Agilent Technologies TriScroll 600, Two-Stage, Dry, Inverter-Driven Pump
Agilent Technologies TriScroll 600, Two-Stage, Dry, Inverter-Driven Pump — Vacuum technologies with Motor power 1 kW, Noise level 50 dB, Ultimate pressure 1.3 mbar.
The TriScroll 600 is a two-stage, dry, inverter-driven scroll pump configured for 220–230 V 50 Hz single-phase operation only, delivering 420 l/min peak pumping speed and 9.3 × 10⁻³ mbar ultimate pressure. It is optimized for helium leak detection and clean inert-gas service, with a 1.0 HP motor and NW25/NW40 inlet. Within the Triscroll series it sits above the TS300 in capacity while sharing the same ultimate pressure.
Agilent Technologies TriScroll Series — Model Comparison & Selection
| Model | Motor power, kW | Storage temperature, °C | Noise level, dB | Ultimate pressure, mbar | Operating speed, rpm | Shipping weight |
|---|---|---|---|---|---|---|
| TriScroll 600, Two-Stage, Dry, Inverter-Driven Pump Current | 1 | -6.67 | 50 | 1.3 | 1450 | 41 |
| TriScroll 300, Two-Stage, Dry, Inverter-Driven Pump | 0.5 | -20...60 | 50 | 1.3 | 334.2 | 33.6 |
Full specifications
| Motor power | 1.0 kW |
| Noise level | 50 dB |
| Ultimate pressure | 1.3 mbar |
| Coverslip type | Clean air. No toxic, corrosive, explosive or particulate forming gases |
| Operating Speed | 1450 RPM |
| Peak pumping speed mfr | 250 hr |
| Exhaust connection mfr | 1 swivel |
| Gas ballast connection mfr | 16 |
| Power requirements mfr | 200 amps |
| Maximum recommended continuous inlet pressure mfr | 50 Torr |
| Maximum outlet pressure mfr | 1.1 atmosphere |
| Gas ballast mfr | 1 NPT |
| Operating voltages mfr | 220 % VAC |
| Motor full load currents mfr | 4.60 amps |
| Motor protection mfr | Type U automatic automatic |
| Vibration level at inlet mfr | 6.3 mm/sec |
| Installation mfr | 1 |
| Safety mfr | 2006 |
| Maximum outlet pressure*** mfr | 1.1 atmosphere |
| Maintenance Tool Kit mfr | 600 pumps |
| Replacement Tip Seal Set mfr | 600 pumps. |
| S4807001 mfr | Exhaust Extension Extension |
| PTS600EXFIL mfr | Exhaust Filter Kit Kit |
| PTSPURGEKIT mfr | Purge Kit Kit |
| Will not start. mfr | Motor thermal protector open open |
| Poor ultimate pressure mfr | System leak leak |
| Pump runs intermittently mfr | Motor thermal protector is cycling open and clos |
| Clearance mfr | 20 cm |
| Hoist capacity mfr | 100 kg |
| Maximum exhaust back pressure mfr | 1.5 psig |
| Peak pumping speed (TS300inv) mfr | 250 hr |
| Peak pumping speed (TS600inv) mfr | 500 h |
| Ultimate pressure (TS600inv) mfr | 9.3 mbar |
| Shipping Weight | 41 lbs |
| Weight | 33 lbs |
| Weight – pump only mfr | 26 kg |
| Weight (TS300inv) mfr | 26 kg |
| Weight (TS600inv) mfr | 31 kg |
| Full Load Motor Current | 4.65 A |
| Maximum inlet pressure at nominal voltage mfr | 1.0 atmosphere |
| Operating voltages** Single phase mfr | 220 |
| Inlet connection | 25 |
| Gas ballast port mfr | 1 NPT |
| Interface dimensions mfr | 355.8 pump |
| Cooling System | Air-cooled |
| Noise level | 50 dB |
| Operating temperature | 5 °C |
| Storage temperature | 20 °F |
| Ambient Relative Humidity mfr | 5 condensing |
| Sound Pressure Levels mfr | 59 dB |
| Altitude mfr | 2000 m |
| Ambient relative humidity mfr | 5 |
| Hammering noise mfr | Pump overheated overheated |
| Relative humidity mfr | 5...95 |
| Peak pumping speed (ISO 21360-1: 2020) (ISO 21360-2: 2020) mfr | 50 hr |
| Ultimate Pressure* (ISO 21360-1: 2020) (ISO 21360-2: 2020) mfr | 7.0 Torr |
| Motor rating: EN 60034-1-2:2007 EN 60034-30 (GB 18613-2020) mfr | 1.0 HP |
| Sound Pressure Levels (EN ISO 2151:2008) (ISO 4871:1996) (ISO 11201:2010) mfr | 59 dB |
| Vibration level at inlet (per ISO 10816-1) mfr | 6.3 mm/sec |
| Noise level per ISO 11201, variable w/frequency mfr | 55 dBA |
Description
Construction and series positioning
- Two-stage dry scroll mechanism with inverter-driven 1.0 HP (0.75 kW) motor; no oil to replace.
- Configured exclusively for 220–230 V ±10%, 50 Hz single-phase; cannot be set for 60 Hz or 100–115 V.
- Sibling TS300 offers lower pumping speed at the same 1.3 mbar ultimate pressure; the 15 and 30 m³/hr models bracket the range below the TS600.
- Inlet NW25 or NW40; exhaust 5/8" hose barb or NW25; gas ballast and bearing purge ports are 1/4" NPT.
Behavior under load and media
- Peak pumping speed 420 l/min (25.2 m³/hr) at 50 Hz; ultimate pressure 9.3 × 10⁻³ mbar (7.0 × 10⁻³ Torr).
- Maximum recommended continuous inlet pressure 50 Torr; maximum outlet pressure 1.1 atm (1.5 psig).
- Water vapor raises ultimate pressure until gas ballast is used; high vapor loads shorten tip-seal life.
- Intended for helium leak detection and air/inert gases; not for toxic, corrosive, explosive, or particulate-forming gases.
Installation, control and operation
- No on/off switch: the pump starts immediately when the cord is plugged in; provide an accessible disconnect.
- Full-load current 4.65 A at 220 V / 4.60 A at 230 V; starting current approximately 31–32 A.
- Auto-reset thermal motor protector restarts the pump automatically after cooling; disconnect mains before service.
- Minimum 20 cm free space on all sides for air circulation; ambient 5–40 °C, 5–95% RH non-condensing.
- Use bellows hose at the inlet for vibration isolation (6.3 mm/sec per ISO 10816-1).
Integration & Operating Notes
The pump is factory-configured for high-voltage 50 Hz operation only and cannot be reconnected for 60 Hz or 100–115 V. Mains fluctuations beyond ±10% risk motor overheating and invalid performance; verify the branch circuit before installation.
There is no on/off control on the pump; applying mains power starts the motor immediately. Install an accessible disconnect device or switched outlet, and confirm inlet and exhaust connections are secure before energizing.
Full-load current is 4.65 A at 220 V and 4.60 A at 230 V, but starting current reaches approximately 31–32 A. Circuit breakers, wiring and any extension cord must be rated accordingly; extension cords are limited to 3-wire, minimum 14 AWG, maximum 25 ft (7.6 m).
Operating continuously above the maximum recommended inlet pressure of 50 Torr can damage the scroll mechanism or shorten tip-seal life. For processes with higher gas loads, throttle the inlet or select a larger pump rather than running the TS600 above this limit.
Maximum outlet pressure is 1.1 atmosphere (1.5 psig). Restricted or undersized exhaust lines raise backpressure and cause overheating; use short, large-diameter exhaust plumbing and avoid filters or silencers that add significant restriction.
The pump is air-cooled and surface temperatures above 80 °C are potentially damaging. Blocking vents or installing in a confined cabinet without airflow leads to overheating and thermal trips; keep at least 20 cm (8 in.) free space around the system.
If ultimate pressure is not reached after initial installation or storage, activate gas ballast for roughly 30 minutes to 24 hours before assuming a fault. Regular gas ballast operation removes condensed water and tip-seal dust and is part of the preventive maintenance plan.
The motor has an auto-reset thermal protector that restarts the pump automatically once the motor cools. Before any maintenance, isolate the mains supply to prevent unexpected restart and injury from the cooling fan.
Selection & Integration Q&A
Pick the TS600 when you need 420 l/min peak speed at 9.3 × 10⁻³ mbar for helium leak detection or inert-gas duty. The TS300 gives the same ultimate pressure at lower throughput; the 15 and 30 m³/hr models suit smaller chambers. All are dry scroll, but only the TS600 50 Hz variant is fixed at 220–230 V.
Inlet is NW25 or NW40; exhaust is 5/8" hose barb or NW25. Use short, large-diameter lines and a bellows hose at the inlet for vibration isolation. The pump is a two-stage dry scroll and does not require a backing pump for its specified ultimate pressure; exhaust must stay below 1.5 psig.
Install indoors at 5–40 °C and 5–95% RH non-condensing, with at least 20 cm free space on all sides for air cooling. The bench must support 33 kg. Continuous inlet pressure must stay at or below 50 Torr, and exhaust backpressure below 1.5 psig; the auto-reset thermal protector will trip if these are exceeded.
Base delivery does not include the Major Maintenance Kit PTSS0600MK, Maintenance Tool Kit PTSS0600TK, or Replacement Tip Seal Set PTSS0600TS, nor the optional exhaust extension, exhaust filter kit, or purge kit. The pump has no oil to change; tip-seal replacement and gas ballast operation form the core service plan.