Eukaryote Total RNA Nano Assay Qualitative range 5 to 500 ng/µL
RNA 6000 Nano LabChip Kit
mRNA Nano Assay* Qualitative range 25 to 250 ng/µL
RNA 6000 Pico LabChip Kit
Eukaryote Total RNA Pico Assay Qualitative range 50 to 5000 pg/µL in water
RNA 6000 Pico LabChip Kit
mRNA Pico Assay* Qualitative range 250 to 5000 pg/µL in water
* The mRNA assays are suitable for analysis of cRNA as well.
Related Microarray Reagents
Description
Vendor and part number
Universal Human Reference RNA
Agilent p/n 740000
Universal Mouse Reference RNA
Agilent p/n 740100
Universal Rat Reference RNA
Agilent p/n 740200
Table 9 Required Reagents
Description
Vendor and part or catalog number
Low Input Quick Amp Labeling Kit, One-Color
Agilent p/n 5190-2305
RNA Spike-In Kit, One-Color
Agilent p/n 5188-5282
Gene Expression Hybridization Kit
Agilent p/n 5188-5242
Gene Expression Wash Buffer Kit
Agilent p/n 5188-5327
DNase/RNase-free distilled water
Invitrogen p/n 10977-015
RNeasy Mini Kit
Qiagen p/n 74104
ethanol (95% to 100% molecular biology grade)
Sigma-Aldrich p/n E7023-6×500ML
Milli-Q water or equivalent
isopropyl alcohol (molecular biology grade)
Table 13 Dilutions of Spike Mix for Cyanine 3-labeling
Starting Amount of RNA — Total RNA (ng)
Starting Amount of RNA — PolyA RNA (ng)
Serial Dilution — First
Serial Dilution — Second
Serial Dilution — Third
Serial Dilution — Fourth
Spike Mix Volume to be used in each labeling reaction (µL)
10
1:20
1:25
1:20
1:10
2
25
1:20
1:25
1:20
1:4
2
50
1:20
1:25
1:20
1:2
2
100
1:20
1:25
1:20
2
200
1:20
1:25
1:10
2
5
1:20
1:25
1:20
2
Engineering Facts
14 facts · every claim linked to a page
Selection
The kit supports a sample input RNA range of 10 ng to 200 ng total RNA or a minimum of 5 ng poly A+ RNA for one-color processing.
Determines whether the kit is compatible with the available sample amount and whether low-input samples can be processed.
Source
“The Low Input Quick Amp Labeling Kit, One-Color generates fluorescent cRNA (complimentary RNA) with a sample input RNA range between 10 ng and 200 ng of total RNA or a minimum of 5 ng of poly A+ RNA for one-color processing.”
Performance
RNA purity must be assessed by A260/A230 ratio; impure RNA must be purified before use to avoid poor results.
Impure template RNA can compromise labeling and hybridization, so a purification step may be required before starting the protocol.
Source
“RNA that is not pure, as measured by A260/A230 ratio, can lead to poor results and must be purified.”
Operation
Spike Mix dilutions must be stored at –70°C to –80°C in a non-defrosting freezer for up to 1 year from receipt.
Proper storage of the Spike-In Kit is necessary to maintain control performance and avoid degradation.
Source
“Store the RNA Spike-In Kit, One-Color at –70°C to –80°C in a non-defrosting freezer for up to 1 year from the date of receipt.”
Operation
The first dilution of the Spike Mix positive controls can be stored for up to 2 months at –70°C to –80°C and must not be freeze/thawed more than eight times.
Repeated freeze-thaw cycles degrade the spike controls, so tracking cycles and storage time is essential for reproducible results.
Source
“Store the first dilution of the Spike Mix positive controls for up to 2 months in a non-defrosting freezer at -70°C to -80°C. Do not freeze/thaw more than eight times.”
Operation
The transcription reaction is incubated in a circulating water bath at 40°C for 2 hours; if not proceeding immediately, samples must be stored at –80°C.
The specified incubation temperature and duration are critical for efficient amplification and Cy3 incorporation.
Source
“Incubate samples in a circulating water bath at 40°C for 2 hours.”
Performance
If the specific activity does not meet the requirements in Table 17, hybridization must not proceed and cRNA preparation must be repeated.
Low specific activity indicates poor labeling and would lead to weak or unreliable microarray signals.
Source
“If the specific activity does not meet the requirements listed in Table 17, do not continue to hybridization. Repeat preparation of cRNA.”
Installation
When loading the hybridization assembly, the pipette tip or hybridization solution must not touch the gasket walls to avoid leakage.
Gasket leakage during hybridization causes sample loss and cross-contamination, invalidating the experiment.
Source
“Do not let the pipette tip or the hybridization solution touch the gasket walls. Allowing liquid to touch the gasket wall greatly increases the likelihood of gasket leakage.”
Operation
Gene Expression Wash Buffer 2 must be prewarmed to 37°C the night before washing arrays.
Prewarming Wash Buffer 2 is required for the elevated-temperature wash step that reduces background and ensures proper stringency.
Source
“Warm the Gene Expression Wash Buffer 2 to 37°C as follows:”
Maintenance
Staining dishes, racks, and stir bars must be solvent-washed with acetonitrile or isopropyl alcohol to avoid wash artifacts on slides and images.
Residual contaminants on wash equipment can produce artifacts that compromise image quality and data extraction.
Source
“Wash staining dishes, racks and stir bars with acetonitrile or isopropyl alcohol to avoid wash artifacts on your slides and images.”
Operation
For Agilent Scanner B or C in environments with ozone levels exceeding 50 ppb, slides must be placed in a slide holder with an ozone-barrier slide cover or processed with Stabilization and Drying Solution.
Ozone degrades Cy3 signals; without mitigation, high-ozone environments can cause signal loss and unreliable results.
Source
“In environments in which the ozone level exceeds 50 ppb, immediately put the slides with active microarray surface (with “Agilent”-labeled barcode) facing up in a slide holder.”
Selection
The Agilent B Scanner does not support G3 microarrays; G3 microarrays require the Agilent C Scanner or SureScan microarray scanner.
Scanner compatibility must be verified before selecting a microarray format to avoid acquiring incompatible hardware.
Source
“The Agilent B Scanner does not support G3 microarrays. For G3 microarrays, use the Agilent C Scanner or SureScan microarray scanner.”
Operation
For 5 µm scans of SurePrint HD formats, Agilent Scan Control software v7.0.03 is recommended.
Using the recommended software version ensures correct scan settings and reliable data acquisition for HD formats.
Source
“Agilent Scan Control software v7.0.03 is recommended for 5 μm scans of SurePrint HD formats.”
Operation
Feature Extraction version 10.7 or higher can automatically download Grid Templates, protocols, and QC metrics from eArray when the eArray server option is enabled.
Automatic download reduces manual setup errors and ensures the correct grid template and QC metrics are used for extraction.
Source
“Feature Extraction version 10.7 or higher can automatically download Grid Templates, protocols and QC metrics (QCM or QCMT).”
Performance
Using a thermocycler instead of water baths can slightly lower the cRNA yield.
Laboratories considering a thermocycler protocol should account for potentially lower cRNA yield when planning input amounts and QC thresholds.
Source
“The use of a thermocycler can slightly lower the yield of cRNA when compared to the use of water baths.”