Contents

  1. Overview pp. 3-4
    Table of contents listing manual sections
  2. Overview pp. 5-6
    Italian translation of the user manual
  3. Overview pp. 7-8
    Chinese translation of the user manual
  4. Overview pp. 9-10
    English translation of the user manual
  5. Warning pp. 13-32
    Italian safety definitions, warnings, and disposal instructions
  6. Warning pp. 33-52
    German safety warnings, proper use, and startup
  7. Warning pp. 53-72
    French safety warnings, proper use, and startup
  8. Warning pp. 73-92
    Spanish safety warnings, proper use, and startup
  9. Warning pp. 93-112
    Chinese safety warnings, proper use, and startup
  10. Warning pp. 113-132
    Japanese safety warnings, proper use, and startup
  11. Warning pp. 133-152
    English safety warnings, proper use, and startup
  12. Overview pp. 153-154
    Description of TriScroll 600 Series 50 Hz pump
  13. Spec pp. 155-158
    Technical specifications and performance data
  14. Procedure pp. 159-160
    Electrical connection instructions and wiring
  15. Procedure pp. 161-166
    Mechanical installation and connection procedures
  16. Procedure pp. 167-168
    Cleaning and preventive maintenance instructions
  17. Procedure pp. 169-170
    Procedure for replacing tip seals
  18. Services pp. 171-172
    Available service options and support
  19. Spec pp. 173-174
    Additional technical data and specifications
  20. pp. 175-178
    Contact information for Vacuum Products Division
  21. Overview p. 179
    Index of manual contents
  22. p. 154 table, 18 rows
  23. p. 170 table, 3 rows
  24. p. 155 table, 8 rows
  25. p. 159 table, 1 rows
  26. p. 174 table, 3 rows
  27. p. 172 table, 16 rows
  28. p. 177 table, 11 rows

Table 1 TriScroll 600 Series 50hz 220V-230V Single-Phase Dry Scroll Pumps Technical Specifications

ParameterTriScroll 600 Series 50hz 220V-230V Single-Phase Dry Scroll Pumps
ModelTriScroll 600 Series 50hz 220V-230V Single-Phase Dry Scroll Pumps
Interface dimensionsH=355.8 mm x W=313.8 mm x L= 481.1 mm (14.01 X 12.35 X18.94)standard pump
Peak pumping speed (ISO 21360-1: 2020) (ISO 21360-2: 2020)50 Hz: 420 l/m, 25.2 m3/hr (14.8 cfm)
MediaClean air. No toxic, corrosive, explosive or particulate forming gases
Ultimate Pressure* (ISO 21360-1: 2020) (ISO 21360-2: 2020)7.0 x 10-3 Torr (9.3 x 10-3 mbar)
Maximum recommended continuous inlet pressure50 Torr
Maximum inlet pressure at nominal voltage1.0 atmosphere (0 psig)
Maximum outlet pressure***1.1 atmosphere (1.5 psig)
Inlet connectionNW25 or NW40
Exhaust connection5/8” Hose barb or NW25 fitting
Gas ballastFemale 1/4" National Pipe Thread (40 Micron sintered filter provided)
Ambient operating temperature5 °C to 40 °C (41 °F to 104 °F)
Ambient Relative Humidity5-95%; non-condensing
Storage temperature–20 °C to 60 °C (–4 °F to 140 °F)
Motor rating: EN 60034-1-2:2007 EN 60034-30 (GB 18613-2020)1.0 HP (0.75 kW)
Operating voltages** Single phase220-230 50Hz VAC +/-10%
Motor full load currents4.60 - 4.65 FLA, starting current 31-32amps
Motor protectionType U automatic

* Ultimate Pressure; ** Operating voltages; *** Maximum outlet pressure

Table 5 Maintenance and Tooling Kits

DescriptionContentsApplicable TriScroll ModelPart Number
Major Maintenance KitAll bearing, bearing seals, bearing lubricant, O-rings and tip seals required to rebuild TriScroll 600 series Dry Scroll Pumps.All TriScroll 600 Series modelsPTSS0600MK
Maintenance Tool KitAll fixtures and tools require to perform any maintenance on TriScroll 600 series pumpsAll TriScroll 600 Series modelsPTSS0600TK
Replacement Tip Seal SetReplacement tip seals and static O-rings for TriScroll 600 series pumps.All TriScroll 600 Series modelsPTSS0600TS

NOTE: the Maintenance Tool Kit is also required for tip seal replacement.

Table 1 TriScroll 600 Series 50hz 220V-230V Single-Phase Dry Scroll Pumps Technical Specifications (continued)

CharacteristicTriScroll 600 Series 50hz 220V-230V Single-Phase Dry Scroll Pumps
Operating speed50 Hz: 1450 RPM
Cooling systemAir-cooled
WeightPump only: 33kg (73 lbs) Shipping weight: Approximately 41kg (87 lbs)
Sound Pressure Levels (EN ISO 2151:2008) (ISO 4871:1996) (ISO 11201:2010)59 ± 2 dB(A)
Vibration level at inlet (per ISO 10816-1)6.3 mm/sec
InstallationIndoor use, Equipment Class 1, Installation Category II, Pollution Degree 2
Altitude2000 m (6562 ft)
SafetyMachinery Directive 2006/42/EC: 2006 EN 1012-2:1996+A1: 2010 EN ISO 12100:2012 Low Voltage Directive 2006/95/EC: 2007 EC 61010-1 Ed. 3.1 b:2018 CSA certified for US and Canada CE, ROHS

* If the pump is new or has been exposed to atmosphere for an extended period of time, it is possible that water can be present which can affect the pump’s ability to achieve ultimate pressure. To remove water from the pump and system, activate the gas ballast using the procedure in the gas ballast paragraph on page 164. Ultimate pressure will vary for different configuration and RPM setting. ** Mains supply voltage fluctuations shall not exceed ±10% of the nominal voltage.

Table 2 Full Load Motor Current, Amp

1 phase motor220V230V
50Hz4.654.60

Table 7 Accessories

Part NumberDescription
S4807001Exhaust Extension
PTS600EXFILExhaust Filter Kit
PTSPURGEKITPurge Kit (contains a flow meter and all necessary valving and tubing.)

Table 6 Troubleshooting Chart

ProblemPossible CausesCorrective Actions
Will not start.Motor thermal protector openAllow the pump to cool. Identify the cause of overload.
Excessive voltage drop in the power cord.Check the size and length of the cable.
Defective motorInspect. Contact Agilent.
Pump jammedInspect. Contact Agilent.
Poor ultimate pressureSystem leakLocate and repair the leak.
Water vapor in the pump or systemFlush pump with air or dry nitrogen. Activate the Gas Ballast.
Water in the pumpActivate the Gas Ballast.
Gas ballast pluggedReplace breather vent. Contact Agilent.
Solvent in the pumpActivate the Gas Ballast. Install trap or filter.
Worn tip sealReplace tip seals.
Poor conductance to pumpReplumb with shorter and/or larger diameter tubing.
Low pumping speed.Replace tip seals.
Hammering noisePump overheatedCheck the ventilation to pump. Check ambient temperature.
Check the ambient temperatureCheck the intake screen. Flush pump. Disassemble the pump head and inspect.
Debris in pumpCheck intake screen. Flush pump. Disassemble pump head and inspect.
Pump runs intermittentlyMotor thermal protector is cycling open and closed. *Allow the pump to cool. Identify the cause of the overload. Check for proper operation of the fan. Ensure there is adequate space for air to circulate at inlets.

* Сноска в тексте не приведена.

Health and safety — The product has been exposed to the following substances: (by selecting ‘YES’ you MUST complete the table to the right)

Substances (please refer to MSDS forms)Trade nameChemical nameChemical SymbolCAS Number
Toxic
Harmful
Corrosive
Reactive
Flammable
Explosive (*)
Radioactive (*)
Biological (*)
Oxidizing
Sensitizer
Other dangerous substances

* Agilent will not accept delivery of any product that is exposed to radioactive, biological, explosive substances or dioxins, PCB’s without written evidence of decontamination.

Engineering Facts

12 facts · every claim linked to a page
Installation

The pump starts immediately when the power cord is plugged in, so the outlet must be treated as the on/off control and kept accessible.

Prevents unexpected startup during installation and requires an accessible disconnect means.

Source
“Pump will turn on immediately upon plugging the power cord to the electrical outlet.”
Installation

The pump is rated for indoor use only in clean, dry environments and is not certified for outdoor operation.

Rules out outdoor siting and requires a protected indoor installation location.

Source
“This pump is not designed or certified for operation outdoors.”
Installation

A minimum 20 cm clearance on all sides is required for ventilation, maintenance access and access to the power cord disconnect.

Sets the minimum installation envelope and prevents overheating from blocked airflow.

Source
“This pump requires a clear space of at least 20 cm (8 in.) on all sides for ventilation,”
Operation

Operating ambient must stay within +5 °C to +40 °C and 5–95% non-condensing relative humidity.

Defines the allowable site conditions for reliable operation and warranty compliance.

Source
“Temperature: +5 °C to +40 °C (41 °F to 104°F)”
Operation

Adsorbed water vapor is the most common cause of poor initial vacuum performance, and gas ballast may need 30 minutes to 24 hours to reach full performance.

Prevents misdiagnosing a new or long-idle pump as faulty during commissioning.

Source
“it can typically take between ~30 minutes to ~24 hours of gas ballast activation to achieve the full vacuum performance.”
Installation

The pump weighs about 33 kg unpacked, so proper lifting aids and techniques are mandatory to avoid injury.

Drives the need for lifting equipment and two-person handling during installation.

Source
“wiegen ohne Verpackung ca. 33 kg (70 Pfund).”
Installation

When unpacking, the pump must be lifted with a hoist hooked to the eyebolt, and the hoist must be rated for 100 kg.

Specifies the minimum lifting capacity required for safe unpacking and positioning.

Source
“Das Hebezeug sollte eine Tragfähigkeit von 100 kg haben.”
Compliance

The pump is configured for 220–230 V 50 Hz only and cannot be run at 60 Hz or at 100–115 V without damage.

Confirms supply compatibility before selection and prevents destructive miswiring.

Source
“cannot be configured to operate at either 60hz or at low voltage, 100V-115V.”
Safety

The product must be grounded; the supplied cord has a grounding wire and plug that must go into a properly grounded outlet.

Grounding is mandatory for electrical safety and reduces shock risk on short circuit.

Source
“This product should be grounded.”
Operation

Exhaust back pressure must not exceed 1.5 psig when an exhaust hose or muffler is used.

Sets the maximum allowable exhaust restriction to avoid overloading the pump.

Source
“排気抵抗によって背圧が1.5 psigを超えることがないようにしてください。”
Maintenance

Before servicing or disconnecting the pump, wait for it to stop completely and cool to room temperature because of sharp edges and hot surfaces.

Prevents burns and cuts during maintenance and system disconnection.

Source
“wait for the pump to come to a complete stop and cool down to room temperature.”
Safety

Do not block the cooling vents; a pump surface temperature above 80 °C is potentially damaging and requires shutdown and inspection.

Gives a concrete over-temperature limit for safe operation and troubleshooting.

Source
“A pump surface temperature in excess of 80 °C (176 °F) is potentially damaging.”