Covered Models & Document Scope

Operating instructions (126 pages, bilingual DE/EN, original) for Pfeiffer Vacuum HiPace 800 H turbopump with electronic drive unit TC 400, covering three variants: HiPace 800 H DN 200 ISO-F, DN 200 ISO-K, and DN 200 CF-F. Related documents: TC 400 standard (PT 0203 BN), TC 400 PB Profibus (PT 0244 BN), TC 400 E74 Semi E74 (PT 0302 BN), TC 400 DN DeviceNet (PT 0352 BN), TC 400 EC EtherCat (PT 0452 BN).

Section-by-Section Technical Breakdown

pp. 72-79
Safety, Limits of Use and Personnel Qualification

Risk assessment per Machinery Directive 2006/42/EC and EN ISO 12100. Covers transport, installation, operation, maintenance and malfunction hazards including rotor burst torques per ISO 27892, electric shock, toxic media and overpressure. Defines permissible ambient conditions and improper use cases. Maintenance is tiered into levels 1-3 with qualification requirements.

  • Mains disconnection device per SEMI-S2 required with circuit breaker interruption rating ≥ 10,000 A (pump/TC 400 have no mains switch)
  • Power supply pack must meet double isolation per IEC 61010-1, IEC 60950-1, IEC 62368-1
  • Never exceed 1500 hPa (absolute) inlet pressure on suction side or venting/sealing gas connection
  • Ambient conditions: 530-1060 hPa air pressure, altitude ≤ 5000 m, ambient temp 5-35°C air-cooled / 5-40°C water-cooled, IP54/Type 12, surrounding magnetic field ≤ 6 mT, irradiated thermal output ≤ 4.2 W
  • Max rotor temperature 90°C; max bakeout temperature at HV flange 120°C
  • Proper use: only for vacuum generation of dry, inert gases with suitable backing pump, indoors
  • Improper use forbidden: explosive/corrosive media, condensing vapors, fluids, dust, overpressure, ionizing radiation, explosive atmospheres
  • Maintenance level 1 = trained customer; level 2 = customer with technical education; level 3 = customer with Pfeiffer service training
pp. 80-82
Product Description

The HiPace 800 H forms a compact unit with TC 400 drive (48 V DC). The 'H' suffix denotes high compression, especially for light gases. Hybrid bearing: wear-free permanent magnetic bearing on the HV side, ball bearing with ceramic balls on the fore-vacuum side lubricated by an operating fluid reservoir.

  • Drive: electronic drive unit TC 400, operating voltage 48 V DC
  • Cooling: water cooling standard; air cooling optional; drive power auto-reduced on excess temperature
  • Rotor bearing: permanent magnetic (HV side, wear-free) + ceramic ball bearing (VV side) with operating fluid reservoir
  • Connections: DN 200 high vacuum flange (ISO-K / ISO-F aluminum; CF-F stainless steel), DN 25 ISO-KF fore-vacuum
  • Sealing gas connection and venting screw on pump bottom part
  • Shipment includes: pump with TC 400, protective covers HV/VV, remote and E74 mating plugs (type-dependent), operating instructions
  • Product naming: HiPace family, size 10-2800, attributes (H = high compression, C = corrosive gas, M = magnetic bearing, etc.)
pp. 83-96
Installation and Mounting

Covers fastening the pump to the mounting plate, designing the counter flange, HV-side connections (ISO-K/ISO-F/CF-F), fore-vacuum connection, cooling water connection and electrical supply/grounding. Rotor burst safety per ISO 27892 governs anchoring requirements; a vibration compensator requires prior consultation with Pfeiffer Vacuum.

  • Fastening: 6× M8 screws strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, tightening torque 25 Nm ± 10%, mounting plate min 3 mm thick, tensile strength > 270 MPa
  • Burst torque (rotor shaft breakage, ISO 27892): max 4200 Nm for HiPace 800
  • Max permissible axial load on HV flange: 1000 N (100 kg), no one-sided load permitted; counter flange flatness ± 0.05 mm
  • Flange material tensile strength: 170 N/mm² at 2.5 × d or 270 N/mm² at 1.5 × d engagement depth
  • ISO-K connections: tighten cross-wise in 3 steps at 5, 15, 25 ± 2 Nm; tools WAF 15, torque wrench factor ≤ 1.6
  • Splinter shield DN 200 reduces pumping speed: H2 10%, He 13%, N2 26%, Ar 28%; protective screen: 3/3/8/9%
  • Oil-sealed backing pumps: fore-vacuum connection vertically downwards, deviation max ± 25°; dry compressing backing pumps: all orientations allowed
  • Vibration compensator: only vertical passage, additional HV-flange securing required, contact Pfeiffer Vacuum first (shear-off risk per ISO 27892)
  • Flexible HV connection required when pump fastened to bottom part to avoid reactive forces
pp. 97-101
Operation

Commissioning, operating modes (without operating unit, via 'remote' multi-function connection, via E74, via Pfeiffer control unit, via field bus), switching on, operation monitoring via LEDs and temperature monitoring, and switching off/venting. Factory settings of the TC 400 and delayed venting settings are tabulated in the manual.

  • Operating modes: stand-alone, 'remote' 26-pole D-Sub connection, E74 connection, Pfeiffer control unit, field bus
  • Operation monitoring: LED operating mode display on TC 400 (behavior/meaning table provided)
  • Temperature monitoring: drive output reduced first, then switched off at excess temperature (intrinsically safe via redundant monitoring)
  • Venting: do not fully unscrew venting screw; factory settings for delayed venting tabulated
  • Do not start pump with open vacuum connections; mating plugs cause automatic run-up — only connect after mechanical installation
pp. 102-107
Maintenance

Describes maintenance intervals and responsibilities, replacement of the operating fluid reservoir (lubrication of fore-vacuum side bearing), and replacement of the electronic drive unit TC 400 including confirmation of the speed specification afterwards. Decontamination of toxic-contaminated parts is mandatory before any maintenance work.

  • Replace operating fluid reservoir (removal/assembly procedures with figures)
  • Replace electronic drive unit TC 400 (removal, installation, then confirm speed specification per characteristic nominal speed table)
  • Decontaminate parts exposed to toxic process media before maintenance; observe operating fluid safety data sheet
  • Fully de-energize before work: mains off, wait f = 0, unplug, secure against restart
pp. 110-117
Malfunctions, Service, Spare Parts and Accessories

Troubleshooting table for turbopumps, Pfeiffer Vacuum service solutions, HiPace 800 spare parts list, and accessory information with ordering guidance.

  • Troubleshooting table (Tbl. 15) for turbopump malfunctions
  • Spare parts list for HiPace 800 (Tbl. 16, Fig. 24)
  • Accessories and ordering information (Tbl. 17)
pp. 118-124
Technical Data and Dimensions

Technical data table for HiPace 800 H, unit conversion tables (pressure and gas throughput), compression ratio curves vs. supply pressure, and dimension diagrams for the three flange variants with TC 400. Includes EC and UK Declarations of Conformity.

  • Technical data table Tbl. 20 for HiPace 800 H
  • Compression ratio to supply pressure at zero gas throughput (Fig. 25)
  • Dimension drawings: HiPace 800 H | TC 400 | DN 200 ISO-F, ISO-K, CF-F (Figs. 26-28)
  • Conversion tables for pressure units and gas throughput units
  • EC and UK Declarations of Conformity included

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal rotor speed is 49,200 1/min (820 Hz) per rating plate; nominal speeds are confirmed after drive unit replacement per Tbl. 14. p.70 ↗ Use as reference for commissioning checks and post-repair speed confirmation. HiPace 800 H with TC 400
Maximum permissible inlet pressure is 1500 hPa (absolute) on the suction side and on the venting/sealing gas connection. p.74 ↗ Process overpressure above this destroys the pump; design pressure relief accordingly. whole series
Fastening to mounting plate: 6× M8 screws, strength class ≥ 8.8, screw-in depth ≥ 1.3 × d, torque 25 Nm ± 10%, plate min 3 mm / >270 MPa. p.83 ↗ Anchoring must absorb rotor burst torques; use calibrated torque wrench (factor ≤ 1.6). HiPace 350-800 series, incl. HiPace 800
Burst torque (rotor shaft breakage per ISO 27892): max 4200 Nm for HiPace 800. p.84 ↗ Mounting structure and counter flange must be dimensioned to absorb this torque on the operator side. HiPace 800
Maximum permissible axial load on the HV flange is 1000 N (100 kg); one-sided load not permitted; counter flange flatness ± 0.05 mm. p.84 ↗ Limit superstructural fittings (screens, valves) mounted on the pump inlet. HiPace 800
Counter flange material tensile strength requirement: 170 N/mm² at 2.5 × d or 270 N/mm² at 1.5 × d screw engagement depth. p.85 ↗ Check flange material and thread depth before using tapped-hole mounting. HiPace 800
ISO-K flange screws tightened cross-wise in 3 steps: 5, 15, 25 ± 2 Nm; tool WAF 15. p.87 ↗ Ensures even sealing compression and safe retention under rotor blockage. DN 200 ISO-K / ISO-F connections
Splinter shield DN 200 reduces pumping speed (H2 10%, He 13%, N2 26%, Ar 28%); protective screen (3%, 3%, 8%, 9%). p.85 ↗ Factor this loss into effective pumping speed sizing when foreign-particle protection is required. DN 200 variants
Ambient temperature limits: 5-35°C with air cooling, 5-40°C with water cooling; reduced backing pressure/throughput allows higher ambient temps. p.77 ↗ Select cooling type per installation environment; derate at higher ambient temperatures. whole series
Max permissible rotor temperature 90°C; max bakeout temperature at HV flange 120°C. p.77 ↗ Limit bakeout heating jackets to 120°C at the flange; redundant temperature monitoring shuts down drive if exceeded. whole series
Drive unit TC 400 operates at 48 V DC; power supply pack must meet double isolation per IEC 61010-1 / IEC 60950-1 / IEC 62368-1. p.80 ↗ Only use approved or equivalent safety-compliant power supplies; non-approved packs risk fatal electric shock. HiPace 800 H with TC 400
No built-in mains switch: a mains disconnection device per SEMI-S2 with circuit breaker interruption rating ≥ 10,000 A must be installed on site. p.73 ↗ Include an external lockable disconnect in the system electrical design. whole series
Permissible ambient conditions: 530-1060 hPa, altitude ≤ 5000 m, IP54/Type 12, magnetic field ≤ 6 mT, irradiated thermal output ≤ 4.2 W, protection class III, overvoltage category II. p.77 ↗ Shield the pump if magnetic fields or radiant heat exceed limits; indoor weatherproof locations only. whole series
Hybrid bearing: permanent magnetic bearing on the HV side (wear-free), ceramic-ball bearing on the fore-vacuum side lubricated by an operating fluid reservoir. p.80 ↗ Operating fluid reservoir replacement is a required maintenance item; observe its safety data sheet. HiPace 800 H
Fore-vacuum connection is DN 25 ISO-KF; HV flange DN 200 in ISO-K/ISO-F (aluminum) or CF-F (stainless steel). p.80 ↗ Match backing pump and flange hardware; CF-F variant for UHV/metal-seal applications. HiPace 800 H variants
With oil-sealed backing pumps, fore-vacuum connection must point vertically downwards, deviation max ± 25°; with dry compressing pumps all orientations are allowed. p.86 ↗ Avoid oil backflow into the turbopump; support piping so no forces act on the pump. whole series
Use of a vibration compensator requires prior consultation with Pfeiffer Vacuum and on-site torque compensation; rotor jamming generates destructive torques per ISO 27892 which may shear the pump off. p.74 ↗ Do not install vibration compensators without dedicated safety measures. whole series
Access to E74/remote mating plugs enables automatic pump run-up when power is applied. p.74 ↗ Only connect mating plugs after mechanical installation and immediately before operation to prevent injury from rotating rotor. whole series
Operating modes: stand-alone, 26-pole D-Sub 'remote' socket, E74 connection, Pfeiffer control unit, field bus (Profibus/DeviceNet/EtherCAT via TC 400 variants). p.98 ↗ Select TC 400 variant (PB/DN/EC/E74) and its dedicated manual per intended control interface. HiPace 800 H with TC 400
Damaging or removing the factory warranty seal voids the warranty; modifications invalidate the Declaration of Conformity and require operator re-evaluation of system conformity. p.70 ↗ Never open sealed areas; any system integration requires a new conformity assessment before commissioning. whole series