Pump power rating differs by model, so the electrical supply and heater circuit must be sized per model.
Determines heater power supply, wiring and breaker sizing at selection.
Source
“VHS-4: 1450 Watts”
| Specification | VHS-4 | VHS-6 | VHS-250 |
|---|---|---|---|
| Power Rating | 1450 Watts | 2200 Watts | 2200 Watts |
| Optimum Operating Range (Torr) | 1x10−3 to < 5x10−9 (1.3 x 10-3 to < 6.6 x 10-9 mbars) | 1x10−3 to < 5x10−9 (1.3 x 10-3 to < 6.6 x 10-9 mbars) | 7x10−4 to < 5x10−9 (9 x 10-4 to < 6 x 10-9 mbars) |
| Maximum Pumping Speed (l/s) | Air – 1200 with std cold cap; 950 with extended cold cap Helium –1500 with std cold cap; 1200 with extended cold cap | Air – 2400 with std cold cap; 1600 with extended cold cap Helium –3000 with std cold cap; 2000 with extended cold cap | Air – 3700 with std cold cap; Helium –4600 |
| Maximum Throughput (Torr-l/s) | 1.2 (1.6 mbar-l/s) in operating range 2.5 (3.2 mbar-l/s) @0.01 Torr | 2.4 (3.2 mbar-l/s) in operating range 3.5 (4.5 mbar-l/s) @0.01 Torr | 2.6 (3.5 mbar-l/s) in operating range 3.5 (4.5 mbar-l/s) @0.01 Torr |
| Maximum Forepressure | No Load – 0.65 Torr (.87 mbars) Full Load – 0.55 Torr (.73 mbars) | No Load – 0.65 Torr (.87 mbars) Full Load – 0.55 Torr (.73 mbars) | No Load – 0.65 Torr (.87 mbars) Full Load – 0.55 Torr (.73 mbars) |
| Symptom | Probable Cause | Repair Action |
|---|---|---|
| Poor system pressure | Leak in system — virtual or real. | Locate and repair. |
| Poor system pressure | High process gas load. | Measure gas load, and eliminate the cause. |
| Poor system pressure | System dirty. | Clean system to reduce outgassing. |
| Poor ultimate pressure | Contaminated pump fluid. | Examine, clean pump, and replace fluid. |
| Poor ultimate pressure | Low heat input. | Check voltage. Check for continuity, burned out element, poor thermal contact. |
| Poor ultimate pressure | Inadequate cooling water flow. | Check water pressure. Check tubing for obstructions or backpressure. |
| Poor ultimate pressure | Excessive or too cold cooling water. | Check temperature, adjust flow. |
| Poor ultimate pressure | High forepressure. | Check for leak in foreline, poor mechanical pump performance, breakdown of mechanical pump fluid. |
| Poor ultimate pressure | Water in Quick Cool Coil. | Check and remove water. |
| Slow pumpdown (prolonged cycle after checking gas load conditions) | Low heat input. | Check heaters. |
| Slow pumpdown (prolonged cycle after checking gas load conditions) | Low oil level. | Add oil. |
| Slow pumpdown (prolonged cycle after checking gas load conditions) | Malfunctioning pump assembly Improperly located jets. Damaged jet system. | Check and repair or replace. |
| Parameter | ASA in | ASA mm | ISO in | ISO mm | Conflat in | Conflat mm |
|---|---|---|---|---|---|---|
| Height | 18.05 | 458 | 18.05 | 458 | 18.05 | 458 |
| Centerline to Centerline | 10.62 | 270 | 10.62 | 270 | 10.62 | 270 |
| Flange Face to Flange Face | 5.03 | 128 | 5.03 | 128 | 5.03 | 128 |
| EIectrlc8I Box Height | 8.63 | 219 | 8.63 | 219 | 8.63 | 219 |
| Sight Glass, Degrees from Foreline | 90° clockwise | 90° clockwise | 90° clockwise | |||
| Electrical Box, Degrees from Foreline | 30° counterclockwise | 30° counterclockwise | 30° counterclockwise | |||
| Inlet Flange, Nominal Size | 4” ASA | ISO -160K | 8” CFF | |||
| Inlet Flange, OD | 9.00 | 229 | 7.09 | 180 | 7.96 | 202 |
| Inlet Flange, ID | 5.91 | 150 | 5.78 | 147 | 5.91 | 150 |
| Inlet Flange, Thickness | 0.50 | 13 | 0.47 | 12 | 0.93 | 24 |
| Inlet Flange, Bolt Circle | 7.50 | 191 | N/A | N/A | 7.12 | 181 |
| Inlet Flange, Number of Holes | 8 | N/A | 20 | |||
| Inlet Range, Hole Size | 0.69 | 18 | N/A | N/A | .33 | 8 |
| O-ring Groove ID | 6.06 | 154 | N/A | N/A | N/A | N/A |
| O-ring Groove Width | 0.18 | 4 | N/A | N/A | N/A | N/A |
| Foreline Flange, Nominal Size | KF40 | KF-40 | 2¾ CFF | |||
| OD | 2.16 | 55 | 2.16 | 55 | 2.73 | 69 |
| ID | 1.38 | 35 | 1.38 | 35 | 1.38 | 35 |
| Thickness | 0.20 | 5 | 0.20 | 5 | 1.38 | 35 |
| Bolt Circle | N/A | N/A | N/A | N/A | 2.31 | 59 |
| Number of Holes | N/A | N/A | 6 | |||
| Hole Size | N/A | N/A | N/A | N/A | 0.27 | 7 |
| O-ring Groove ID | N/A | N/A | N/A | N/A | N/A | N/A |
| O-ring Groove Width | N/A | N/A | N/A | N/A | N/A | N/A |
| ASA in | ASA mm | ISO in | ISO mm | Conflat in | Conflat mm | |
|---|---|---|---|---|---|---|
| Height | 21.92 | 557 | 21.92 | 557 | 21.92 | 557 |
| Centerline to Centerline | 13.38 | 340 | 13.38 | 340 | 13.38 | 340 |
| Flange Face to Flange Face | 8.62 | 219 | 8.62 | 219 | 8.62 | 219 |
| EIectrlc8I Box Height | 9.12 | 232 | 9.12 | 232 | 9.12 | 232 |
| Sight Glass, Degrees from Foreline | 90° clockwise | 90° clockwise | 90° clockwise | |||
| Electrical Box, Degrees from Foreline | 30° counterclockwise | 30° counterclockwise | 30° counterclockwise | |||
| Inlet Flange, Nominal Size | 6” ASA | ISO -200K | 10” CFF | |||
| Inlet Flange, OD | 11.00 | 279 | 9.45 | 240 | 9.97 | 253 |
| Inlet Flange, ID | 7.88 | 200 | 7.72 | 196 | 7.88 | 200 |
| Inlet Flange, Thickness | 0.75 | 19 | 0.47 | 12 | 0.96 | 24 |
| Inlet Flange, Bolt Circle | 9.50 | 241 | N/A | N/A | 9.12 | 232 |
| Inlet Flange, Number of Holes | 8 | N/A | 24 | |||
| Inlet Range, Hole Size | 0.81 | 21 | N/A | N/A | .33 | 8 |
| O-ring Groove ID | 8.20 | 208 | N/A | N/A | N/A | N/A |
| O-ring Groove Width | 0.18 | 4 | N/A | N/A | N/A | N/A |
| Foreline Flange, Nominal Size | 1½" ASA | KF-50 | 33/8" CFF | |||
| OD | 5.00 | 127 | 2.05 | 75 | 3.38 | 86 |
| ID | 1.95 | 50 | 1.95 | 50 | 1.95 | 50 |
| Thickness | 0.63 | 16 | 0.24 | 6 | 0.62 | 16 |
| Bolt Circle | 3.88 | 99 | N/A | N/A | 2.85 | 72 |
| Number of Holes | 4 | N/A | N/A | 8 | ||
| Hole Size | 0.62 | 16 | N/A | N/A | 0.33 | 8 |
| O-ring Groove ID | 2.22 | 56 | N/A | N/A | N/A | N/A |
| O-ring Groove Width | 0.30 | 8 | N/A | N/A | N/A | N/A |
Значения приведены дословно, включая опечатки оригинала (EIectrlc8I, Inlet Range).
| ISO | in | mm | |
|---|---|---|---|
| Height | 22.79 | 579 | |
| Centerline to Centerline | 13.38 | 340 | |
| Flange Face to Flange Face | 9.56 | 243 | |
| Electrical Box Height | 9.12 | 232 | |
| Sight Glass, Degrees from Foreline | 90° clockwise | ||
| Electrical Box, Degrees from Foreline | 30° counterclockwise | ||
| Inlet Flange, Nominal Size | ISO - 250F | ||
| Inlet Flange, OD | 13.19 | 335 | |
| Inlet Flange, ID | 10.75 | 273 | |
| Inlet Flange, Thickness | 0.63 | 16 | |
| Inlet Flange, Bolt Circle | 12.205 | 310 | |
| Inlet Flange, Number of Holes | 12 | ||
| Inlet Range, Hole Size | 0.44 | 11 | |
| O-ring Groove ID | 10.95 | 278 | |
| O-ring Groove Width | 0.14 | 4 | |
| Foreline Flange, Nominal Size | 1½ ASA | ||
| OD | 5.00 | 127 | |
| ID | 1.88 | 48 | |
| Thickness | 0.62 | 16 | |
| Bolt Circle | 3.88 | 99 | |
| Number of Holes | 4 | ||
| Hole Size | 0.62 | 16 | |
| O-ring Groove ID | 2.22 | 56 | |
| O-ring Groove Width | 0.30 | 8 |
| Pump | Voltage | Part Number | Flange Type |
|---|---|---|---|
| Standard cold cap | 120 V | 86460301 | ASA |
| Standard cold cap | 120 V | L6256301 | ISO |
| Standard cold cap | 120 V | L6188301 | Conflat |
| Extended cold cap | 120 V | 86460311 | ASA |
| Extended cold cap | 120 V | L6256311 | ISO |
| Extended cold cap | 120 V | L6188311 | Conflat |
| Standard cold cap | 208 V | 86460306 | ASA |
| Standard cold cap | 208 V | L6256306 | ISO |
| Standard cold cap | 208 V | L6188306 | Conflat |
| Extended cold cap | 208 V | 86460316 | ASA |
| Extended cold cap | 208 V | L6256316 | ISO |
| Extended cold cap | 208 V | L6188316 | Conflat |
| Standard cold cap | 240 V | 86460302 | ASA |
| Standard cold cap | 240 V | L6256302 | ISO |
| Standard cold cap | 240 V | L6188302 | Conflat |
| Extended cold cap | 240 V | 86460312 | ASA |
| Extended cold cap | 240 V | L6256312 | ISO |
| Extended cold cap | 240 V | L6188312 | Conflat |
| Pump | Voltage | Part Number | Flange Type |
|---|---|---|---|
| Standard cold cap | 120 V | K0543301 | ISO |
| Standard cold cap | 208 V | K0543306 | ISO |
| Standard cold cap | 240 V | K0543302 | ISO |
| Specification | Value |
|---|---|
| Backstreaming Rate at Pump Inlet | 5x10−4 mg/cm2/minute (with std cold cap) |
| Electrical Requirements | 120, 208, 240; 50/60 Hz; single phase |
| Warm-up Time | 10 minutes |
| Cooldown Time (using Quick Cool Coil) | 10 minutes |
| Fluid Charge | VHS-4: 300 cc; VHS-6 and VHS-250: 500 cc |
| Cooling Water Requirements | Maximum inlet temperature – 60/80 °F (15 - 26 °C); Maximum outlet temperature at foreline – 120 °F (48.9 °C); Flow rate – VHS-4: 0.15 gpm; VHS-6 and VHS-250: 0.25 gpm |
| Pressure drop across coils | VHS-4: 4 psi; VHS-6 and VHS-250: 15 psi |
| Backing Pump Size | VHS-4: ≥ 10 cfm for maximum throughput; VHS-6 and VHS-250: 17 cfm for maximum throughput |
| Recommended Jet Assembly | 4-stage, self-aligning, stainless steel |
| Foreline Baffle | Stacked half moon |
| Cold Cap | Nickel-plated copper |
| Water Connections | 1/8" FPT |
| Thermal Switches | Manual reset at 300 °F (149 °C) |
| Environmental | Maximum ambient temperature 113 °F (45 °C) |
| Installation | Indoor use, Installation Category 2, Pollution Degree 2 |
| Altitude | 6562’ (2000 m) |
| Specification | VHS-4 | VHS-6 and VHS-250 |
|---|---|---|
| Heater circuit resistance (ohms) | ||
| Heater circuit resistance (ohms) — 120 V | 9.4 | 6.2 |
| Heater circuit resistance (ohms) — 208 V | 28.3 | 18.7 |
| Heater circuit resistance (ohms) — 240 V | 37.7 | 24.9 |
| Heater power | 1450 watts | 2200 watts |
| Clearance (for heater removal) | 6" minimum | 6" minimum |
| Materials — Body, Flanges, Foreline, Baffle | stainless steel | stainless steel |
| Materials — Jet Assembly | stainless steel | stainless steel |
| Materials — Body Cooling Coils | copper | copper |
| Materials — Quick Cooling Coil | stainless steel | stainless steel |
| Materials — Cold Cap | nickel-plated copper | nickel-plated copper |
| Shipping Weight | 55 pounds | 75 pounds |
Pump power rating differs by model, so the electrical supply and heater circuit must be sized per model.
Determines heater power supply, wiring and breaker sizing at selection.
“VHS-4: 1450 Watts”
A minimum clearance is required below the pump to allow heater removal.
Affects mechanical layout and serviceability of the heater.
“Clearance (for heater removal)”
The pump body must be mounted vertical and plumb, and the inlet mating flange horizontal within ±1°.
Prevents uneven fluid level and boiler plate overheating.
“Ensure that the pump inlet mating flange on the system is horizontal within ±1°.”
Cooling water must be supplied continuously at model-specific flow and 60 to 80 °F.
Defines cooling water utility requirements and prevents overheating.
“Connect the inlet water fitting (near the inlet flange at the top of the pump) to a continuously running water supply at 0.15 gpm (VHS-4) or .25 gpm (VHS-6 and 250) and at a temperature of 60 to 80 °F.”
The thermal switch must be wired in series with the heater to cut power on overtemperature.
Critical interlock wiring to protect personnel and pump from overtemperature.
“Connect the thermal switch in series with the heater.”
The pump must be rough-pumped below 0.5 Torr before the heater is energized.
Operating above this pressure risks explosion and pump damage.
“The pressure must be reduced to less than 0.5 Torr (.66 mbars).”
In humid environments heaters must be started at reduced voltage to expel absorbed moisture.
Prevents heater damage on startup in humid conditions.
“start pumps at a low line voltage,e.g., 50% of line voltage, for a period or time to slowly expel any moisture.”
Cooling water must keep flowing until the body cools to about 130 °F before venting.
Prevents fluid degradation and unsafe venting of a hot pump.
“Allow cooling water to flow through the diffusion pump until the pump body temperature, located just above the boiler plate, has cooled to a temperature of approximately 130 °F.”
New O-rings must be used whenever fill plugs or the sight glass are replaced.
Prevents leaks at fill/drain and sight glass during servicing.
“Always use new O-rings when replacing fill plugs or the sight glass.”
The heater clamp bolt must be tightened to a specific torque during heater replacement.
Ensures proper thermal contact and prevents heater damage.
“Tighten the heater clamp bolt to a torque of 250 inch-pounds.”
The boiler stud should be coated with anti-seize compound before heater installation.
Prevents seizing and eases future heater servicing.
“Coat the boiler stud with an anti-seize compound such as FEL-PRO C5A or common milk of magnesia.”
Fluid charge is model-specific and must be replenished to the FULL/COLD mark.
Correct fluid charge is required for proper pumping and thermal switch protection.
“300 cc for the VHS-4”
If gas load exceeds maximum throughput, the diffusion pump stops functioning and only the backing pump acts.
Guides throughput sizing and diagnosis of poor performance.
“If the gas load (Q) exceeds the maximum throughput capability of the diffusion pump, the diffusion pump will not function and the pumping action will essentially be performed by the mechanical backing pump.”
Hydrocarbon-based pumping oils must not be used due to explosion risk.
Fluid selection directly affects explosion safety of the system.
“Do not use hydrocarbon-based pumping oils.”