Induztr
Engineering Digest Document Viewer Related Models

HiPace 800 H with TC 400 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Brochure for the Pfeiffer Vacuum HiPace 800 H turbomolecular vacuum pump. Single model covered (HiPace 800 H); mentions available accessories and options (electrical adapter, mains/connecting cables, fieldbus systems, power supply kit and display unit, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper). No ordering codes or ordering matrix are given.

Section-by-Section Technical Breakdown

pp. 1
Product Highlights

Positions the HiPace 800 H as a high-performance turbomolecular pump with best-in-class hybrid bearing technology and a patented laser balancing process. Emphasizes suitability for high and ultra-high vacuum applications and low noise/vibration operation.

  • High performance: best compression ratio for light gases, suited for high and ultra-high vacuum applications
  • Reliable: best-in-class hybrid bearing technology, long life cycle, maximum uptime
  • Quiet: low noise and vibration levels thanks to patented laser balancing technology
  • Accessories/options: electrical adapter, mains cable, connecting cable, various fieldbus systems, power supply kit and display unit, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper
pp. 2-3
Technical Data

Lists key performance and mechanical data for the HiPace 800 H: pumping speeds and compression ratios for N₂, H₂, He and Ar, gas throughput, maximum forevacuum pressure, run-up time, noise level and weight. Also specifies standard communication interfaces and flange connections.

  • Pumping speed: N₂ 790 l/s, H₂ 580 l/s, He 700 l/s, Ar 780 l/s
  • Compression ratio: N₂ 1·10¹³, H₂ 2·10⁷, He 1·10⁹, Ar 1·10¹³
  • Gas throughput at final rotational speed: N₂ 3.5 hPa·l/s, H₂ 14 hPa·l/s, He 18 hPa·l/s, Ar 2 hPa·l/s
  • Max. forevacuum pressure (N₂): 22 hPa
  • Run-up time: 2 min
  • Sound pressure level (EN ISO 2151): 50 dBA
  • Weight: approx. 12.8 kg
  • Communication interfaces standard: RS-485, remote
  • Gas inlet: DN 200 ISO-K/ISO-F/CF-F; outlet: DN 25 ISO-KF / G 1/4"
  • Mounting orientation: arbitrary

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed for N₂ is 790 l/s; for H₂ 580 l/s; for He 700 l/s; for Ar 780 l/s p.2 ↗ Size the pump for the actual process gas; light-gas performance stays high (H₂ still 580 l/s), which matters for hydrogen-loaded UHV processes HiPace 800 H
Compression ratio for N₂ and Ar is 1·10¹³; for He 1·10⁹; for H₂ 2·10⁷ p.2 ↗ Suitable for ultra-high vacuum: with a typical forevacuum of ~10⁻² hPa, ultimate pressures well below 10⁻¹⁰ hPa are achievable for N₂/Ar; H₂ limits the ultimate pressure HiPace 800 H
Gas throughput at final rotational speed: N₂ 3.5 hPa·l/s, H₂ 14 hPa·l/s, He 18 hPa·l/s, Ar 2 hPa·l/s p.2 ↗ Defines the maximum continuous gas load the pump can handle at full speed; size fore pumps and process gas dosing accordingly HiPace 800 H
Maximum forevacuum pressure for N₂ is 22 hPa p.2 ↗ The backing pump must maintain ≤22 hPa at the forevacuum port under full gas load; allows use of relatively small backing pumps HiPace 800 H
Run-up time is 2 minutes p.2 ↗ Process start sequences should allow ~2 min from start to full rotational speed; also relevant for power-failure recovery planning HiPace 800 H
Sound pressure level is 50 dBA per EN ISO 2151 p.2 ↗ Suitable for laboratory environments with low noise requirements; no additional acoustic enclosure needed for typical installations HiPace 800 H
Weight is approximately 12.8 kg p.2 ↗ Mounting flanges and support structures can be designed for a light pump; single-person handling is practical HiPace 800 H
Gas inlet is DN 200 ISO-K/ISO-F/CF-F; outlet is DN 25 ISO-KF / G 1/4" p.3 ↗ Vacuum chamber must provide a DN 200 compatible flange (ISO-K, ISO-F, or CF-F); foreline plumbing sized for DN 25 ISO-KF or G 1/4" connection HiPace 800 H
Mounting orientation is arbitrary p.3 ↗ The pump can be installed in any orientation, simplifying integration into existing chamber layouts HiPace 800 H
Standard communication interfaces are RS-485 and remote p.3 ↗ Control/monitoring integration via RS-485 is possible out of the box; other fieldbus systems are available as options HiPace 800 H
A power failure venting unit is available as an accessory p.1 ↗ For safety-critical processes, include automatic venting on power loss to protect the chamber and pump HiPace 800 H (optional accessory)
Hybrid bearing technology with patented laser balancing is used, giving low vibration levels p.1 ↗ Bearing type is hybrid (permanently lubricated magnetic/mechanical hybrid per Pfeiffer design); a vibration damper accessory is available for vibration-sensitive applications HiPace 800 H
Safety valve and venting accessories are offered as options p.1 ↗ For processes with hazardous gas loads, add the safety valve at the forevacuum side and venting accessories for controlled pump shutdown HiPace 800 H (optional accessories)
A display unit and power supply kit are available, along with various fieldbus systems p.1 ↗ Standalone operation with local display is possible; plant-wide integration via fieldbus requires ordering the appropriate fieldbus option HiPace 800 H (optional accessories)