MVP 020-3 AC — product leaflet mvp 020 070 global
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum MVP 020-070 series diaphragm vacuum pumps, covering six variants: MVP 020-3 AC, MVP 030-3 DC, MVP 030-3 C DC (corrosive gas C-version), MVP 040-2 AC, MVP 070-3 AC, and MVP 070-3 C AC (corrosive gas version). Accessories/options include RS-485 interface cable, USB RS-485 converter, OmniControl with/without integrated power supply, handheld programming unit, silencer G1/8", service kit, optional gas-ballast valve, variable speed drive, C-version, and cleanroom option.
Section-by-Section Technical Breakdown
The document opens with three qualitative benefit claims (oil- and particle-free vacuum for process safety, quiet low-vibration operation, energy-efficient low-cost operation). It then lists available accessories and optional equipment including connectivity, purge gas, control, and special-version options.
- Oil- and particle-free vacuum for contamination-free, stable process conditions
- Low noise and low vibration for comfortable lab work and stable analytical conditions
- Energy-efficient operation with minimal wear and long service intervals
- Accessories: connecting cable, RS-485 interface cable, USB RS-485 converter, purge gas nozzle with filter, flange and hose connection, OmniControl (with or without integrated power supply), handheld programming unit, silencer G1/8", service kit
- Options: gas-ballast valve (optional), variable speed drive, C-version (corrosive gas), cleanroom option
Dimensional drawing plus performance and electrical data for the small pumping-speed models. MVP 020-3 AC is 50/60 Hz AC powered; MVP 030-3 DC and MVP 030-3 C DC are 24 V DC with RS-485 I/O, variable speed (600–2400 min⁻¹) and natural convection cooling.
- Pumping speed: MVP 020-3 AC 1.2 m³/h (50 Hz) / 1.4 m³/h (60 Hz); MVP 030-3 DC and 030-3 C DC 1.8 m³/h (DC, speed-dependent)
- Ultimate pressure: 2.0 hPa (MVP 020-3 AC and MVP 030-3 DC), 3.0 hPa (MVP 030-3 C DC)
- Rotational speed: 1500/1800 min⁻¹ (MVP 020-3 AC); 600–2400 min⁻¹ variable (DC models)
- Supply: 100–115 / 220–230 V AC 50/60 Hz (also 120 V/60 Hz) or 24 V DC ±10 % for DC models
- Max current: 1.7 A (MVP 020-3 AC); 7.0 A (DC models)
- I/O: RS-485 on both DC models; none on MVP 020-3 AC
- Cooling: air forced convection (AC model), natural convection (DC models)
- Max intake and exhaust pressure: 1100 hPa for all
- Sound pressure level: 48 dB(A) (MVP 020-3 AC); 45 dB(A) (DC models)
- Weight: 7.3 kg (AC model); 4.3 kg (DC models)
- Dimensions: 324 x 151 x 216 mm (AC); 223 x 143 x 217 mm (030-3 DC); 238 x 142 x 223 mm (030-3 C DC)
- Gas inlet/outlet: G 1/8" / G 1/8" with silencer for all three
- Operating ambient 10–40 °C; cTUVus certified per UL/CSA 61010-1
- Gas ballast: optional on MVP 020-3 AC and MVP 030-3 DC; standard (Yes) on MVP 030-3 C DC
Performance and dimensional data for the larger AC-powered models. All run at 1500/1800 min⁻¹ (50/60 Hz) on AC mains with forced-air cooling; the C-version of the MVP 070 trades pumping speed for corrosion resistance and uses ISO-KF/hose connections instead of G 1/4".
- Pumping speed 50/60 Hz: MVP 040-2 AC 2.3/2.5 m³/h; MVP 070-3 AC 3.8/4.3 m³/h; MVP 070-3 C AC 3.4/3.8 m³/h
- Ultimate pressure: 4.0 hPa (MVP 040-2 AC); 1.0 hPa (MVP 070-3 AC); 1.5 hPa (MVP 070-3 C AC)
- Gas ballast: standard on MVP 040-2 AC and MVP 070-3 C AC; not available on MVP 070-3 AC
- Rotational speed: 1500 min⁻¹ (50 Hz) / 1800 min⁻¹ (60 Hz) for all
- Supply voltage: 100–115 / 220–230 V AC, 50/60 Hz (also 120 V/60 Hz); max current 3.4 A (MVP 040-2 AC) and 5.7 A (MVP 070 models)
- Max intake and exhaust pressure: 1100 hPa for all
- Cooling: air (forced convection); ambient range 10–40 °C; cTUVus certified per UL/CSA 61010-1
- Sound pressure level: 53 dB(A) (MVP 040-2 AC), 52 dB(A) (MVP 070-3 AC), 50 dB(A) (MVP 070-3 C AC)
- Weight: 11.4 kg / 16.4 kg / 14.3 kg respectively
- Dimensions: 246 x 238 x 198 mm (MVP 040-2 AC); 328 x 238 x 198 mm (MVP 070-3 AC); 328 x 242 x 198 mm (MVP 070-3 C AC)
- Gas inlet/outlet: G 1/4" / G 1/4" with silencer (MVP 040-2 AC and MVP 070-3 AC); DN 16 ISO-KF / 10 mm hose connector (MVP 070-3 C AC)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| All six MVP variants have a maximum intake pressure and maximum exhaust pressure of 1100 hPa (mbar). p.20 ↗ | Pumps can start and operate at atmospheric pressure and can discharge against up to ~1.1 bar absolute without a pressure restriction — no pre-vacuum or exhaust-side pressure stage is required. | whole series |
| MVP 070-3 AC achieves the lowest ultimate pressure of the series at 1.0 hPa; the C-version MVP 070-3 C AC is 1.5 hPa. p.20 ↗ | Selecting the corrosive-gas C-version costs 0.5 hPa in ultimate pressure and ~10 % pumping speed — factor this in when corrosive media must still reach low pressures. | MVP 070-3 AC / MVP 070-3 C AC |
| MVP 030-3 DC and MVP 030-3 C DC operate from 24 V DC (±10 %) with max current 7.0 A and offer RS-485 I/O. p.20 ↗ | Suitable for integration into 24 V control cabinets/mobile systems; RS-485 (plus optional USB RS-485 converter) allows remote monitoring/control via OmniControl or handheld programming unit. | MVP 030-3 DC / MVP 030-3 C DC |
| DC models allow variable rotational speed from 600 to 2400 min⁻¹ (variable speed drive option). p.20 ↗ | Pumping speed can be tuned in-process; above 60 Hz-equivalent speed yields higher effective pumping speed than fixed AC models (1.8 m³/h vs 1.2–1.4 m³/h). | MVP 030-3 DC / MVP 030-3 C DC |
| AC models are supplied at 100–115 / 220–230 V AC, 50/60 Hz (also 120 V/60 Hz) with max currents 1.7 A (MVP 020-3), 3.4 A (MVP 040-2), 5.7 A (MVP 070-3). p.20 ↗ | Circuit/fuse sizing: a single-phase outlet is sufficient; worst case is 5.7 A for the MVP 070-3. | all AC models |
| MVP 030-3 DC models use natural convection cooling; all AC models use forced-air (air) cooling, and all require ambient temperatures of 10–40 °C. p.20 ↗ | DC models need no fan clearance but can only run in 10–40 °C environments; ensure ventilation for AC versions despite forced convection. | whole series |
| Gas ballast is standard on MVP 030-3 C DC, MVP 040-2 AC and MVP 070-3 C AC, optional on MVP 020-3 AC and MVP 030-3 DC, and not available on MVP 070-3 AC. p.20 ↗ | For condensable vapor handling, avoid MVP 070-3 AC or order the optional gas-ballast valve on the small models; the C-versions and MVP 040-2 include it as standard. | whole series (variant-specific) |
| The MVP 070-3 C AC uses DN 16 ISO-KF intake with 10 mm hose outlet instead of the G 1/4" threaded connections of the MVP 040-2 AC and MVP 070-3 AC. p.20 ↗ | Piping/fitting selection differs for the corrosive version; plan for ISO-KF clamping hardware rather than G 1/4" threads. | MVP 070-3 C AC |
| Small models (MVP 020/030) have G 1/8" inlet/outlet, both with silencer. p.20 ↗ | Use G 1/8" fittings and keep silencers fitted to meet the 45–48 dB(A) noise levels quoted. | MVP 020-3 AC, MVP 030-3 DC, MVP 030-3 C DC |
| Sound pressure levels range 45–48 dB(A) for MVP 020/030 and 50–53 dB(A) for MVP 040/070. p.20 ↗ | All variants are quiet enough for desktop/lab placement without acoustic enclosures. | whole series |
| Weights are 4.3–7.3 kg for the small models and 11.4–16.4 kg for MVP 040/070. p.20 ↗ | Portable/benchtop mounting is feasible for all; mounting plates for MVP 070 should be sized for ~16.5 kg. | whole series |
| 60 Hz AC versions deliver slightly higher pumping speeds (e.g. MVP 070-3 AC: 3.8 m³/h at 50 Hz vs 4.3 m³/h at 60 Hz). p.20 ↗ | For maximum throughput in 60 Hz regions, specify nominal speeds from the 60 Hz column; check line frequency when dimensioning systems. | all AC models |
| All models are cTUVus certified according to UL/CSA 61010-1. p.20 ↗ | Directly usable in North American lab/analytical installations without additional certification work for the pump itself. | whole series |
| MVP 030-3 C DC has a higher ultimate pressure (3.0 hPa) than the standard MVP 030-3 DC (2.0 hPa). p.20 ↗ | For corrosive applications at the 1.8 m³/h class, ultimate pressure degrades by 1 hPa — verify it still meets process requirements. | MVP 030-3 C DC |
| Optional purge gas nozzle with filter and optional cleanroom option are available across the series. p.20 ↗ | For corrosive/dusty environments order the purge kit; for cleanroom installations (e.g. particle-sensitive labs) specify the cleanroom option at ordering time. | whole series |