Wash Procedure without Stabilization and Drying Solution. "Step 4. Wash the microarray slides" on page 39
No
> 5 ppb and < 10 ppb
Wash Procedure without Stabilization and Drying Solution. "Step 4. Wash the microarray slides" on page 39
Yes
> 10 ppb
Wash Procedure with Stabilization and Drying Solution. See "Step 1. Prepare the Stabilization and Drying Solution" on page 65 and "Step 2. Wash with Stabilization and Drying Solution" on page 66.
Yes
Related Microarray Reagents
Description
Vendor and part number
Universal Human Reference RNA
Agilent p/n 740000
Universal Mouse Reference RNA
Agilent p/n 740100
Universal Rat Reference RNA
Agilent p/n 740200
Table 10 Dilutions of Spike A Mix for cyanine 3 and Spike B Mix for cyanine 5 labeling
Starting Amount of RNA Total RNA (ng)
Serial Dilution First
Serial Dilution Second
Serial Dilution Third
Serial Dilution Fourth
Spike Mix Volume to be used in each labeling reaction (µL)
25
1:20
1:40
1:16
1:8
2
50
1:20
1:40
1:16
1:4
2
100
1:20
1:40
1:16
1:2
2
Table 11 WT Primer MasterMix
Component
Volume (µL) per reaction
Volume (µL) per 5 reaction
Volume (µL) per 10 reactions
WT Primer
1
5
10
Diluted Spike A Mix and Spike B Mix
2
10
20
Total Volume
3
15
30
Table 12 cDNA Master Mix
Component
Volume (µL) per reaction
Volume (µL) per 5 reaction
Volume (µL) per 10 reactions
5× First Strand Buffer (green cap)
2
10
20
0.1 M DTT (white cap)
1
5
10
10 mM dNTP Mix (green cap)
0.5
2.5
5
Affinity Script RNase Block Mix (violet cap)
1.2
6
12
Total Volume
4.7
23.5
47
Engineering Facts
14 facts · every claim linked to a page
Selection
The protocol requires the Low Input Quick Amp WT Labeling Kit, Two-Color (p/n 5190-2944), not the standard Low Input Quick Amp Labeling kit.
Using the wrong labeling kit will invalidate the workflow and Agilent will not guarantee performance.
Source
“Low Input Quick Amp WT Labeling Kit, Two-Color”
Selection
Agilent Exon microarrays are only available on SurePrint G3 formats, so a SureScan or Agilent C scanner is mandatory.
Scanner selection must be confirmed before committing to the Exon microarray platform.
Source
“A SureScan or Agilent C microarray scanner is required for SurePrint G3 formats.”
Selection
The starting total RNA input is 25–100 ng, with 50 ng recommended for 2-, 4-, and 8-pack arrays and 100 ng for 1-pack arrays.
Sample input capacity determines whether low-input samples are compatible with the chosen array format.
Source
“The starting input for the Low Input Quick Amp WT Labeling Kit, Two-Color ranges from 25 ng to 100 ng of total RNA.”
Installation
After opening the microarray pouch foil, slides must be stored at room temperature in the dark under vacuum desiccation or nitrogen purge, never in open air.
Improper storage after opening causes slide degradation and invalidates experiments.
Source
“Do not store microarray slides in open air after breaking foil.”
Installation
All unused wells in the SureHyb gasket slide must be filled with 1× Hybridization Buffer because empty wells cause hybridization failure.
Filling unused wells is a mandatory assembly step to prevent run failure.
Source
“Make sure all wells contain sample or 1× Hybridization Buffer. Empty wells can cause failure in hybridization.”
Operation
The array slide must be placed active side down with the Agilent barcode facing down and the numeric barcode facing up, and must not be dropped onto the gasket.
Correct orientation and gentle placement prevent cross-contamination between gasket wells.
Source
“Do not drop the array slide onto the gasket. Doing so increases the chances of samples mixing between gasket wells.”
Operation
Gene Expression Wash Buffer 2 must be prewarmed to 37°C the night before washing arrays.
Wash Buffer 2 temperature directly affects wash stringency and signal quality.
Source
“Tightly cap the sterile storage bottle and put in a 37°C water bath the night before washing arrays.”
Operation
The array-gasket sandwich must be pried open from the barcode end only, and only the barcode portion or edges of the microarray slide may be touched.
Opening from the wrong end or touching the array area damages features and contaminates the slide.
Source
“pry the sandwich open from the barcode end only”
Operation
For uniform washing, no more than eight microarray slides should be processed per disassembly/wash group.
Exceeding eight slides per group causes non-uniform washing and inconsistent data.
Source
“For uniform washing, do up to a maximum of eight disassembly procedures yielding eight microarray slides.”
Maintenance
Staining dishes, racks, and stir bars must be solvent-washed with acetonitrile (if exposed to Stabilization and Drying Solution) or isopropyl alcohol (if not) to avoid wash artifacts.
Residual contaminants on glassware produce image artifacts that corrupt feature extraction.
Source
“Wash staining dishes, racks and stir bars with acetonitrile or isopropyl alcohol to avoid wash artifacts on your slides and images.”
Performance
The two-color platform is robust only at ozone levels of 5 ppb or less; above 10 ppb the Stabilization and Drying Solution wash procedure is required.
Lab ozone level dictates which wash procedure and ozone-barrier accessories must be used.
Source
“The Agilent two-color platform is robust in environments where the ozone level is 5 ppb (approximately 10 μg/m3) or less.”
Performance
The recommended specific activity for hybridization is 15 pmol Cy3 per μg cRNA.
Specific activity is a key QC acceptance criterion for labeled cRNA before hybridization.
Source
“The recommended Specific Activity for hybridization is increased to 15 pmol Cy3 per μg cRNA.”
Cumplimiento
Agilent does not guarantee microarray performance or provide technical support when non-Agilent protocols are used to process Agilent microarrays.
Deviating from the protocol voids Agilent performance guarantees and support.
Source
“Agilent cannot guarantee microarray performance and does not provide technical support to those who use non-Agilent protocols in processing Agilent microarrays.”
Cumplimiento
Agilent guarantees data quality only when Agilent microarrays are scanned on Agilent scanners.
Using a non-Agilent scanner removes the data quality guarantee and requires compatibility verification.
Source
“Agilent can guarantee the quality of data only if the data comes from Agilent microarrays scanned on Agilent scanners.”