Mounting, magnet unit removal and installation warnings
Procedurepp. 28-29
Wiring and electrical connection instructions
Procedurepp. 30-33
Operating notes and gauge adjustment procedures
Procedurepp. 34-43
Removal, cleaning and disassembly of FRG-700/702
Procedurepp. 44-45
Fault diagnosis and spare parts
Orderingpp. 46-51
Spare parts list, technical info and contact details
p. 16table, 7 rows
p. 17table, 31 rows
p. 45table, 5 rows
p. 18table, 31 rows
p. 15table, 24 rows
p. 29table, 3 rows
p. 46table, 5 rows
p. 5table, 11 rows
Part numbers
Part number
Description
FRG-700 (FPM sealed)
FRG-702 (all-metal)
FRG700KF25
DN 25 ISO-KF
FRG702KF25
DN 25 ISO-KF
FRG700CF35
DN 40 CF-C
FRG702KF40
DN 40 ISO-KF
FRG702CF35
DN 40 CF-C
Table 1 Technical Specifications
Specifications
Value
Measuring range (air, N2)
5×10-9 ... 1000 mbar
Accuracy (N2)
≈±30%
in the range 1×10-8 ... 100 mbar
Reproducibility
≈±5%
in the range 1×10-8 ... 100 mbar
Gas type dependence
See page 21
Output signal (measuring signal)
Voltage range
0 ... +10.5 V
Measuring range
1.82 ... 8.6 V
Voltage vs. pressure
logarithmic, 0.6 V / decade
Error signal
<0.5 V no supply
>9.5 V Pirani measurement element
defective (filament rupture)
Output impedance
2×10 Ω
Minimum loaded impedance
10 kΩ, short-circuit proof
Response time (pressure dependent)
p > 10-6 mbar
<10 ms
p = 10-8 mbar
≈1000 ms
Gauge identification
85 kΩ referenced to supply common
Status
p > 10-2 mbar
Pirani-only mode
p < 10-2 mbar
Cold cathode not ignited
Pirani-only mode
p < 10-2 mbar
Cold cathode ignited
Combined Pirani /
cold cathode mode
Lamp
Pin 6
Low = 0 V
Low = 0 V
High = 15 ... 30 VDC
High voltage on (LED on)
Table 4 Spare Parts for FRG-700
Pos.
Description
Ordering number
12 13a 11 10a
Maintenance kit, consisting of: 1× support ring 1× O-ring FPM ø3.69×1.78 1× O-ring FPM ø10.82×1.78 3× ignition aid
FRG700MAINT
13 12 13a 11 10a 10
Repair kit, consisting of: 1× Pirani element 1× support ring 1× O-ring FPM ø3.69×1.78 1× O-ring FPM ø10.82×1.78 3× ignition aid 1× anode, complete
FRG700REPR
10a
Set of ignition aids, comprising: 10× ignition aid
FRG700IGN
Mounting tool for ignition aid
FRG700IGNT
Measuring system DN 25 ISO-KF flange DN 40 CF-F flange
FRG700MEAS25 FRG700MEAS40
Table 1 Technical Specifications (continued)
Specifications
Value
Supply
WARNING
The gauge may only be connected to power supplies, instruments or control devices that conform to the requirements of a grounded extra-low voltage (SELV). The connection to the gauge has to be fused1).
Supply voltage at the gauge
15.0 ... 30.0 VDC (ripple ≤1 Vpp)
Power consumption
≤2 W
Fuse1)
≤1 AT
The minimum voltage of the power supply must be increased proportionally to the length of the sensor cable.
Voltage at the supply unit with maximum line length
16.0 ... 30.0 VDC (ripple ≤1 Vpp)
Adjustment
<HV> potentiometer at <10-4 mbar
<ATM> potentiometer at atmospheric pressure
Electrical connection
Sensor cable
FCC68 female, 8 poles
Line length
8 conductors plus shielding ≤50 m (8x0.14 mm²)
Operating voltage
≤3.3 kV
Operating current
≤500 µA
Grounding concept
Vacuum connection-signal common
see paragraoh Electrical Connection (page 28)
Supply common-signal common
connected via 10 kΩ (max. voltage differential with respect to safety ±50 V with respect to accuracy ±10 V)
conducted separately
Materials exposed to vacuum
Vacuum connection
stainless steel
Measuring chamber
stainless steel
Feedthrough
ceramic
Internal seals
FPM 75
Anode
Ag, Cu, soft solder (Sn, Ag)
Ignition aid
Mo
Pirani measuring tube
stainless steel
Pirani filament
Ni, Au
FRG-700
FRG-702
W
1) Agilent controller fulfill these requirements.
Содержание
Раздел
Страница
Technical Information
Product Description
16
Technical Specifications
17
FRG-700 and FRG-702 Outline
20
Installation
24
Vacuum Connection
24
Removing the Magnet Unit (only for Gauges with CF Flanges)
27
Electrical Connection
28
Use With an Agilent Controller
28
Use With Another Control Device
28
Operation
29
Adjusting the Gauge
32
Deinstallation
34
Cleaning FRG-700
35
Replacing the Pirani element
36
Cleaning FRG-700, Replacing Parts
36
Disassembling FRG-700
37
Reassembling FRG-700
38
Cleaning FRG-702
39
Cleaning FRG-702, Replacing Parts
40
Disassembling FRG-702
41
Reassembling FRG-702
42
Troubleshooting
44
Spare Parts
45
Table 2
Pressure
Lamp on the gauge
Operating mode
Pin 6
p > 1x10-2 mbar
Pirani-only mode
Low = 0 V
p < 1x10-2 mbar
Cold cathode not ignited
Pirani-only mode
Low = 0 V
p < 1x10-2 mbar
Cold cathode ignited
Combined Pirani / cold cathode mode
High = 15 ... 30 VDC
Table 5 Spare Parts for FRG-702
Pos.
Description
Ordering number
11
12
10a
Maintenance kit, consisting of:
1× seal HN 100 (9x1.6)
1× centering ring
3× ignition aid
FRG702MAINT
13
13a
13b
9a
10
10a
11
12
Repair kit, consisting of::
1× Pirani element with glass feedthrough
1× screw fitting
1× copper seal
1× anode extension piece
1× anode, complete
3× ignition aid
1× seal HN 100 (9x1.6)
1× centering ring
FRG702REPR
10a
Set of ignition aids, comprising:
10× ignition aid
FRG700IGN
Mounting tool for ignition aid
FRG700IGNT
Measuring system
DN 25 ISO-KF flange
DN 40 CF-F flange
FRG702MEAS25
FRG702MEAS40
Instructions for Use
Раздел
Страница
About this manual
6
Validity
6
Definitions and terminology
7
Definition of Caution, Warning and Note
7
Warning Symbols
8
Safety
10
Proper use
10
Improper use
11
Safety guideline for FRG-700 and FRG-702
12
Disposal
13
Service
14
Engineering Facts
14 facts · every claim linked to a page
Selection
The gauge covers a very wide measuring range from 5×10-9 to 1000 mbar, so one sensor can span UHV to rough vacuum.
Determines whether a single gauge can replace separate Pirani and cold-cathode heads across the whole process range.
Source
“Measuring range (air, N2) 5×10-9 ... 1000 mbar”
Selection
Accuracy is only about ±30% in the 1×10-8 to 100 mbar range, so it is not a precision measurement device.
Sets expectations for process control and whether a higher-accuracy gauge is needed.
Source
“≈±30% in the range 1×10-8 ... 100 mbar”
Installation
The gauge must be powered from a grounded SELV supply and the connection must be fused at ≤1 AT.
Defines the electrical supply and protection requirements before wiring the sensor cable.
Source
“The gauge may only be connected to power supplies, instruments or control devices that conform to the requirements of a grounded extra-low voltage (SELV).”
Installation
Supply voltage at the gauge must be 15.0 to 30.0 VDC with ripple ≤1 Vpp, and must be raised for long cables.
Prevents undervoltage on long sensor cable runs up to 50 m.
Source
“15.0 ... 30.0 VDC (ripple ≤1 Vpp)”
Installation
The sensor cable is limited to 50 m using 8×0.14 mm² conductors with FCC68 female connector.
Constrains cable routing and controller placement during installation.
Source
“≤50 m (8x0.14 mm²)”
Installation
For KF connections a conductive metallic clamping ring is required to achieve protective grounding.
Ensures protective earth continuity required by EN 61010 for safety.
Source
“For gauges with KF connections, use a conductive metallic clamping ring.”
Installation
KF flange connections with elastomer seals cannot withstand overpressure above 2.5 bar, so O-rings with outer centering ring must be used.
Prevents seal failure and process media leakage in pressurized systems.
Source
“KF flange connections with elastomer seals (e.g. O-rings) cannot withstand such pressures.”
Installation
The gauge may be mounted in any orientation, but horizontal to upright is preferred to keep condensates and particles out.
Guides mounting position to reduce contamination and extend maintenance intervals.
Source
“The gauge may be mounted in any orientation.”
Installation
The gauge must be installed so the <HV> and <ATM> trimmer potentiometers remain accessible with a screwdriver for adjustment.
Allows field adjustment after installation without removing the gauge.
Source
“be sure to install it so that the <HV> and <ATM> trimmer potentiometers can be accessed with a screw driver.”
Operation
A stabilizing time of about 10 minutes is required, and the gauge should be left on permanently regardless of pressure.
Affects startup procedure and power/control design for continuous operation.
Source
“Allow for a stabilizing time of approx. 10 min. Once the gauge has been switched on, permanently leave it on irrespective of the pressure.”
Operation
The cold cathode circuit only activates below 1×10-2 mbar; above that the gauge runs in Pirani-only mode.
Explains the status output behavior and the crossover point between measuring principles.
Source
“The cold cathode measuring circuit is controlled by the Pirani circuit and is activated only at pressures <1×10-2 mbar”
Maintenance
Continuous operation in the 10-4 to 10-2 mbar range causes severe contamination and reduced uptime, while below 1×10-6 mbar the gauge can run over a year without cleaning.
Informs process operating point choice and maintenance planning.
Source
“Continuous operation in the range of 10-4 mbar ... 10-2 mbar can cause severe contamination as well as reduced up-time and maintenance cycles.”
Safety
Helium used in leak testing can cause electric arcing that damages the electronics, so the product must be switched off and the electronics unit removed before tightness tests.
Prevents electronics damage during helium leak testing of the installed system.
Source
“Before performing any tightness tests put the product out of operation and remove the electronics unit.”
Maintenance
Gauge failures due to contamination or wear and expendable parts such as the filament are not covered by warranty.
Affects lifecycle cost and spare-part stocking decisions.
Source
“Gauge failures due to contamination or wear and tear as well as expendable parts (e.g. filament) are not covered by the warranty.”