- Accessory ordering information (p.60-62)
General data, detailed technical data table, substances in contact with pumped media, and dimension drawings for HiPace 700 with TC 400 in DN 160 ISO-K, DN 160 CF-F and DN 160 ISO-F variants. EU and UK Declarations of Conformity close the manual.
- Dimension drawings: DN 160 ISO-K (p.68), DN 160 CF-F and DN 160 ISO-F (p.69)
- Substances in contact with pumped media listed (p.68)
- EU Declaration of Conformity (p.70) and UK Declaration of Conformity (p.71)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Ambient temperature limits: 5–35 °C with air cooling, 5–40 °C with water cooling p.10 ↗ | Select water cooling if cabinet/process ambient exceeds 35 °C; higher temperatures are possible only with reduced backing pressure or gas throughput | HiPace 700, both cooling variants |
| Maximum permissible bakeout temperature at the high vacuum flange is 120 °C; maximum rotor temperature 90 °C p.10 ↗ | Limit UHV bakeout cycles to 120 °C at the flange; pump self-protects via redundant temperature monitoring (derates then shuts off) | HiPace 700 whole series |
| Permissible radial magnetic field max. 6 mT and max. irradiated thermal output 4.2 W p.10 ↗ | Install magnetic and thermal shielding if process equipment (heaters, magnets) exceeds these values | HiPace 700 whole series |
| Fastening to mounting plate: M8 thread, 6 screws, screw-in depth ≥ 1.3 × d, torque 25 Nm ± 10%; plate min. 3 mm thick with > 270 MPa tensile strength, screws class ≥ 8.8 p.17 ↗ | Mounting plate must absorb rotor-block torques; undersized plates or screws risk pump shearing off on rotor crash | HiPace 700 |
| Maximum axial load on the high vacuum flange: 1000 N (100 kg), one-sided load not permitted; counter flange flatness ± 0.05 mm p.18 ↗ | Chamber/superstructure weight must be externally supported; uneven counter flanges cause housing stress and leaks | HiPace 700, DN 160 flange variants |
| Theoretical burst torque per ISO 27892: 4200 Nm p.18 ↗ | System design (mounting, flange connections, vibration compensators) must absorb this torque on the operator side; note ISO-K connections may twist on rotor blockage but remain leak-tight | HiPace 700 whole series |
| Splinter shield DN 160 reduces pumping speed: H2 −6%, He −9%, N2 −20%, Ar −23%; protective screen DN 160: H2 −1%, He −2%, N2 −6%, Ar −7% p.19 ↗ | Size the pump for effective pumping speed after screen/shield losses, especially for nitrogen/argon processes | HiPace 700 DN 160 |
| With oil-sealed backing pumps, mount the pump with forevacuum connection downwards vertical, permissible deviation ± 25°; with dry compressing backing pumps all orientations are allowed p.20 ↗ | Prevents oil backflow from the forevacuum line into the turbo; support piping so no forces act on the pump | HiPace 700 with oil-sealed backing pumps |
| ISO-K / ISO-K to ISO-F flange connections: tighten crosswise in 3 steps at 5, 15, 25 ± 2 Nm (WAF 15, calibrated torque wrench, tightening factor ≤ 1.6) p.21 ↗ | Following the staged torque sequence prevents flange stress and leaks; use only approved Pfeiffer mounting kits | HiPace 700 DN 160 ISO-K / ISO-F mating |
| ISO-F to ISO-F flange connections: crosswise tightening in 3 steps at 5, 15, 22 ± 2 Nm p.24 ↗ | Lower final torque than ISO-K mating — do not exceed 22 Nm on ISO-F/ISO-F joints | HiPace 700 DN 160 ISO-F |
| Minimum tensile strength of counter flange material: 170 N/mm² at 2.5 × d screw-in depth or 270 N/mm² at 1.5 × d p.18 ↗ | Verify flange material and screw-in depth when using tapped holes; aluminum counter flanges need deeper engagement | HiPace 700 high vacuum connection |
| Rotor bearing is hybrid: wear-free permanent magnetic bearing on high vacuum side, ceramic-ball bearing on forevacuum side lubricated by an operating fluid reservoir p.13 ↗ | The operating fluid reservoir is a wear item requiring periodic replacement (section 7.3); plan this in maintenance schedules | HiPace 700 |
| Forevacuum connection DN 25 ISO-KF; high vacuum connection DN 160 (CF-F standard design, ISO-K/ISO-F variants) p.13 ↗ | Backing pump and lines must be matched to DN 25 forevacuum sizing and deliver the required maximum forevacuum pressure | HiPace 700 all variants |
| Permissible protection degree IP54 / Type 12 per UL 50E; protection class III, overvoltage category II; air pressure 530–1060 hPa, altitude max. 5000 m p.10 ↗ | Verify enclosure rating and site altitude before installation; use only in weatherproof indoor locations | HiPace 700 with TC 400 |
| Interface options for operation: without operating unit, 26-pole D-Sub 'remote' connection, 'E74' connection, Pfeiffer control unit, or field bus; mating plugs for remote and E74 are supplied type-dependent p.14 ↗ | Plan system I/O integration around the remote D-Sub or E74 interface; field bus available for automated systems | HiPace 700 with TC 400 |
| Before any work on the high vacuum connection, wait until rotor has fully stopped (f = 0); never start the pump with open vacuum connections; never disconnect plugs during operation p.9 ↗ | Rotor run-down time must be respected in interlock logic and service procedures to avoid injury from rotating rotor | HiPace 700 whole series |
| Use only for evacuation of dry, inert gases; misuse includes pumping explosive/corrosive media, condensing vapors, fluids, dust, and operation in potentially explosive areas or ionizing radiation p.10 ↗ | Process gas compatibility must be verified against these restrictions before pump selection | HiPace 700 whole series |
| Damaging or removing the factory warranty seal results in loss of warranty p.6 ↗ | Do not open sealed assemblies; repairs beyond maintenance level 1 must go to trained/authorized personnel or Pfeiffer service | HiPace 700 whole series |
| Intended use requires operation with a suitable backing pump capable of delivering up to the required maximum forevacuum pressure; pump is for closed indoor areas only p.10 ↗ | Backing pump selection is a mandatory part of system design; check forevacuum pressure limits of the datasheet | HiPace 700 whole series |
| Keep lines and cables away from hot surfaces > 70 °C; relative humidity limit 80% at T < 31 °C, decreasing to 50% at T < 40 °C p.9 ↗ | Route cabling with thermal separation; verify site humidity against the derated limits at elevated temperatures | HiPace 700 with TC 400 |