Determination of Pb, Cd, Cr and Ni in

Contents

  1. Procedure p. 2
    Sample preparation and measurement conditions for the analysis
  2. p. 3
    Presentation and discussion of analytical results
  3. Spec p. 4
    Calibration data and range tables for the elements
  4. Overview p. 5
    Summary of findings and method performance
  5. p. 6
    Agilent website and contact details
  6. p. 4 table, 9 rows
  7. p. 2 table, 16 rows
  8. p. 3 table, 2 rows
  9. p. 5 table, 1 rows

Table 3. Furnace program for the analysis of Pb, Cd, Cr, and Ni in wheat flour.

StepPb Temp (°C)Pb Time (s)Cd Temp (°C)Cd Time (s)Cr Temp (°C)Cr Time (s)Ni Temp (°C)Ni Time (s)Read
1Drying8510.08510.08510.08510.0-
2Drying9525.09525.09525.09530-
3Drying1205.01205.01205.01205.0-
4Ashing5005.04005.013005.012005.0-
5Ashing50010.04005.013005.012005.0-
6Ashing5002.04002.013002.012002.0-
7Atomization21000.818000.726500.628000.8Yes
8Atomization21001.018001.026501.028001.0Yes
9Clean out27002.023002.026502.028002.0-

Table 1. Agilent 280Z GFAAS instrument operating conditions.

ParameterPbCdCrNi
ElementPbCdCrNi
Instrument modeAbsorbanceAbsorbanceAbsorbanceAbsorbance
Calibration modeConcentrationConcentrationConcentrationConcentration
Measurement modePeak heightPeak heightPeak HeightPeak Height
Wavelength (nm)283.3228.8357.9232.0
Lamp typeUltrAA lampHollow cathode lampHollow cathode lampUltrAA Lamp
Lamp current (mA)10477
Slit width (nm)0.50.50.20.2
Replicates2222
Calibration standards (μg/L)10, 30, 501, 2, 32, 4, 810, 25, 50
Sample volume (μL)10101010
Sample introductionAuto-mixAuto-mixAuto-mixAuto-mix
Background correctionOnOnOnOn
ModifierPremixed 2000 μg/mL NH4H2PO4 + 120 µg/mL Mg(NO3)22000 mg/L NH4H2PO4 300 mg/L Pd150 mg/L Pd
Modifier injection TypeCo injectionCo injectionCo injectionCo injection
Modifier volume (μL)5555

Table 2. Parameters used for microwave digestion (where t is the time in minutes, and T1 and T2 are the programmed initial and final vessel temperatures).

Stept (min)T1 (°C)T2 (°C)Power (W)
15.5Ambient2001500
24.52002001500

Table 6. %RSD of a 5 ppb Pb standard measured periodically over 7 hours.

SampleMeasured value, (mean, n=26) (ppb)RSD (%)
Wheat Flour5.221.6

Engineering Facts

13 facts · every claim linked to a page
Selection

Zeeman background correction GFAAS is chosen for trace Pb, Cd, Cr and Ni in grain because of high sensitivity, low running costs and accurate correction of complex backgrounds.

Defines the core technology selection criterion for trace metals in grain matrices.

Source
“Features such as high sensitivity, low running costs, and accurate correction for complex backgrounds make GFAAS with Zeeman background correction suitable for the determination of trace levels of Pb, Cd, Cr and Ni in grain.”
Selection

The autosampler must support at least 135 solutions and perform automatic standard preparation, modifier addition and overrange dilution.

Sets the minimum autosampler capacity and automation features needed for this method.

Source
“The Agilent PSD 120 provides the capacity for up to 135 solutions and performs automatic standard preparation, modifier addition and overrange dilution.”
Performance

Longitudinal graphite tube heating and Constant Temperature Zone design extend tube lifetime and give uniform heating needed for low-level complex samples.

Justifies furnace design requirements for accurate low-level analysis and tube lifetime.

Source
“In addition to prolonging the graphite tube lifetime, these features provide consistent, uniform heating, which is essential for the accurate low-level determination of complex samples.”
Operation

A furnace viewing camera is used to set probe dispensing height and confirm optimum drying conditions, simplifying method development.

Explains how to set up and verify drying parameters during method development.

Source
“Still images and videos can be recorded to set the probe dispensing height and confirm optimum drying conditions, simplifying method development.”
Operation

Chemical modifiers are required to reduce matrix interferences and stabilize analyte signals.

Mandates modifier use for reliable signal stability in grain digests.

Source
“Chemical modifiers were used to reduce matrix interferences and to stabilize analyte signals.”
Operation

The autosampler rinse solution is 0.01% v/v non-ionic surfactant with 2% v/v high purity nitric acid.

Specifies the rinse chemistry needed to prevent carryover and maintain baseline.

Source
“The PSD 120 autosampler rinse solution comprised a 0.01 % v/v non-ionic surfactant and 2% v/v of high purity nitric acid.”
Operation

Microwave digestion uses nitrogen pressurization at 40 bar, so vials do not need sealing.

Defines the digestion vessel sealing and pressurization requirement.

Source
“Sealing of the vials was not required since the SRC was pressurized using nitrogen gas at 40 bar.”
Operation

Wet digestion for Cr and Ni uses a hot block at 80 °C, then 120 °C for 1 hour with reflux, then 150 °C for 1 hour to boil off water.

Gives the required hot-block temperature profile for Cr and Ni sample preparation.

Source
“The temperature was increased to 120 °C for 1 hour and the test tubes were covered with a watch glass to allow reflux.”
Operation

Calibration standards must be matrix matched to 20% HNO3 to match the acid content of digested samples.

Prevents calibration bias from acid mismatch between standards and digests.

Source
“The bulk standards were matrix matched to 20% HNO3 to ensure consistency with the acid content of the digested samples.”
Operation

Inert high-purity gas must be used throughout all furnace programs, with default SpectrAA gas flow rates.

Specifies gas purity and flow settings required for the furnace programs.

Source
“Inert high-purity gas was used throughout all furnace programs.”
Compliance

Method detection limits obtained with the 280Z Zeeman GFAAS are lower than those specified in the GB methods.

Confirms the instrument meets the GB method MDL requirements for compliance.

Source
“The MDLs for both elements obtained using the Agilent 280Z Zeeman GFAAS are lower than the MDLs specified in the GB methods.”
Performance

Instrument stability over a seven-hour period showed 1.6% RSD and only 4% deviation from the initial reading.

Demonstrates long-term stability needed for unattended batch runs.

Source
“The %RSD was 1.6 over a seven-hour period.”
Selection

The 280Z GFAAS is suited to food testing labs needing accurate, sensitive and cost-effective instrumentation with long tube life and low argon use.

Summarizes the selection criteria for food testing labs including cost and gas consumption.

Source
“With its excellent precision, faster analysis times when used in conjunction with the PSD 120 autosampler, long graphite tube lifetime and low argon gas usage, the 280Z GFAAS is ideally suited to meet the needs of food testing labs.”