HiPace 800 M with TM 700 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiPace 800 M turbomolecular vacuum pump with magnetically levitated rotor. Document covers a single model; lists available accessories and options (adapter, mains/connecting cables, fieldbus systems, power supply kit and display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit). No ordering matrix or variant codes given.
Section-by-Section Technical Breakdown
Key selling points of the HiPace 800 M: maintenance-free operation, quiet running, and reliability. These stem from the electromagnetic rotor suspension and automatic sensor calibration.
- Maintenance-free: electromagnetic suspension protects rotor from premature wear; continuous rotor imbalance compensation
- Quiet: low noise and vibration levels
- Reliable: hydrocarbon-free operation, automatic sensor calibration during operation
- Available accessories/options: electrical adapter, mains cable, connecting cable, various fieldbus systems, power supply kit and display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit
Pumping speeds, compression ratios and gas throughputs for N₂, H₂, He and Ar, plus forevacuum limit, run-up time and noise level. These characterize the pump's vacuum performance envelope.
- Pumping speed: N₂ 790 l/s, H₂ 500 l/s, He 625 l/s, Ar 775 l/s
- Compression ratio: N₂ 1·10¹¹, H₂ 2·10⁵, He 1·10⁷, Ar 1·10¹¹
- Gas throughput at final rotational speed: N₂ 13 hPa·l/s, Ar 8 hPa·l/s
- Gas throughput at 0.1 hPa HV-pressure: 16 hPa·l/s for He, H₂ and N₂
- Max. forevacuum pressure for N₂: 8 hPa
- Run-up time: 4 min
- Sound pressure level (EN ISO 2151): 45 dBA
Physical and connectivity specifications: weight, flange connections, mounting orientation, and standard communication interfaces.
- Weight: approx. 17.2 kg
- Gas inlet: DN 200 ISO-K/ISO-F/CF-F; outlet: DN 25 ISO-KF / G 1/4"
- Mounting orientation: arbitrary
- Communication interfaces (standard): RS-485, remote
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed for N₂ is 790 l/s p.2 ↗ | Size backing pumps and conductance upstream accordingly for high-vacuum applications around ~800 l/s class | HiPace 800 M |
| Pumping speed for H₂ is 500 l/s and for He 625 l/s p.2 ↗ | Light-gas applications (hydrogen/helium) see ~30-40% lower effective speed than N₂; account for this in pump-down calculations | HiPace 800 M |
| Compression ratio for H₂ is 2·10⁵ (vs 1·10¹¹ for N₂) p.2 ↗ | Ultimate pressure for hydrogen-bearing systems is strongly limited by forevacuum pressure; keep foreline pressure low for H₂ | HiPace 800 M |
| Maximum forevacuum pressure for N₂ is 8 hPa p.2 ↗ | Backing pump must maintain ≤8 hPa at the turbo outlet to avoid rotor overload; verify backing pump curve at this pressure | HiPace 800 M |
| Gas throughput at final rotational speed: 13 hPa·l/s (N₂), 8 hPa·l/s (Ar) p.2 ↗ | Limit process gas load to these throughputs at full speed to stay in safe operating range | HiPace 800 M |
| Gas throughput at 0.1 hPa HV-pressure is 16 hPa·l/s for He, H₂ and N₂ p.2 ↗ | Defines permissible gas load when operating at 0.1 hPa inlet pressure | HiPace 800 M |
| Run-up time is 4 min p.2 ↗ | Allow ~4 minutes to full speed after start before process vacuum is at spec; relevant for cycling sequences | HiPace 800 M |
| Sound pressure level is 45 dBA per EN ISO 2151 p.2 ↗ | Suitable for laboratory/quiet environments; no special acoustic enclosure typically required | HiPace 800 M |
| Weight is approx. 17.2 kg p.3 ↗ | Ensure mounting flange/support and handling equipment are rated for ~17 kg; two-person lift advisable | HiPace 800 M |
| Gas inlet is DN 200 ISO-K/ISO-F/CF-F; forevacuum outlet is DN 25 ISO-KF / G 1/4" p.3 ↗ | Order matching ISO-K/ISO-F/CF inlet hardware and a DN 25 ISO-KF (or G 1/4") foreline; a small foreline keeps conductance acceptable given 8 hPa forevacuum limit | HiPace 800 M |
| Mounting orientation is arbitrary p.3 ↗ | Pump can be installed horizontally or vertically (e.g., on chambers or manifolded systems) without derating | HiPace 800 M |
| Standard communication interfaces are RS-485 and remote p.3 ↗ | Native digital control via RS-485; additional fieldbus options (e.g., for PLC integration) are available as accessories | HiPace 800 M |
| The pump is magnetically levitated (electromagnetic suspension) with continuous rotor imbalance compensation p.1 ↗ | No lubricant in pump; hydrocarbon-free vacuum suitable for contamination-sensitive processes; no bearing wear maintenance | HiPace 800 M |
| A power failure venting unit is available as an accessory p.1 ↗ | Recommended to include automatic venting on power loss to bring chamber to atmosphere safely (magnetic-levitation rotor coasts down when unpowered) | HiPace 800 M installation |