Agilent Technologies VHS-250 Diffusion Pump
Agilent Technologies VHS-250 Diffusion Pump — Vacuum technologies with Motor power 1.45 kW, Pumping speed 4320 m³/h.
The VHS-250 is a 4-stage, self-aligning stainless-steel jet diffusion pump with a standard cold cap, rated for 3700 L/s air and 4600 L/s helium. It is the largest of the VHS-4/VHS-6/VHS-250 series, requiring a 17 cfm backing pump and 0.25 gpm cooling water at 15 psi. It serves high-gas-load vacuum processes where a 10-minute warm-up and 0.65 Torr maximum forepressure are acceptable.
Agilent Technologies Small Diffusion Pumps Series — Model Comparison & Selection
| Model | Motor power, kW | Pumping speed, m³/h | Cooldown time, minutes | Cooling water pressure, psi | Diffusion pump pumping speed, L/s | Warm up time, minutes |
|---|---|---|---|---|---|---|
| VHS-250 Diffusion Pump Current | 1.45 | 4320 | 10 | 15 | 50 | 10 |
| VHS-4 Diffusion Pump | 1.45 | 4320 | 10 | 15 | 50 | 10 |
| VHS-6 Diffusion Pump | 1.45 | 4320 | 10 | 15 | 50 | 10 |
Full specifications
| Motor power | 1450 Watts |
| Pumping speed | 1200 l/s |
| Cooldown Time | 10 minutes |
| Cooldown Time (using Quick Cool Coil) | 10 minutes |
| Diffusion Pump Pumping Speed | 50 L/s |
| Extended Cold Cap | 120 V |
| Heater Power | 1450 watts |
| Warm-up Time | 10 minutes |
| Vacuum gauge measurement range mfr | 6 pressure |
| Dry scroll pump typical size mfr | 3...30 h |
| Turbomolecular pump pumping speed mfr | 70...3 L/s |
| Optimum Operating Range mfr | 9 mbars |
| Maximum Forepressure mfr | 0.65 Torr |
| Backstreaming Rate at Pump Inlet mfr | 5 minute |
| Backing Pump Size mfr | -4 throughput |
| Recommended Jet Assembly mfr | 4 steel |
| Maximum Throughput mfr | 1.2 l/s |
| Optimum Operating Range (Torr) mfr | 1 |
| Maximum Throughput (Torr-l/s) mfr | 1.2 Torr |
| Poor system pressure mfr | System dirty. dirty. |
| Poor ultimate pressure mfr | Water in Quick Cool Coil. Coil. |
| Slow pumpdown (prolonged cycle after checking gas load conditions) mfr | Malfunctioning pump assembly Improperly located |
| Electrical Requirements mfr | 120 phase |
| Pressure drop across coils mfr | -4 psi |
| Foreline Baffle mfr | Stacked half moon moon |
| Water Connections mfr | 1 FPT |
| Installation mfr | 2 |
| Heater circuit resistance (ohms) — 120 V mfr | 9.4 |
| Heater circuit resistance (ohms) — 208 V mfr | 28.3 |
| Heater circuit resistance (ohms) — 240 V mfr | 37.7 |
| Clearance (for heater removal) mfr | 6 minimum |
| Materials — Body, Flanges, Foreline, Baffle mfr | stainless steel steel |
| Materials — Jet Assembly mfr | stainless steel steel |
| Materials — Cold Cap mfr | nickel-plated copper copper |
| Rough pump pressure before startup mfr | 0.5 Torr |
| Heater clamp bolt torque mfr | 250 pounds |
| Shipping Weight | 55 pounds |
| Turbomolecular pump controller protection class mfr | 54 |
| Inlet Flange, Nominal Size mfr | 250 F |
| Foreline Flange, Nominal Size mfr | 1 ASA |
| Fluid Charge mfr | -4 cc |
| VHS-4 oil charge mfr | 300 cc |
| VHS-6 and VHS-250 oil charge mfr | 500 cc |
| VHS-4 fluid charge mfr | 300 cc |
| VHS-6 and VHS-250 fluid charge mfr | 500 cc |
| Cooling Water Pressure | 15 psi |
| Cooling Water Requirements mfr | 60 °F |
| Cooling Water Flow mfr | 0.25 gpm |
| Thermal Switches mfr | 300 °F |
| Environmental mfr | 113 °F |
| Altitude mfr | 6562 |
| Materials — Body Cooling Coils mfr | copper copper |
| Materials — Quick Cooling Coil mfr | stainless steel steel |
| VHS-4 cooling water flow mfr | 0.15 gpm |
| VHS-6 and VHS-250 cooling water flow mfr | 25 gpm |
| Standard cold cap mfr | 240 |
Description
Construction and series positioning
- The VHS-250 shares the VHS series architecture: a 4-stage, self-aligning stainless-steel jet assembly and a standard cold cap for backstreaming suppression.
- Within the series, the VHS-250 is the high-throughput member: its air pumping speed is 3700 L/s with the standard cold cap, versus 1200 L/s for the VHS-4. The VHS-10 is larger still at 5300 L/s.
- The VHS-250 uses a 17 cfm backing pump, while the VHS-4 uses 10 cfm; cooling water flow is 0.25 gpm at 15 psi for both the VHS-6 and VHS-250.
Operating behavior and limits
- Maximum forepressure (no load) is 0.65 Torr (0.87 mbar); the backing pump must keep the foreline below this value.
- Warm-up and cooldown are each 10 minutes; thermal switches are manual-reset at 300 °F (149 °C).
- Prolonged operation at high throughput above 10⁻³ Torr (1.33×10⁻³ mbar) can cause high chamber contamination and should be avoided.
- Cooling water inlet temperature must be 60–80 °F (15–26 °C); maximum ambient temperature is 113 °F (45 °C).
Installation and operation specifics
- Leak-check inlet and foreline connections, drain and fill plugs, compression fittings, and threaded connections first if leakage is suspected.
- Gas load can be estimated as Q = V × ΔP/Δt after isolating the system and measuring pressure rise.
- The pump will not start if safety circuits or protective devices prevent the contactor from staying closed; check utilities, flow switches, interlocks, and thermostat operation.
Integration & Operating Notes
The maximum forepressure (no load) is 0.65 Torr (0.87 mbar). If the backing pump cannot hold the foreline below this value, the jet assembly cannot maintain its compression ratio and the pump will stall or backstream. Size the backing line and pump accordingly.
Cooling water flow for the VHS-250 is 0.25 gpm at 15 psi, with a maximum inlet temperature of 60/80 °F (15–26 °C). Insufficient flow or excessive inlet temperature will overheat the pump and may trip the 300 °F thermal switches, forcing a manual reset after cooldown.
The pump will not start if safety circuits and/or protective devices prevent the contactor from staying closed. Verify that flow switches, interlocks, and thermostat contacts are closed and that all utilities are present before troubleshooting the heater circuit.
Thermal switches are manual reset at 300 °F (149 °C). After a trip, allow the pump to cool, correct the cause (usually cooling water flow or temperature), and manually reset the switch; automatic restart is not provided.
Prolonged operation at high throughput at pressure above 10⁻³ Torr (1.33×10⁻³ mbar) can cause high chamber contamination. Use the pump within its optimum operating range and avoid sustained high-pressure operation to protect process cleanliness.
Selection & Integration Q&A
Choose the VHS-250 for the highest gas load in the VHS series: 3700 L/s air and 4600 L/s helium with standard cold cap, versus 1200 L/s for the VHS-4. It requires a 17 cfm backing pump and 0.25 gpm cooling water; the VHS-4 needs only 10 cfm. Select the VHS-4 for smaller chambers and lower throughput.
Use a backing pump of at least 17 cfm for maximum throughput and ensure the foreline pressure stays below the 0.65 Torr maximum forepressure (no load). The backing line must be sized to handle the gas load without exceeding this limit, otherwise pumping action is lost.
The manual covers cold cap and jet assembly removal/installation, heater replacement, and pump fluid charge; these are service actions, not included accessories. Leak checks should target inlet and foreline connections, drain and fill plugs, compression fittings, and threaded connections. Order replacement parts per the manual's parts list; cables and controllers are not listed as included.