MOC-120H Temperature Calibration Kit - Instruction Manual
Technical Reference Summary
Covered Models & Document Scope
This document covers the Temperature Calibration Kit for MOC-120H (part number 321-78320, dated Nov. 2018). It includes the hand-held digital alarm thermometer supplied with the kit and the procedure for installing and using the kit to calibrate the MOC-120H moisture analyzer. The optional printer for producing calibration records is referenced but not described in detail.
Section-by-Section Technical Breakdown
Electrical and metrological specifications of the hand-held digital alarm thermometer supplied with the calibration kit. It covers measurement range, accuracy bands, resolution, sampling rate, battery type and life, and auto power-off behavior.
- Measurement range -50 °C to 300 °C
- Accuracy ±1.5 °C from -50 to -10 °C, ±0.5 °C from -10 to 100 °C, ±1.5 °C from 100 to 200 °C, ±2.5 °C from 200 to 300 °C
- Display resolution 0.1 °C; alarm resolution 1 °C
- Sampling rate 1 second
- Battery: 2 × 1.5 V AAA/UM4, battery life 400 hours (standard); included battery is for operation check
- Auto power off: None
Steps for unpacking the thermometer and installing the batteries with correct polarity. A self-test with all display segments on for about 2 seconds occurs after battery installation.
- Unpack the thermometer
- Pull out the battery cover
- Install the battery with correct polarity position
- Put back the battery cover
- Self-test displays all segments for about 2 seconds, then normal temperature display
Basic measurement operation including probe insertion depth and reading stabilization. Also describes the data hold, maximum/minimum memory, temperature alarm setting, and Celsius/Fahrenheit selection functions.
- Insert probe into test substance about 25 mm / 1 inch; read when stable
- HOLD button retains display and flashes HOLD icon; press again to release
- MAX/MIN button cycles maximum, minimum, and normal display; clear memory by switching off and on
- SET button enters HI and LO alarm limits (default OFF); UP/DOWN adjust; ON/OFF toggles limit on/off
- Alarm sounds when temperature is outside set limit; any button stops sound but HI/LO icon keeps flashing until within limit
- Press and hold C/F button for about 4 seconds to change temperature unit
- Thermometer cannot be powered off while MAX/MIN or HOLD is active
- Clean probe immediately after each measurement to avoid cross contamination
- Test probe cannot be placed inside the oven
Assembly and placement of the temperature calibration kit on the MOC-120H. The sensor probe is inserted into the holder brackets, the kit is placed over the sample pan, and the heater cover must be completely closed with the probe fitting in the notch.
- Insert sensor probe into brackets on the holder; place probe end in the bracket at holder center
- Open heater cover and place kit over the sample pan; fit hole of holder to stay of main unit
- Confirm heater cover is completely closed and sensor probe fits in the notch of heater cover
- Caution: do not touch heater cover, sensor probe, or holder during calibration as they become very hot
Procedure for performing temperature calibration on the MOC-120H using the kit. Calibration can only be entered when 'ST' mode is selected as the temperature display mode, indicated by a blinking °C symbol. The calibration setpoint should be about 20 °C lower than the desired sample position temperature.
- Verify °C symbol is blinking, indicating 'ST' mode is selected
- Install temperature calibration kit and switch on digital thermometer in advance
- Select a temperature about 20 °C lower than the sample position temperature (e.g., set 80 °C if 100 °C at sample position is desired)
- Calibration record can be automatically produced with optional printer; refer to MOC-120H instruction manual for details
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Measurement range is -50 °C to 300 °C. p.2 ↗ | The thermometer is suitable for the temperature range encountered in MOC-120H calibration but cannot be used beyond 300 °C. | Hand-held digital alarm thermometer |
| Accuracy is ±0.5 °C between -10 °C and 100 °C, ±1.5 °C between -50 °C and -10 °C, ±1.5 °C between 100 °C and 200 °C, and ±2.5 °C between 200 °C and 300 °C. p.2 ↗ | Calibration uncertainty depends on the setpoint; best accuracy is in the -10 to 100 °C band. | Hand-held digital alarm thermometer |
| Display resolution is 0.1 °C over the full range; alarm resolution is 1 °C. p.2 ↗ | Temperature readings can be resolved to 0.1 °C, but alarm limits are set in 1 °C increments. | Hand-held digital alarm thermometer |
| Sampling rate is 1 second. p.2 ↗ | The display updates once per second; allow time for the reading to stabilize after probe insertion. | Hand-held digital alarm thermometer |
| Battery: 2 × 1.5 V AAA/UM4 or equivalent; battery life 400 hours (standard). The included battery is for operation check. p.2 ↗ | Use two AAA batteries; replace with the same type and leakage-free battery when display becomes dim. | Hand-held digital alarm thermometer |
| Auto power off: None. p.2 ↗ | The thermometer will not turn off automatically; manual power-off is required to conserve battery. | Hand-held digital alarm thermometer |
| Insert the probe into the test substance about 25 mm / 1 inch and read when the reading becomes stable. p.2 ↗ | Insufficient immersion depth may cause inaccurate readings; ensure at least 25 mm immersion. | Hand-held digital alarm thermometer |
| Press and hold the C/F button for about 4 seconds to change the temperature unit. p.3 ↗ | Unit selection requires a deliberate long press; accidental short presses will not change units. | Hand-held digital alarm thermometer |
| The thermometer cannot be powered off while MAX/MIN reading is displayed or HOLD function is activated. p.3 ↗ | Release all active functions before attempting to power off. | Hand-held digital alarm thermometer |
| The test probe cannot be placed inside the oven. p.3 ↗ | Do not use the probe in the MOC-120H oven directly; use the calibration kit as intended. | Hand-held digital alarm thermometer |
| Temperature calibration can be entered only when 'ST' mode is selected as temperature display mode, indicated by a blinking °C symbol. p.5 ↗ | Verify the °C symbol is blinking before starting calibration; otherwise the calibration procedure cannot be accessed. | MOC-120H with temperature calibration kit |
| Select a temperature about 20 °C lower than the sample position temperature (e.g., set 80 °C if 100 °C at the sample position is desired). p.5 ↗ | The calibration setpoint must be offset by approximately 20 °C below the target sample temperature. | MOC-120H with temperature calibration kit |
| Install temperature calibration kit and switch on digital thermometer in advance before calibration. p.5 ↗ | Ensure the kit is installed and the thermometer is powered on before entering the calibration procedure. | MOC-120H with temperature calibration kit |
| The calibration record can be automatically produced with the optional printer; refer to the instruction manual of MOC-120H for details. p.4 ↗ | An optional printer is required for automatic calibration records; consult the MOC-120H manual for printer operation. | MOC-120H with temperature calibration kit |
| Do not touch heater cover, sensor probe, or holder of the temperature calibration kit during temperature calibration as they become very hot. p.4 ↗ | Burn hazard; allow components to cool before handling. | MOC-120H with temperature calibration kit |