Induztr
Engineering Digest Document Viewer Related Models

HiPace 80 Neo with TC 80 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Original operating instructions (DE/EN) for the Pfeiffer Vacuum HiPace 80 Neo turbopump with integrated electronic drive unit TC 80. Variants (Tbl. 1) differ by high vacuum flange (DN 63 ISO-K, DN 63 CF-F, DN 40 ISO-KF), sealing gas connection (with/without, G 1/4" fore-vacuum thread), and cooling (convection standard; air or water cooling optional). Companion document: operating instructions for electronic drive unit 'TC 80 standard' (PT 0659 BN).

Section-by-Section Technical Breakdown

pp. 74-81
Safety, Limits of Use and Personnel Qualification

Risk-based safety instructions per Machinery Directive 2006/42/EC and EN ISO 12100 covering transport, installation, operation, maintenance and malfunction hazards. Defines permissible ambient conditions, proper use (vacuum generation of dry, inert gases indoors) and foreseeable improper use. Defines three maintenance qualification levels.

  • Never exceed 1500 hPa (absolute) at suction side, venting and sealing gas connections
  • Mains disconnection device per SEMI-S2 required; circuit breaker interruption rating ≥ 10,000 A
  • Power supply pack must meet double insulation requirements (IEC 61010-1/60950-1/62368-1)
  • Surface temperature can exceed 70 °C — touch protection required if accessible
  • Rotor jamming generates destructive torques per ISO 27892; vibration compensator use requires prior Pfeiffer Vacuum contact
  • Improper use includes pumping explosive/corrosive media, condensing vapors, fluids, dust, use in explosive atmospheres or ionizing radiation
pp. 86-91
Installation and Flange Connections

Covers fastening the pump bottom part to a mounting plate, high vacuum side connection (counter flange dimensioning, earthquake protection, splinter shields, vibration compensator, mounting orientation) and flange-specific installation for ISO-KF, ISO-K and CF flanges. Mechanical installation must always precede electrical connection.

  • Bottom part fastening: M5 screws, 4×, strength ≥ 8.8, engagement depth ≥ 1.3×d, torque 6 Nm ±10%, mounting plate ≥2 mm thick, >270 MPa tensile strength
  • Counter flange: max burst torque 620 Nm; max axial load on HV flange 200 N (20 kg), one-sided load not permitted; flatness ±0.05 mm; 170 N/mm² at 2.5×d, 270 N/mm² at 1.5×d
  • ISO-KF clamping ring screws: 3.7 Nm tightening torque
  • Splinter shield DN 63 reduces N2 pumping speed by ~15%, protective screen DN 63 only ~4%
  • Oil-sealed backing pumps: mount with fore-vacuum connection pointing downwards, deviation max ±25°
  • Use calibrated torque wrench (tightening factor ≤ 1.6)
  • Mechanical installation before electrical; never start with open vacuum connections
pp. 95-99
Sealing Gas and Electrical Connections (structure per table of contents)

Sealing gas (option) is connected either directly or via external sealing gas valve/throttle. Electrical connection includes grounding the turbopump via the functional earth connector and connecting the electronic drive unit to the power supply pack.

  • Ground the turbopump via the ground terminal / functional earth (protection class III)
  • Never disconnect plug connections during operation
pp. 100-104
Operation, Monitoring and Venting

Commissioning, operating modes (without operating unit, via X3 multi-function connection, or via Pfeiffer Vacuum control unit), switching on/off, LED status display, temperature monitoring and venting with factory-set delayed venting. The drive unit automatically reduces power then shuts off if the rotor temperature limit is exceeded.

  • LED operating mode display on the electronic drive unit
  • Redundant temperature monitoring makes the pump intrinsically safe
  • Venting screw must not be fully unscrewed during venting
pp. 105-110
Maintenance and Service

Maintenance intervals per checklist, replacement of the operating fluid reservoir (which may contain toxic substances from pumped media — decontaminate and observe its safety data sheet), replacement of the electronic drive unit, and confirming the speed specification. Maintenance levels 1–3 define who may perform which work.

  • Level 1: trained customer; Level 2: technically educated customer/Pfeiffer technician; Level 3: Pfeiffer-trained customer/service technician
  • Secure the pump against restart; wait for f = 0 before all work
  • Decontaminate parts exposed to toxic media before maintenance
pp. 111-115
Decommissioning, Disposal and Malfunctions

Shutting down for longer periods and recommissioning procedures, general and pump-specific disposal information, and a troubleshooting table (Tbl. 13) for malfunctions.

  • Dispose of operating fluid reservoir according to applicable regulations
  • Troubleshooting table provided for pump malfunctions
pp. 117-121
Spare Parts and Accessories

Overview of available spare parts for HiPace 80 Neo (Tbl. 14) and accessory ordering information (Tbl. 15). Only accessories and spare parts listed in these instructions may be used.

  • Use only original accessories or Pfeiffer-approved fixing material
pp. 122-126
Technical Data and Dimensions

Technical data for HiPace 80 Neo (Tbl. 18) with optional output connection flanges, materials in contact with process media (Tbl. 19), pressure and gas throughput unit conversion tables, and dimensional drawings for DN 40 ISO-KF, DN 63 ISO-K and DN 63 CF-F versions.

  • Rated speed 90,000 min⁻¹ / 1500 Hz
  • Mass 1.7 kg
  • Dimensional drawings for all three HV flange variants

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Rated rotor speed is 90,000 min⁻¹ (1500 Hz) and pump mass is 1.7 kg (rating plate data). p.72 ↗ High-speed rotor demands strict adherence to fastening torques and counter-flange rigidity; low mass simplifies mounting but burst torque is significant. whole series
Bottom part fastening: 4× M5 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3×d, tightening torque 6 Nm ±10%, mounting plate min. 2 mm thick with >270 MPa tensile strength. p.87 ↗ Specifies exact mounting plate and screw requirements for safe anchoring against rotor-jam torques. HiPace 80 Neo (table also lists HiPace 30/60/80, SplitFlow 50/80 family)
Maximum burst torque on the high vacuum flange is 620 Nm (per ISO 27892, theoretically calculated; not reached in any experimental test). p.88 ↗ Customer-side structure must absorb 620 Nm in case of rotor shaft breakage. HiPace 80 Neo
Maximum permissible axial load on the high vacuum flange is 200 N (20 kg), one-sided load not permitted; counter flange flatness ±0.05 mm. p.88 ↗ Limits superstructural weight on the HV flange and requires precise counter-flange machining to avoid housing stress/leaks. HiPace 80 Neo
Flange material tensile strength requirement: 170 N/mm² at 2.5×d engagement, 270 N/mm² at 1.5×d engagement. p.88 ↗ Defines minimum counter-flange material strength relative to screw engagement depth. HiPace 80 Neo
Never exceed 1500 hPa (absolute) inlet pressure at the suction side or at venting/sealing gas connections. p.76 ↗ Overpressure destroys the pump with ejection risk; process overpressure must be blocked from reaching the pump. whole series
Maximum permissible rotor temperature 90 °C; maximum bakeout temperature at the HV flange 120 °C. p.79 ↗ Constrains bakeout procedures; exceeding rotor temperature causes automatic drive power reduction then shutdown. whole series
Ambient temperature limits: 5–30 °C with convection cooling (no gas throughput), 5–35 °C with air cooling, 5–40 °C with water cooling. p.79 ↗ Cooling choice directly extends the permissible ambient temperature range. HiPace 80 Neo
Permissible surrounding magnetic field: 3.1 mT (air cooling) / 3.7 mT (water cooling); max irradiated thermal output 0.9 W / 1.2 W respectively. p.79 ↗ Magnetic shielding or thermal shielding needed near magnets, heaters or high-current equipment. HiPace 80 Neo
Protection degree IP54 / Type 12 (UL 50E), protection class III, overvoltage category II, installation altitude max 5000 m, air pressure 530–1060 hPa. p.79 ↗ Defines housing/environment requirements and electrical supply category. whole series
ISO-KF clamping ring screws must be tightened evenly to 3.7 Nm. p.90 ↗ Ensures correct flange sealing without overloading the aluminium HV flange. DN 40 ISO-KF variant
A mains disconnection device per SEMI-S2 and a circuit breaker with interruption rating ≥ 10,000 A must be installed; pump and drive unit have no built-in mains switch. p.75 ↗ Integrator must supply the disconnection hardware; the drive unit alone is not a safe isolation point. whole series
Power supply pack must satisfy double insulation between mains input and output per IEC 61010-1 / IEC 60950-1 / IEC 62368-1. p.75 ↗ Only approved or equivalent power supplies may be used; otherwise risk of fatal electric shock. whole series
Splinter shield DN 63 reduces pumping speed by 15% (N2), protective screen DN 63 by 4%; DN 40 splinter shield reduces N2 speed by 17%. p.89 ↗ Must be included in effective pumping speed calculations when foreign-matter protection is used. HiPace 80 Neo
With oil-sealed backing pumps, fore-vacuum connection must point vertically downwards with max deviation ±25°; with dry compressing backing pumps all orientations are permitted. p.90 ↗ Mounting orientation selection depends on backing pump type; prevents oil backflow into the pump. HiPace Neo series
Rotor bearing is hybrid: wear-free permanent magnetic bearing on the HV side, ceramic-ball bearing lubricated by an operating fluid reservoir on the fore-vacuum side. p.83 ↗ The reservoir is a maintenance item (replaceable) and the bearing type is relevant to lubrication/magnetic-field catalog parameters. whole series
Sealing gas connection (option, G 1/4" fore-vacuum thread variants) has an integrated throttle with sinter filter and cannot be retrofitted. p.71 ↗ Sealing gas capability must be chosen at ordering time. HiPace 80 Neo sealing gas variants
HV flange variants: DN 63 ISO-K (aluminium), DN 63 CF-F (stainless steel), DN 40 ISO-KF (aluminium); fore-vacuum 1/4" thread. p.71 ↗ Determines chamber interface and corrosion compatibility (CF = stainless steel, metal-sealed per ISO 3669). HiPace 80 Neo variants
Vibration compensators may only be installed vertical, with prior consultation with Pfeiffer Vacuum and additional HV-flange securing, due to ISO 27892 shear-off risk. p.89 ↗ Do not fit vibration compensators ad hoc; on-site torque compensation measures are mandatory. whole series
Surface temperature of the pump can exceed 70 °C during operation; cables/lines must be kept >70 °C away and touch protection fitted if accessible. p.76 ↗ Thermal protection and cable routing must be planned in the system layout. whole series