Induztr
Engineering Digest Document Viewer Related Models

HiPace 700 H with TC 400 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiPace 700 H turbomolecular vacuum pump product leaflet. Covers a single model (HiPace 700 H) with its performance data, interface/mounting specifications, and the range of available accessories and options (adapter, mains/connection cables, fieldbus systems, power supply kit, display unit, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper).

Section-by-Section Technical Breakdown

pp. 1
Product Features and Accessories / Options

Highlights three key characteristics: high performance with best compression ratio for light gases suited to high and ultra-high vacuum, reliable hybrid bearing technology for long life and maximum uptime, and quiet operation via patented laser balancing technology. Lists available accessories, spare parts, and options.

  • Best compression ratio for light gases; suited for high and ultra-high vacuum applications
  • Best-in-class hybrid bearing technology, long life cycle, maximum uptime
  • Low noise and vibration levels thanks to patented laser balancing technology
  • Accessories/options: electrical adapter, mains cable, connection cable, various fieldbus systems, power supply kit and display unit, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper
pp. 2
Pumping Performance Data

Pumping speeds, compression ratios, and gas throughputs at final rotational speed for N₂, H₂, He, and Ar, plus maximum forevacuum pressure. Shows very high compression ratios for N₂ and Ar and strong H₂ pumping (555 l/s).

  • Pumping speed: N₂ 685 l/s, H₂ 555 l/s, He 655 l/s, Ar 665 l/s
  • Compression ratio: N₂ 1·10¹³, H₂ 2·10⁷, He 1·10⁹, Ar 1·10¹³
  • Gas throughput at final rotational speed: N₂ 3.5 hPa·l/s, H₂ 14 hPa·l/s, He 18 hPa·l/s, Ar 2 hPa·l/s
  • Max. forevacuum pressure (N₂): 22 hPa
pp. 2-3
Operational, Interface, and Mechanical Data

Standard communication interfaces, mounting orientation, flange connections, run-up time, sound pressure level, and weight. Confirms arbitrary mounting orientation and DN 160 ISO-K/ISO-F/CF-F inlet.

  • Communication interfaces standard: RS-485, remote
  • Mounting orientation: arbitrary
  • Gas inlet / outlet: DN 160 ISO-K/ISO-F/CF-F / DN 25 ISO-KF / G 1/4"
  • Run-up time: 2 min
  • Sound pressure level (EN ISO 2151): 50 dBA
  • Weight approx.: 11.8 kg

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed is 685 l/s for N₂ and 555 l/s for H₂. p.2 ↗ Strong hydrogen pumping makes the pump suitable for UHV systems with H₂-dominated residual gas; size forepump accordingly. HiPace 700 H
Compression ratio for H₂ is 2·10⁷ and for He 1·10⁹, versus 1·10¹³ for N₂ and Ar. p.2 ↗ Even for light gases the compression is high enough that base pressure is largely determined by the forevacuum side; select a good forepump for UHV work. HiPace 700 H
Maximum permissible forevacuum pressure for N₂ is 22 hPa. p.2 ↗ Forepump and backing-line design must keep backing pressure below 22 hPa during operation, including during gas load events. HiPace 700 H
Gas throughput at final rotational speed: 14 hPa·l/s for H₂ and 18 hPa·l/s for He, but only 3.5 hPa·l/s for N₂ and 2 hPa·l/s for Ar. p.2 ↗ The pump tolerates much higher continuous gas loads of light gases than heavy gases; process gas composition strongly affects permissible throughput. HiPace 700 H
Run-up time to final speed is 2 minutes. p.2 ↗ Process sequencing and interlocks (e.g., opening the high-vacuum valve) should account for ~2 min start-up delay. HiPace 700 H
Sound pressure level is 50 dBA per EN ISO 2151. p.2 ↗ Low noise level permits installation near workstations without additional acoustic enclosure in most cases. HiPace 700 H
Weight is approximately 11.8 kg. p.2 ↗ Mounting flanges and chamber interfaces must support ~12 kg; relatively easy manual handling for installation. HiPace 700 H
Gas inlet is DN 160 ISO-K/ISO-F/CF-F; outlet is DN 25 ISO-KF / G 1/4". p.3 ↗ High-vacuum side requires DN 160 flange hardware; forevacuum side connects via DN 25 ISO-KF or a G 1/4" threaded port. HiPace 700 H
Mounting orientation is arbitrary. p.3 ↗ The pump can be installed in any orientation, simplifying integration into constrained system layouts (verify with full installation manual for oil/bearing details). HiPace 700 H
Standard communication interfaces are RS-485 and remote. p.3 ↗ Integration into control systems uses RS-485 as standard; fieldbus options are available as accessories for other protocols. HiPace 700 H
Hybrid bearing technology with patented laser balancing is used. p.1 ↗ Reduced vibration and long bearing life; a vibration damper accessory is available for vibration-sensitive applications. HiPace 700 H
A power failure venting unit and venting accessories are offered. p.1 ↗ For unattended or safety-critical systems, specify automatic/power-failure venting to protect the pump and process on shutdown. HiPace 700 H (optional accessories)
Fieldbus systems, power supply kit, and display unit are available as options. p.1 ↗ Non-RS-485 plant networks require ordering the appropriate fieldbus option and power supply/display kit for local monitoring. HiPace 700 H (optional accessories)