MVP 010-3 DC Neo - Cleanroom — product leaflet mvp 010 070 cleanroom global
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum MVP 010-070 Cleanroom diaphragm vacuum pump series datasheet covering four variants: MVP 010-3 DC Neo - Cleanroom, MVP 020-3 AC - Cleanroom, MVP 030-3 DC - Cleanroom, and MVP 070-3 AC - Cleanroom, with AC and DC supply options, optional gas-ballast valve, variable speed drive, RS-485 interface, and OmniControl accessories.
Section-by-Section Technical Breakdown
Overview of the MVP 010-070 Cleanroom diaphragm pump series emphasizing oil- and particle-free operation suitable up to cleanroom class ISO 8, low-maintenance diaphragm design, and quiet, low-vibration operation for lab environments.
- Oil- and particle-free vacuum for contamination-free operation up to cleanroom class ISO 8
- Durable diaphragm pump design minimizes service effort and maintenance interventions
- Low noise and low vibration for comfortable lab work and stable analytical conditions
Lists available accessories, spare parts, and options for the MVP 010-070 Cleanroom series, including connectivity items, gas-ballast equipment, control units, and service kits.
- Connecting cable; RS-485 interface cable; USB RS-485 converter
- Gas-ballast valve kit; optional gas-ballast valve; purge gas nozzle with filter
- Flange and hose connection
- OmniControl (with or without integrated power supply); handheld programming unit
- Service kit; overhaul kit; variable speed drive
Performance and electrical specifications for all four variants, including pumping speed, ultimate pressure, rotational speed, pressure limits, supply voltage, current, cooling, sound level, dimensions, weight, and connection sizes. 60 Hz AC versions have slightly higher pumping speeds.
- Pumping speed: MVP 010-3 DC Neo 0.6 m³/h; MVP 020-3 AC 1.2 m³/h (50 Hz) / 1.4 m³/h (60 Hz); MVP 030-3 DC 1.8 m³/h; MVP 070-3 AC 3.8 m³/h (50 Hz) / 4.3 m³/h (60 Hz)
- Ultimate pressure: 1.0 hPa for MVP 010-3 DC Neo and MVP 070-3 AC; 2.0 hPa for MVP 020-3 AC and MVP 030-3 DC
- Max intake pressure 1100 hPa for all; max exhaust pressure 1300 hPa for MVP 010-3 DC Neo, 1100 hPa for others
- Supply: 24 V DC (±10%) for DC variants (max current 2.5 A / 7.0 A); 100–115 / 220–230 V AC 50/60 Hz or 120 V/60 Hz for AC variants (max current 1.7 A / 5.7 A)
- RS-485 I/O interface on DC variants (MVP 010-3 DC Neo, MVP 030-3 DC); none on AC variants
- Rotational speeds: DC variants variable (540–3000 and 600–2400 min⁻¹); AC variants 1500 min⁻¹ (50 Hz) / 1800 min⁻¹ (60 Hz); MVP 010-3 DC Neo nominal 1800 min⁻¹
- Cooling: natural convection (010, 030); forced air convection (020, 070)
- Ambient temperature: 5–40 °C (MVP 010-3 DC Neo); 10–40 °C (others)
- Sound pressure level: 50 / 48 / 45 / 52 dB(A); weight: 2.0 / 7.3 / 4.3 / 16.4 kg
- Dimensions (L×W×H): 173×105×103, 324×151×216, 223×143×217, 328×238×198 mm
- Gas inlet/outlet: G 1/8" for MVP 010/020/030; G 1/4" for MVP 070
- Cleanroom class ISO 8 (DIN EN ISO 14644-1) for all variants
- Gas ballast: not available on MVP 010-3 DC Neo and MVP 070-3 AC; optional on MVP 020-3 AC and MVP 030-3 DC
- Certification: cTUVus per UL / CSA 61010-1 for all variants
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| All four MVP 010-070 Cleanroom variants are certified for cleanroom class ISO 8 per DIN EN ISO 14644-1. p.10 ↗ | Pumps can be placed directly inside ISO 8-classified cleanrooms without external enclosures or exhaust scrubbing for particles. | whole series |
| Pumping speed ranges from 0.6 m³/h (MVP 010-3 DC Neo) to 3.8 m³/h at 50 Hz / 4.3 m³/h at 60 Hz (MVP 070-3 AC). p.10 ↗ | Select the 070 variant when ~4 m³/h throughput is needed; 60 Hz sites get ~13% higher speed on AC models. | MVP 010-3 DC Neo, MVP 020-3 AC, MVP 030-3 DC, MVP 070-3 AC |
| Ultimate pressure is 1.0 hPa for MVP 010-3 DC Neo and MVP 070-3 AC; 2.0 hPa for MVP 020-3 AC and MVP 030-3 DC. p.10 ↗ | Applications needing the lowest attainable pressure should choose the 010 DC Neo or 070 AC, not the 020/030. | whole series (per-variant values) |
| Maximum intake pressure is 1100 hPa and maximum exhaust pressure is 1100 hPa (1300 hPa for MVP 010-3 DC Neo). p.10 ↗ | Pumps may start at atmospheric pressure; the DC Neo tolerates slightly higher back-pressure on the exhaust side. | whole series |
| DC variants (MVP 010-3 DC Neo, MVP 030-3 DC) run on 24 V DC ±10% with maximum currents of 2.5 A and 7.0 A respectively. p.10 ↗ | Power supply sizing must cover 7 A for the MVP 030-3 DC; ±10% tolerance defines acceptable supply voltage window (21.6–26.4 V). | MVP 010-3 DC Neo, MVP 030-3 DC |
| AC variants accept 100–115 / 220–230 V AC, 50/60 Hz, or 120 V / 60 Hz, with max current 1.7 A (MVP 020-3 AC) and 5.7 A (MVP 070-3 AC). p.10 ↗ | Single worldwide AC input compatibility; circuit/fuse sizing per model current draw. | MVP 020-3 AC, MVP 070-3 AC |
| DC variants offer variable speed control: 540–3000 min⁻¹ (MVP 010-3 DC Neo) and 600–2400 min⁻¹ (MVP 030-3 DC); AC variants run fixed at 1500 min⁻¹ (50 Hz) / 1800 min⁻¹ (60 Hz). p.10 ↗ | Process throughput can be modulated on DC models; AC models cannot be speed-controlled without an external variable speed drive option. | whole series (per-variant) |
| RS-485 I/O interface is present only on the DC variants (MVP 010-3 DC Neo and MVP 030-3 DC). p.10 ↗ | Remote monitoring/control via RS-485 requires choosing a DC model; accessories include RS-485 cable and USB RS-485 converter. | MVP 010-3 DC Neo, MVP 030-3 DC |
| Gas ballast is not available on MVP 010-3 DC Neo and MVP 070-3 AC; it is optional on MVP 020-3 AC and MVP 030-3 DC. p.10 ↗ | For condensable vapor handling, specify gas ballast option with the 020 or 030 models (gas-ballast valve kit available as accessory). | MVP 020-3 AC, MVP 030-3 DC |
| Ambient temperature range is 5–40 °C for MVP 010-3 DC Neo and 10–40 °C for the other three variants. p.10 ↗ | Installation below 10 °C is only permissible for the DC Neo model; consider pre-heating for others in cold environments. | whole series |
| Cooling is natural convection on MVP 010-3 DC Neo and MVP 030-3 DC; forced air convection on MVP 020-3 AC and MVP 070-3 AC. p.10 ↗ | Forced-air models need clearance for fan airflow; natural convection models benefit from fan-free, particle-minimizing operation in cleanrooms. | whole series |
| Gas connections are G 1/8" inlet/outlet on MVP 010/020/030 and G 1/4" on MVP 070-3 AC. p.10 ↗ | Line/hose fittings and flange accessories must match thread size per model; the 070 needs G 1/4" hardware. | whole series |
| Sound pressure levels are 45–52 dB(A): 50 dB(A) (010), 48 dB(A) (020), 45 dB(A) (030), 52 dB(A) (070). p.10 ↗ | All models are quiet enough for desk-side lab placement; the 030 DC is the quietest, the 070 the loudest. | whole series |
| Weights range from 2.0 kg (MVP 010-3 DC Neo) to 16.4 kg (MVP 070-3 AC); largest footprint 328×238×198 mm. p.10 ↗ | The 010 Neo is portable/integrable in compact analyzers; mounting panels must support up to ~16.4 kg for the 070. | whole series |
| All variants are cTUVus certified according to UL / CSA 61010-1. p.10 ↗ | Suitable for North American lab-equipment safety compliance without additional certification effort. | whole series |
| Oil- and particle-free diaphragm operation; service kits and overhaul kits are offered as accessories. p.10 ↗ | No oil management/oil capacity in maintenance planning; preventive maintenance scheduled via Pfeiffer service/overhaul kits. | whole series |