HiPace 800 P with TC 400 — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Operating instructions (PT 0319 BEN/F, 1211) for the Pfeiffer Vacuum HiPace 800 P process-version turbopump, operated with the electronic drive unit TC 400 (standard, PB/Profibus, E74/SEMI E74, DN/DeviceNet panel variants) at 48 V DC ±5%. High vacuum flange variants: DN 200 ISO-K and ISO-F (aluminium) and DN 200 CF-F (stainless steel). Scope of delivery: pump with TC 400, mating plugs for 'remote' and 'E74', protective flange covers, operating instructions.
Section-by-Section Technical Breakdown
The HiPace 800 P is a process turbopump forming a complete unit with the TC 400 drive. Rotor bearing is hybrid: permanent magnetic bearing on the high vacuum side, ceramic ball bearing on the fore-vacuum side. Cooling is water (standard) or air (optional). Certified to CAN/CSA-C22.2 No. 61010-1, TÜV ID 0000021320.
- Model designation: HiPace (family) 800 (size/pumping speed class) P (Process)
- HV flange options: DN 200 ISO-K / ISO-F (aluminium), DN 200 CF-F (stainless steel)
- Hybrid bearing: maintenance-free permanent magnetic bearing (HV side), ceramic ball bearing (fore-vacuum side)
- Supply voltage 48 V DC ±5%; only Pfeiffer power supplies (TPS/DCU) allowed
- On excess temperature the TC 400 automatically reduces drive power
Installation is weather-protected indoor, protection category IP 54, protection class III, overvoltage category II, pollution degree 2. Ambient temperature limits depend on cooling method; the pump is intrinsically safe by redundant temperature monitoring and reduces then cuts drive power on overheating.
- Ambient temperature: +5 °C to +35 °C (air cooling), +5 °C to +40 °C (water cooling)
- Relative humidity: max 50% at 40 °C, up to max 80% at 31 °C
- Atmospheric pressure 75-106 kPa; installation altitude max 2000 m
- Permissible surrounding magnetic field ≤6 mT; shielding required above 6 mT
- Higher ambient temperatures permissible with reduced fore-vacuum pressure or gas throughput
Details set-up, HV flange mounting (ISO-K claw clamps, ISO-F screws/studs, CF M8 studs/hex screws), flange loading limits, and torque values. Sudden rotor blocking can generate torques up to 4200 Nm, so fastening must absorb these; mixed flange-type combinations are prohibited. Counter-flange flatness must be ≤0.05 mm.
- Max axial load on HV flange: 1000 N (100 kg), one-sided load not permitted
- ISO-K DN 200: 8 claw clamps, tighten crosswise 5 / 15 / 25 ±2 Nm
- ISO-F DN 200: 12 hex screws or stud screws, tighten crosswise 10 / 20 / 38 ±3 Nm; screw-in depth 1.5 d (>270 N/mm² flange) or 2.5 d (170-270 N/mm²)
- CF DN 200: 24 M8 screws/studs, circular tightening 22 ±2 Nm, re-check torque after seal flow
- Counter flange evenness max 0.05 mm over whole surface; flange material strength min 170 N/mm²
- Sudden rotor blocking torques up to 4200 Nm — use only Pfeiffer original mounting components, never mix flange types
- With oil-sealed backing pumps, fore-vacuum flange must point vertically downward (±25°)
- Splinter shield DN 200 reduces pumping speed: H2 7%, He 11%, N2 23%, Ar 26%; protection screen: 2/3/7/8%
- Max rotor temperature 90 °C; radiated heat input must not exceed 4.2 W; chamber flange ≤120 °C with casing heating + water cooling
- Vibration damper does not compensate rotor-blocking torques; max temperature at vibration damper 100 °C
Covers power supply, remote plug, accessory connections (4 preconfigured M12 ports), backing pump, air cooling, venting valve, heating jacket, sealing gas and water cooling connections. TC 400 factory presets: A1 air cooling unit, A2 backing pump, B1 venting valve, B2 casing heating; Y-connectors allow two units per port.
- Remote mating plug bridges Pin 2, 3, 4, 14 to Pin 1 (supply) enabling stand-alone operation — automatic start on power-up
- Backing pump must achieve ≤5 hPa; connected via relay box, default accessory port A2
- Air cooling permissible up to +35 °C ambient; default accessory port A1
- Venting valve: max connection pressure 150 kPa absolute; default accessory port B1; intake G 1/8"
- Sealing gas: mandatory above 50% of max gas throughput (rotor cooling); flow rate 17.5-20 sccm; max connection pressure 150 kPa absolute
- Heating jacket only permissible with stainless steel (CF-F) HV flange; min heating duration 4 h; fixing screw torques M5: 6/7/7 Nm, M6: 11/12/12 Nm (cold/heat-up/after cooling)
- Water cooling: hose OD 8 mm / ID 6 mm, fore-line overpressure max 600 kPa, pH 7-9, chloride max 100 mg/kg, water hardness max 10 °dH (178 ppm CaCO3), oxygen max 4 mg/kg
- Water cooling required whenever casing heating is used or pump connected to heated chamber
Commissioning, operation modes (stand-alone, remote, RS-485/DCU/HPU, fieldbus), LED status indication, temperature monitoring and switching off/venting. Factory parameters of the TC 400 are listed; gases with molecular mass >80 require consultation with Pfeiffer Vacuum.
- Factory settings: P:027 gas mode 0 (heavy gases); P:700 max run-up time 8 min; P:701 rotation speed switchpoint 80% nominal; P:707 rotation speed setting 65% nominal; P:708 power consumption 100%; P:720 venting rotation speed 50% nominal; P:721 venting time 3600 s
- LED display: green = pumping station on/off, yellow = warning, red = malfunction; differentiated display only via DCU/HPU
- Switching off requires venting to avoid backstreaming contamination from fore-vacuum side
- Rapid venting: 20 s at pressure rise rate max 15 hPa/s, valve cross-section adapted to chamber size; then vent with any additional valve
- Delayed venting: valve opens at 50% nominal rotation speed or on mains failure, for 3600 s (1 h); restoring mains power aborts venting
- Never operate with open HV flange; wrong gas mode with heavy gases can destroy the pump
User-serviceable items are the operating fluid reservoir and the electronic drive unit. Operating fluid reservoir and bearings must be changed at least every 4 years (bearing via Pfeiffer Service). External cleaning with lint-free cloth and little industrial alcohol; no cleaning fluids inside.
- Change operating fluid reservoir at least every 4 years; change bearing at least every 4 years (Pfeiffer Service)
- Reservoir replacement: 3 Allen screws on end cover, 9 Porex rods replaced with tweezers, felt side of reservoir toward splash nut
- End cover tightening torque: 2.5 Nm with new O-ring
- Do not add additional operating fluid — reservoir is pre-filled
- Decontaminate parts contaminated by pumped media before maintenance; dispose of reservoir per regulations
- Shorter intervals for extreme loads/impure processes to be clarified with Pfeiffer Service
Shutdown for longer periods, restart and disposal procedures, malfunction rectification and service contact. (Content in provided extract limited to table of contents; detailed procedures on pages 29-31.)
- Sections 8-10 cover shutdown, restart, disposal, rectifying malfunctions and service
Spare parts lists for HiPace 800 P and approved accessories with order numbers are given in sections 11 and 12; use of original parts only.
- Only accessories/spare parts named in the manual may be used (otherwise improper use)
Technical data for the HiPace 800 P 48 V DC variant and dimension drawings, plus declaration of conformity. The rating plate example shows S(N2) 260 l/s, 60,000 1/min (1000 Hz), weight 6.7 kg (example plate values).
- Data tables on pages 35-37 cover general data, HiPace 800 P 48 V DC variant and dimensions
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Supply voltage is 48 V DC ±5%, protection class III, only original Pfeiffer power supplies (TPS 400 / DCU 400) permitted p.10 ↗ | Power architecture must use Pfeiffer OEM supplies; third-party supplies void compliance guarantees | HiPace 800 P with TC 400 |
| Ambient temperature limits: +5 to +35 °C with air cooling, +5 to +40 °C with water cooling; use water cooling if ambient >35 °C p.11 ↗ | Cooling selection depends on installation environment; air cooling only for ≤35 °C | HiPace 800 P with TC 400 |
| Sudden rotor blocking can generate torques up to 4200 Nm capable of tearing the pump off p.12 ↗ | HV flange fastening must be rated for this torque; only original Pfeiffer mounting components | HiPace 800 P, all flange variants |
| ISO-K DN 200: 8 claw clamps tightened crosswise in three steps at 5, 15, 25 ±2 Nm p.14 ↗ | Defines correct clamp installation procedure and torque sequence | DN 200 ISO-K variant |
| ISO-F and stud-screw DN 200 connections: 12 screws, crosswise 10 / 20 / 38 ±3 Nm; screw-in depth 1.5 d for flange >270 N/mm², 2.5 d for 170-270 N/mm² p.14 ↗ | Counter flange material strength dictates screw engagement depth | DN 200 ISO-F variant |
| CF DN 200: 24 M8 screws tightened circularly at 22 ±2 Nm, re-check torque after seal flow p.15 ↗ | Copper seal creep requires torque re-check to preserve sealing | DN 200 CF-F variant |
| Counter flange evenness must not exceed 0.05 mm over the whole surface; flange material strength min 170 N/mm² p.12 ↗ | Specify counter-flange machining tolerances and material in chamber design | All HV flange variants |
| Max axial loading of the HV flange is 1000 N (100 kg), one-sided load not permitted p.13 ↗ | Limit attached hardware mass on the inlet flange | HiPace 800 P |
| Max rotor temperature 90 °C; radiated heat input must not exceed 4.2 W; chamber flange ≤120 °C with casing heating and water cooling p.12 ↗ | Hot processes require heat screening; design radiation shields if needed | HiPace 800 P |
| Splinter shield DN 200 reduces pumping speed by 7% (H2), 11% (He), 23% (N2), 26% (Ar); protection screen by 2/3/7/8% p.13 ↗ | Include speed derating in sizing calculations when particle protection is fitted | DN 200 variants |
| Backing pump must generate ≤5 hPa fore-vacuum pressure; default accessory port A2 via relay box p.18 ↗ | Backing pump sizing/selection criterion for the pumping station | HiPace 800 P with TC 400 |
| Sealing gas mandatory above 50% of max gas throughput (rotor cooling); flow 17.5-20 sccm; max supply pressure 150 kPa absolute p.21 ↗ | Process setup must include sealing gas supply for high-throughput/unclean processes | HiPace 800 P |
| Cooling water: OD 8 mm / ID 6 mm hoses, fore-line overpressure max 600 kPa, pH 7-9, chloride ≤100 mg/kg, hardness ≤10 °dH (178 ppm CaCO3), oxygen ≤4 mg/kg p.22 ↗ | Water quality spec for facility cooling loop; dirt trap recommended in fore-line | Water-cooled HiPace 800 P |
| Factory TC 400 parameters: gas mode 0 (heavy gases), max run-up time 8 min, speed switchpoint 80% nominal, set speed 65% nominal, venting speed 50% nominal, venting time 3600 s p.23 ↗ | Baseline configuration for commissioning and parameter comparison | TC 400 drive |
| Delayed venting opens the venting valve for 3600 s at 50% nominal speed or on mains failure; restoring mains aborts venting p.25 ↗ | Defines power-failure venting behavior for process chamber protection | TC 400 with Pfeiffer venting valve |
| Rapid venting: max pressure rise rate 15 hPa/s for 20 s, valve cross-section adapted to chamber size p.25 ↗ | Prevents particle backstreaming and pump/chamber damage during fast venting | HiPace 800 P systems |
| Operating fluid reservoir and pump bearings must be changed at least every 4 years p.26 ↗ | Preventive maintenance scheduling requirement; bearing change via Pfeiffer Service | HiPace 800 P |
| End cover tightening torque 2.5 Nm with new O-ring; 9 Porex rods and pre-filled reservoir replaced during fluid change p.26 ↗ | DIY reservoir replacement procedure; no additional fluid to be added | HiPace 800 P |
| Circuit breaker with interruption rating ≥10,000 A required per SEMI-S2 (pump has no built-in power disconnection device) p.5 ↗ | User must install compliant disconnection device in the supply circuit | TC 400 installations |
| Fore-vacuum flange must point vertically downward (±25°) with oil-sealed backing pumps; storage temperature -25 to +55 °C; magnetic field ≤6 mT p.8 ↗ | Mounting orientation, storage and site EM environment constraints | HiPace 800 P |