Agilent Technologies TwisTorr 305 FS
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies TwisTorr 305 FS

Agilent Technologies TwisTorr 305 FS — Turbomolecular Vacuum Pumps with Motor power 0.3 kW.

The TwisTorr 305 FS is a hybrid turbomolecular pump combining 8 bladed stages with 3 TwisTorr drag stages, driven by a high-frequency induction motor at up to 60600 rpm. It delivers 250 l/s N2 pumping speed and reaches 1 Torr base pressure with the recommended forepump. This FS modification is the flange-configurable platform, available in ISO 100, ISO 160, CFF 6", and CFF 8" high-vacuum connections for high and ultrahigh vacuum systems.

Motor power
0.3 kW

Direct drop-in replacements

Pfeiffer Vacuum
HiPace 300 with TC 110
43%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Motor power 300 VA
Base Pressure With Recommended Forepump 1 mbar
Pumping Speed Ar 250 l/s
Pumping Speed H2 220 l/s
Pumping Speed He 255 l/s
Pumping Speed N2 250 l/s
Rotational speed 60600 rpm
Run-up time 3 minutes
Max gas flow rate (N2) mfr 250 sccm
Compression ratio N2 mfr 1
Compression ratio He mfr 1
Compression ratio H2 mfr 1.5
Compression ratio Ar mfr 1
Max Rotation Speed mfr 60600 rpm
Operating position mfr Any Any
Max foreline pressure tolerance N2 mfr 12 mbar
Lubricant mfr Permanent lubrication lubrication
Installation category mfr II II
Maximum magnetic field strength axial mfr 100 Gauss
Maximum magnetic field strength transversal mfr 50 Gauss
Max rotation speed mfr 60600 rpm
Max magnetic field axial mfr 100 Gauss
Max magnetic field transversal mfr 50 Gauss
Max gas flow rate (N2): mfr 250 sccm
Vent Valve Extension cable (5 m) mfr 9699941
X3506-64130 mfr 305
X3506-64131 mfr 305 Profibus
X3513-64008 mfr 305 COOLING
X3513-64009 mfr 305 COOLING
X3513-64010 mfr 305 COOLING
X3513-64011 mfr 305 COOLING
X3513-64012 mfr 305 COOLING
X3513-64013 mfr 305 COOLING
X3513-64014 mfr 305 COOLING
X3513-64015 mfr 305 COOLING
shelf life mfr 12 months
ISO-K clamp torque mfr 22 Nm
ConFlat screw torque mfr 20 Nm
Dimensions, Weight & Mechanical Design
Weight Pump CFF 6" mfr 8.16 kg
Weight Pump CFF 8" mfr 10.43 kg
Electrical, Power & Drive Specifications
Pump stand-by average power mfr 10 W
Pump max operating power mfr 150 W
Pwr supply : mfr 300 W
X3507-64130 mfr 305 V
X3507-64131 mfr 305 V
Interfaces, Connections & Control
Inlet flange 8 O.D.
Foreline flange mfr 16 NW
Turbopump extension cable (for remote controller only) mfr 9699942
Vent Valve N.O. 0,5 mm Orifice (for remote controller only) mfr 9699844
Vent Valve N.O. 1,2 mm Orifice (for remote controller only) mfr 9699845
Vent Valve N.C. 0,5 mm Orifice (for remote controller only) mfr 9699847
Vent Valve N.C. 1,2 mm Orifice (for remote controller only) mfr 9699846
Mains Cable CHINA plug (3 m long) mfr 8121
high-vacuum flange torque mfr 1000 Nm
Materials, Media & Process Compatibility
Coolant Water Minimum Flow 50 l/h
Coolant water max pressure mfr 5 bar
Operating Conditions, Environment & Protection
Bakeout Temperature 75 °C
Noise Pressure Level 41 dB
Operating Altitude 3000 m
Operating temperature 5 °C
Storage temperature -40 °C
Relative humidity of air mfr 13
Air cooling requirements mfr Natural convection Forced air air
Pollution degree mfr 2
Small TMP air cooling kit ext cable (5 m) mfr 9699940
maximum ambient pressure mfr 2 pressure
Standards, Safety & General Data
Weight Pump ISO 100 K mfr 5.84 kg
Weight Pump ISO 160 K mfr 6.28 kg

Description

Construction and series positioning

Behavior across gases and operating modes

Installation, connection, and control

Integration & Operating Notes

Mounting
Anchor the pump against 1000 Nm rotor-failure torque

The installation must withstand a torque of 1000 Nm around the pump axis in the event of rotor failure. Use only high-resistance-class clamps or bolts (≥ 8.8) and follow the specified tightening torques; the pump can be mounted in any position, but the fixing is not optional.

User Manual, Installation, p.200-201
Process Limits
Keep foreline below 12 mbar N2 for rated compression

The maximum foreline pressure tolerance for N2 is 12 mbar, defined as the pressure at which the turbopump still produces a compression of 100. Size and operate the forepump so the foreline stays below this limit under the actual gas load, otherwise pumping performance degrades.

User Manual, Technical Specification, p.219
Control & I/O
Force soft start on first run and after 60 days idle

Soft start mode is mandatory at the first ramp-up and after 60 days or more of inactivity to ensure proper bearing lubrication. If a new pump is paired with a used controller, set soft start manually and wait for full speed before stopping the pump.

User Manual, Soft Start, p.208
Electrical
Provide a separate earth conductor below 0.1 Ω

Install a dedicated earth conductor using the M5 x 10 screw and shake-proof washer tightened to 2 Nm. The impedance between the pump body and the earth point must be less than 0.1 Ω; this is a safety requirement, not a performance tuning step.

User Manual, Earth (Ground) Connection, p.233
Cooling & Media
Water cooling needs clean, low-conductivity coolant

Water cooling kits require flow >50 l/h, water temperature +15 to +30 °C, maximum pressure 5 bar, and conductivity ≤500 µs/cm. In closed circuits with higher conductivity, use up to 20% ethyl glycol; tighten the kit into the pump body to 5 Nm.

User Manual, Water Cooling Kit Connection, p.240-241
Process Limits
Purge bearings when pumping corrosive or oxidizing gases

For corrosive and oxidizing gases, inert purge gas must flow around the upper bearing to the forevacuum line at 0.1 to 0.8 mbar l/s, with a minimum of 10 sccm (0.17 mbar l/s). Install the purge valve with 2.5 Nm torque after switching off the pump.

User Manual, Pump Used with Corrosive and Oxidizing Gases, p.245
Maintenance
Track ramp cycles and return the pump at 405000 cycles

Rotor fatigue resistance is 405000 cycles of ramp up from stop to full speed and back to stop; Agilent also recommends return after 17 years of use. Log cycle counts and calendar age to plan maintenance or replacement, since the pump itself requires no routine maintenance.

User Manual, Rotor life, p.207
Environment
Respect magnetic field and ambient temperature limits

Maximum allowed magnetic field is 100 Gauss (10 mT) axial and 50 Gauss (5 mT) transversal. Operating ambient temperature is +5 °C to +35 °C; if the pump is stored below 5 °C, wait until it reaches operating temperature before switching on.

User Manual, Technical Specification, p.220-221

Selection & Integration Q&A

When should I choose the TwisTorr 305 FS rather than another flange variant of the same series?

All eight models share the same rotor and performance; the choice is the high-vacuum flange. Pick ISO 100 or ISO 160 for elastomer-sealed ISO-K systems with claw clamps at 22 Nm; pick CFF 6" or CFF 8" for metal-sealed ConFlat systems needing 16 or 20 screws at 20 Nm and bakeout to 100 °C instead of 75 °C.

What forepump and controller arrangement does this pump require?

Recommended forepumps are DS102, DS302, IDP-3 (no gas flow), IDP-7, and IDP10, connected to the NW 16 foreline flange. Controllers are onboard 24 V (X3507-64130), onboard 110/230 V (X3507-64131), or AG Remote RS232/485 (X3506-64130) and Profibus (X3506-64131). Max input power is 300 VA.

Can the pump be mounted in any orientation, and what environment limits apply?

Yes, any operating position is allowed, but the fixing must withstand 1000 Nm around the axis in case of rotor failure. Keep ambient at +5 to +35 °C, magnetic field below 100 Gauss axial / 50 Gauss transversal, and observe bakeout limits of 75 °C (ISO) or 100 °C (CFF) at the inlet flange.

What maintenance and accessories are needed beyond the base pump delivery?

The pump needs no routine maintenance, but plan return at 405000 ramp cycles or 17 years. Accessories ordered separately include air cooling kits X3500-68010/X3500-68011, water cooling kits 9699337/9699347, vent valves (e.g. 9699834, 9699844), purge valves, inlet screens, vibration isolators (9699344/9699345/9699334/9699335), and the KF25 foreline flange.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
TwisTorr 305 FS Turbomolecular Pump User Manual
Manual ·258 pp.
View PDF