HiPace 800 with TC 400 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Original operating instructions (128 pages) for the Pfeiffer Vacuum HiPace 800 turbopump with electronic drive unit TC 400. Covers six variants: DN 200 ISO-K, DN 200 CF-F and DN 200 ISO-F high vacuum flanges, each available with 24 V DC or 48 V DC drive voltage. Related TC 400 controller manuals (standard PT 0203 BN, Profibus PT 0244 BN, Semi E74 PT 0302 BN, DeviceNet PT 0352 BN, EtherCat PT 0452 BN) are referenced separately. Note: provided document excerpt covers pages 1-2 and 67-90 only (manual, safety, product description, transport, installation); later chapters exist per table of contents.

Section-by-Section Technical Breakdown

pp. 82-83
Product Description

The HiPace 800 forms a compact unit with the TC 400 electronic drive unit. It is a hybrid-bearing turbopump: permanent magnetic bearing on the high vacuum side, ceramic ball bearing lubricated by an operating fluid reservoir on the fore-vacuum side. Cooling is water (standard) or air (optional), with automatic drive power reduction on excessive temperature.

  • High vacuum connection DN 200 (ISO-K / ISO-F aluminum flange, CF-F stainless steel flange)
  • Fore-vacuum connection DN 25 ISO-KF
  • Rotor: rating plate example shows 49,200 1/min (820 Hz)
  • Sealing gas connection and venting screw present on pump housing
  • Shipment includes pump with drive unit, protective covers, mating plugs (type-dependent) and manual
pp. 85-90
Installation

Covers fastening to the mounting plate, high vacuum flange connections (ISO-K, ISO-F, CF-F), fore-vacuum connection, cooling water and safety considerations. ISO 27892-tested crash and burst safety dictates fastening requirements; all burst torques must be absorbed on the operator side.

  • Fastening: 6× M8 screws, strength class ≥ 8.8, tightening torque 25 Nm ±10%, screw-in depth ≥ 1.3×d, mounting plate ≥ 3 mm thick with > 270 MPa tensile strength
  • Maximum burst torque (rotor shaft breakage, ISO 27892 calc.): 4200 Nm
  • Max. permissible axial load on HV flange: 1000 N (100 kg), one-sided load not permitted; counter flange flatness ± 0.05 mm; flange material min. 170 N/mm² at 2.5×d or 270 N/mm² at 1.5×d engagement depth
  • Splinter shield DN 200 reduces pumping speed: H2 −10%, He −13%, N2 −26%, Ar −28%; protective screen DN 200: H2 −3%, He −3%, N2 −8%, Ar −9%
  • ISO-K/ISO-F flange screws tightened crosswise in 3 steps: 5, 15, 25 ±2 Nm; torque wrench with tightening factor ≤ 1.6
  • With oil-sealed backing pumps: fore-vacuum connection vertically downward, ±25° max deviation
  • Mechanical installation always before electrical connection; never start pump with open vacuum connections

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Six variants covered: HiPace 800 with DN 200 ISO-K, DN 200 CF-F or DN 200 ISO-F high vacuum flange, each with TC 400 drive unit in 24 V DC or 48 V DC version. p.71 ↗ Select flange type and drive voltage per chamber interface and available power supply; CF-F version has stainless steel flange, ISO versions aluminum. All HiPace 800 variants
Inlet pressure at suction side, venting and sealing gas connections must never exceed 1500 hPa absolute. p.76 ↗ Overpressure gas entry destroys the pump and ejects fragments; process overpressures must be interlocked upstream. All HiPace 800 variants
Fastening to mounting plate: 6× M8 screws, strength class ≥ 8.8, washer DIN EN ISO 7090/7092, screw-in depth ≥ 1.3×d, torque 25 Nm ±10%, plate ≥ 3 mm thick with > 270 MPa tensile strength. p.86 ↗ Mounting plate and screws must meet these minimums so burst torques are absorbed operator-side in a rotor-jam event. HiPace 800 (shared spec with HiPace 350-700)
Maximum calculated burst torque (rotor shaft breakage, ISO 27892): 4200 Nm. p.87 ↗ System anchoring and flange connections must withstand this torque; consider earthquake/safety connections on the chamber. HiPace 800
Maximum axial load on the high vacuum flange: 1000 N (100 kg); one-sided loading not permitted; counter flange flatness ± 0.05 mm. p.86 ↗ Limits superstructural fittings (screens, gauges, valves) mounted on the pump inlet. HiPace 800
Counter flange material tensile strength: ≥ 170 N/mm² at 2.5×d screw engagement depth or ≥ 270 N/mm² at 1.5×d. p.87 ↗ Dimension tapped-hole counter flanges accordingly to avoid flange failure in a rotor-jam event. HiPace 800
DN 200 splinter shield reduces pumping speed by 10% (H2), 13% (He), 26% (N2), 28% (Ar); DN 200 protective screen by 3%, 3%, 8%, 9% respectively. p.87 ↗ Include this derating when sizing effective pumping speed for the process gas. HiPace 800, DN 200 flanges
ISO-K and ISO-F flange fasteners are tightened crosswise in 3 steps: 5, 15, 25 ±2 Nm, using a calibrated torque wrench (tightening factor ≤ 1.6), WAF 15. p.89 ↗ Deviating torque risks housing stress, leaks or altered running characteristics. ISO-K/ISO-F high vacuum connections
With oil-sealed backing pumps, fore-vacuum connection must point vertically downward with max ±25° deviation, tubing supported so no piping forces act on the pump. p.88 ↗ Prevents oil backflow from the fore-vacuum range into the turbopump. HiPace 800 with oil-sealed backing pumps
Vibration compensator use requires prior consultation with Pfeiffer Vacuum; rotor jam torque per ISO 27892 would likely shear the pump off with a compensator installed. p.76 ↗ Do not install vibration compensators without on-site torque-absorption safety measures and manufacturer approval. All HiPace 800 variants
Mains disconnection device per SEMI-S2 and a circuit breaker with interruption rating ≥ 10,000 A are required at installation; pump and drive unit have no built-in mains switch. p.75 ↗ Must be provided by the system integrator before commissioning. All HiPace 800 / TC 400 installations
Power supply pack must satisfy double isolation between mains input and output per IEC 61010-1 / IEC 60950-1 / IEC 62368-1. p.75 ↗ Non-approved supply packs risk lethal electric shock; use original or equivalent-certified supplies. TC 400, 24 V and 48 V versions
Permissible ambient conditions: air pressure 530-1060 hPa, altitude ≤ 5000 m, protection class III, overvoltage category II, IP54/Type 12, pollution degree 2, ambient temperature 5-35°C (air cooling) or 5-40°C (water cooling), surrounding magnetic field ≤ 6 mT, irradiated thermal output ≤ 4.2 W. p.79 ↗ Site selection must account for temperature, magnetic fields and heat radiation; shielding needed otherwise. All HiPace 800 variants
Max permissible rotor temperature 90°C; max bakeout temperature at high vacuum flange 120°C. p.79 ↗ Bakeout jackets must be limited to 120°C at the HV flange; the drive unit reduces then cuts drive power if 90°C is exceeded. All HiPace 800 variants
Relative air humidity: max 80% at T < 31°C, up to max 50% at T < 40°C. p.79 ↗ Constrains indoor installation environment; dehumidification may be needed at higher temperatures. All HiPace 800 variants
Hybrid bearing: wear-free permanent magnetic bearing on high vacuum side, ceramic-ball bearing lubricated by an operating fluid reservoir on fore-vacuum side. p.82 ↗ Operating fluid reservoir replacement is a required maintenance task (see manual section 7.3); oil type per sticker on product. All HiPace 800 variants
Fore-vacuum connection is DN 25 ISO-KF; drive unit TC 400 available with 24 V DC or 48 V DC operating voltage; cooling is water (standard) or air (optional). p.82 ↗ Size backing pump line and select cooling type and supply voltage during system design. All HiPace 800 variants
Proper use restricted to vacuum generation with suitable backing pump, indoors, dry and inert gases; improper use includes pumping explosive, corrosive or condensing media, fluids, dust and operation in ionizing radiation or potentially explosive areas. p.79 ↗ Corrosive applications require the C (corrosive gas) variant selection approach; standard pump must not be used for such media. All HiPace 800 variants
The rating plate example shows nominal rotor speed 49,200 1/min (820 Hz); product designation system includes feature letters (U overhead, C corrosive, M magnetic bearing, H high compression, etc.). p.72 ↗ Verify variant letter codes when ordering and always quote full rating plate data to Pfeiffer service. HiPace series designation scheme
Before any maintenance work: switch off mains, wait until rotation speed f = 0, remove mains plug, secure against restart; decontaminate parts exposed to toxic media. p.77 ↗ Pump is only fully de-energized when mains plug is disconnected and rotor is at standstill. All HiPace 800 variants