321-78426

Technical Reference Summary

Covered Models & Document Scope

This manual covers the Shimadzu UW/UX Series top-loading balances with aluminum UniBloc sensors. Model naming: UW/UX + capacity (220, 420, 620, 820, 1020, 2200, 4200, 6200, 8200 g) + H (high resolution) or S (standard resolution) + optional V (verified balance as a legal measuring instrument). Lineup includes small pan models (UW/UX220H, 420H, 620H, 820H, 1020H, 420S, 820S) and large pan models (UW/UX2200H, 4200H, 6200H, 4200S, 8200S). UW Series includes built-in weight functions (PSC, Clock-CAL, PCAL); UX Series does not.

Section-by-Section Technical Breakdown

pp. 17-21
Name and Function of Components

Describes the physical components of large pan and small pan models, including pan, pan supporter caps, windbreak set (standard on 0.001 g minimum display models), level screws, display panel, operation keys, and rear connectors (RS-232C, DATA I/O, DC IN, AUX, KEY, ground terminal, below-weigh hook cap). Explains key functions during weighing and menu selection, including short press vs. 3-second hold behavior.

  • Small pan models with minimum display 0.001 g include a windbreak set as standard.
  • Rear connectors: RS-232C, DATA I/O, AUX, DC IN, KEY, ground terminal, below-weigh hook cap.
  • [POWER] switches operation/standby; [CAL] enters span calibration or menu; [O/T] tares; [UNIT] changes unit; [PRINT] sends data.
  • Holding [CAL] for about 3 seconds recalls the last menu item set.
  • Holding [PRINT] for about 3 seconds sends date and time to a peripheral device.
pp. 22-27
Installation

Covers choosing the installation site, unpacking and delivery inspection, and mechanical installation. Installation includes removing transportation screws on UW models, installing pan supporter caps, pan, windbreak (0.001 g models), and leveling the balance. Lists standard packed items by model type.

  • Avoid air flow, vibration, direct sunlight, extreme temperature/humidity changes, corrosive or flammable gases, dust, electromagnetic waves, and magnetic fields.
  • Install on a strong and stable flat table or floor; large capacity balances require a sturdy floor/table supporting total load.
  • UW models: turn two transportation screws counterclockwise until they tighten again before operation.
  • UX models: start installation at step 3 (no transportation screws).
  • Standard accessories include pan supporter caps (4), pan, AC adapter, in-use protective cover, and instruction manual.
pp. 16-19, 33-44
Turning on the Power and Span Calibration

Describes power-on procedure and span calibration. UW Series supports span calibration using the built-in weight; external weights can be used for all models. Calibration settings include PSC fully-automatic calibration, Clock-CAL, and PCAL for the built-in weight (UW only).

  • UW Series can perform span calibration using the built-in weight.
  • External weight span calibration is available for all models.
  • PSC fully-automatic calibration and Clock-CAL are UW Series only.
  • PCAL calibrates the built-in weight (UW Series only) and has a password setting.
  • Calibration report settings support GLP/GMP/ISO compliant measurement management.
pp. 20-21
Basic Operation

Covers basic weighing and changing the unit display. Taring/zeroing behavior differs for verified balances: taring occurs at a weight exceeding 2.0% of capacity, zeroing within 2.0% of capacity. Only gram (g), percent (%), and piece counting (PCS) are set up before shipment.

  • Press [O/T] to tare the balance (displays zero).
  • Press [UNIT] to change weighing unit or select specific gravity measurement.
  • Units other than g, %, and PCS must be set up before use.
  • On verified balances, negative weight values may not be output or printed.
  • On verified balances, zero symbol indicates exact zero within ±0.25e (e = verification scale interval).
pp. 22-29
Menu Item Selection and Settings

Explains the menu system, menu map, item selection procedure, numeric value entry, last menu recall, menu reset, menu lock, and Windows Direct function setting. Menu lock prevents unauthorized menu changes.

  • Menu lock can be enabled to prevent menu item selection.
  • Last menu recall displays the last menu item that was set by holding [CAL] for about 3 seconds.
  • Menu reset returns settings to default.
  • Windows Direct function setting is available for direct communication with a PC.
  • Numeric values are entered using [UNIT] to increment and [PRINT] to move to the next digit.
pp. 30-31
Built-in Clock Set-up

Covers setting the date, date output style, time, and display during standby. The clock is used for calibration reports and date/time output.

  • Date and time can be output to a peripheral device by holding [PRINT] for about 3 seconds.
  • Date output style is configurable.
  • Display during standby is configurable.
  • Clock-CAL uses the built-in clock for scheduled calibration (UW Series only).
pp. 32
Display Selection

Covers bar graph display and changing the minimum display digit. The minimum display resolution can be decreased by one decimal place (10d display).

  • Bar graph display can be enabled.
  • Minimum display digit can be changed; 10d display decreases resolution by one decimal place.
  • 10d display is not applicable to verified balances as legal measuring instruments in using region.
pp. 33-44, 52
Calibration

Details calibration types and execution: span calibration using built-in weight (UW only), calibration check using built-in weight (UW only), span calibration using external weights, and calibration check using external weights. Includes calibration settings: selecting calibration type, PSC fully-automatic calibration (UW only), Clock-CAL (UW only), PCAL built-in weight calibration (UW only), PCAL password setting (UW only), and maintaining calibration reports for GLP/GMP/ISO.

  • Span calibration using built-in weight is UW Series only.
  • Calibration check using built-in weight is UW Series only.
  • External weight calibration is available for all models.
  • PSC fully-automatic calibration and Clock-CAL are UW Series only.
  • PCAL calibrates the built-in weight and has a password setting (UW Series only).
  • Calibration report settings and balance ID setting support GLP/GMP/ISO compliant measurement management.
pp. 45-48
Environment Settings

Covers stability and response (averaging), stability detection and settings, stability detection band, timing of stability mark illumination and data output, and tracking. These settings adjust the balance response to environmental conditions.

  • Stability and response can be adjusted via averaging settings.
  • Stability detection band is configurable.
  • Timing of stability mark illumination and data output can be set.
  • Tracking can be enabled to compensate for zero point drift.
  • Pouring mode is available and the right-most part of [O/T] key functions as environmental condition setting switch.
pp. 49-50, 108
Unit Display Set-up

Covers unit selection and percentage (%) conversion. Units other than gram, percent, and piece counting must be set up before use.

  • Only gram (g), percent (%), and piece counting (PCS) are set up before shipment.
  • Other units must be set up in the menu.
  • Percentage conversion is available.
  • Unit conversion coefficients are listed in Appendix 6.
pp. 51-61
Enhancing Productivity

Covers checkweighing and target display (comparator display types 1 and 2, target mode), piece counting (PCS), automatic printing and output (Auto Print), automatic compensation of zero point drift (Auto Zero), sample loading and unloading determination (Zero Range), taring/printing at stability, and registering container weight (Pretare value).

  • Checkweighing comparator display types 1 and 2 are available.
  • Target mode is available for target weighing.
  • Piece counting (PCS) is available.
  • Auto Print automatically outputs data.
  • Auto Zero compensates for zero point drift.
  • Zero Range determines sample loading/unloading.
  • Taring/printing at stability can be enabled.
  • Pretare value allows registering container weight.
pp. 62-77
Application Functions

Covers solid specific gravity measurement, liquid density measurement, peak hold, interval timer output, add-on mode for measuring many small samples, animal weighing mode, and formulation mode.

  • Solid specific gravity measurement is available.
  • Liquid density measurement is available.
  • Peak hold detects and holds peak values.
  • Interval timer outputs at fixed times.
  • Add-on mode allows measuring a large number of small samples.
  • Animal weighing mode is available.
  • Formulation mode is available.
pp. 78-92
Connections and Communications with External Equipment

Covers connecting printers, personal computers, PLCs and other serial communications equipment, RS-232C cable connections, data format, command codes, and communication settings (handshaking, format, speed, parity/bit length, stop bit, delimiter, decimal point symbol).

  • RS-232C connector is used for serial communication.
  • DATA I/O connector is available for external equipment.
  • Command codes are provided for integration with software.
  • Communication settings include handshaking, format, speed, parity/bit length, stop bit, and delimiter.
  • Decimal point symbol in output data is configurable.
  • Windows Direct function setting allows direct PC connection.
pp. 93, 9
Maintenance and Transportation

Covers daily maintenance and moving the balance. When moving the balance, do not hold the windbreak set; hold the main body from the bottom with both hands. For transportation during repairs, remove the pan and pan supporter.

  • Daily inspections and part replacement are described.
  • When moving the balance, do not hold the windbreak set; securely hold the main body from the bottom using both hands.
  • Always remove the pan and pan supporter when transporting the balance for repair.
  • Unplug the power cord from the AC adapter during inspections, maintenance, and when replacing parts.
  • Use only parts specified in the instruction manual when replacing parts.
pp. 94-96, 110-112
Troubleshooting

Provides general display messages, error display codes with descriptions and countermeasures, troubleshooting symptoms with probable causes and remedies, and LCD check procedure. Error displays include d OVER, d UNDER, and various calibration and communication errors.

  • General display messages indicate wait states, decimal place changes, date/time output, aborted operations, calibration check errors, menu lock status, and power failure reset.
  • Error display table lists error codes, descriptions, and countermeasures.
  • Troubleshooting table lists symptoms, probable causes, and countermeasures.
  • LCD check can be performed using menu items 37 (E8-StoP) or 38 (E8-Cont).
  • If all LCD segments do not display in full lit mode, contact service company.
pp. 97-110
Specifications

Appendix 1 provides specifications for the UW/UX Series models. Appendix 5 provides RS-232C connector specifications. Appendix 6 provides unit conversion coefficients. Appendix 7 provides performance checks. Appendix 8 provides below-weigh hook dimensions.

  • Specifications are listed in Appendix 1.
  • RS-232C connector specifications are in Appendix 5.
  • Unit conversion coefficients are in Appendix 6.
  • Performance checks are described in Appendix 7.
  • Below-weigh hook dimensions are in Appendix 8.
pp. 105-106
Standard Accessories and Optional Accessories

Appendix 3 lists standard accessories and maintenance parts. Appendix 4 lists optional accessories. Standard packed items vary by model type (large pan, small pan 0.01 g, small pan 0.001 g).

  • Standard accessories include pan supporter caps, pan, AC adapter, in-use protective cover, and instruction manual.
  • Windbreak set is standard on small pan models with minimum display 0.001 g.
  • Optional accessories are listed in Appendix 4.
  • Maintenance parts list is provided in Appendix 3.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
On verified balances, taring occurs at a weight exceeding 2.0% of capacity; zeroing occurs within 2.0% of capacity. p.19 ↗ Operators must understand that pressing [O/T] performs different functions depending on load, affecting measurement workflow on verified models. Verified balance as a legal measuring instrument in using region (V suffix models)
On verified balances, the zero symbol indicates exact zero within ±0.25e (e = verification scale interval). p.21 ↗ Zero indication tolerance is tight; ensure stable conditions before zeroing. Verified balance as a legal measuring instrument in using region (V suffix models)
Only gram (g), percent (%), and piece counting (PCS) units are set up before shipment. p.19 ↗ Other weighing units must be configured in the menu before use. Whole series
UW models have two transportation screws that must be turned counterclockwise until they tighten again before operation. p.27 ↗ Failure to loosen transportation screws can cause 'OL' or '-OL' display and prevent proper weighing. UW Series
When moving the balance again, turn the two transportation screws clockwise until they tighten. p.27 ↗ Transportation screws must be re-tightened before moving to protect the mechanism. UW Series
Small pan models with minimum display 0.001 g include a windbreak set as standard. p.18 ↗ These models require windbreak installation to prevent air flow effects on weighing. UW/UX220H, 420H, 620H, 820H, 1020H
The design standard period of usage for this product is seven years. p.8 ↗ Using beyond seven years may result in inability to maintain performance or malfunctions. Whole series
Replacement parts are supplied up until seven years after termination of manufacture. p.12 ↗ Plan maintenance and spare parts procurement within this period. Whole series
The [d xxx] display shows estimated drift of the balance display when a weight close to weighing capacity is placed. For UW4200S (capacity 4,200 g, minimum display 0.1 g), [d -0.3 g] indicates 3,999.7 g would be displayed for a 4 kg calibration weight. p.52 ↗ Drift value helps assess calibration accuracy near full capacity. UW4200S (example); concept applies to all models
Error code [d ouEr] (d OVER) indicates the 'd' value is 1000 counts or more; [d UndEr] (d UNDER) indicates the 'd' value is -1000 counts or less. p.52 ↗ These codes indicate excessive drift beyond ±1000 counts during calibration. Whole series
Menu lock can be enabled to prevent menu item selection; release via menu item 4.5.3. p.28 ↗ Prevents unauthorized parameter changes; must be released to change settings. Whole series
LCD check modes: menu item 37 [E8-StoP] holds full lit display until [O/T] pressed; menu item 38 [E8-Cont] automatically switches to mass display after approximately 0.5 seconds of full lit display. p.96 ↗ Allows verification of all LCD segments during startup or maintenance. Whole series
When the unit is set to 10d, the resolution of the minimum display is decreased by one decimal place. p.19 ↗ 10d display provides lower resolution but may improve stability; not applicable to verified balances in using region. Whole series (not applicable to verified balances in using region)
In Pouring mode, the right-most part of [O/T] key marked with a circle functions as the switch for environmental condition setting. p.19 ↗ Provides quick access to environment settings during pouring applications. Whole series
Negative weight values may not be output or printed in all output functions on a verified balance as a legal measuring instrument in using region. p.19 ↗ Data output limitations apply to verified models; plan data handling accordingly. Verified balance as a legal measuring instrument in using region (V suffix models)
The balance complies with EN61326-1 class B for emissions and basic electromagnetic environment for immunity; UW□□□V Series complies with industrial electromagnetic environment. p.10 ↗ Class B suitable for residential and low-voltage network locations; V models for industrial environments. Whole series; UW□□□V Series for industrial environment
When installing in industrial locations, locate the product away from devices emitting strong electromagnetic noise and use a separate power source. p.10 ↗ Reduces risk of measurement errors due to electromagnetic interference. Whole series
Unplug the power cord from the AC adapter during inspections, maintenance, and when replacing parts. p.8 ↗ Safety requirement to prevent electric shock or short-circuits. Whole series
Always remove the pan and pan supporter when transporting the balance for repair. p.9 ↗ Prevents damage to the main body during transport. Whole series
When moving the balance, do not hold the windbreak set; securely hold the main body from the bottom using both hands. p.9 ↗ Prevents windbreak detachment and potential injury. Models with windbreak set (0.001 g minimum display models)

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
d ouEr (d OVER) p.52 ↗ The 'd' value is 1000 counts or more. Not specified in manual; indicates excessive drift during calibration. Section 7, page 52
d UndEr (d UNDER) p.52 ↗ The 'd' value is -1000 counts or less. Not specified in manual; indicates excessive drift during calibration. Section 7, page 52
Trouble in weight loading mechanism p.111 ↗ Transportation screws not loosened or mechanism issue. Check transportation screws. (Refer to 2.3) Section 14.2
The load on the pan is unstable at calibration p.111 ↗ Wind or vibration affecting the balance. Avoid wind and vibration. Section 14.2
The drift of zero point is large at calibration. p.111 ↗ Pan not installed properly or load on pan. Install the pan properly. Unload the pan. Section 14.2
The drift is large at the time of PCAL p.111 ↗ Incorrect weight used. Use correct weight. Section 14.2
The drift is large at span calibration p.111 ↗ Incorrect weight used. Use correct weight. Section 14.2
Calibration weight is wrong p.111 ↗ Incorrect calibration weight used. Use correct weight. Section 14.2
Received command code is not correct p.111 ↗ Incorrect command characters or delimiter. Check command characters and delimiter etc. (Refer to 12.3) Section 14.2
Integer of the displayed unit exceeded 7 digits p.111 ↗ Load exceeds display capacity. Decrease the load. Section 14.2
Failure in the balance. p.111 ↗ Internal balance failure. Contact the service company. Section 14.2
PCAL password error p.111 ↗ Incorrect PCAL password entered. Check the password. (Refer to 7.3.5) Section 14.2
Attempted to set improper numeric value. p.111 ↗ Invalid numeric value or decimal places. Review the numeric value and decimal places. Section 14.2
Supply voltage is abnormal. p.111 ↗ Abnormal supply voltage. Check the supply voltage. Section 14.2
Nothing is displayed. p.112 ↗ The AC adapter is disconnected. The breaker of the room is off. The voltage is wrong. Check the power and connect AC adapter correctly. Section 14.3
"OL" or "-OL" is displayed. p.112 ↗ Transportation screws haven't been loosened. (UW only) The pan is displaced. Pan supporter caps are not installed. The load on the pan is too large. Turn them anti-clockwise until they stop. Place the pan properly. Install pan supporter caps. Use balance within its capacity. Section 14.3
Display does not change when pan is loaded. p.112 ↗ Pan is displaced. Place the pan properly. Section 14.3
The display fluctuates and it is difficult to get the stability symbol to display. p.112 ↗ Affected by vibration or air flow. Protective in-use cover touches the pan. Install the balance at a proper site. Try changing environment settings. Adhere the cover firm to the balance body. Section 14.3
The weighed result is not accurate. p.112 ↗ Span calibration has not been done. Balance was not zeroed before measurement. Calibrate it properly. Press the [O/T] key to zero the mass display before measuring. Section 14.3
Does not display the unit desired to use. p.112 ↗ The unit has not been set up. Set it up in advance. Section 14.3
Menu item selection is rejected. p.112 ↗ Menu Lock is ON. Release Menu Lock. (See 4.5.3) Section 14.3
An error code is shown. p.112 ↗ Various errors. See "Error Display" (14.2). Section 14.3