Induztr
Engineering Digest Document Viewer Related Models

HiPace 800 M with TM 700 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiPace 800 M turbomolecular vacuum pump with magnetically levitated rotor. Document covers a single model; lists available accessories and options (adapter, mains/connecting cables, fieldbus systems, power supply kit and display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit). No ordering matrix or variant codes given.

Section-by-Section Technical Breakdown

pp. 1
Product Highlights

Key selling points of the HiPace 800 M: maintenance-free operation, quiet running, and reliability. These stem from the electromagnetic rotor suspension and automatic sensor calibration.

  • Maintenance-free: electromagnetic suspension protects rotor from premature wear; continuous rotor imbalance compensation
  • Quiet: low noise and vibration levels
  • Reliable: hydrocarbon-free operation, automatic sensor calibration during operation
  • Available accessories/options: electrical adapter, mains cable, connecting cable, various fieldbus systems, power supply kit and display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit
pp. 2
Pumping Performance Data

Pumping speeds, compression ratios and gas throughputs for N₂, H₂, He and Ar, plus forevacuum limit, run-up time and noise level. These characterize the pump's vacuum performance envelope.

  • Pumping speed: N₂ 790 l/s, H₂ 500 l/s, He 625 l/s, Ar 775 l/s
  • Compression ratio: N₂ 1·10¹¹, H₂ 2·10⁵, He 1·10⁷, Ar 1·10¹¹
  • Gas throughput at final rotational speed: N₂ 13 hPa·l/s, Ar 8 hPa·l/s
  • Gas throughput at 0.1 hPa HV-pressure: 16 hPa·l/s for He, H₂ and N₂
  • Max. forevacuum pressure for N₂: 8 hPa
  • Run-up time: 4 min
  • Sound pressure level (EN ISO 2151): 45 dBA
pp. 3
Mechanical and Interface Data

Physical and connectivity specifications: weight, flange connections, mounting orientation, and standard communication interfaces.

  • Weight: approx. 17.2 kg
  • Gas inlet: DN 200 ISO-K/ISO-F/CF-F; outlet: DN 25 ISO-KF / G 1/4"
  • Mounting orientation: arbitrary
  • Communication interfaces (standard): RS-485, remote

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed for N₂ is 790 l/s p.2 ↗ Size backing pumps and conductance upstream accordingly for high-vacuum applications around ~800 l/s class HiPace 800 M
Pumping speed for H₂ is 500 l/s and for He 625 l/s p.2 ↗ Light-gas applications (hydrogen/helium) see ~30-40% lower effective speed than N₂; account for this in pump-down calculations HiPace 800 M
Compression ratio for H₂ is 2·10⁵ (vs 1·10¹¹ for N₂) p.2 ↗ Ultimate pressure for hydrogen-bearing systems is strongly limited by forevacuum pressure; keep foreline pressure low for H₂ HiPace 800 M
Maximum forevacuum pressure for N₂ is 8 hPa p.2 ↗ Backing pump must maintain ≤8 hPa at the turbo outlet to avoid rotor overload; verify backing pump curve at this pressure HiPace 800 M
Gas throughput at final rotational speed: 13 hPa·l/s (N₂), 8 hPa·l/s (Ar) p.2 ↗ Limit process gas load to these throughputs at full speed to stay in safe operating range HiPace 800 M
Gas throughput at 0.1 hPa HV-pressure is 16 hPa·l/s for He, H₂ and N₂ p.2 ↗ Defines permissible gas load when operating at 0.1 hPa inlet pressure HiPace 800 M
Run-up time is 4 min p.2 ↗ Allow ~4 minutes to full speed after start before process vacuum is at spec; relevant for cycling sequences HiPace 800 M
Sound pressure level is 45 dBA per EN ISO 2151 p.2 ↗ Suitable for laboratory/quiet environments; no special acoustic enclosure typically required HiPace 800 M
Weight is approx. 17.2 kg p.3 ↗ Ensure mounting flange/support and handling equipment are rated for ~17 kg; two-person lift advisable HiPace 800 M
Gas inlet is DN 200 ISO-K/ISO-F/CF-F; forevacuum outlet is DN 25 ISO-KF / G 1/4" p.3 ↗ Order matching ISO-K/ISO-F/CF inlet hardware and a DN 25 ISO-KF (or G 1/4") foreline; a small foreline keeps conductance acceptable given 8 hPa forevacuum limit HiPace 800 M
Mounting orientation is arbitrary p.3 ↗ Pump can be installed horizontally or vertically (e.g., on chambers or manifolded systems) without derating HiPace 800 M
Standard communication interfaces are RS-485 and remote p.3 ↗ Native digital control via RS-485; additional fieldbus options (e.g., for PLC integration) are available as accessories HiPace 800 M
The pump is magnetically levitated (electromagnetic suspension) with continuous rotor imbalance compensation p.1 ↗ No lubricant in pump; hydrocarbon-free vacuum suitable for contamination-sensitive processes; no bearing wear maintenance HiPace 800 M
A power failure venting unit is available as an accessory p.1 ↗ Recommended to include automatic venting on power loss to bring chamber to atmosphere safely (magnetic-levitation rotor coasts down when unpowered) HiPace 800 M installation