Waters Xevo MRT Product Range Site Preparation Guide

Technical Reference Summary

Covered Models & Document Scope

Covers the Waters Xevo MRT and Xevo MRT P10 mass spectrometers, including the external embedded PC (EPC), vacuum pump variants (Xevo MRT and Xevo MRT P10), data system, optional ionBench, optional WRDS, optional DESI XS source, and optional nitrogen generator. Document is the Site Preparation Guide (715008881 VER. 09).

Section-by-Section Technical Breakdown

pp. 6-8
Power Requirements

Defines AC mains tolerances, power cord compatibility, socket requirements, and electrical safety for the system. Includes UPS recommendation and component electrical specifications.

  • AC main voltage fluctuations must not exceed ±10% of nominal device voltage.
  • Power cords are shipped as ordered and must only be used with the Waters system.
  • Vacuum pump and mass spectrometer power cords are 2 m long.
  • Vacuum pump is normally in continuous operation; supply should not be inadvertently switched off.
  • AC supply must be wired with protective ground and fused or fitted with a circuit breaker.
  • RCDs recommended; GFCIs may be required by local codes.
pp. 7
Uninterruptible Power Supply (UPS)

Recommends a UPS to prevent local power instability from impacting reliability and performance. Waters supplies UPS units configured for Waters MS systems that step up single-phase line voltage to 230 V AC.

  • UPS units are sized to protect LC, MS, and data system hardware.
  • Do not connect additional components (nitrogen generators, water chillers, gas chromatographs) to UPS output.
  • North America: UPS requires one L6-30, 30-amp wall socket.
  • Other areas: UPS typically connects using the standard MS instrument power cord and wall socket.
  • Approximate maximum UPS dimensions: 838 mm (33 in) height, 813 mm (32 in) width, 305 mm (12 in) depth.
pp. 8
Electrical Specifications

Table of nominal voltage, rated current/power, typical power consumption, and power cord connector for each system component. Also covers transformer requirements for nitrogen generators.

  • Xevo MRT/Xevo MRT P10: 200-240 V, 50/60 Hz, 700 W rated, 650 W typical, IEC 60320 C19.
  • External embedded PC (EPC): 100-240 V, 50/60 Hz, 12 to 6 A, 350 W typical, IEC 60320 C13.
  • PC: 100-240 V, 50/60 Hz, 10 to 5 A, 110 W typical, IEC 60320 C13.
  • Monitor: 100-240 V, 50/60 Hz, 1.5 A, 25 W typical, IEC 60320 C13.
  • Vacuum pump: 200-240 V, 50/60 Hz, 11.1 A, 460 W typical, IEC 60320 C19.
  • If supply voltages do not meet operating range under all conditions, use a transformer.
  • Nitrogen generator: do not run continuously below 220 V for extended periods (except US 120-V Peak Scientific generator).
  • If main supply runs continuously between 194 and 219 V, install Peak Scientific Dual Tap Transformer (06-3200).
pp. 9
Gas Requirements

Specifies nitrogen gas purity, pressure regulation, consumption, and connection hardware for API gas and collision gas. Includes customer-supplied regulator and tubing requirements.

  • API nitrogen: dry, oil-free, purity at least 95%, regulated at 6.75 ±0.25 bar (98 ±4 psi) outlet pressure.
  • Regulator outlet range example: 0 to 11 bar (0 to 160 psi).
  • Customer must provide regulator with adapter for 6-mm push-in fitting OD FEP tube (not applicable when nitrogen generator supplied).
  • Typical API nitrogen usage: 900 to 1800 L/h at atmospheric pressure.
  • Collision gas nitrogen: dry, high purity >99.0%.
  • If main nitrogen supply >99.0% purity, Waters supplies a tee regulator for collision gas.
  • If main supply <99.0% purity and separate collision gas supply, regulate at 0.75 ±0.25 bar (11 ±4 psi) using high-purity regulator (e.g., 0 to 2 bar / 0 to 29 psi range).
  • Typical collision gas usage: 0.30 mL/min.
  • If nitrogen purity <99%, customer provides regulator with adapter for 1/8-inch Swagelok type fitting.
  • API gas supply: 5-m (16.4-ft) FEP tube, 6-mm OD supplied; customer supplies high-purity nitrogen through 6-mm adapter.
  • Collision gas supply: 5-m (16.4-ft) stainless steel tubing, 1/8-inch OD supplied; customer supplies high-purity nitrogen through 1/8-inch adapter (Swagelok recommended).
pp. 10-13
Extraction and Waste Collection Requirements

Covers liquid waste collection, exhaust outlet requirements, source exhaust connection types (open vs. closed), vacuum pump exhaust, and waste connection hardware.

  • Source exhaust line must feed into ducted laboratory fume hood or laboratory exhaust system.
  • Pressure at source exhaust outlet must never exceed atmospheric pressure.
  • Open exhaust connection requires exhaust flow rate of at least 5000 L/hour.
  • Open exhaust connection preferred for high aqueous LC flow (≥60% water at ≥0.5 mL/min).
  • Open exhaust connection should be horizontal and not more than 1.2 m above floor level.
  • Closed exhaust connection requires approximately 2000 L/hour.
  • 3 m (9.8 ft) of 12-mm OD FEP tubing supplied for source exhaust; extend lab exhaust system, not source exhaust tubing.
  • Vacuum pump exhaust: 5-m (16-ft) PVC tube, 12-mm ID supplied; customer supplies industrial vent or fume hood.
  • Liquid waste: 2-m (6.6-ft) PTFE corrugated tubing, 3/8 inch ID supplied; customer supplies waste bottle, 1 L minimum.
pp. 14
Shipping Container Handling Requirements

Provides storage conditions, moving guidelines, and crate dimensions/weights for the shipping containers. Emphasizes not unpacking before lifting or moving.

  • Do not unpack equipment before lifting or moving.
  • Storage temperature: -30 to 55 °C (-22 to 131 °F); humidity <80%, non-condensing.
  • Store away from heavy machinery generating excessive floor vibration.
  • Use forklift truck for transporting the system.
  • Largest crate (Package 1): 1050 mm W x 1620 mm D x 1330 mm H, 560 kg (1234.6 lbs).
  • Package 2: 1000 mm W x 1200 mm D x 840 mm H, 113 kg (249.1 lbs).
  • Move largest crate with pallet truck forks through the middle of the longest side.
pp. 15-23
Space and Load Requirements

Specifies bench space, clearances, lifting equipment, dimensions, weights, and positioning for the mass spectrometer, EPC, vacuum pump, LC system, data system, WRDS, and DESI XS.

  • System must be installed on a flat surface level to within ±1° in all directions.
  • Maximum sound pressure level at operator position: 59.2 dB(A), including MS and external vacuum pump.
  • Minimum bench depth: 900 mm.
  • Mass spectrometer dimensions: 705 mm W x 1205 mm D x 813 mm H; weight 248 kg (546.7 lbs).
  • EPC dimensions: 215 mm W x 462 mm D x 455 mm H; weight 6.8 kg (14.9 lbs).
  • Vacuum pump (Xevo MRT): 324 mm W x 584 mm D x 450 mm H; weight 65 kg (141 lbs).
  • Vacuum pump (Xevo MRT P10): 356 mm W x 713 mm D x 467 mm H; weight 95 kg (209.4 lbs).
  • Data system (PC and monitor) weight: <18 kg (40 lbs).
  • During service, minimum clearance of 60 cm (23.6 in) required on all sides of instrument.
  • Additional 10 cm (4 in) recommended behind workbench for vacuum tubing.
  • Lifting table required: capacity 300 kg (661 lbs), lowered height ≤250 mm (9.8 in), raised height at least bench height.
  • Vacuum pump must be within 2.5 m (100 in) of instrument rear panel (control cable 2.5 m, vacuum hoses 1.5 m each).
  • Vacuum pump clearance: minimum 10 cm on two short sides and 5 cm on two long sides; ambient temperature around pump must not exceed 40 °C.
  • EPC must be within 2 m (79 in) of instrument rear panel; do not place on floor.
  • ionBench part number: BCH87123-MRT.
pp. 24-25
Environmental Requirements

Defines laboratory ventilation, positioning, temperature, humidity, altitude, vibration, magnetic field, and RF emission limits for the installation site.

  • Laboratory environment must be pollution degree 2 (only non-conductive pollution allowed).
  • Do not locate in direct sunlight; indoor use only.
  • Air-conditioning units must not be located directly above the system; airflow from diffusers must not be directed on the system.
  • Ambient temperature range: 15 to 28 °C (59 to 82 °F).
  • Temperature stability: less than 2 °C or 3.5 °F peak-to-peak in 1.5 hours.
  • Relative humidity: 20% to 80%, non-condensing.
  • Altitude: below 2000 m (6562 ft).
  • Avoid magnetic fields greater than 10 Gauss (e.g., from NMR or magnetic sector mass spectrometers).
  • Do not place within RF field greater than 1.0 V/m.
pp. 26-27
Chemicals and Labware Requirements

Lists customer-supplied chemicals, sample preparation equipment, glassware cleaning precautions, ultrasonic bath requirements, and test sample storage conditions.

  • Customer supplies LC-MS-grade water and LC-MS-grade acetonitrile.
  • Sample preparation equipment: calibrated pipettes 1 μL to 1 mL, measuring cylinders 100 mL to 1 L, 100-mL volumetric flask, nitrile gloves.
  • Do not use surfactants for cleaning glassware or components.
  • Ultrasonic bath for cleaning MS parts must be at least 300 mm x 150 mm x 100 mm deep.
  • Cleaning vessels must be surfactant-free glass, diameter at least 120 mm and approximately 120 mm high.
  • Test sample kits and storage conditions: 700002741 room temperature; 186007091 -20 °C; 186002325 room temperature; 186002328 room temperature; 700002642 room temperature; 700002669 room temperature; 700002857 room temperature; 186008430 -20 °C.
pp. 28
Informatics Software Requirements

Requires completion and sign-off of the waters_connect/UNIFI Site Preparation Guide checklist (715008018). Lists applicable checklists for installation types.

  • waters_connect Base Kit with UNIFI v3.6.0 (and later) Site Preparation Guide applies to workstation migration from UNIFI, workstation migration from waters_connect/UNIFI, and new workstation installation.
pp. 29-31
Appendix: Regional Power Cords, Piping Materials, Gas Regulators

Provides regional power cord connector types, piping material requirements for stainless steel and copper, and gas regulator recommendations.

  • Regional power cords: US/Canada NEMA 5-15P (C13) and NEMA L6-15P (C19); Europe CEE 7/7 16 A; UK 13 A; etc.
  • Avoid threaded joints and Teflon tape where possible; do not apply pipe thread compound where Teflon tape is used.
  • Pressure-test and purge all piping before use; leave capped and charged with inert gas.
  • Stainless steel: type 316L, compression fittings or orbital welded joints, medical grade, chemically cleaned and passivated or electropolished.
  • Copper: no soldered joints, no fluxes; refrigerant-grade or medical-grade copper with brazed joints; internally cleaned, degreased, dried to ASTM B280.
  • Brazing: flux-free copper-phosphorous-silver rod under constant flow of oil-free dry nitrogen until cooled.
  • Dual-stage regulators recommended for gas cylinders; single-stage suitable for central laboratory supply systems with stable pressure.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
AC main power supply voltage fluctuations must not exceed ±10% of the nominal device voltage. p.6 ↗ Verify site power quality before installation; use transformer or conditioner if outside range. Whole system
Xevo MRT/Xevo MRT P10 mass spectrometer requires 200-240 V, 50/60 Hz, 700 W rated, IEC 60320 C19 connector. p.8 ↗ Provide a C19 outlet rated for at least 16 A on a 200-240 V circuit. Xevo MRT/Xevo MRT P10
Vacuum pump requires 200-240 V, 50/60 Hz, 11.1 A, 460 W typical, IEC 60320 C19 connector. p.8 ↗ Provide a dedicated C19 outlet; pump runs continuously. Vacuum pump
Nitrogen API gas must be dry, oil-free, purity at least 95%, regulated at 6.75 ±0.25 bar (98 ±4 psi). p.9 ↗ Select regulator with outlet range covering 6.75 bar, e.g., 0-11 bar (0-160 psi). Xevo MRT/Xevo MRT P10
Typical API nitrogen consumption is 900 to 1800 L/h at atmospheric pressure. p.9 ↗ Size nitrogen supply or generator accordingly; a large cylinder may empty during long runs. Xevo MRT/Xevo MRT P10
Collision gas nitrogen must be dry and >99.0% purity; if separate supply, regulate at 0.75 ±0.25 bar (11 ±4 psi). p.9 ↗ Use high-purity regulator with 0-2 bar (0-29 psi) range and 1/8-inch Swagelok adapter if main supply <99%. Xevo MRT/Xevo MRT P10
Open source exhaust connection requires exhaust flow rate of at least 5000 L/hour; closed connection requires approximately 2000 L/hour. p.11 ↗ Size laboratory exhaust system accordingly; open connection preferred for high aqueous flow. Xevo MRT/Xevo MRT P10
Source exhaust pressure at outlet must never exceed atmospheric pressure. p.11 ↗ Design exhaust connection to ensure negative pressure at the outlet to avoid API source pressure monitoring faults. Xevo MRT/Xevo MRT P10
System must be installed on a flat surface level to within ±1° in all directions. p.15 ↗ Verify bench levelness before installation. Whole system
Mass spectrometer dimensions: 705 mm W x 1205 mm D x 813 mm H; weight 248 kg (546.7 lbs). p.21 ↗ Bench must support at least 248 kg and provide minimum depth 900 mm. Xevo MRT/Xevo MRT P10
Vacuum pump (Xevo MRT P10) dimensions: 356 mm W x 713 mm D x 467 mm H; weight 95 kg (209.4 lbs). p.21 ↗ Provide floor space near instrument rear; ensure clearance and ventilation. Xevo MRT P10
During service visits, minimum clearance of 60 cm (23.6 in) required on all sides of the mass spectrometer. p.22 ↗ Use mobile bench or provide clearance for service access. Xevo MRT/Xevo MRT P10
Vacuum pump must be positioned within 2.5 m (100 in) of instrument rear panel; control cable is 2.5 m and vacuum hoses are 1.5 m each. p.22 ↗ Plan pump location accordingly; ensure ambient temperature around pump does not exceed 40 °C. Whole system
EPC must be positioned within 2 m (79 in) of instrument rear panel; do not place on floor. p.23 ↗ Place EPC on bench, separate desk, or underneath instrument with open ventilation. External embedded PC (EPC)
Ambient temperature range: 15 to 28 °C (59 to 82 °F); stability less than 2 °C peak-to-peak in 1.5 hours. p.24 ↗ Air-conditioned laboratory required; avoid drafts and direct sunlight. Whole system
Relative humidity must be 20% to 80%, non-condensing. p.24 ↗ Monitor lab humidity; avoid condensation conditions. Whole system
System designed and tested to operate below 2000 m (6562 ft) altitude. p.24 ↗ Confirm site altitude is within limit. Whole system
Do not place system within magnetic fields greater than 10 Gauss or RF fields greater than 1.0 V/m. p.25 ↗ Locate away from NMR/magnetic sector instruments and RF sources. Whole system
Lifting table required: capacity 300 kg (661 lbs), lowered height ≤250 mm (9.8 in), raised height at least bench height. p.20 ↗ Provide suitable lifting equipment before Waters engineer arrives; do not use A-frame. Whole system
If main power supply runs continuously between 194 and 219 V, install Peak Scientific Dual Tap Transformer (06-3200) for nitrogen generator. p.8 ↗ Protect nitrogen generator from low voltage damage. Nitrogen generator option