Evaluating the Performance of the LotusStream Gas-Liquid Separator for Preparative Supercritical Fluid Chromatography

Technical Reference Summary

Covered Models & Document Scope

Technical report C190-E250 evaluating the LotusStream gas-liquid separator in a Nexera UC Prep preparative supercritical fluid chromatography system; covers recovery rate, carryover, and contamination tests. No product model/size variants or ordering codes are given.

Technical Data

Parameter Value Condition
System Nexera UC Prep Stacked fraction system p.2 ↗ Recovery rate test
Flow rate 125 mL/min p.2 ↗ Recovery rate test
Detection wavelength 272 nm p.2 ↗ Caffeine, recovery rate test
BPR pressure 10 MPa p.2 ↗ Recovery rate test
BPR temperature 50 °C p.2 ↗ Recovery rate test
Sample concentration 10 mg/mL p.2 ↗ Caffeine, recovery rate test
Injection volume 1 mL p.2 ↗ Recovery rate test
Modifier Methanol p.1 ↗ LotusStream separator test
Modifier concentration 20 % p.1 ↗ LotusStream separator test
Modifier flow rate 100 mL/min p.1 ↗ LotusStream separator test
System Nexera UC Prep Multi-fraction system p.3 ↗ Contamination test
Makeup Methanol p.3 ↗ Contamination test
Makeup flow rate 15 mL/min p.3 ↗ Contamination test
BPR pressure 10 MPa p.3 ↗ Contamination test
Sample Linalool 1 % p.3 ↗ Contamination test
Fraction collector FRC-40 SF p.3 ↗ Contamination test
Column Silica column (250 mm L. × 20 mm I.D.) p.4 ↗ Carryover test
Makeup flow rate 10 mL/min p.4 ↗ Carryover test
Sample concentration 30 mg/mL p.4 ↗ Hydrocortisone, carryover test
Modifier Methanol p.4 ↗ Carryover test
Flow rate 3 mL/min p.4 ↗ Carryover test

Operating Limits

  • Contamination (Linalool): 0.006 % max. p.3 ↗
  • Carryover (Hydrocortisone): 0.024 % p.4 ↗