Induztr
Engineering Digest Document Viewer Related Models

HiPace 800 P with TC 400 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiPace 800 P turbomolecular vacuum pump — a high-performance pump designed for coating processes with an optimized rotor/stator concept. The document is a short brochure/datasheet (3 pages) covering technical data, accessories and options; no ordering codes are given.

Section-by-Section Technical Breakdown

pp. 1
Product Features

Highlights the pump's positioning for coating applications: high performance, reliability and easy maintenance. Emphasizes hybrid bearing technology and on-site bearing change with tolerance to process dust.

  • Specially designed for coating processes, optimized rotor/stator concept
  • Best-in-class hybrid bearing technology, long life cycle, maximum uptime
  • On-site bearing change, insensitive to process dust
pp. 2-3
Technical Data

Pumping speeds, compression ratios and gas throughputs for N₂, H₂, He and Ar, plus mechanical and operating data. Defines forevacuum limit, run-up time, noise level and weight.

  • Pumping speed: N₂ 785 l/s, H₂ 320 l/s, He 595 l/s, Ar 770 l/s
  • Compression ratio: N₂ 1·10⁶, H₂ 4·10², He 3·10³, Ar 1·10⁷
  • Gas throughput at final rotational speed: N₂ 2.5 hPa·l/s, H₂ 8 hPa·l/s, He 5 hPa·l/s, Ar 2 hPa·l/s
  • Max. forevacuum pressure for N₂: 1.1 hPa
  • Run-up time: 1.6 min
  • Sound pressure level (EN ISO 2151): 50 dBA
  • Weight approx. 12.7 kg
  • Communication interfaces standard: RS-485, remote
  • Mounting orientation: arbitrary
  • Gas inlet DN 200 ISO-K/ISO-F/CF-F; outlet DN 25 ISO-KF / G 1/4"
pp. 1
Accessories, Spare Parts and Options

Lists available accessories and options for the HiPace 800 P, covering electrical, fieldbus, mounting, venting and vibration items.

  • Electrical adapter, mains cable, connecting cable
  • Various fieldbus systems
  • Power supply kit and display unit
  • Mounting kit, safety valve
  • Venting accessories, power failure venting unit
  • Vibration damper

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed for N₂ is 785 l/s at final rotational speed. p.2 ↗ Size the forevacuum pump and process chamber conductance so effective speed near 785 l/s is preserved at the DN 200 inlet. HiPace 800 P
Pumping speed for H₂ is 320 l/s, for He 595 l/s and for Ar 770 l/s. p.2 ↗ For hydrogen-containing coating processes expect ~40% of N₂ speed; consider gas-type effects when estimating pump-down of light gases. HiPace 800 P
Compression ratio for N₂ is 1·10⁶ and for Ar 1·10⁷. p.2 ↗ Very high compression for heavy gases supports clean high-vacuum base pressures when backed adequately. HiPace 800 P
Compression ratio for H₂ is only 4·10² (He: 3·10³). p.2 ↗ Ultimate pressure for hydrogen is limited by forevacuum pressure; a low-H₂-content backing pump is needed for low total pressure with hydrogen. HiPace 800 P
Maximum admissible forevacuum pressure for N₂ is 1.1 hPa. p.2 ↗ Select and size the backing pump to maintain foreline pressure below 1.1 hPa at full gas throughput to avoid loss of compression/back-migration. HiPace 800 P
Gas throughput at final speed is 2.5 hPa·l/s for N₂ and 8 hPa·l/s for H₂. p.2 ↗ Check backing pump throughput against these values when specifying continuous gas loads in coating processes. HiPace 800 P
Run-up time is 1.6 min to final rotational speed. p.2 ↗ Allow ~2 minutes from start to full pumping performance in process automation timing; consider power failure venting unit if process cannot tolerate this delay after restart. HiPace 800 P
Gas inlet flange is DN 200 ISO-K/ISO-F/CF-F; outlet is DN 25 ISO-KF or G 1/4". p.3 ↗ Plan chamber interface with DN 200 hardware and select matching foreline connections; CF option available for UHV-tight systems. HiPace 800 P
Mounting orientation is arbitrary. p.3 ↗ The pump can be installed horizontally or vertically, simplifying integration into coating systems without orientation constraints. HiPace 800 P
Standard communication interfaces are RS-485 and remote. p.3 ↗ Plan control wiring for RS-485; additional fieldbus systems are available as options for higher-level control integration. HiPace 800 P
Sound pressure level is 50 dBA per EN ISO 2151. p.2 ↗ Suitable for installation in close proximity to operators without special acoustic enclosure. HiPace 800 P
Weight is approximately 12.7 kg. p.2 ↗ Single-person handling feasible; mounting brackets/supports must be sized for this mass plus attached cabling and vibration damper. HiPace 800 P
Hybrid bearing technology allows on-site bearing change. p.1 ↗ Maintenance can be performed in situ, reducing downtime versus returning the pump to the factory; schedule bearing replacement around process campaigns. HiPace 800 P
The rotor/stator concept is insensitive to process dust and optimized for coating processes. p.1 ↗ Suitable for sputtering/evaporation environments with particulate load without additional particle protection in many cases. HiPace 800 P
A power failure venting unit and venting accessories are available as options. p.1 ↗ Specify the power failure venting unit to automatically vent the pump on mains failure, protecting the process chamber from oil/backstreaming-related contamination paths. HiPace 800 P