Product identification and general description of the CS 25 unit
Specp. 2
Technical data according to DIN 12876 for the CS 25
p. 1table, 31 rows
Technical data according to DIN 12876 — Huber CS 25 (VDH1100664)
Parameter
Value
Operating temperature range
-20...15 °C
temperature set point / display
colour LED Touchscreen
Internal temperature sensor
Pt100
Temperature stability at -10°C
1,5 K
Interface digital
RS232
Safety classification
I / NFL
Cooling power at ambient temperature 20°C
at 15°C
3,5 kW
at 10°C
2,5 kW
at 0°C
1,6 kW
at -10°C
0,98 kW
at -20°C
0,7 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Refrigerant quantity
2 kg
CO2 equivalent
2,79 t
Circulation pump: max. delivery
50 l/min
max. delivery pressure
4,5 bar
Pump connection
G1/2 male
min. filling capacity
24 l
max. filling capacity
37 l
Overall dimensions WxDxH **
580x660x820 mm
Net weight
121 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
208V 2~ 60Hz
max. current
12,5 A
Fuse
2x20 A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
32 °C
from Serial-No.:
1.0/20
** Overall dimensions WxDxH. Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
13 facts · every claim linked to a page
Selection
The chiller's cooling capacity is rated at 20°C ambient and will drop if ambient temperature rises, so sizing must account for actual room conditions.
Cooling capacity must be derated for higher ambient temperatures to avoid undersizing.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires a 208V two-phase 60Hz supply with 12.5A max current and 2x20A fuses, so the electrical supply must match this configuration.
Electrical supply and protection must be provisioned accordingly before installation.
Source
“Power supply requirement 208V 2~ 60Hz”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Supply quality must be verified to avoid invalid tolerance combinations that could damage the unit.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices under 63A are supplied with cable but without plug, so a suitable plug must be provided on site.
Planning must include procurement and wiring of a compatible plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The chiller must be installed in an ambient temperature between 5°C and 32°C.
Site conditions outside this range will affect performance and may void warranty.
Source
“min. ambient temperature 5 °C”
Operation
The recommended thermofluid is a 50:50 water-monoethylene glycol mixture.
Using the correct fluid ensures proper heat transfer and freeze protection.
Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Operation
The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.
Operators need to know how to regulate flow to match process requirements.
Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280 (pressurized gas), requiring appropriate handling precautions.
Personnel must be aware of the refrigerant hazard for safe installation and maintenance.
Source
“R-449A (A1, H280)”
Safety
The unit has safety classification I / NFL according to DIN 12876.
This classification determines the required safety measures and permissible fluids.
Source
“Safety classification I / NFL”
Cumplimiento
The equipment emits Class A, Group 1 disturbances per EN55011, so it may cause radio interference in residential areas.
Installation in residential environments may require additional EMC mitigation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Cumplimiento
The equipment is compliant with US-SNAP for industrial process refrigeration and all applicable EU laws.
Regulatory compliance must be confirmed for the intended market and application.
Source
“This equipment is compliant to US-SNAP and all applicable EU laws.”
Maintenance
The refrigerant charge is 2 kg with a CO2 equivalent of 2.79 t, which must be recorded for leak checks and F-gas regulations.
Refrigerant quantity determines leak inspection intervals and environmental reporting obligations.
Source
“Refrigerant quantity 2 kg”
Performance
The circulation pump delivers up to 50 l/min at 4.5 bar, which must be sufficient for the process loop pressure drop.
Pump performance must match the external circuit resistance to ensure adequate flow.