HiPace 800 M with TM 700 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiPace 800 M turbopump with active 5-axis magnetic bearing, integrated TM 700 electronic drive unit (48 V DC). Three variants covered: DN 200 ISO-K (aluminum flange), DN 200 ISO-F (aluminum flange), DN 200 CF-F (stainless steel flange). Applicable drive unit documents: TM 700 Standard (PT 0329 BN), TM 700 PB Profibus (PT 0354 BN), TM 700 E74 (PT 0355 BN), TM 700 DN DeviceNet (PT 0356 BN), TM 700 EC EtherCAT (PT 0503 BN).

Section-by-Section Technical Breakdown

pp. 18-19
Product Description

HiPace M series uses active 5-axis magnetic bearings, making it maintenance-free and oil-free (no hydrocarbon contamination risk). Compact unit with TM 700 electronic drive unit (48 V DC). Cooling options: convection, water, optional air cooling. Scope of delivery includes pump with drive unit, venting valve, protective caps, and manual.

  • 5-axis active magnetic bearing → maintenance-free for entire service life
  • Pre-assembled venting valve for automatic venting after switch-off or power failure
  • Variant flange materials: ISO-K and ISO-F aluminum; CF-F stainless steel
  • Drive: TM 700 electronic drive unit, 48 V DC operating voltage
pp. 21-34
Installation

Mechanical installation must precede electrical connection. High vacuum connection requires customer-side counter flange dimensioning (torque/flatness/load limits per ISO 27892 testing), optional splinter shield/protective screen, vibration compensator (only after consulting Pfeiffer Vacuum), and specific ISO-K/ISO-F/CF flange bolting procedures with defined torque sequences. Also covers fore-vacuum connection, cooling water, venting valve, accessories, grounding, and electrical supply.

  • ISO-K flange bolts: tighten cross-wise in 3 steps, 5 / 15 / 25 ± 2 Nm, WAF 15, torque wrench factor ≤ 1.6
  • Counter flange flatness ± 0.05 mm; max burst torque 6500 Nm; max axial load on HV flange 500 N (50 kg), one-sided load not permitted
  • Flange material minimum tensile strength: 170 N/mm² at 2.5×d or 270 N/mm² at 1.5×d screw engagement
  • Splinter shield DN 200 reduces pumping speed: H2 −10%, He −13%, N2 −26%, Ar −28%; protective screen: H2/He −3%, N2 −8%, Ar −9%
  • Mounting orientations: all orientations with dry backing pumps; with oil-sealed pumps align fore-vacuum connection downward within ± 25°
  • Vibration compensator: install only with vertical passage and additionally secure pump to HV flange; contact Pfeiffer Vacuum first
pp. 35-40
Operation

Commissioning, operating modes (stand-alone, remote multi-pin D-Sub, E74, Pfeiffer control unit, field bus), switch-on, operation monitoring (LEDs, temperature, safety bearing wear, balance), and switching off with electric brake, emergency power supply and venting.

  • Operation monitoring: LED operating mode display, temperature monitoring, safety bearing wear, balance
  • If max permissible operating temperature exceeded, drive unit first reduces output then switches off
  • Delayed venting factory settings defined for the turbopump (Tbl. 12)
pp. 41-45
Maintenance

General maintenance information, intervals and responsibilities per qualification level, and replacement of the electronic drive unit TM 700 (removal, installation, confirming speed specification). Magnetic bearings make the pump itself maintenance-free.

  • Electronic drive unit replacement includes confirming nominal rotation speed specification (Tbl. 13)
  • Maintenance levels 1–3 define who may perform which tasks
pp. 46-47
Decommissioning, Recycling and Disposal

Shutting down for longer periods and recommissioning procedures, plus general disposal information and turbopump-specific disposal. Decontamination required for toxic-media-exposed parts before service or disposal.

  • Decontaminate parts exposed to toxic process media before maintenance or disposal
pp. 48-55
Malfunctions and Service Solutions

Troubleshooting table for turbopump malfunctions and overview of Pfeiffer Vacuum service solutions.

  • Troubleshooting table (Tbl. 14) on page 49
  • Spare parts overview for HiPace 800 M (Tbl. 15) and accessories with ordering information (Tbl. 16)
pp. 56-61
Technical Data and Dimensions

Technical data for HiPace 800 M with TM 700, materials in contact with process media, pressure/gas throughput unit conversion tables, and dimension drawings for DN 200 ISO-K, ISO-F and CF-F versions. Includes EC and UK Declarations of Conformity.

  • Dimension drawings for all three flange variants (Figs. 23–25)
  • Materials in contact with media table (Tbl. 20); FKM used as elastomer

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal rotor speed: 49200 rpm (820 Hz); pump mass 18 kg; protection rating IP54 / Type 12 (UL 50E). p.8 ↗ Anchor pump securely for high-speed rotor energy; install in protected indoor location with IP54-equivalent enclosure. All HiPace 800 M variants
Maximum permissible inlet pressure on suction side and venting/sealing gas connection: 1500 hPa absolute. p.12 ↗ Interlock process to prevent overpressure entering the pump; exceeding it destroys the pump with ejection risk. All variants
Maximum burst torque per ISO 27892: 6500 Nm (HiPace 800 M); counter flange flatness tolerance ± 0.05 mm; minimum flange material tensile strength 170 N/mm² (2.5×d engagement) or 270 N/mm² (1.5×d). p.22 ↗ Customer counter flange and safety connections must absorb 6500 Nm and respect flatness/strength specs to avoid leakage or pump shearing. HiPace 800 M (also listed for HiPace 700 M)
Maximum permissible axial load on the high vacuum flange: 500 N (≈50 kg), one-sided load not permitted. p.22 ↗ Limit attached superstructure (screens, heaters, chambers) weight and support it independently. HiPace 800 M
ISO flange bolts tightened cross-wise in 3 steps: 5, 15, 25 ± 2 Nm using WAF 15 and calibrated torque wrench (factor ≤ 1.6). p.24 ↗ Follow this sequence to avoid housing stress and leaks; use only Pfeiffer approved mounting kits. ISO-K and ISO-F connections
Splinter shield DN 200 reduces pumping speed (H2 10%, He 13%, N2 26%, Ar 28%); protective screen DN 200 (H2 3%, He 3%, N2 8%, Ar 9%). p.23 ↗ Account for pumping-speed loss when sizing the system; use screens for particle protection from the chamber. DN 200 flange variants
With oil-sealed backing pumps, fore-vacuum connection must point downward vertically with max deviation ± 25°; all orientations permitted with dry compressing backing pumps. p.24 ↗ Determines mounting orientation and piping layout to prevent oil backflow. All variants
Ambient temperature limits: 5–30 °C (convection cooling, no gas throughput), 5–35 °C (air cooling), 5–40 °C (water cooling). p.15 ↗ Select cooling type based on installation environment; reduce gas throughput/backing pressure to allow higher ambient temperatures. All variants
Permissible surrounding magnetic field: 5.0 mT; maximum irradiated thermal output: 2.4 W; max rotor temperature 110 °C; max bakeout temperature at HV flange 120 °C. p.15 ↗ Shield the pump from magnetic fields and radiant heat; bakeout limited to 120 °C at the flange. All variants
Installation altitude up to 5000 m (with sufficient cooling), air pressure 530–1060 hPa, rel. humidity max 80 % at T < 31 °C (up to 50 % at T < 40 °C), protection class III, overvoltage category II. p.15 ↗ Defines permissible installation sites; high-altitude or humid installations need derating/sealing measures. All variants
Power supply pack must satisfy double isolation between mains input and output per IEC 61010-1 / IEC 60950-1 / IEC 62368-1; mains disconnection device per SEMI-S2 with circuit breaker interruption rating ≥ 10,000 A required. p.11 ↗ Specifying a non-compliant PSU or omitting the disconnect device is a safety violation and shock hazard. All variants with TM 700
Operating voltage of electronic drive unit TM 700: 48 V DC. p.19 ↗ Dimension the 48 V DC supply and cabling accordingly; Pfeiffer TPS power supply packs recommended. All variants
Rotor bearing is an active 5-axis magnetic bearing → maintenance-free over the entire service life and hydrocarbon-free (oil-free). p.18 ↗ No lubrication schedule; suitable for contamination-sensitive processes; bearing type = magnetic. HiPace M series
Mains disconnection: device is only fully de-energized when mains plug is removed and rotor is at standstill (f = 0); wait for complete stop before working on the pump. p.13 ↗ Lockout procedure must include rotor standstill verification, not just power-off. All variants
Vibration compensators must only be installed with vertical passage, with additional pump securing to the HV flange, and only after consulting Pfeiffer Vacuum; sudden rotor jamming generates destructive torques per ISO 27892 that can shear off the pump. p.23 ↗ On-site torque compensation measures are mandatory when using a vibration compensator. All variants
Fore-vacuum connection is DN 25 ISO-KF; venting valve is pre-assembled for automatic venting on switch-off or power failure; sealing gas connection present. p.18 ↗ Size fore-vacuum lines and venting gas supply to match these connections. All variants
Cooling options: convection, water, optional air cooling; drive unit automatically reduces drive power on excessive temperature. p.18 ↗ Cooling water/air supply must be planned per variant; overtemperature causes automatic power reduction then shutdown. All variants
Improper use explicitly prohibited: pumping explosive, corrosive, condensing, fluid or dusty media; operation in ionizing radiation, explosive areas, magnetic fields > 5.0 mT, or with wrong gas mode/venting rates. p.15 ↗ Use the C (corrosive) version or other variants for aggressive processes; violating limits voids warranty/liability. Standard HiPace 800 M
Electronic drive unit options with separate manuals: TM 700 Standard (PT 0329 BN), PB Profibus (PT 0354 BN), E74 (PT 0355 BN), DN DeviceNet (PT 0356 BN), EC EtherCAT (PT 0503 BN); operation possible via 'remote' D-Sub, E74, Pfeiffer control unit, or field bus. p.7 ↗ Select the drive unit/communication variant matching the fab/tool control architecture; parameters [P:xxx] control configuration. TM 700 drive unit, all variants