This operator's guide covers the Waters Positive Pressure-96 Processor, a laboratory device for positive-pressure solid-phase extraction (SPE) and filtration in 96-well plates and 1 cc flangeless cartridges. It applies to the base processor (part number 186006961) and associated accessories and spare parts listed in the spare parts table, including the 96-place sealing gasket (186005522), µElution positive pressure spacer (405006528), gas supply adapter (186005524), and various 96-well plates and collection plates.

Section-by-Section Technical Breakdown

pp. 1-2
System Overview

The Waters Positive Pressure-96 Processor provides positive pressure operation for SPE and filtration in 96-well formats. Each of the 96 manifold holes is restricted to maintain constant pressure even if not all wells are filled. The system offers reproducible analyte recoveries, improved flow for viscous samples, and compatibility with most 96-well plate formats.

  • Positive pressure processing with 96 independent restricted ports for uniform extraction.
  • Gas pressure range 60 to 100 psi, optimal 80 psi.
  • Self-adjusting to plate height for most 96-well plate formats.
  • Two toggle switches for compression mechanism with safety feature requiring simultaneous pressing.
  • SPE flow control via rotometer and Max Flow regulator.
pp. 2-3
Installation

The processor must be installed indoors in a hood vent whenever possible, with an inert gas supply of 60 to 100 psig. Site requirements include maximum altitude 2000 m, ambient temperature 5 to 40 °C, and humidity limits. The installation kit includes polyethylene tubing and connectors for 1/8-inch or 1/4-inch NPT fittings.

  • Indoor use only; maximum altitude 2000 m.
  • Ambient temperature range 5 to 40 °C.
  • Maximum relative humidity 80% for temperatures up to 31 °C, decreasing linearly to 50% at 40 °C.
  • Installation Category II, Pollution Degree 2.
  • Recommended gas source: compressed nitrogen, oil-free.
  • Installation kit includes 6 feet of 1/8-inch i.d. (1/4-inch o.d.) polyethylene tubing and connectors to 1/8-inch NPT or 1/4-inch NPT.
pp. 3-4
Using the Waters Positive Pressure-96 Processor

This section describes configuring plates, plate compression, operating the manifold, controlling positive pressure with flow restrictors, and adjusting extraction flow. The processor accommodates integrated and modular plates, µElution plates with spacer, and 1 cc flangeless cartridges. Compression is activated by two toggle switches pressed simultaneously; the mechanism automatically opens if released too soon.

  • Plate assembly sits on slide tray platform and slides under manifold until stop.
  • Both toggle switches must be pressed simultaneously to open or close compression mechanism.
  • Compression speed and pressure are pre-set and not adjustable.
  • Manifold holes restricted to maintain pressure; flow through a hole at 25 psig is approximately 0.75 ±0.05 scfh.
  • Two-level gas delivery: low flow via rotometer (0 to 2.5 SCFH) and max flow via regulator (0 to 80 psig).
  • Selector valve must be in off position during compression/decompression and when not in use.
pp. 4-5
Maintenance

Maintenance includes replacing and cleaning the 96-place sealing gasket, storage requirements, and cleaning the device exterior. The gasket must be stored at 4 °C and has a 6-month shelf life from shipping date. Cleaning the gasket uses methanol; exterior cleaning uses water applied to a cloth only.

  • Replace 96-place sealing gasket using 3/16-inch hex key and two 1/4-20 screws.
  • Clean silicone rubber sealing gasket with soft cloth dampened with methanol.
  • Store gasket at 4 °C; shelf life 6 months from shipping date.
  • Clean exterior with soft cloth dampened with water; never spray water directly on device.
  • Device constructed of anodized aluminum, stainless steel, and plastics; clean solvent spills immediately.
pp. 8
Spare Parts

The spare parts table lists part numbers for the processor, plates, gasket, spacer, gas supply adapter, and waste collection plate. These are the authorized replacement and consumable items.

  • Waters Positive Pressure-96 Processor: 186006961.
  • 96-well plate, polypropylene, 2 mL: 186002482 (50/pk).
  • 96-well 1 mL collection plate: 186002481 (50/pk).
  • Polypropylene cap mat for 1 mL plate: 186002483 (50/pk).
  • 96-flangeless cartridges holder, 1 cc: 186005523.
  • 96-place sealing gasket: 186005522 (store at 4 °C, 6-month shelf life).
  • µElution positive pressure spacer: 405006528.
  • Gas supply adapter, includes 1/8-inch to 1/4-inch NPT fitting, 6 feet of 1/4-inch tubing: 186005524.
  • 10 mL 24 waste collection plate: 186005586.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Optimal gas source pressure is 80 psig; minimum 60 psig; maximum 100 psig. p.3 ↗ Set the regulated gas supply within this range for proper operation; 80 psig is recommended for best performance. Whole series
Flow through a manifold hole at 25 psig is approximately 0.75 ±0.05 scfh when no cartridges or empty wells are present. p.3 ↗ This restricted flow maintains manifold pressure and ensures uniform processing even with empty positions. Whole series
Low flow rotometer adjustable range: 0 to 2.5 SCFH. p.4 ↗ Use the rotometer needle valve for precise low-flow control during SPE; do not use it as an on/off valve. Whole series
Max Flow regulator pressure range: 0 to 80 psig. p.4 ↗ Use Max Flow setting for viscous samples or drying wells/cartridges; adjust regulator to achieve desired flow. Whole series
Compression mechanism requires both toggle switches pressed simultaneously for 2 to 3 seconds until manifold stops moving. p.3 ↗ If released too soon, the manifold automatically opens as a safety feature; ensure full compression for proper sealing. Whole series
96-place sealing gasket must be stored at 4 °C and has a shelf life of 6 months from shipping date. p.5 ↗ Observe storage conditions and track shelf life to ensure proper sealing and prevent leaks. Whole series
Site requirements: maximum altitude 2000 m; ambient temperature 5 to 40 °C; maximum relative humidity 80% for temperatures up to 31 °C, decreasing linearly to 50% at 40 °C. p.2 ↗ Install in an environment meeting these conditions to ensure safe and reliable operation. Whole series
Installation kit includes 6 feet of 1/8-inch i.d. (1/4-inch o.d.) polyethylene tubing and connectors to 1/8-inch NPT or 1/4-inch NPT. p.2 ↗ Use provided tubing and fittings to connect to a regulated oil-free gas source; ensure compatibility with gas supply. Whole series
The processor does not include a filter on the source gas input. p.3 ↗ A clean, oil-free gas source is essential to prevent sample contamination; consider adding an external filter if necessary. Whole series
Compression speed and pressure are pre-set and not adjustable. p.3 ↗ Do not attempt to adjust compression parameters; rely on factory settings for proper sealing. Whole series
Selector valve must be in off position during compression and decompression cycles and when not in use. p.4 ↗ Prevent unintended gas flow and ensure safety during plate loading/unloading. Whole series
Replacement of 96-place sealing gasket requires a 3/16-inch hex key and two 1/4-20 screws. p.4 ↗ Keep these tools available for maintenance; follow procedure to align gasket using screws before adhering. Whole series
Clean the silicone rubber sealing gasket with a soft cloth dampened with methanol. p.5 ↗ Use methanol only; avoid other solvents that may damage the gasket. Whole series
Clean the device exterior with a soft cloth dampened with water; never spray or apply water directly to any surface. p.5 ↗ Prevent liquid ingress and damage to internal components. Whole series
The device is constructed of anodized aluminum, stainless steel, and plastics. p.4 ↗ Clean solvent spills and overflow immediately to prevent damage to materials. Whole series
The Waters Positive Pressure-96 Processor is for laboratory use only and is not intended for diagnostic applications. p.9 ↗ Use only in research settings; not for clinical diagnostics. Whole series