Technical Reference Summary

Covered Models & Document Scope

This manual covers the Trapezium HITS (HITS series software) for high-speed impact testing, consisting of testing software and data processing software. It applies to systems using Windows 7/10 (and mentions Windows 10/11 in trademarks), with either an NI PCI-6132 or ADM670 AD expansion board, and optionally a 4870-type controller. Software version-dependent feature availability is noted.

Section-by-Section Technical Breakdown

pp. 13-14
System Configuration

The system consists of a personal computer connected to the controller via RS232C. Measurement data are collected and converted with an AD expansion board in the PC. The software comprises testing software and data processing software.

  • PC connects to controller with RS232C.
  • Measurement data collected and converted with AD expansion board in PC.
  • Data processing software can be operated even while a test is being executed.
pp. 22-29
Software Installation

Describes the required operating environment and installation procedures for the software, including AD board driver installation for NI PCI-6132 and ADM670 boards. Configuration files can be backed up and restored for re-installation.

  • OS: Windows 7/10; CPU: 1 GHz or higher; Memory: 256 MB or more; HDD: 8 GB or more.
  • Display: 1920 × 1200, 65535 colors; at least one empty full-size PCI slot; RS232C communication; CD-ROM and color printer required.
  • Configuration files HTM_System.fil, HTM.ini, HTMTable.ini store basic settings; back up regularly.
  • NI PCI-6132 driver requires Microsoft .NET Framework 4.6.2; installation may take about 10 minutes.
  • ADM670 driver installed via Device Manager; confirm 'ADM-670PCI (win_2k_xp)' under MSCIENCE.
pp. 35-37
System Setting

Configures sensor information, controller usage, loading direction, measurement frequency, and external trigger settings. Sensor information is compared at startup and before testing to detect changes.

  • Up to three sensors each for puncture and tensile tests can be registered, distinguished by ID.
  • Sensor name limited to <20 characters>; value up to 4 digits; units kN, N, tonf, kgf.
  • CH1 filter cutoff: 20 kHz or 100 kHz; CH2 filter: 2 kHz or 20 kHz; CH3 filter: 20 kHz or 100 kHz.
  • External trigger only usable with 4870-type controller; not usable if sampling interval >10 msec or sampling time <10 sec.
  • Trigger level set via STROKE indicator on 4870 controller.
pp. 38-41
Calibration

Describes calibration settings including span, range calibration values, velocity calibration values, and temperature calibration values. These are critical for correct measurement.

  • Span coefficient converts sensor signal to voltage around 5 V; range calibration value corrects hardware error at range changing.
  • Velocity calibration value set for 1 m/sec to 20 m/sec; range <100, 4095>; not required below 1 m/sec due to feedback control.
  • Temperature calibration at reference temperatures -30, 0, 80, and 130 °C against span at 20 °C; linear interpolation for other temperatures.
  • Calibration value input range: -32767 to +32767 (integer); span: 0 to 65535 (integer).
pp. 42-?
Condition Setting

Covers creation and editing of test condition files using the Condition Setting Wizard, including system info, sample info, range info/data sampling, data processing, and report settings.

  • Condition settings file created via Condition Setting from Home screen.
  • Condition Setting Wizard guides through System Info., Sample Info., Range Info./Data Sampling, Data Processing, and Report screens.
pp. 31-50
Testing

Describes the testing mode, monitor display, status display, start/stop test, completion, graph display, auto calibration, and return to home. Includes safety warnings about actuator movement.

  • Before starting test, verify nobody is near the load machine.
  • Do not allow body parts in actuator range; actuator moves abruptly during waveform output.
  • Upon completion of test, sampled data are saved.
  • Auto Calibration function available from Testing screen.
pp. 8, 14
Data Processing Software

Covers data processing functions including graph display (load-stroke, load-stroke-time, load-velocity-time, load-energy-stroke), statistics processing, overlay processing, CSV output, and report printing.

  • Graph types: Load-Stroke, Load•Stroke-Time, Load•Velocity-Time, Load•Energy-Stroke, Option Graph.
  • ISO 22183:2023 judgment available.
  • Statistics processing includes standard deviation toggle and total CSV output.
  • Overlay processing allows comparison of multiple data files.
pp. 129-130
Troubleshooting

Provides troubleshooting for computer and software issues. Hardware-related problems are referred to separate manuals for the computer, controller, testing machine, hydraulic pump, and Windows.

  • Software troubleshooting covers system activation, test execution, data saving, and limit value errors.
  • For hardware issues, refer to instruction manuals of respective components.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Operating system: Windows 7/10; CPU: 1 GHz or higher; memory: 256 MB or more; HDD: 8 GB or more. p.22 ↗ Ensure PC meets these minimum requirements for software installation and operation. Whole series
Display resolution and color: 1920 × 1200, 65535 colors. p.22 ↗ Use a monitor capable of this resolution and color depth for proper display. Whole series
At least one empty slot for full-size PCI expansion bus required. p.22 ↗ PC must have a free full-size PCI slot for the AD expansion board. Whole series
Communication: RS232C. p.22 ↗ PC must have an RS232C port to connect to the controller. Whole series
Configuration files HTM_System.fil, HTM.ini, HTMTable.ini store basic configuration items such as calibration information. p.25 ↗ Back up these files to CD-R or other external media for recovery after PC failure or re-installation. Whole series
NI PCI-6132 driver installation requires Microsoft .NET Framework 4.6.2; installation screen may display for approximately 10 minutes. p.26 ↗ Ensure .NET Framework 4.6.2 is installed before driver installation; allow sufficient time. Systems with NI PCI-6132 board
ADM670 driver installation: confirm 'ADM-670PCI (win_2k_xp)' is displayed in 'MSCIENCE' in Device Manager. p.29 ↗ Verify correct driver installation for ADM670 board. Systems with ADM670 board
Up to three sensors for puncture test and three for tensile test can be registered, distinguished by ID. p.35 ↗ Plan sensor inventory accordingly; each sensor has unique ID. Whole series
Sensor name limited to <20 characters>; value up to 4 digits; units selectable from kN, N, tonf, kgf. p.35 ↗ Enter sensor information within these limits. Whole series
CH1 filter cutoff frequency selectable: 20 kHz or 100 kHz. p.35 ↗ Select appropriate filter based on test requirements. CH1 sensor
CH2 filter cutoff frequency selectable: 2 kHz or 20 kHz; CH3: 20 kHz or 100 kHz. p.36 ↗ Select appropriate filter for CH2 and CH3 sensors. CH2, CH3
External trigger can be used only with 4870-type controller; not usable if sampling interval >10 msec or sampling time <10 sec. p.37 ↗ Ensure sampling settings meet these conditions when using external trigger. Systems with 4870-type controller
Velocity calibration value set for 1 m/sec to 20 m/sec; input range <100, 4095>. p.41 ↗ Calibrate velocity for high-speed tests; not required below 1 m/sec due to feedback control. Whole series
Temperature calibration reference temperatures: -30, 0, 80, and 130 °C against span at 20 °C; linear interpolation for other temperatures. p.39 ↗ Perform temperature calibration at these points for accurate force measurement across temperature range. Systems with temperature calibration
Calibration value input range: -32767 to +32767 (integer); span: 0 to 65535 (integer). p.39 ↗ Enter calibration values within these limits. Whole series
To replace the sensor, exit software or move to the Home screen; otherwise incorrect measurement value is displayed. p.34 ↗ Always replace sensors in a safe software state and confirm connection state on System Setting screen. Whole series
Software checks sensor connection state before displaying Testing screen and before starting test; if changed, System Setting screen is forcibly displayed. p.37 ↗ Be aware that sensor changes trigger system setting verification. Whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
System cannot be activated. p.130 ↗ Software not properly installed. 1. Check if software is properly installed. 2. Reinstall software. How to reinstall, see Section 2.2.1 Installation Procedures. Section 5.1.1
Test cannot be executed. p.130 ↗ Test condition file not created or not in Testing mode. 1. Create the test condition file and switch to Testing mode. Section 5.1.1
Data cannot be saved after completion of the test. p.130 ↗ Save location issue. 1. Try to save the data to other folder or disk. Section 5.1.1
An error message is displayed when setting the upper or lower limit value. p.130 ↗ Upper limit value < lower limit value. 1. If the upper limit value < lower limit value, an error message is displayed. Section 5.1.1