Induztr
Engineering Digest Document Viewer Related Models

TwisTorr 305 FS Turbomolecular Pump User Manual

Contents

  1. Overview pp. 1-6
    Cover, contents and document introduction
  2. Procedure pp. 16-42
    Italian safety, installation, operation and maintenance instructions
  3. Procedure pp. 43-72
    German safety, installation, operation and maintenance instructions
  4. Procedure pp. 73-99
    French safety, installation, operation and maintenance instructions
  5. Procedure pp. 100-126
    Spanish safety, installation, operation and maintenance instructions
  6. Procedure pp. 127-156
    Chinese safety, installation, operation and maintenance instructions
  7. Procedure pp. 157-183
    Japanese safety, installation, operation and maintenance instructions
  8. Procedure pp. 184-210
    English safety, installation, operation and maintenance instructions
  9. Spec pp. 211-249
    Technical data, graphs, dimensions and operating parameters
  10. Ordering pp. 250-252
    Accessories, spare parts and ordering information
  11. pp. 253-257
    Vacuum Products Division contact and legal information
  12. p. 219 table, 17 rows
  13. p. 211 table, 16 rows
  14. p. 220 table, 13 rows
  15. p. 251 table, 11 rows
  16. p. 255 table, 5 rows
  17. p. 188 table, 8 rows
  18. p. 205 table, 4 rows
  19. p. 214 table, 8 rows
  20. p. 235 table, 2 rows
  21. p. 257 table, 11 rows

Tab. 1 Technical Specification

Technical SpecificationISO 100/ CF 6" / ISO 160 / CF 8"
Pumping speed N2250 l/s
Pumping speed He255 l/s
Pumping speed H2220 l/s
Pumping speed Ar250 l/s
Max gas flow rate (N2):250 sccm
Compression ratio N2>1 × 1011
Compression ratio He>1 × 108
Compression ratio H21.5 × 106
Compression ratio Ar> 1 × 1011
Base pressure with recommended forepump1 x 10-10 mbar (1 x 10-10 Torr)
Inlet flangeCFF 8" O.D. / ISO 160 / CFF 6" O.D. / ISO 100
Foreline flangeKF16 NW (KF25 – optional)
Max Rotation Speed60600 rpm (1010 Hz driving frequency)
Start-up time3 minutes (it can be longer in case of soft start active)
Recommended forepumpMechanical pump: DS102, DS302 Dry Pump: IDP-3 (no gas flow), IDP-7, IDP10
Operating positionAny
Max foreline pressure tolerance N212 mbar

According to standard DIN 28 428, the base pressure is that measured in a leak-free test dome, 48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange and using the recommended pre-vacuum pump. Note: foreline Tolerance defined as the pressure at which the turbopump still produces a compression of 100. For continuous operation, water cooling is recommended (water temperature between 15°C and 20°C). Note: value refer to water-cooling pump version with: water temperature between 15°C and 20°C (non condensing, refer to Figure 13); backing pump with pumping speed equal or above 5 m3/h

Содержание

Стр.Описание
213Description of the TwisTorr 305 FS
219Technical Specification
222TwisTorr 305 FS Outline
233Interconnections
233Earth (Ground) Connection
234Pump Operations
234Inlet screen installation
237Air Cooling Kit Installation
239Water Cooling Kit Connection (only for water cooling pumps)
242Vent Accessories
245Pump Used with Corrosive and Oxidizing Gases
247Purge Valve Installation
248Pump Used in Presence of Magnetic Fields Vibration
248Isolator Installation
249Controller Operations
250Accessories and Spare Parts

Technical Specification

ParameterValue
Operating ambient temperature+5 °C to +35 °C
Relative humidity of airFrom graph in Figure 13 (non condensing) **
Bakeout temperatureISO flange: 75°C at inlet flange max CFF flange: 100°C at inlet flange max Note: Measure a point close to the sealing element.
LubricantPermanent lubrication
Air cooling requirementsNatural convection (only with no gas load) Forced air (5- 35 °C ambient temperature)
Coolant waterMinimum flow: 50 l/h (0.22 GPM) Temperature: +15 °C to +30 °C Max pressure: 5 bar (75 psi)
Noise Pressure level41 dB(A) at 1 meter
Installation categoryII
Pollution degree2
Storage temperature-40° C to +70° C
Weight kg (lbs): (*)Pump ISO 100 K Pump CFF 6’’ Pump ISO 160 K Pump CFF 8’’ Note: versions with water cooling kit ISO 100 K 5.84 (12.87) CFF 6" 8.16 (17.98) ISO 160 K 6.28 (13.84) CFF 8" 10.43 (22.99)
Pwr supply :Max input pwr: Pump stand-by average pwr: Pump max oper. pwr: 300 VA 10 W 150 W
Max operating altitude3000 m

Note: mean values based on a significative sample (Ar and N2 compression ratio estimated); standard deviation per test: pumping speed: below ± 7%; noise pressure level: ± 10%” (only pump). The maximum magnetic field strengths allowed for Agilent turbo pumps are: • 100 Gauss (10 mT) in the axial direction • 50 Gauss (5 mT) in the transveral direction

Table

DESCRIPTIONPART NUMBER
Turbopump extension cable (for remote controller only)9699942M005
Turbopump extension cable (for remote controller only)9699942M004
Turbopump extension cable (for remote controller only)9699942M015
Small TMP air cooling kit ext cable (5 m)9699940
Vent Valve Extension cable (5 m)9699941
Vent Valve N.O. 0,5 mm Orifice (for remote controller only)9699844
Vent Valve N.O. 1,2 mm Orifice (for remote controller only)9699845
Vent Valve N.C. 0,5 mm Orifice (for remote controller only)9699847
Vent Valve N.C. 1,2 mm Orifice (for remote controller only)9699846
5 m Turbopump extension cable (for remote controller only)9699949
Mains Cable CHINA plug (3 m long)8121-0723

The communication S/W between the controller and the pump (A-Plus) is available on the Agilent website: https://www.agilent.com/en/products/vacuum-technologies/vacuum-leak-detection-software/a-plus-software The connection is provided through serial communication or via RS232 SERIAL to B/T Adapter (X3514-68003).

RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION

EUROPENORTH AMERICAPACIFIC RIM
Fax:00 39 011 9979 3301 781 860 9252
Fax Free:00 800 345 345 00
Toll Free:00 800 234 234 00800 882 7426, Option 3
vpt-customercare@agilent.comvpl-ra@agilent.comPlease visit our website for individual office information
http://www.agilent.com

Warning Symbols

SymbolDescription
Protective Ground Terminal
Do Not Touch
J1Main connector
CE certification
CSA certification
Restriction of Hazardous Substances certification
RoHS China certification
Waste Electrical and Electronic Equipment

Символы нанесены на прибор; изображения символов в тексте не переданы.

Controller MPNs

Controller MPNsController DescritpionRef.Instruction Manual
X3507-64130Twistorr 305 FS onboard 24V87-901-060-01
X3507-64131Twistorr 305 FS onboard 110/230V87-901-060-01
X3506-64130Twistorr 305 FS AG Remote RS232/48587-901-059-01
X3506-64131Twistorr 305 FS AG Remote Profibus87-901-059-01

The TwisTorr 305 FS models are:

ModelDescription
X3513-64008TWISTORR 305-K FS ISO100K STD AIR COOLING
X3513-64009TWISTORR 305-K FS CFF6" STD AIR COOLING
X3513-64010TWISTORR 305-K FS ISO160K STD AIR COOLING
X3513-64011TWISTORR 305-K FS CFF8" STD AIR COOLING
X3513-64012TWISTORR 305-K FS ISO100K STD WATER COOLING
X3513-64013TWISTORR 305-K FS CFF6" STD WATER COOLING
X3513-64014TWISTORR 305-K FS ISO160K STD WATER COOLING
X3513-64015TWISTORR 305-K FS CFF8" STD WATER COOLING

Reduction in pumping speed (%)

ArN2HeH2
X3500-68000
X3500-6800120%18%7%5%

Health and safety

Substances (please refer to MSDS forms)Trade nameChemical nameChemical SymbolCAS Number
ToxicYESNO
HarmfulYESNO
CorrosiveYESNO
ReactiveYESNO
FlammableYESNO
Explosive (*)YESNO
Radioactive (*)YESNO
Biological (*)YESNO
OxidizingYESNO
SensitizerYESNO
Other dangerous substancesYESNO

* Agilent will not accept delivery of any product that is exposed to radioactive, biological, explosive substances or dioxins, PCB’s without written evidence of decontamination.

Engineering Facts

13 facts · every claim linked to a page
Installation

The pump must be mounted so that the high-vacuum flange connection can withstand a torque of 1000 Nm around its axis.

The mounting hardware and chamber flange must be rated for this torque to prevent the pump from detaching if the rotor seizes.

Source
“with a connection capable of withstanding a torque of 1000 Nm around its axis”
Installation

Only high-resistance class clamps or bolts of grade 8.8 or higher may be used for the high-vacuum flange.

Lower-grade fasteners may fail under the sudden torque generated by a rotor stop, causing the pump to tear off.

Source
“Use only high resistance class clamps (or bolts) ≥ 8.8.”
Installation

Mixing different flange types is not permitted; the pump must be fastened to flanges recommended by Agilent.

Non-recommended flange combinations can lead to the pump twisting or tearing off if the rotor suddenly stops.

Source
“The combination of different flange types is not allowed.”
Installation

For ISO-K flanges, use 4 claw clamps for ISO 100 or ISO 160 and tighten to 22 Nm.

Correct clamp count and torque ensure a secure vacuum seal and mechanical retention.

Source
“Tighten the claw clamps with a torque of 22 Nm.”
Installation

For ConFlat flanges, use 16 screws for CFF 6-inch or 20 screws for CFF 8-inch and tighten to 20 Nm.

Proper screw count and torque are required to compress the copper gasket and hold the flange under rotor-failure torque.

Source
“Tighten the screws with a torque of 20 Nm.”
Operation

The pump must always be soft-started on first use and after 60 days of inactivity.

Soft start prevents excessive wear and stress on the rotor bearings after long idle periods.

Source
“the turbomolecular pumps must be always soft-started after a period of 60 days of inactivity of the pump.”
Operation

The pump can only pump non-corrosive and non-oxidizing gases or gas compounds.

Using the pump with corrosive or oxidizing gases without proper purge flow voids warranty and damages the pump.

Source
“puede bombear cualquier tipo de gas o compuesto de gas no corrosivo y oxidante”
Operation

The maximum rotational speed of the TwisTorr 305 FS is 60,600 rpm.

Knowing the maximum speed helps assess rotor energy and the required flange torque rating.

Source
“max. 60600 rpm”
Operation

Emergency stop by removing power must be used only in an emergency; improper use can damage the pump and void warranty.

Frequent emergency stops can degrade bearings and electronics, so normal stop commands should be used.

Source
“This mode is to be used only in an emergency. The use of this stop mode could lead to faults and/or damages at the pump.”
Safety

The pump must be installed indoors only and must not be exposed to rain, ice, snow, dust, aggressive gases, or explosive environments.

Outdoor or hazardous environments can cause condensation, corrosion, or ignition risk, leading to pump failure or safety incidents.

Source
“Install this device indoors only, and do not use the pump in an environment exposed to atmospheric agents (rain, ice, snow), dust, aggressive gases, or in explosive environments or those with a high fire risk.”
Safety

Operating environment must be kept at +5 °C to +35 °C and 0–90 % relative humidity non-condensing to avoid condensation.

Condensation inside the pump can cause electrical shorts and corrosion; the specified range prevents this.

Source
“temperature: from +5 °C to +35 °C”
Maintenance

The rotor fatigue life is estimated at 405,000 start-stop cycles; Agilent recommends returning the pump for maintenance after 17 years of use.

This gives a preventive maintenance interval to avoid unexpected rotor failure and costly downtime.

Source
“305 FS涡轮分子泵转子的抗疲劳性为从停止到全速再回到停止的405000次循环。”
Maintenance

The shelf life of a turbomolecular pump is 12 months from the shipping date; exceeding it requires return to factory.

Long storage can degrade bearings and lubricants; knowing the shelf life prevents commissioning a degraded pump.

Source
“the shelf life of a turbomolecular pump is 12 months from the shipping date.”